The Aesculap Slim Clip Applier forceps are used for opening, closing, and applying Aesculap YASARGIL aneurysm clips.
Device Story
Manual surgical instrument; stainless steel forceps; used by neurosurgeons in OR. Function: opening, closing, and applying YASARGIL aneurysm clips. Input: manual force applied by surgeon to handles. Transformation: mechanical leverage transfers force to jaws to manipulate clip. Output: controlled placement/release of aneurysm clip. Features: various lengths (50-130mm), jaw angulations (straight, 5-45° up/down, 5-30° left/right), and optional latches. Identification: laser marking and identification plug in handle to match applier to specific clip type. Benefit: enables precise occlusion of cerebral aneurysms.
Clinical Evidence
Bench testing only. No clinical data. Testing included functional benchmarks (interface, opening width, clip release, latch functionality), predicate comparison, and usability testing (identification, legibility, visualization, handling). All tests passed.
Technological Characteristics
Materials: Stainless steel (body and jaw). Form factor: Manual forceps, 50-130mm lengths, various jaw angulations (straight, up/down, left/right). Features: Optional latch, laser-marked clip designation, identification plug. Sterilization: Non-sterile, intended for user sterilization. Connectivity: None.
Indications for Use
Indicated for use in the opening, closing, and application of Aesculap YASARGIL aneurysm clips during neurosurgical procedures.
Regulatory Classification
Identification
An aneurysm clip applier is a device used by the surgeon for holding and applying intracranial aneurysm clips.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
April 17, 2018
Aesculap, Inc. Paul Amudala Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K180914
Trade/Device Name: Aesculap Slim Clip Applier Regulation Number: 21 CFR 882.4175 Regulation Name: Aneurysm Clip Applier Regulatory Class: Class II Product Code: HCI Dated: April 6, 2018 Received: April 9, 2018
Dear Paul Amudala:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/MedicalDevices/ DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image contains the text "Carlos L. Pena - S". The text is in a simple, sans-serif font and is left-aligned. The letters are dark, contrasting with the white background. There is a faint logo of the FDA in the background.
Carlos L. Peña. PhD. MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180914
Device Name Aesculap Slim Clip Applier
#### Indications for Use (Describe)
The Aesculap Slip Clip Applier forceps are used for opening, closing, and applying Aesculap Y ASARGIL aneurysm clips.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|--|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CER 801 Subpart D) | |
| Over-The-Counter Use (21 CFR 801 Subpart C) | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) SUMMARY - K180914
# Aesculap Slim Clip Applier April 06, 2018
| COMPANY: | Aesculap®, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 2916714 |
|--------------------|--------------------------------------------------------------------------------------------------------------------|
| CONTACT: | Paul Amudala<br>610-984-9303 (phone)<br>610-791-6882 (fax)<br>paul.amudala@aesculapimplants.com |
| TRADE NAME: | Aesculap Slim Clip Applier |
| COMMON NAME: | Aneurysm Clip Applier |
| REGULATION NUMBER: | 21CFR882.4175 – Applier, Aneurysm Clip |
| PRODUCT CODE: | HCI |
| REVIEW PANEL: | Neurology |
### SUBSTANTIAL EQUIVALENCE
Aesculap. Inc. believes that the Aesculap Slim Clip Applier forceps are substantially equivalent to Clip Appliers cleared in Primary Predicate K173271 and Reference Predicate K131500 (including K043041, K003519, K002871, K984109, K983758, K970050, K922272, K913765, K833652, K833651, K833650, and K772200).
### DEVICE DESCRIPTION
The Aesculap Slim Clip Appliers are manufactured from stainless steel (body & jaw). The previously cleared appliers are available in lengths ranging from 50 mm to 110 mm in 10 mm increments. In addition to the straight jaw configuration (0%), the Clip Appliers are available with jaws angled up & down (5° to 15°), left & right (5° to 15°), in increments of 5° in all lengths. This submission intends to add additional lengths of 120 mm and 130 mm in these configurations. This submission intends to add additional jaw angulations of 20° to 45° up and down and 20° to 30° left and right in 5° increments and lengths ranging from 50-130 mm in 10 mm increments. In addition to these configurations, clip appliers with latches will be available in lengths ranging from 70 mm to 110 mm increments with jaw angulations ranging from straight, 5° to 45° up and down and 5° to 30° left and right in 5° increments.
{4}------------------------------------------------
Each Clip Applier is individually laser marked with the type of clip designated for use. In addition, at least one identification plug is located within one of the handles to aid in identifying the aneurysm clip designation.
## INDICATIONS FOR USE
The Aesculap Slim Clip Applier forceps are used for opening, closing, and applying Aesculap YASARGIL aneurysm clips.
## TECHNOLOGICAL CHARACTERISTICS (Compared to the Primary Predicate)
The Aesculap Slim Clip Applier forceps are substantially equivalent to the predicate Aesculap Slim Clip Applier forceps and YASARGIL Aneurysm Clip Applier forceps. The intended use, fundamental scientific principles, and base materials of the Clip Appliers (body and jaw) remain unchanged since last clearance. The additional Aesculap Slim Clip Applier forceps are offered with optional latches and additional sizes and angulations but function the same as the Primary and Reference predicates.
| | Principle Device | Primary Predicate<br>K173271 | Reference Predicate<br>K131500<br>(including K043041, K003519,<br>K002871, K984109, K983758,<br>K970050, K940970, K922272,<br>K913765, K833652, K833651,<br>K833650 and K772200)<br>YASARGIL Aneurysm Clips<br>and Clip Appliers (Clip<br>Appliers only) |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications | The Aesculap Slim<br>Clip Applier forceps<br>are used for opening,<br>closing, and applying<br>Aesculap YASARGIL<br>aneurysm clips. | The Aesculap Slim Clip<br>Applier forceps are used<br>for opening, closing, and<br>applying Aesculap<br>YASARGIL aneurysm<br>clips. | The YASARGIL Aneurysm<br>Clips are intended for<br>occlusion of cerebral<br>aneurysms in either a<br>temporary or permanent<br>manner. They are applied with<br>Aesculap Clip Appliers, which<br>contain titanium alloy or<br>phynox jaws. |
| Clip Applier: | | | |
| Body Material<br>(Jaw & Body) | Stainless Steel | Stainless Steel | Titanium Alloy (Ti6AI6V) or<br>Stainless Steel |
| Length | 50mm to 110mm,<br>120mm and 130mm | 50mm to 110mm | 50mm, 90mm, 70mm, and<br>110mm |
| Jaw Angulation | Straight, up & down,<br>left & right | Straight, up & down, left<br>& right | Straight, angled, bayonet |
| Optional Latch | Yes | No | Yes |
| Non-Sterile | Yes | Same | Same |
| Clip Designation<br>Feature | Yes | Same | Same |
{5}------------------------------------------------
#### TESTING
All required testing was completed for the Aesculap Slim Clip Applier forceps. Bench testing results demonstrate that the Aesculap Slim Clip Applier forceps are substantially equivalent when compared to the Aesculap Clip Appliers currently on the market. Bench testing was performed to demonstrate that the Aesculap Slim Clip Applier forceps perform as intended and are safe, as effective, and perform as well as the predicate devices.
{6}------------------------------------------------
| Performance Test<br>completed | Methodology | Results | Conclusions |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Benchmark Functional<br>Test | (1) Meet required interface<br>and opening width of the<br>aneurysm clip.<br>(2) Successful release of<br>designated aneurysm clip in<br>the craniotomy.<br>(3) Latch functionality | Pass | The devices interfaced successfully<br>with the aneurysm clips while<br>providing the correct opening width.<br>In addition the devices successfully<br>released the aneurysm clips within<br>the craniotomy. |
| Predicate/Subject<br>Device Comparison<br>Test | Comparison between<br>predicate and subject device<br>ensuring:<br>(1) Required interface and<br>opening width of designated<br>aneurysm clip.<br>(2) Successful performance of<br>the clip applier with the<br>aneurysm clip.<br>(3) Latch functionality | Pass | The performance testing<br>demonstrated that the subject<br>devices are substantially equivalent<br>to the predicate devices. |
| Usability Test | (1) Ensuring identification of<br>designated clip type for the<br>specific clip applier.<br>(2) Legibility of additional<br>Clip Applier identification<br>features, i.e. company name,<br>article number, and UDI<br>information.<br>(3) Ensuring visualization of<br>surgical site.<br>(4) Ability to grasp, hold, and<br>apply YASARGIL Aneurysm<br>Clip.<br>(5) Latch usability test | Pass | The usability test demonstrated that<br>the subject devices were properly<br>identified and met the usability<br>requirements. |
STERILIZATION: The Aesculap Slim Clip Applier forceps will continue to be provided nonsterile similar to the predicate devices. They are intended to be sterilized prior to use.
BIOCOMPATIBILITY: There is no change to the patient contacting material of the Aesculap Slim Clip Applier forceps since they are manufactured from the same Stainless Steel as the predicate devices.
# CONCLUSION
Aesculap believes that the Aesculap Slim Clip Applier forceps presented in this submission are substantially equivalent in design, materials, intended use, and perform as safe and effective as Aesculap's currently marketed devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.