K183319 · AliveCor, Inc. · DXH · May 10, 2019 · Cardiovascular
Device Facts
Record ID
K183319
Device Name
Triangle System
Applicant
AliveCor, Inc.
Product Code
DXH · Cardiovascular
Decision Date
May 10, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2920
Device Class
Class 2
Attributes
AI/ML
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Cardiac arrhythmia detection
—
—
100% (n=44) diagnostic equivalence for assessing cardiac arrhythmias; minimum correlation of 0.96 and maximum RMS error of 47 uV for 6-lead ECG output.
—
—
Clinical study of 44 subjects (healthy volunteers and arrhythmia patients).
2 (electrophysiologists)
Indications for Use
The Triangle System is intended to record, store and two-channel electrocardiogram (ECG) rhythms. In single channel mode, the Triangle System can record Lead-I. In two channel mode, the Triangle System can record Lead-I and Lead-II simultaneously and derive Lead-II and unipolar limb leads aVR, aVF and aVL. The Triangle System also displays ECG rhythms and output of ECG analysis from AliveCor's KardiaAl platform including detecting the presence of normal sinus rhythm, atral fibrillation, bradycardia, and others. The Triangle System is intended for use by healthcare professionals, patients with known or suspected heart conditions and health conscious individuals. The device has not been tested and is not intended for pediatric use.
Device Story
Triangle System is a trans-telephonic ECG event recorder; records/stores/transfers single or two-channel ECG rhythms. Hardware consists of portable wireless device with three dry electrodes; user completes circuit via skin contact (fingers/legs). Device transmits ECG signals via Bluetooth Low Energy to Kardia app on iOS/Android smartphone (Mobile Computing Platform). App displays ECG waveforms and provides automated analysis via KardiaAI platform (detecting normal sinus rhythm, atrial fibrillation, bradycardia, tachycardia). Used by healthcare professionals, patients, or health-conscious individuals in ambulatory settings. Output allows clinicians to identify cardiac arrhythmias; aids in monitoring heart conditions. Benefits include convenient, portable access to diagnostic-quality ECG rhythm strips for arrhythmia detection.
Clinical Evidence
Clinical study (n=44) compared Triangle System 6-lead ECGs to gold-standard 12-lead ECGs. Qualitative assessment by two electrophysiologists found 100% diagnostic equivalence. Quantitative analysis showed median beat cross-correlation ≥ 0.96 and RMS error ≤ 47 µV. No adverse events reported.
Technological Characteristics
Three dry electrodes; 16-bit resolution; 300 samples/second; 0.5Hz-40Hz frequency response. Bluetooth Low Energy connectivity. Powered by 1 Lithium Manganese Dioxide coin cell (100-hour life). Dimensions: 90x30x7mm; weight: 24g. Software: iOS/Android app. Standards: ISO 10993-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-47.
Indications for Use
Indicated for healthcare professionals, patients with known/suspected heart conditions, and health-conscious individuals to record, store, and transfer 1- or 2-channel ECG rhythms. Not for pediatric use.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
AliveCor, Inc. % Prabhu Raghavan Vice President of Regulatory Quality AliveCor 224 Airport Parkway, Suite 250 San Jose, California 95110
Re: K183319
Trade/Device Name: Triangle System Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter And Receiver Regulatory Class: Class II Product Code: DXH, DPS, QDA Dated: April 9, 2019 Received: April 10, 2019
Dear Prabhu Raghavan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Acting Assistant Director External Heart Rhythm and Rate Devices Team Division of Cardiac. Electrophysiology. Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183319
Device Name Triangle System
#### Indications for Use (Describe)
The Triangle System is intended to record, store and two-channel electrocardiogram (ECG) rhythms. In single channel mode, the Triangle System can record Lead-I. In two channel mode, the Triangle System can record Lead-I and Lead-II simultaneously and derive Lead-II and unipolar limb leads aVR, aVF and aVL. The Triangle System also displays ECG rhythms and output of ECG analysis from AliveCor's KardiaAl platform including detecting the presence of normal sinus rhythm, atral fibrillation, bradycardia, and others. The Triangle System is intended for use by healthcare professionals, patients with known or suspected heart conditions and health conscious individuals. The device has not been tested and is not intended for pediatric use. and and in the
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Notification K183319
# GENERAL INFORMATION [807.92(a)(1)]
# Applicant:
AliveCor, Inc. 444 Castro Street, Suite 600 Mountain View, CA 94041 Phone: 650-396-8553 Fax: 650-282-7932
#### Contact Person:
Prabhu Raghavan Vice President of Regulatory and Quality Phone: 650-396-8553 Email: RA@alivecor.com.
### Date Prepared:
April 8, 2019
### DEVICE INFORMATION [807.92(a)(2)]
#### Trade Name:
Triangle System
#### Generic/Common Name:
Telephone electrocardiograph transmitter and receiver
#### Classification:
21 CFR\$870.2920, Transmitters and Receivers, Electrocardiograph, Telephone, Class II
#### Product Code:
DXH, DPS, QDA
{4}------------------------------------------------
# PREDICATE DEVICE(S) [807.92(a)(3)]
K142672 - AliveCor Heart Monitor (currently marketed as "KardiaMobile")
K181823 - AliveCor KardiaAI
# DEVICE DESCRIPTION [807.92(a)(4)]
The Triangle System is a trans-telephonic (transmission by smartphone) electrocardiogram (ECG) event recorder that records, stores, transfers, and analyzes single-channel or two-channel ECG rhythm recordings. The Triangle System provides output of one or six ECG leads by recording Lead-I and Lead-II simultaneously and derive Lead-III and unipolar limb leads aVR, aVF, and aVL for identifying cardiac arrhythmias. The device utilizes the computing power of iOS-based or Android-based smartphones (referred to as "Mobile Computing Platforms" (MCP) within this submission) to obtain and analyze ECG signals. The device consists of the Triangle Hardware (portable small wireless hardware with electrodes) and the Kardia app, which is installed on an MCP (i.e., iPhone or Android phone). The Triangle System is a new addition to AliveCor's ambulatory ECG devices that are intended to be used by lay users to record their ECG and obtain ECG analysis of Atrial Fibrillation. Normal Sinus Rhythm, Bradycardia, Tachycardia, or unclassified file. The Triangle System hardware transmits the ECG signal from the electrodes to the Kardia phone app on the MCP to be analyzed and presented to the user. All ECGs are synced with the user's account. The device is available for Over-the-Counter (OTC) purchase and for purchase with clinician prescription.
### INDICATIONS FOR USE [807.92(a)(5)]
The Triangle System is intended to record, store and transfer one- and two-channel electrocardiogram (ECG) rhythms. In single channel mode, the Triangle System can record Lead-I. In two channel mode, the Triangle System can record Lead-II simultaneously and derive Lead-III and unipolar limb leads aVR, aVF and aVL. The Triangle System also displays ECG rhythms and output of ECG analysis from AliveCor's KardiaAI platform including detecting the presence of normal sinus rhythm, atrial fibrillation, bradycardia, tachycardia, and others. The Triangle System is intended for use by healthcare professionals, patients with known or suspected heart conditions and health conscious individuals. The device has not been tested and is not intended for pediatric use.
{5}------------------------------------------------
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE[807.92(a)(6)]
| Feature | Triangle System<br>(subject device) | AliveCor Heart Monitor<br>K142672 |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Product Code | DXH, DPS QDA | No difference |
| Mechanism of Action | User completes circuit with skin contact<br>and hardware transmits ECG signal to<br>MCP to convert and display ECG<br>waveform | No difference |
| Where used (intended use) | Mobile/active users at rest (ambulatory) | No difference |
| Anatomical sites | Left hand fingers to right hand fingers<br>Left hand fingers to right hand fingers<br>and to left leg/knee | Left hand fingers to right hand fingers<br>Right hand fingers to left leg/knee |
| Available Algorithms | Atrial Fibrillation<br>Noise Algorithm<br>Normal Sinus Rhythm<br>Tachycardia<br>Bradycardia | Atrial Fibrillation |
| Data Acquisition:<br>Frequency Response | 0.5Hz - 40Hz | No difference |
| No. of ECG electrodes | Three (3) dry electrodes | Two (2) dry electrodes |
| No. of ECG channels | Single-channel and two-channel | Single channel only |
| Resolution | 16 bits | No difference |
| Sample Rate | 300 Samples/Second | No difference |
| Power Supply:<br>Battery<br>Battery Life (typical) | 1 Lithium Manganese Dioxide Coin<br>Cells<br>100 hours operational | No difference |
| Leads | Lead-I, or Lead-I, Lead-II, Lead-III,<br>aVL, aVR and aVF. | Lead-I or Lead-II |
| Hardware interface | Three-electrode sensor | Two-electrode sensor |
| Software interface | Apple iOS-based or Google<br>Android-based software | No difference |
| Communications | Bluetooth Low Energy | Ultrasonic Acoustics acquired by phone |
| Physical Specs:<br>Dimensions | 90 x 30 x 7mm | 82 x 32 x 3.5mm |
| Weight | 24 grams | 15 grams |
### SUBSTANTIAL EQUIVALENCE
The Triangle System subject device has the same intended use and similar technological characteristics as the AliveCor Heart Monitor predicate device. The differences in technological characteristics have been analyzed and addressed through performance testing. The testing results showed that differences between the subject and predicate device do not raise different questions of safety or effectiveness.
{6}------------------------------------------------
# PERFORMANCE DATA [807.92(b)]
All necessary non-clinical and clinical performance testing was conducted on the Triangle System to support a determination of substantial equivalence to the predicate device.
# [807.92(b)(1)] Nonclinical Testing Summary:
All necessary performance testing was conducted on the Triangle System with passing results supporting the determination of substantial equivalence to the predicate device. This testing included the following:
- Verification of the device's specification -
- Testing to software level of concern requirements -
- Validation of KardiaAI integration -
- Validation of BLE ECG transmission -
- -Human factors evaluation
- Evaluation to the following standards: -
- ISO 10993-1:2009 (CORR 2010), Biological Evaluation of Medical O Devices - Part 1: Evaluation and Testing Within a Risk Management Process
- IEC 60601-1:2012, Medical Electrical Equipment Part 1: General O Requirements For Basic Safety And Essential Performance,
- IEC 60601-1-2:2014, Medical Electrical Equipment Part 1-2: General о Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests,
- IEC 60601-1-11:2015, Medical electrical equipment -- Part 1-11: General O requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment, and
- IEC 60601-2-47:2012, Medical Electrical Equipment -- Part 2-47: O Particular Requirements For The Basic Safety And Essential Performance Of Ambulatory Electrocardiographic Systems.
# [807.92(b)(2)] Clinical Study Summary:
The performance of the Triangle System for recording a 6-lead ECG was validated in a clinical study. Overall, 44 subjects participated in the study, that included nearly equal numbers of healthy volunteers and arrhythmia patients. ECG recordings were taken simultaneously with both the Triangle device and gold standard 12-lead ECG device for comparison. Qualitative and quantitative analyses of equivalence were performed on the 44 pairs of ECGs.
For qualitative assessment, two board-certified electrophysiologists compared 6-lead ECG rhythm strips acquired from the Triangle device and the corresponding leads from the reference standard 12-lead ECG device for diagnostic equivalence. All paired recordings (100%, n=44 subjects), were deemed equivalent for assessing cardiac arrhythmias by both electrophysiologists. The results of the assessment determined that
{7}------------------------------------------------
ALIVECOR, INC.
the subject device records a 6-lead ECG that is qualitatively equivalent to the recordings of corresponding leads from a commercial gold standard 12-lead ECG device.
For quantitative equivalence, median beat cross correlation for Lead I and II and RMS error for all 6 limb leads were computed between the paired ECGs for each subject. This analysis was conducted on the unfiltered ECG output as well as the enhanced filtered (EF) ECG output. Triangle ECGs had a minimum correlation of 0.96 and a maximum RMS error of 47 uV as compared to the corresponding lead of the 12-lead ECG. The results of the quantitative analysis of the ECG recordings further confirmed that the Triangle device ECG has equivalent output to that of the gold standard 12-lead ECG device.
During this clinical study, no adverse events were observed.
### SUMMARY
The Triangle System has the same intended use and similar technological characteristics as the AliveCor Heart Monitor predicate device. The differences in technological characteristics have been analyzed and addressed through performance testing which included non-clinical and clinical testing. The testing results show that differences between the subject and predicate device do not raise different questions of safety or effectiveness. The Triangle System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.