K211572 · Aesculap, Inc. · HCI · Dec 13, 2021 · Neurology
Device Facts
Record ID
K211572
Device Name
Aesculap Slim Clip Applier
Applicant
Aesculap, Inc.
Product Code
HCI · Neurology
Decision Date
Dec 13, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4175
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The slim clip applier is used to open, close and apply permanent/temporary Aesculap YASARGIL titanium aneurysm clips.
Device Story
The Aesculap Slim Clip Applier is a manual surgical instrument used by neurosurgeons to manipulate and apply titanium aneurysm clips during intracranial procedures. The device consists of stainless steel forceps available in various lengths and jaw angulations, featuring optional latches and identification plugs to ensure compatibility with specific clip types. The current iteration introduces a surface coating to the previously cleared design. The surgeon operates the device manually to open, position, and close the clip on an aneurysm; the device provides the mechanical force necessary to secure the clip, thereby facilitating the occlusion of the aneurysm. The device is provided non-sterile and must be sterilized by the user prior to clinical application.
Clinical Evidence
Bench testing only. Performance testing verified that the minimum opening width of aneurysm clips meets requirements. Biocompatibility testing per ISO 10993-5 and ISO 10993-11 confirmed the device is non-cytotoxic, non-sensitizing, non-pyrogenic, non-irritating, non-toxic, and non-hemolytic.
Technological Characteristics
Stainless steel forceps; various lengths (70mm, 90mm, 110mm) and jaw angulations; optional latch mechanism; identification plugs (POM-C or Propylux); laser marking for clip designation; surface coating added. Manual operation. Provided non-sterile; intended for sterilization prior to use.
Indications for Use
Indicated for opening, closing, and applying permanent or temporary Aesculap YASARGIL titanium aneurysm clips in patients requiring neurosurgical aneurysm repair.
Regulatory Classification
Identification
An aneurysm clip applier is a device used by the surgeon for holding and applying intracranial aneurysm clips.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
December 13, 2021
Aesculap, Inc. Paul Amudala Senior Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K211572
Trade/Device Name: Aesculap Slim Clip Applier Regulation Number: 21 CFR 882.4175 Regulation Name: Aneurysm Clip Applier Regulatory Class: Class II Product Code: HCI Dated: November 9, 2021 Received: November 10, 2021
Dear Paul Amudala:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan, Ph.D., Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211572
Device Name Aesculap Slim Clip Applier
#### Indications for Use (Describe)
The slim clip applier is used to open, close and apply permanent/temporary Aesculap Y ASARGIL titanium aneurysm clips.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="text-align: left;"> <svg height="15" width="15"> <path d="M2 2 L2 13 L13 13 L13 2 Z M3 3 L12 3 L12 12 L3 12 Z" fill="white" stroke="black"></path> <path d="M4 4 L11 11" stroke="black"></path> <path d="M11 4 L4 11" stroke="black"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="text-align: left;"> <svg height="15" width="15"> <path d="M2 2 L2 13 L13 13 L13 2 Z M3 3 L12 3 L12 12 L3 12 Z" fill="white" stroke="black"></path> </svg> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
[as required by 21 CFR §807.92(c)]
# K211572
| DATE PREPARED: | 13 December 2021 |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| SUBMITTER: | Aesculap®, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>+1 800-258-1946 |
| CONTACT: | Paul Amudala<br>Sr. Regulatory Affairs Specialist<br>610-984-9303 (phone)<br>610-417-0839 (mobile)<br>610-791-6882 (fax) |
| TRADE NAME: | Aesculap Slim Clip Applier |
| COMMON NAME: | Aneurysm Clip Applier |
| DEVICE CLASSIFICATION: | Class II |
| REGULATION NUMBER: | 21 CFR 882.4175 |
| PRODUCT CODE: | HCI |
| REVIEW PANEL: | Neurology |
| PRIMARY PREDICATE<br>DEVICE: | Aesculap Slim Clip Applier (K180914; April 17, 2018) |
| REFERENCE DEVICES: | Aesculap Slim Clip Applier (K173271; November 10, 2017) |
| | Yasargil, Caspar, Vario Clip Appliers (K940970; January<br>24, 1995) |
## DEVICE DESCRIPTION
The Aesculap Slim Clip Appliers are manufactured from stainless steel and are available in various lengths, jaw angulations, with or without latches. This submission intends to add a coating to the previously cleared clip appliers. Each clip applier is individually laser marked with the type of clip designated for use. In addition, at least one identification plug is located within one of the handles to aid in identifying the aneurysm clip designation.
## INDICATIONS FOR USE
The slim clip applier is used to open, close and apply permanent/temporary Aesculap YASARGIL titanium aneurysm clips.
{4}------------------------------------------------
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The Aesculap Slim Clip Applier forceps are substantially equivalent to the predicate Aesculap Slim Clip Applier forceps. The intended use, fundamental scientific principles, and base materials of the clip appliers remain unchanged since last clearance. This submission intends to add a coating to the previously cleared clip appliers.
| | Subject Device<br>Aesculap Slim Clip<br>Applier | Predicate<br>Aesculap Slim Clip<br>Applier | Reference Device<br>Aesculap Slim Clip<br>Applier | Reference Device<br>Yasargil, Caspar,<br>Vario Clip Appliers |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| | K211572 | K180914 | K173271 | K940970 (Vario<br>functionality reference<br>only) |
| Indications for<br>Use | The slim clip applier<br>is used to open, close<br>and apply permanent/<br>temporary Aesculap<br>YASARGIL titanium<br>aneurysm clips. | The Aesculap Slim<br>Clip Applier forceps<br>are used for<br>opening, closing,<br>and applying<br>Aesculap<br>YASARGIL<br>aneurysm clips. | The Aesculap Slim<br>Clip Applier forceps<br>are used for<br>opening, closing,<br>and applying<br>Aesculap<br>YASARGIL<br>aneurysm clips. | The Aesculap Titanium<br>Aneurysm clip appliers<br>are used for holding and<br>applying intracranial<br>titanium aneurysm clips. |
| Clip Applier: | | | | |
| Base Material | Stainless Steel | Same | Same | Titanium |
| Type | Standard & Mini &<br>Vario | Standard & Mini | Standard & Mini | Standard & Mini &<br>Vario |
| Length | 70mm, 90mm, 110mm | 50mm to 110mm,<br>120mm and 130mm | 50mm to 110mm | 90mm |
| Jaw Angulation | Vario, straight, up &<br>down, left & right | Straight, up & down, left<br>& right | Straight, up & down, left<br>& right | Vario & various |
| Optional Latch | Yes | Latch or no-latch | No Latch | Latch or no-latch |
| Designated<br>clips | YASARGIL Titanium<br>Clips only<br>(Standard/Mini) | YASARGIL Titanium<br>or Phynox clips<br>(Standard/Mini) | YASARGIL Titanium<br>or Phynox clips<br>(Standard/Mini) | YASARGIL Titanium<br>Clips only<br>(Standard/Mini) |
| Non-Sterile | Yes | Same | Same | Same |
| Clip<br>Designation<br>Feature | Yes (POM-C or<br>Propylux Plug & Laser<br>marking on spring) | Same | Same | POM-C |
| Coating | Yes | No coating | No coating | No coating |
# TESTING
There is no change to the base design, functionality or intended use of the clip appliers with added coating. Bench testing results demonstrate that the Aesculap Slim Clip Applier is substantially equivalent to the Aesculap Slim Clip Appliers currently on the market. The bench testing of clip appliers demonstrates that the Aesculap Slim Clip Applier forceps perform as intended and are as safe, effective, and perform as well as the predicate devices.
{5}------------------------------------------------
| Performance Test Completed | Test purpose | Results | Conclusion |
|---------------------------------------|------------------------------------------------------------------------------------|---------|-----------------------------------------------------------------------------|
| Verification test for clip<br>opening | To verify the minimum opening<br>width of aneurysm clips meets the<br>requirement. | Pass | The Aesculap Slim<br>Clip Appliers meet<br>the test acceptance<br>criteria. |
STERILIZATION: The Aesculap Slim Clip Applier forceps will continue to be provided nonsterile similar to the predicate devices. They are intended to be sterilized prior to use.
BIOCOMPATIBILITY: Testing confirmed that the Aesculap Slim Clip Appliers with added coating are biocompatible.
| Test | Summary | Conclusion |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Cytotoxicity | Test articles were extracted for 24 hours per ISO 10993-5. All test<br>articles exhibited a reactivity grade of 0. Positive and negative<br>controls behaved as anticipated. | Non-cytotoxic |
| Sensitization | The test article extracts showed no evidence of causing delayed<br>dermal contact sensitization in the guinea pig. The test article was<br>not considered a sensitizer in the guinea pig maximization test. | Non-sensitive |
| Pyrogenicity | The purpose of this assessment is to determine if a test article<br>extract induces a pyrogenic response following intravenous<br>injection in rabbits (temperature rise below 0.5 °C<br>indicates the absence of pyrogens). Based on the assessment<br>per ISO 10993-11, the materials are non-pyrogenic. | Non-pyrogenic |
| Irritation | All animals appeared normal throughout the study. The test article<br>met the requirements of the test since the difference between each<br>test article extract overall mean score and corresponding control<br>extract overall mean score was 0.0 for both the SC and SO test<br>article extracts. | Normal |
| Systemic Toxicity | There was no mortality or evidence of systemic toxicity from the<br>extracts injected into mice. Each test article extract met the<br>requirements of the study. | Non-Toxic |
| Hemolysis | The test article extract was tested in vitro. The hemolytic index for<br>the test article extract was 0.0% and the test article extract was<br>considered non-hemolytic. | Non-Hemolytic |
# CONCLUSION
Aesculap Slim Clip Applier forceps presented in this submission are substantially equivalent in design, intended use, and performance to the predicate and reference devices. The added coating does not raise new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.