Small Foam Kit, Medium Foam Kit, Large Foam Kit, Extra Large Foam Kit
K211571 · Zhejiang Longterm Medical Technology Co., Ltd. · OMP · May 31, 2022 · General, Plastic Surgery
Device Facts
Record ID
K211571
Device Name
Small Foam Kit, Medium Foam Kit, Large Foam Kit, Extra Large Foam Kit
Applicant
Zhejiang Longterm Medical Technology Co., Ltd.
Product Code
OMP · General, Plastic Surgery
Decision Date
May 31, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Longterm NPWT Foam Dressing Kit is intended to be used along with VCare 1000 -300S pump for wound management via the application of negative pressure to the wound, in order for the removal of fluids, including wound exudates, irrigation fluids, body fluids and infectious materials. The system is intended for patients with chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts. Longterm NPWT foam dressing kit is for use in professional healthcare facilities only.
Device Story
Accessory kit for VCare 1000-300S negative pressure wound therapy (NPWT) pump; used in professional healthcare facilities. Components include hydrophobic reticulated polyurethane foam dressing, polyurethane film drape, and suction bell. Foam packs wound bed; drape creates sealed environment; suction bell acts as conduit for fluid transfer to pump canister. System applies negative pressure to remove wound exudates, irrigation fluids, and infectious materials. Clinician applies dressing and monitors wound; output is fluid removal and wound environment management. Benefits include promotion of wound healing via vacuum-assisted drainage.
Clinical Evidence
Bench testing only. Evaluated foam dimensions, fluid removal rates, pressure distribution, and pressure differentials under continuous/intermittent modes over 72 hours. Mechanical properties tested per ASTM D3574. Biocompatibility testing performed per ISO 10993 standards (cytotoxicity, irritation, sensitization, pyrogenicity, acute/subacute/subchronic toxicity, muscle implantation). Porcine wound healing study conducted. No clinical testing required.
Technological Characteristics
Materials: Hydrophobic reticulated polyurethane ester foam (with 0.1% carbon black), polyurethane film drape with acrylic adhesive, PVC suction bell, polypropylene pinch clamp. Dimensions: Foam (4 sizes), Drape (2 sizes). Energy: Passive accessory to powered suction pump. Sterilization: Ethylene Oxide or Gamma Irradiation (SAL 10^-6). Mechanical testing per ASTM D3574.
Indications for Use
Indicated for patients with chronic, acute, traumatic, subacute, and dehisced wounds, partial-thickness burns, ulcers (diabetic or pressure), flaps, and grafts requiring negative pressure wound therapy for fluid removal.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
KCI USA Inc-V.A.C.® ATS, mini VAC, VAC Freedom (K032310)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 31, 2022
Zhejiang Longterm Medical Technology Co., LTD Claudia Zsang Director of Regulatory Affairs No. 493 North Huancheng Road Mogan Mountain National High-Tech District Deging, Zhejiang 313200 China
Re: K211571
Trade/Device Name: Small Foam Kit. Large Foam Kit. Extra Large Foam Kit. Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: November 29, 2021 Received: December 7, 2021
Dear Claudia Zsang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211571
Device Name Longterm NPWT Foam Dressing Kit
### Indications for Use (Describe)
Longterm NPWT Foam Dressing Kit is intended to be used along with VCare 1000 -300S pump for wound management via the application of negative pressure to the wound, in order for the removal of fluids, including wound exudates, irrigation fluids, body fluids and infectious materials. The system is intended for patients with chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts. Longterm NPWT foam dressing kit is for use in professional healthcare facilities only.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K211571
## 510(k) SUMMARY
### 1 The 510(k) summary information is being submitted in accordance with the requirements of 21 CFR 807. 92(c).
| Date Prepared: | May 03, 2022 |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Zhejiang Longterm Medical Technology Co., LTD<br>No. 493 North Huancheng Road, Mogan Mountain National High-<br>Tech District, Deqing Zhejiang, CHINA 313200 |
| Official Correspondent: | Claudia Zsang<br>Claudia.zsang@gmail.com |
| Phone Number: | 1-416-276-9555 |
| Device Name: | Longterm NPWT Foam Dressing Kit |
| Common Name: | NPWT System Accessories |
| FDA Panel: | General & Plastic surgery |
| Product Code: | OMP |
| Class: | II |
| Predicate Devices: | VCare 1000-300S Pump, VCare 1000-300S System, Perme-<br>foam Dressing (K162159) (Primary)<br>KCI USA Inc-V.A.C.® ATS, mini VAC, VAC Freedom<br>(K032310) |
#### 2 Device Description:
Longterm NPWT Foam dressing kit is an accessory to VCare 1000-300S pump, manufactured by VR Medical Technology Co., Ltd. The dressing kit is composed of black foam dressing, polyurethane (PU) film drape and suction bell.
The foam dressing is used to pack the wound bed, the PU film drape is used to cover the packed wound bed and peri-wound area to create and maintain a sealed environment, the suction bell is served as a conduit between the wound bed and the negative pressure pump to transfer the wound fluid to the canister attached to the pump.
Longterm NPWT Foam dressing kit consists of hydrophobic, reticulated Polyurethane foam, polyurethane film drape coated with acrylic adhesive, suction bell composed of a polyvinyl chloride (PVC) drainage tubing with a polypropylene pinch clamp, a bell shape dome made of polyvinyl chloride, a transparent polyurethane film coated with acrylic
{4}------------------------------------------------
## K211571
adhesive. Each component is individually packaged and sterilized. The foam dressing should be changed every 2-3 days depending on the wound condition or according to the local protocol.
The suction bell is comprised of a polyvinyl chloride drainage tubing with a polypropylene pinch clamp, a bell shape dome made of polyvinyl chloride, a transparent poly urethane film coated with acrylic adhesive.
Foam dressing is available in 4 sizes; small (10cmx7.5cmx2.5cm), medium (18cmx12.5cmx2.5cm), large (26cmx15cmx2.5cm) and extra large (45cmx30cmx2.5cm). PU film drape is available in 2 sizes; 30cmx30cm and 45cmx30cm.
#### 3 Indications for Use:
Longterm NPWT Foam Dressing kit is intended to be used along with VCare 1000-300S pump for wound management via the application of negative pressure to the wound, in order for the removal of fluids, including wound exudates, irrigation fluids, body fluids and infectious materials. The system is intended for patients with chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts.
Longterm NPWT foam dressing kit is for use in professional healthcare facilities only
{5}------------------------------------------------
#### 4 Substantial Equivalence
| Parameter | Subject Device | Predicate Device #1 (Primary) | Predicate Device #2 |
|-----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)# | K211571 | K162159 | K032310 |
| Device name | Longterm NPWT Foam Dressing Kit | VCare 1000-300S Pump, VCare 1000-300S<br>System, Perme-foam Dressing (K162159) | KCI-V.A.C. ATS, mini VAC, VAC Freedom |
| Note | Accessory to VCare pump (510 (k) K162159) | - | Only foam dressing is used as the predicate |
| Classification<br>Regulation | 878.4780 | Same | Same |
| Product Code | OMP | Same | Same |
| Skin Contact<br>Materials | - Polyurethane ester foam with 0.1% carbon<br>black colorant<br>- Polyurethane film with acrylic adhesive | - Polyurethane ester foam;<br>- Polyurethane drape | - Black, reticulated, polyurethane foam |
| Non-Skin<br>Contact<br>Materials | Suction Bell comprised of a polypropylene<br>drainage tubing with a polypropylene pinch<br>clamp, a bell shape dome made of polyvinyl<br>chloride, a transparent poly urethane film<br>coated with acrylic adhesive. | Suction Bell with connecting (drainage) tube<br>and clamp | Non-specific |
| Mechanism of<br>Action | The foam dressing is used to pack the wound<br>bed, the PU film drape is used to cover the<br>packed wound bed and peri-wound area to<br>create and maintain a sealed environment, the<br>suction bell is served as a conduit between the<br>wound bed and the negative pressure pump to<br>transfer the wound fluid to the canister<br>attached to the pump. | Same | Same |
| Intended Use | The dressing kit is intended to be used along<br>with VCare 1000-300S pump for wound<br>management via the application of negative<br>pressure to the wound, in order for the removal<br>of fluids, including wound exudates, irrigation<br>fluids, body fluids and infectious materials.<br>The system is intended for patients with<br>chronic, acute, traumatic, subacute and<br>dehisced wounds, partial-thickness burns,<br>ulcers (such as diabetic or pressure), flaps and<br>grafts. | The VR Medical VCare 1000-300S Negative<br>Pressure Wound Therapy System is an<br>integrated wound management system,<br>indicated for wound management via the<br>application of negative pressure to the wound,<br>in order for the removal of fluids, including<br>wound exudates, irrigation fluids, body fluids<br>and infectious materials. The system is intended<br>for patients with chronic, acute, traumatic,<br>subacute and dehisced wounds, partial-thickness | The V.A.C. family of devices are feedback-<br>controlled negative pressure devices used<br>to help promote wound healing, through means<br>including vacuum assisted drainage and<br>removal of infectious material or other fluids, under<br>the influence of continuous and or<br>intermittent suction pressures, particularly for<br>patients with chronic, acute, traumatic,<br>subacute and dehisced wounds, partial-thickness<br>burns, ulcers (such as diabetic or pressure), flaps and |
| Parameter | Subject Device | Predicate Device #1 (Primary) | Predicate Device #2 |
| 510(k)# | K211571 | K162159 | K032310 |
| | | burns, ulcers (such as diabetic or pressure), flaps<br>and grafts. | grafts. Feedback control is achieved by measuring the<br>level of negative pressure at the wound site. |
| Use environment | The foam dressing kit is for use in professional<br>healthcare facilities only. | The device is for use in professional<br>healthcare facilities only. | Non-specific |
| Biocompatibility<br>(breached skin,<br>prolonged,<br>>24hr to 30days) | Non cytotoxic, non-irritant, non-sensitizing,<br>non-toxic, none- pyrogenic | Non cytotoxic, non-irritant, non-sensitizing,<br>non-toxic, none- pyrogenic | Biocompatible |
| Sterilization | Each kit components individually sterilized<br>by Ethylene Oxide or Gamma Irradiation | Gamma Irradiation | Gamma Irradiation |
| Sterility<br>Assurance Level | SAL of 10-6 | SAL of 10-6 | SAL of 10-6 |
{6}------------------------------------------------
{7}------------------------------------------------
#### ਟ Non-clinical Testing Summary:
The following tests were performed to support the substantial equivalence of the subject
## device: Biocompatibility Testing:
- Cytotoxicity - ISO 10993-5:2009
- Irritation ISO 10993-10:2010 ●
- Sensitization ISO 10093-10:2010 ●
- . Pyrogenicity -ISO 10993-11:2006/USP39-NF34 <151>
- Acute toxicity ISO 10993-11:2017 ●
- Subacute toxicity ISO 10993-11:2017/ ISO 10993-6:2016 ●
- Subchronic toxicity ISO 10993-11: 2017 / ISO 10993-2016 ●
- Muscle implantation ISO 10993-6:2016 .
## Bench testing:
Longterm NPWT Foam Dressing Kit was evaluated under of bench tests to ensure the dressing kit can be used along with VCare 1000-300S pump regarding performance and functionality requirements.
- . The foam dressing dimension test under various negative pressure after 72 hours at both continuous and intermittent modes;
- The fluid removal rate under various negative pressure at 72 hours at both continuous and intermittent modes;
- The pressure distribution under the foam dressing in 72 hours at both continuous and ● intermittent modes;
- The pressure difference under the foam dressing in 72 hours at both continuous and ● intermittent modes.
- Mechanical properties (tensile strength and elongation, ASTM D3574)
## Animal testing:
Porcine wound healing study
#### Clinical Testing: 6
No clinical testing was required to support substantial equivalence.
#### 7 Conclusion:
The non-clinical tests demonstrate that the subject device is as safe, as effective and performs as well as the legally marked predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.