The VR Medical VCare 1000-300S Negative Pressure Wound Therapy System is an integrated wound management system, indicated for wound management via the application of negative pressure to the wound, in order for the removal of fluids, including wound exudates, irrigation fluids, body fluids and infectious materials. The system is intended for patients with chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts.
Device Story
VCare 1000-300S is an AC-powered portable suction pump with battery backup; used with Perme-Foam Dressing kit (polyurethane foam, suction bell, drainage tube, clamp, polyurethane drape). Device applies localized negative pressure to wounds to remove exudates, irrigation fluids, and infectious materials. Operated by clinicians in clinical settings. System features continuous and intermittent (variable/pulse) pressure modes (-20 to -200 mmHg). Healthcare providers monitor system performance via alarms and pressure settings. Therapy promotes wound healing by managing fluid and infectious debris. Dressing is single-use, gamma-sterilized, and intended for up to 72 hours of continuous use.
Clinical Evidence
Bench testing only. Verification included negative pressure delivery in continuous and intermittent modes, system performance with foam dressings, alarm functionality, and battery performance. Biocompatibility testing per ISO 10993 (cytotoxicity, sensitization, reactivity, systemic toxicity, genotoxicity, implantation, hemolysis, pyrogenicity) confirmed safety for prolonged contact.
Indicated for patients with chronic, acute, traumatic, subacute, and dehisced wounds, partial-thickness burns, ulcers (diabetic or pressure), flaps, and grafts requiring negative pressure wound therapy for fluid removal.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
Predicate Devices
ExtriCare 2400 NPWT System with ExtriCare 2400 Pump and ExtriCare NPWT Foam Dressing Kit (K140634)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three heads, representing the department's focus on health, human services, and the well-being of the population. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 28, 2017
VR Medical Technology Company, Ltd. Lee Pan CEO 90 Gaoxin Rd. Kunshan, CN 215325 Jiangsu
Re: K162159
Trade/Device Name: VCare 1000-300S Pump, VCare 1000-300S System, Perme-foam Dressing Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: March 27, 2017 Received: March 28, 2017
Dear Lee Pan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162159
Device Name
VCare 1000-300S Pump, VCare 1000-300S System, Perme-foam Dressing
Indications for Use (Describe)
The VR Medical VCare 1000-300S Negative Pressure Wound Therapy System is an integrated wound management system, indicated for wound management via the application of negative pressure to the wound, in order for the removal of fluids, including wound exudates, irrigation fluids, body fluids and infectious materials. The system is intended for patients with chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size: 10pt;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> |
|-------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 10pt;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
#### I. Submitter Information
VR Medical Technology Co., Ltd. 90 Gaoxin Rd., Zhouzhuang Kunshan, Jiangsu, CHINA 215325
- Contact Person: Lee Pan, Ph.D. lpan@vrmedical.net +86 512 5720 0500
Date Prepared: April 22, 2017
#### II. Device Information
| Name of Device: | VCare 1000-300S Pump<br>VCare 1000-300S System<br>Perme-foam Dressing |
|-------------------------|--------------------------------------------------------------------------------|
| Common Names: | Negative Pressure Wound Therapy Pump<br>Negative Pressure Wound Therapy System |
| Classification Name(s): | Powered Suction Pump (21 CFR 878.4780) |
| Regulatory Class: | II |
| Product Code: | OMP |
#### III. Predicated Device
- Primary: K140634 Devon Medical Products Inc., ExtriCare 2400 NPWT System with ExtriCare 2400 Pump and ExtriCare NPWT Foam Dressing Kit This predicate has not been subject to a design-related recall
K112458 Medica-Rents Co., Prospera Pro-II This predicate has not been subject to a design-related recall
K132252 Devon Medical Inc., ExtriCare Negative Pressure Wound Therapy 100cc, 250cc, and 300cc canisters This predicate has not been subject to a design-related recall
#### IV. Device Description
The VR Medical VCare 1000-300S System is an AC-Powered portable suction device, with battery backup, that provides localized negative pressure when used with the VR Medical Perme-Foam Dressing to remove fluid, irrigation solutions and infectious materials from the wound. The systems consist of a powered suction pump device with a built-in placement holder for the disposable fluid collection canister, and a foam dressing kit. The kit consists of polyurethane foam, suction bell with connecting (drainage) tube and clamp, and
{4}------------------------------------------------
polyurethane drape. The single-use foam dressing kit is packaged in a PE/PET peel pouch bag, which is gamma sterilized.
The Foam Dressing Kit (VR Medical Perme-Foam Dressing) can be sold alone or as a part of the VCare 1000-300S System in two sizes:
- . Large foam dressing kit
- Medium foam dressing kit
- . The Foam Dressing can be used for up to 72 hours of continuous use.
# V. Intended Use(s)
The VR Medical VCare 1000-300S Negative Pressure Wound Therapy System is an integrated wound management system, indicated for wound management via the application of negative pressure to the wound, in order for the removal of fluids, including wound exudates, irrigation fluids, body fluids and infectious materials. The system is intended for patients with chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts.
# VI. Technological Characteristics
The manufacturer believes that the technological characteristics of the VCare 1000-300S System are substantially similar to those of the predicate devices. The subject and predicate devices are based on the following same technical elements:
- Continuous and Intermitted negative pressure treatment modes -
- -Selectable Pressure Settings from -20 mmHg to -200 mmHg
- -AC and rechargeable battery with 48 hour battery life
# VII. Performance Data
## Non-Clinical Tests (Bench)
Testing has been conducted to verify that the VR Medical VCare 1000-300S System demonstrates substantial equivalence to the predicate devices. The list below summarizes the testing conducted for the VR Medical VCare 1000-300S device.
- . Verification that the device delivers negative pressure wound therapy in continuous and intermittent (variable & pulse) operating modes identical to the predicate devices using simulated exudate
- . Verification of system performance with foam dressing kits and accessories identical to predicate devices
- Verification of system alarms
- Verification of battery performance
# Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the VR Medical VCare 1000-300S System. The system complies with the IEC 60601-1 and IEC 60601-1-11 standards for safety and the IEC 60601-1-2 standard for EMC.
{5}------------------------------------------------
#### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by the FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern, since a failure or latent flaw in the software could directly result in minor injury to the patient or operator.
#### Biocompatibility
The VCare 1000-300S System consists of both the VCare 1000-300S Pump and the VR Medical Perme-Foam Dressing. While the pump has no direct body contact when used as indicated, the Perme-Foam Dressing does have direct body contact. Per ISO-10993 B-prolonged contact (24h-30days) requirements, the Perme-Foam Dressing samples were tested for the following items and all tests were successfully passed.
- In Vitro Cytotoxicity Test per ISO 10993-5:2009
- Skin Sensitization Test per ISO 10993-10:2010
- . Intracutaneous Reactivity Test per ISO 10993-10:2010
- Acute Systemic Toxicity Test per ISO 10993-11:2006 ●
- In Vitro Mammalian Cell Gene Mutation test per ISO 10993-3:2009
- Ames Test per ISO 10993-3:2014
- In Vitro Mammalian Chromosome Aberration Test per ISO 10993-3:2014
- Muscle Implantation Test per ISO 10993-6:2007
- In Vitro Hemolytic Test per ISO 10993-4:2002/Amd.1: 2006 ●
- Subchronic Systemic Toxicity Test per ISO 10993-11:2006
- Pyrogen Test per ISO 10993-11:2006 and USP39-NF34<151>
## VIII. Conclusion
The non-clinical tests, electrical safety and EMC, software verification & validation, and biocompatibility, demonstrate that the VR Medical VCare 1000-300S System is equivalent to the predicate device for the intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.