K211526 · Cooler Heads Care, Inc. · PMC · Oct 21, 2021 · General, Plastic Surgery
Device Facts
Record ID
K211526
Device Name
Portable Scalp Cooling System
Applicant
Cooler Heads Care, Inc.
Product Code
PMC · General, Plastic Surgery
Decision Date
Oct 21, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4360
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Portable Scalp Cooling System (PSCS) is indicated to reduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors. The PSCS is indicated for adult patients only.
Device Story
Portable Scalp Cooling System (PSCS) is a wearable device for adult cancer patients undergoing chemotherapy for solid tumors. Device aims to reduce chemotherapy-induced alopecia. System circulates a water/isopropyl alcohol (IPA) mixture through a cooling cap worn on the patient's head. Operated by user or caregiver in treatment centers, during travel, and in home settings. Provides continuous cooling therapy during transition from infusion center to home. Powered by wall outlet, internal rechargeable lithium battery, or car charger. Differences from predicate include portability, power options, and use of water/IPA/ice coolant mixture versus proprietary coolant. Healthcare providers/caregivers monitor therapy; device usage intended to mitigate hair loss, potentially improving patient quality of life.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety (IEC 60601-1, 1-6, 1-11), electromagnetic compatibility (IEC 60601-1-2), software V&V (IEC 62304), performance bench testing (scalp cooling performance, cooling stability), human factor testing (IEC 62366-1), and biocompatibility (ISO 10993-1).
Technological Characteristics
Wearable scalp cooling system. Materials meet ISO 10993-1 for prolonged skin contact. Cooling via circulating water/70% isopropyl alcohol/ice mixture. Power: 100-240 VAC, internal rechargeable lithium battery, or 11-15 VDC car charger. Connectivity: Portable/mobile. Software V&V per IEC 62304.
Indications for Use
Indicated for adult cancer patients with solid tumors to reduce the likelihood of chemotherapy-induced alopecia.
Regulatory Classification
Identification
A scalp cooling system to reduce the likelihood of chemotherapy-induced alopecia is a prescription device intended to reduce the frequency and severity of alopecia during chemotherapy in which alopecia-inducing chemotherapeutic agents are used.
Special Controls
*Classification* —Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must demonstrate that the device meets all design specifications and performance requirements, and that the device performs as intended under anticipated conditions of use. This information must include testing to demonstrate accuracy of the temperature control mechanism.
(2) Performance testing must demonstrate the electromagnetic compatibility and electrical safety of the device.
(3) Software verification, validation, and hazard analysis must be performed.
(4) The patient contacting components of the device must be demonstrated to be biocompatible. Material names must be provided.
(5) Labeling must include the following:
(i) A statement describing the potential risk of developing scalp metastasis.
(ii) Information on the patient population and chemotherapeutic agents/regimen for which the device has been demonstrated to be effective.
(iii) A summary of the non-clinical and/or clinical testing pertinent to use of the device.
(iv) A summary of the device technical parameters, including temperature cooling range and duration of cooling.
(v) A summary of the device- and procedure-related adverse events pertinent to use of the device.
(vi) Information on how the device operates and the typical course of treatment.
(6) Patient labeling must be provided and must include:
(i) Relevant contraindications, warnings, precautions, and adverse effects/complications.
(ii) Information on how the device operates and the typical course of treatment.
(iii) Information on the patient population for which there is clinical evidence of effectiveness.
(iv) The potential risks and benefits associated with use of the device.
(v) Postoperative care instructions.
(vi) A statement describing the potential risk of developing scalp metastasis.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of a human figure. To the right of the symbol is a blue square containing the acronym "FDA" in white letters. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
Cooler Heads Care, Inc. % Pierre Bounaud Principal Consultant Acknowledge Regulatory Strategies 2251 San Diego Avenue, Suite B-257 San Diego, California 92110
October 21, 2021
## Re: K211526
Trade/Device Name: Portable Scalp Cooling System Regulation Number: 21 CFR 878.4360 Regulation Name: Scalp Cooling System to reduce the likelihood of Chemotherapy-Induced Alopecia Regulatory Class: Class II Product Code: PMC Dated: October 1, 2021 Received: October 4, 2021
## Dear Pierre Bounaud:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Neil R.P. Ogden Assistant Director, THT4A4 DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K211526
Device Name Portable Scalp Cooling System
Indications for Use (Describe)
The Portable Scalp Cooling System (PSCS) is indicated to reduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors. The PSCS is indicated for adult patients only.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☑</span> Beneficial Use (Part 31 CFR 901.2, subpart D) </span> |
|----------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☐</span> Own-The-Counter Use (31 CFR 901.2, subpart C) </span> |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
DATE PREPARED
May 13, 2021
#### MANUFACTURER AND 510(k) OWNER
Cooler Heads Care, Inc. 2801 B Street, #209, San Diego, CA 92102, USA Telephone: Official Contact: Kate Dilligan, CEO
#### REPRESENTATIVE/CONSULTANT
Pierre Bounaud, Ph.D. Allison C. Komiyama, Ph.D., R.A.C. AcKnowledge Regulatory Strategies, LLC Telephone: +1 (619) 458-9547 Email: pbounaud@acknowledge-rs.com
#### DEVICE INFORMATION
| Proprietary Name/Trade Name: | Portable Scalp Cooling System (PSCS) |
|------------------------------|-------------------------------------------|
| Common Name: | Scalp Cooling System |
| Regulation Number: | 21 CFR 878.4360 |
| Class: | II |
| Product Code: | PMC |
| Premarket Review: | OPEQ/OHT4/General Surgery Devices (DHT4A) |
| Review Panel: | General & Plastic Surgery |
#### PREDICATE DEVICE IDENTIFICATION
The PSCS is substantially equivalent to the following predicate:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
|---------------|------------------------------------------|-------------------|
| K173032 | Paxman Scalp Cooler / Paxman Coolers Ltd | ✓ |
The predicate device has not been subject to a design related recall.
#### DEVICE DESCRIPTION
The Portable Scalp Cooling System (PSCS) is a wearable device for people undergoing chemotherapy treatment for solid tumor cancer who want to avoid chemotherapy-induced hair loss, or alopecia. Treatment with the PSCS is administered by a user or caregiver in a treatment center, during travel time to the patient's home and in the home setting. It is designed for continual use during the transition from the infusion center to the patient's home. The portable device provides the cooling therapy by circulating a water/isopropyl alcohol (IPA) mixture through a cooling cap worn securely on the patient's head.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the words "Cooler Heads" stacked on top of each other. The words are written in a sans-serif font and are a light blue color. The background is white. The text is slightly blurred, giving it a soft appearance.
#### 510(k) Summary
## INDICATIONS FOR USE
The Portable Scalp Cooling System (PSCS) is indicated to reduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors. The PSCS is indicated for adult patients only.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Cooler Heads Care, Inc. believes that the PSCS is substantially equivalent to the predicate device based on the information summarized here:
The subject device has the same intended use as the device cleared in K173032. Both devices are scalp cooling systems intended to reduce the likelihood of chemotherapy-induced alopecia. The subject device is indicated for home use, while the predicate device is indicated for healthcare facilities. This difference has undergone human factor testing to ensure the device is as safe and effective as the predicate.
The subject device has similar, albeit more compact design, similar materials, and similar technological characteristics as the device cleared in K173032. The main technological differences of the subject device, as compared to the predicate device cleared in K173032, are:
- The type of coolant. The subject device uses a mixture of readily available 70% isopropyl o alcohol, water, and ice. The predicate device uses a proprietary cooling mixture (OrbisC).
- . The power source. The subject device was designed to be a mobile device and includes three different power sources: wall outlet (100-240 VAC – 50/60 Hz), internal rechargeable lithium battery, and a car charger (11-15 VDC 10A). The predicate device is intended to be plugged into a wall outlet only (100-120 VAC).
These technological differences have undergone testing to ensure the subject device is as safe and effective as the predicate device.
#### SUMMARY OF NON-CLINICAL TESTING
No FDA performance standards have been established for the PSCS. The following tests were performed to demonstrate safety based on current industry standards:
- o Electrical safety per IEC 60601-1, IEC 60601-1-6, and IEC 60601-1-11
- Electromagnetic compatibility per IEC 60601-1-2 o
- . Software V&V per IEC 62304
- . Performance bench testing (scalp cooling performance, cooling stability between power sources)
- . Human factor testing per IEC 62366-1
The results of these tests indicate that the PSCS is substantially equivalent to the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the words "Cooler Heads" stacked on top of each other. The words are written in a sans-serif font and are a light blue color. The background is white, which makes the text stand out. The text is slightly blurred.
## 510(k) Summary
PSCS meets the requirements of ISO 10993-1 and FDA's Guidance for Industry and Food and Druq Administration Staff - Use of International ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", for a surface device with cumulative prolonged contact duration with intact skin.
#### CONCLUSION
Based on the testing performed, including electrical safety, electromagnetic compatibility, software V&V, performance bench testing, and human factor testing, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. The similar indications for use, technological characteristics, and performance characteristics for the proposed PSCS are assessed to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.