K211668 · AliveCor, Inc. · DXH · Nov 30, 2021 · Cardiovascular
Device Facts
Record ID
K211668
Device Name
KardiaMobile Card
Applicant
AliveCor, Inc.
Product Code
DXH · Cardiovascular
Decision Date
Nov 30, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2920
Device Class
Class 2
Attributes
AI/ML
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Atrial Fibrillation Detection
—
—
—
—
—
—
—
Bradycardia Detection
—
—
—
—
—
—
—
Normal Sinus Rhythm Detection
—
—
—
—
—
—
—
Indications for Use
The KardiaMobile Card System is intended to record, store and transfer single-channel electrocardiogram (ECG) rhythms. The KardiaMobile Card System also displays ECG rhythms and output of ECG analysis from AliveCor's KardiaAl platform including detecting the presence of normal sinus rhythm, atrial fibrillation, bradycardia, and others. The KardiaMobile Card System is intended for use by healthcare professionals, patients with known or suspected heart conditions and health-conscious individuals. The device has not been tested and is not intended for pediatric use.
Device Story
Ambulatory single-channel ECG device; consists of hardware with stainless steel electrodes and Kardia phone app on iOS/Android mobile computing platforms (MCPs). User touches electrodes to record ECG; hardware transmits signal via Bluetooth Low Energy (BLE) to app. App displays ECG and provides analysis via KardiaAI platform; detects normal sinus rhythm, atrial fibrillation, bradycardia, and others. Used in ambulatory settings by clinicians, patients, or health-conscious individuals. Data synced to user account. Output assists healthcare providers in clinical decision-making regarding heart conditions. Benefits include convenient, accessible cardiac monitoring.
Clinical Evidence
Bench testing only. Verified ECG acquisition/transmission per IEC 60601-2-47:2012; Bluetooth coexistence testing; biocompatibility per ISO 10993-1:2018; electrical safety/EMC per IEC 60601-1:2012, IEC 60601-1-2:2014, and IEC 60601-1-11:2015. System-level verification confirmed performance against design specifications.
Technological Characteristics
Single-channel ECG; two stainless steel dry electrodes. Chassis material: PVC. Data transmission: Bluetooth Low Energy (BLE). Power: Battery-operated hardware. Connectivity: Wireless to iOS/Android MCPs. Software: Kardia app/KardiaAI platform. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-47, ISO 10993-1.
Indications for Use
Indicated for healthcare professionals, patients with known/suspected heart conditions, and health-conscious individuals to record, store, transfer, and analyze single-channel ECG rhythms. Not intended for pediatric use.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 30, 2021
AliveCor, Inc. Shani Frenkel Regulatory Affairs Manager 444 Castro Street, Suite 600 Mountain View, California 94041
Re: K211668
Trade/Device Name: KardiaMobile Card Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter And Reciever Regulatory Class: Class II Product Code: DXH, DPS, QDA Dated: November 3, 2021 Received: November 4, 2021
Dear Shani Frenkel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211668
Device Name KardiaMobile Card System
#### Indications for Use (Describe)
The KardiaMobile Card System is intended to record, store and transfer single-channel electrocardiogram (ECG) thythms. The KardiaMobile Card System also displays ECG rhythms and output of ECG analysis from AliveCor's KardiaAl platform including detecting the presence of normal sinus rhythm, atrial fibrillation, bradycardia, and others. The KardiaMobile Card System is intended for use by healthcare professionals, patients with known or suspected heart conditions and health-conscious individuals. The device has not been tested and is not intended for pediatric use.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Notification K211668
## GENERAL INFORMATION [807.92(a)(1)]
## Applicant:
AliveCor, Inc. 444 Castro Street, Suite 600 Mountain View, CA 94041 Phone: 650-396-8553 Fax: 650-282-7932
### Contact Person:
Shani Frenkel Regulatory Affairs Manager AliveCor, Inc. 444 Castro Street, Suite 600 Mountain View, CA 94041 Phone: 650-396-8553 Email: ra@alivecor.com
## Date Prepared: May 27, 2021
DEVICE INFORMATION [807.92(a)(2)]
## Trade/Proprietary Name:
KardiaMobile Card System
### Generic/Common Name:
Telephone electrocardiograph transmitter and receiver
### Classification:
21 CFR§870.2920, Telephone electrocardiograph transmitter and receiver, Class II
# Product Code:
DXH, DPS, QDA
Establishment Registration: 3009715978
## PREDICATE DEVICE(s) [807.92(a)(3)]
K191406 – AliveCor KardiaMobile System (Primary predicate) K183319 – AliveCor Triangle System (Reference device)
{4}------------------------------------------------
## DEVICE DESCRIPTION [807.92(a)(4)]
The AliveCor KardiaMobile Card System is a single-channel ambulatory electrocardiogram (ECG) device that is intended to record, store, transfer, display, and analyze single-channel ECG rhythms in an ambulatory setting. The device utilizes the computing power of Apple iOS- and Google Android-based smartphones to obtain and analyze single-channel ECGs. These smartphones are termed Mobile Computing Platforms (MCPs). The device consists of the hardware (that has the electrodes), and the Kardia phone app (installed on an MCP). The same software is implemented in the iOS and Android MCP. In either configuration, the same hardware is used to sense the ECG. The KardiaMobile Card Hardware transmits the ECG signal from the electrode to the Kardia phone app on the MCP to be analyzed and presented to the user. All ECGs are synced with the user's account.
## INDICATIONS FOR USE [807.92(a)(5)]
The KardiaMobile Card System is intended to record, store and transfer single-channel electrocardiogram (ECG) rhythms. The KardiaMobile Card System also displays ECG rhythms and output of ECG analysis from AliveCor's KardiaAl platform including detecting the presence of normal sinus rhythm, atrial fibrillation, bradycardia, and others. The KardiaMobile Card System is intended for use by healthcare professionals, patients with known or suspected heart conditions and health-conscious individuals. The device has not been tested and is not intended for pediatric use.
## TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED WITH THE PREDICATE DEVICE [807.92(a)(6)]
The KardiaMobile Card System device is similar to the primary predicate device in terms of principle of operation, technological and performance characteristics (mechanism of action, sterilization and shelf life), materials, software and anatomical site.
The design modifications of the KardiaMobile Card System that prompted this Special 510(k) relate to:
- Change in the Technological Characteristics: There is change in the technological characteristics related to the BLE data transmission method (instead of ultrasonic acoustics as utilized by the primary predicate device, KardiaMobile System K191406).
- AliveCor's KardiaMobile 6L System (also known as Triangle System, K183319), which o has also been determined to be substantially equivalent to the KardiaMobile System (K191406), will be used as a reference device for the BLE components related to the data transmission method of the KardiaMobile Card system.
- Change in the Hardware Material and Dimensional Specifications: There is a change in the hardware material for the device chassis from Acrylonitrile butadiene styrene (ABS) plastic to polyvinyl chloride (PVC). However, there is no change in electrode material. The KardiaMobile Card System will employ two dry electrodes made of Stainless Steel, which is the same material as used by the primary predicate device. There is a change in the subject device dimensional specifications.
These modifications do not raise different questions of safety or effectiveness. The KardiaMobile Card System has the same intended use and similar technological characteristics as the primary predicate device, the AliveCor KardiaMobile System (K191406), and utilizes the same Kardia app and KardiaAl platform cleared for the predicate device. No modifications were made to the Kardia app software's clinical functionalities with respect to ECG acquisition, display, and analysis because of this change since its prior clearance under K191406, K182396 and K201985.
{5}------------------------------------------------
## PERFORMANCE DATA [807.92(b)]
The following bench testing were conducted, and results of the collective testing demonstrated that differences between the subject device and the predicate device do not raise different questions of safety or effectiveness and that the KardiaMobile Card System is substantially equivalent to the primary predicate KardiaMobile System.
- ECG acquisition and transmission were verified with applicable clauses of IEC 60601-2-47:2012 to assess potential differences in ECG measurement and transmission which may arise as a result of the use of BLE for ECG transmission. Bluetooth Coexistence Testing was also conducted to verify the performance of the KardiaMobile Card System with respect to ECG measurement and MCP connectivity in the intended use environment.
- . A biocompatibility evaluation of the KardiaMobile Card hardware materials was conducted per the requirements of ISO 10993-1:2018 and in consideration of the FDA Guidance Document titled, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process" (September 4, 2020).
- . Electrical safety and electromagnetic compatibility testing were performed to verify that the KardiaMobile Card hardware complies with the requirements of:
- O IEC 60601-1:2012 and ANSI/AAMI ES60601-1: A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Medical electrical equipment - Part 1: General requirements for basic safety and essential performance),
- IEC 60601-1-2:2014 (Medical electrical equipment -Part 1-2: General requirements for O basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests),
- o IEC 60601-2-47:2012 (Medical electrical equipment - Part 2-47: Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems), and
- IEC 60601-1-11:2015 (Medical electrical equipment Part 1-11: General requirements o for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment).
- General hardware and system-level verifications were conducted to demonstrate that the subject device performs in accordance with its design specifications.
### SUMMARY
AliveCor, Inc. considers the KardiaMobile Card device to be substantially equivalent to the legally marketed predicate KardiaMobile device cleared under K191406. The testing, performed in agreement with recommendations in relevant FDA guidance documents and compliance standards, demonstrates that the differences in hardware between the modified device and the predicate device do not raise different questions of safety or effectiveness and that the modified KardiaMobile Card device is substantially equivalent to the predicate KardiaMobile device for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.