K211217 · Candela Corporation · GEI · Dec 16, 2021 · General, Plastic Surgery
Device Facts
Record ID
K211217
Device Name
Profound Matrix
Applicant
Candela Corporation
Product Code
GEI · General, Plastic Surgery
Decision Date
Dec 16, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PROFOUND MATRIX SYSTEM is intended for dermatological procedures, as follows: The MATRIX PRO applicator is indicated for general dermatological procedures for electrocoagulation and hemostasis. The SUBLATIVE RF applicator is indicated for dermatological procedures requiring ablation and resurfacing of the skin, and for the treatment of facial wrinkles. At higher energy levels greater than 62mJ/pin, the Sublative RF applicator is limited to Fitzpatrick Skin Types I-IV. The SUBLIME applicator is indicated for non-invasive wrinkle treatment. At higher energy levels, greater than 100 J/ cm3, the Sublime applicator is limited to Fitzpatrick Skin Types I-IV.
Device Story
Profound Matrix System is a multi-applicator electrosurgical device for dermatological procedures. It utilizes three applicators: Matrix Pro (bipolar RF microneedling for electrocoagulation/hemostasis), Sublative RF (bipolar RF for skin ablation/resurfacing/wrinkles), and Sublime (combined IR light and bipolar RF for non-invasive wrinkle treatment via bulk dermal heating). Operated by physicians in clinical settings, the system uses a touchscreen interface to adjust parameters like needle depth, RF energy, and power density. The Matrix Pro applicator features impedance-based energy control to deliver precise thermal doses. The device transforms electrical energy into thermal energy at the skin surface or within the dermis to stimulate tissue healing and collagen remodeling. Clinical benefits include improved skin appearance and wrinkle reduction. The system is intended for professional use.
Clinical Evidence
No clinical or animal studies were conducted. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1, 60601-1-2, 60601-2-2), biocompatibility (ISO 10993), software validation, and ex-vivo thermal damage evaluation on excised abdominal skin. Ex-vivo testing confirmed that the Matrix Pro produces thermal effects (collagen denaturation, fat cell membrane disruption) consistent with the predicate Infini device.
Technological Characteristics
System console with touchscreen interface. Applicators: Matrix Pro (49 microneedles, 1 MHz bipolar RF), Sublative RF (multi-electrode pin array, 1 MHz bipolar RF), Sublime (IR 700-2000 nm + 1 MHz bipolar RF). Materials: Biocompatible patient-contacting components per ISO 10993. Energy: Bipolar RF and IR. Connectivity: Standalone console. Sterilization: Sterile microneedle tips. Software: Controlled by impedance sampling for energy delivery.
Indications for Use
Indicated for dermatological procedures including electrocoagulation, hemostasis, skin ablation, resurfacing, and treatment of facial wrinkles. Sublative RF and Sublime applicators at higher energy levels are restricted to Fitzpatrick Skin Types I-IV.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Syneron Medical Ltd eTwo Skin Treatment System (K141507, K110672)
Syneron Candela Corporation Profound System (K161043)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 16, 2021
Candela Corporation Rina Ordonez 251 Locke Dr Marlborough, Massachusetts 01752
Re: K211217
Trade/Device Name: Profound Matrix Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: November 19, 2021 Received: November 22, 2021
Dear Rina Ordonez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements,
{1}------------------------------------------------
including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and. if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations. please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH Learn (https://www.fda.gov/training-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE website assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211217
Device Name Profound Matrix
Indications for Use (Describe) The PROFOUND MATRIX SYSTEM is intended for dermatological procedures, as follows:
The MATRIX PRO applicator is indicated for general dermatological procedures for electrocoagulation and hemostasis.
The SUBLATIVE RF applicator is indicated for dermatological procedures requiring ablation and resurfacing of the skin, and for the treatment of facial wrinkles. At higher energy levels greater than 62mJ/pin, the Sublative RF applicator is limited to Fitzpatrick Skin Types I-IV.
The SUBLIME applicator is indicated for non-invasive wrinkle treatment. At higher energy levels, greater than 100 J/ cm3, the Sublime applicator is limited to Fitzpatrick Skin Types I-IV.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <input checked="true" type="checkbox"/> <span>Executive Use (Part 91, 125) Private Pleasure</span> </div> | <div> <input type="checkbox"/> <span>For-Hire, Commercial (Part 135, 121)</span> </div> |
|-----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary — K211217 Profound Matrix System
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92
#### 1. DATE PREPARED
April 21, 2021
#### 2. SUBMITTER
Candela Corporation 251 Locke Drive Marlborough MA 01752 USA
### 3. OFFICIAL CORRESPONDENT
Rina Ordonez Senior Regulatory Affairs Specialist Candela Corporation 251 Locke Dr. Marlborough, MA 01752 Office: 508-834-6072 Email: rinao@candelamedical.com
#### 4. DEVICE INFORMATION
| Proprietary Name: | Profound Matrix™ |
|------------------------|------------------------------------------------------------------------------------------|
| Common/Usual Name: | Electrosurgical coagulation device and accessory |
| Classification Name: | Electrosurgical cutting and coagulation device and<br>accessories (21 CFR Part 878.4400) |
| Product Code: | GEI |
| Device Classification: | Class II |
#### 5. PREDICATE DEVICE
Primary Predicate: Lutronic Infini (K121481) Reference Devices: Syneron Medical Ltd eTwo Skin Treatment System (K141507, K110672), Syneron Candela Corporation Profound System (K161043)
### 6. INTENDED USE/INDICATION FOR USE
The PROFOUND MATRIX SYSTEM is intended for dermatological procedures, as follows:
{4}------------------------------------------------
## 510(k) Summary — K211217
The MATRIX PRO applicator is indicated for general dermatological procedures for electrocoagulation and hemostasis.
The SUBLATIVE RF applicator is indicated for dermatological procedures requiring ablation and resurfacing of the skin, and for the treatment of facial wrinkles. At higher energy levels greater than 62 mJ/pin, the Sublative RF applicator is limited toFitzpatrick Skin Types I-IV.
The SUBLIME applicator is indicated for non-invasive wrinkles treatment. At higher energy levels greater than 100 J/cm³, the Sublime applicator is limited to Fitzpatrick Skin Types I-IV.
### DEVICE DESCRIPTION 7.
The Profound Matrix System combines three existing technologies and applications for dermatological procedures. Each respective technology is used in one of the three applicators designed for the Profound Matrix System. These three applicators are named: Matrix Pro, Sublative RF and Sublime.
The Matrix Pro applicator is designed to deliver bipolar, non-ablative radiofrequency energy through a 7x7 matrix/grid of sterile microneedles to the skin in a fractional manner. The microneedles can be set up to pulse a maximum of three depths during each insertion of the matrix/grid of microneedles with a maximum depth of up to 3.5mm.
The Sublative RF applicator delivers bipolar radiofrequency (RF) to the skin surface viaan array of multi-electrode pin tips. The applicator delivers bipolar RF energy to the skin, which results in heating of demarcated spots to temperatures that ablate and resurface at contact points of the multi-electrode pins.
The Sublime applicator uses a combination of an infrared (IR) light source and bipolar radiofrequency (RF) to bulk heat the dermis and affect dermal collagen in order to treat wrinkles. The bulk heating of the dermal layers refers to the gradual and gentle accumulation of heat with each additional pass performed.
#### 8. TECHNOLOGICAL CHARACTERISTICS
The technological characteristics comparison between the Profound Matrix System and its predicates is done per applicator and system console, as each applicator incorporates different technology, and the console enables each technology.
The following table (Table 8-1) compares the key technological characteristics and features of the Profound Matrix System to the predicate device. The tables that follow compare the key technical characteristics and features of the different applicators with predicate and reference devices (Tables 8-2, 8-3, and 8-4).
{5}------------------------------------------------
| Predicate and Reference Devices | | | Reference Device | Reference Device | | Reference Devices | | |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|----------------------------------------------------|----------------------------------------------------------------|--------------------------------------------------------------|
| Specifications | Subject Device | Predicate Device | | | Specifications | Profound Matrix-<br>Matrix Pro<br>(Subject Device) | Infini Infini Handpiece<br>K121481<br>(Predicate Device) | Profound Profound<br>Dermal K161043<br>(Reference Device #2) |
| | | Profound Matrix | Infini<br>Radiofrequency<br>System (K121481) | eTwo<br>(K141507) | Applicator Type | Short Pulse<br>Microneedling | Short Pulse<br>Microneedling | Long Pulse<br>Microneedling |
| Indications for Use | The PROFOUND<br>MATRIX SYSTEM is<br>Intended for<br>dermatological<br>procedures, as<br>follows:<br><br>The MATRIX PRO<br>applicator is<br>intended for use in<br>dermatologic,<br>procedures for<br>electrocoagulation<br>and hemostasis.<br><br>The SUBLATIVE RF<br>applicator is<br>indicated for<br>dermatological<br>procedures<br>requiring ablation<br>and resurfacing of<br>the skin, and for<br>the treatment of<br>facial wrinkles. At<br>higher energy<br>levels greater than<br>62 mJ/pin, the<br>Sublative RF<br>applicator is<br>limited to<br>Fitzpatrick skin<br>types I-IV.<br><br>The SUBLIME<br>applicator is<br>indicated for non-<br>invasive wrinkles<br>treatment At | The INFINI<br>Radiofrequency<br>System is intended<br>for use in<br>dermatologic and<br>general surgical<br>procedures for<br>electrocoagulation<br>and hemostasis,<br>and the<br>percutaneous<br>treatment of facial<br>wrinkles. | The eTwo Skin<br>Treatment<br>System is<br>intended for<br>dermatological<br>procedures.<br><br>The Sublative<br>RF applicator is<br>indicated for<br>dermatological<br>procedures<br>requiring<br>ablation and<br>resurfacing of<br>the skin, and<br>for the<br>treatment of<br>facial wrinkles.<br>At higher<br>energy levels<br>greater than 62<br>mJ/pin, the<br>Sublative RF<br>applicator is<br>limited to skin<br>types I-IV.<br>The Sublime<br>applicator is<br>indicated for<br>non-invasive<br>wrinkles<br>treatment. At<br>higher energy<br>levels greater<br>than 100<br>J/cm3, the<br>Sublime<br>applicator is | The Profound<br>System is<br>indicated for use<br>in dermatologic<br>and general<br>surgical<br>procedures for<br>electrocoagulation<br>and hemostasis.<br>Specifically, the<br>25º Dermal<br>handpiece and<br>cartridge are used<br>for percutaneous<br>treatment of facial<br>wrinkles, and the<br>75º SubQ<br>handpiece and<br>cartridge are used<br>to improve the<br>appearance of<br>cellulite in<br>patients with<br>Fitzpatrick skin<br>types I-III as<br>supported by<br>long- term clinical<br>data (6 months). | Applicator Dimensions | 14.4 x 14.3 cm / 5.7 x<br>5.6" [W x H] | Unknown | 19.4 x 13.9 cm / 7.6 x<br>5.5" |
| Specifications | Subject Device | Predicate Device | Reference Device | Reference Device | Energy Source | Matrix Pro: Radio<br>Frequency | Infini handpiece: Radio<br>Frequency | Profound Dermal<br>Cartridge: Radio<br>Frequency |
| | Profound Matrix | Infini<br>Radiofrequency<br>System (K121481) | eTwo<br>(K141507) | Profound System<br>(K161043) | RF Mode of operation | Bipolar | Bipolar | Bipolar |
| | higher energy<br>levels greater than<br>100 J/cm3, the<br>Sublime applicator<br>is limited to<br>Fitzpatrick skin<br>types I-IV | | limited to skin<br>types I-IV. | | RF Frequency | 1 MHz | 1 MHz | 460±5 kHz |
| Device Dimensions | 39.4 x 57.4 x 135.6<br>cm | 36.2 x 40.9 x 171.3<br>cm | 30 x 23 x 53 cm | 46.5x 44.5 x 125<br>cm | Disposable Tip Area<br>Coverage & Injection<br>Pattern | 1 cm² Square | 49 needle Tip<br>1 cm²<br><br>16 needle Tip 0.36 cm²<br>Square | 1.4 cm x 0.16 cm Pseudo-<br>Linear |
| Weight<br>(Pounds, Kg) | 50, 22.5 | 61.7, 28 | 18.7, 8.5 | 55, 23 | RF Pulse duration | Up to 280 ms | 10 to 1000 ms | 3-5 seconds |
| Power Supply | Yes | Yes | Yes | Yes | # Of needles | 49 (7 x 7) | 49 (7 x 7) or 16 (4 x 4) | 10 (5 pairs of 2) |
| ElectricalRating | Single phase 100-240V, 3A,50/60Hz | Single phase<br>AC220-230V, 50-60Hz<br>Power<br>consumption:<br>500VA<br>(Fuse: 250V/6.3A) | Single phase<br>100-240 VAC,<br>3A, 50-60Hz | Single phase 100-240 V, 2.5A,50-60Hz | # Of treatment levels<br>per<br>insertion | 1, 2 or 3 | 1 | 1 |
| User Interface | Graphics with<br>touchscreen | Graphics with<br>touchscreen | Graphics with<br>touchscreen | Graphics with<br>touchscreen | # Of treatment zones<br>per<br>insertion | 49, 98, or 147 | 49 | 5 |
| Applicators | Matrix Pro (RF<br>microneedles)<br>Sublative Sublime | RF microneedles | Sublative<br>Sublime | Dermal RF<br>Subcutaneous RF | Maximum RF power | 50 W | 50 W | 15 W |
| Treatment Modes | Matrix Pro<br>- Manual Sublative<br>- Manual Sublime<br>- Manual | Manual and Auto | Sublative –<br>Manual<br>Sublime –<br>Manual | Dermal – Manual<br>Subcutaneous –<br>Manual | Energy delivered per<br>treatment<br>level | 1 - 4 J | 25 mJ – 50 J<br>1 – 20 Levels* | Unknown |
| RF Energy Activation | Fingerswitch or<br>Footswitch | Fingerswitch or<br>Footswitch | Fingerswitch | Fingerswitch | Energy delivered per<br>treatment<br>zone | 20 – 82 mJ | 0.51 mJ – 1.02 J | Unknown |
| Treatment Settings | Matrix Pro –<br>needle depth and<br>RF energy<br>Sublative - RF<br>energy per pin<br>Sublime - RF<br>power density and<br>IR fluence | Manual mode –<br>needle depth, RF<br>level (power), and<br>RF pulse<br>duration. Auto<br>mode — Location<br>(Lower face<br>Submental, Neck,<br>Forehead,<br>Periorbital), Pass<br>(1 to 3), Intensity<br>(Low, Medium, | Sublative - RF<br>energy per pin<br>Sublime – RF<br>power density<br>and IR fluence | Dermal,<br>Subcutaneous –<br>Treatment<br>temperature and<br>treatment time | Controlled energy<br>delivered | Yes, Controlled by<br>sampling impedance | No<br>Not controlled | No<br>Not controlled |
| Controlled thermal dose | No | No | Yes, PID temperature<br>control 50-75°C ±10°C<br>with treatment duration<br>of 3-5 seconds | | | | | |
| Repetition<br>Rate** | No existing setting | No existing setting | No existing setting | | | | | |
| Needle uncoated | 0.6 mm | 0.3 mm | 3 mm | | | | | |
## Table 8-1: SYSTEM LEVEL - Technical Comparison of Subject System Device with Predicate and Reference Devices
{6}------------------------------------------------
# 510(k) Summary – K211217
{7}------------------------------------------------
## 510(k) Summary – K211217
## Table 8-2: HANDPIECE AND CARTRIDGE LEVEL - Technical Comparison of Subject Device with Predicate and Reference Devices
{8}------------------------------------------------
# 510(k) Summary – K211217
| Specifications | Profound Matrix-<br>Matrix Pro<br>(Subject Device) | Infini Infini Handpiece<br>K121481<br>(Predicate Device) | Profound Profound<br>Dermal K161043<br>(Reference Device #2) | |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|
| Needle insertion angle<br>to<br>tissue | 90° | 90° | 25° | |
| | Insertion depth range | 0.8 - 3.5 mm<br>Steps of 0.8 mm | 0.5 - 3.5 mm<br>Recommended steps of<br>at least 0.5 mm | 2 mm |
| | Accuracy of insertion<br>depth | ±0.4 mm | Unknown | Unknown |
| Needle Thickness /<br>Diameter | | 34 gauge<br>160 μm | 32 gauge<br>200 μm | 32 gauge<br>200 μm |
| Treatment Duration | Up to 4.5 sec | 10-400 ms (10 ms steps)<br>400-1000 ms (50 ms<br>steps) | 5 sec ±0.2 sec<br>increments | |
| Rated Voltage | 275 Vrms | Unknown | 84 Vrms | |
| Mode of Operation | Microneedles are<br>inserted into the dermis<br>where they emit bipolar<br>RF energy from their<br>electrode needle tips,<br>thermally damaging<br>surrounding tissue and<br>stimulating a<br>tissue healing process. | Microneedles are<br>inserted into the dermis<br>where they emit bipolar<br>RF energy from their<br>electrode needle tips,<br>thermally damaging<br>surrounding tissue and<br>stimulating a<br>tissue healing process | Microneedles are<br>inserted into the dermis<br>where they emit bipolar<br>RF energy from their<br>electrode needle tips,<br>thermally damaging<br>surrounding tissue and<br>stimulating a<br>tissue healing process | |
*Level is defined by Intensity/Rate of Heating (Level 5 is equivalent to 9.58 W with a 300-ohm impedance) **Repetition Rate is defined as preset period between needle withdrawal and the next needle insertion
{9}------------------------------------------------
| Specifications | Profound Matrix<br>Sublative RF<br>(Subject device) | eTwo Skin Treatment System<br>Sublative RF<br>K141507<br>(Reference Device #1) |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Energy Type | RF | RF |
| Applicator Dimensions | 15 x 16 cm / 5.9 x 6.3"<br>[W x H] | 15 x 16 cm / 5.9 x 6.3"<br>[W x H] |
| Applicator Weight | 0.5 kg. / 1.1 lbs. | 0.5 kg. / 1.1 lbs. |
| RF Frequency | 1 MHz | 1 MHz |
| Power Range | 14-55 W | 14-55 W |
| Treatment Area | 64 pin Tip 12 x 12 mm<br>-or-<br>44 pin Tip 3 x 11 mm | 64 pin Tip 12 x 12 mm<br>-or-<br>44 pin Tip 3 x 11 mm |
| Pulse Duration | Up to 500 ms | Up to 500 ms |
| Total RF Energy | Up to 100mJ/pin | Up to 100mJ/pin |
| Applicator Dimensions | 15 x 16 cm | 15 x 16 cm |
| Disposable Tip Matrix | 44 or 64 matrix pins | 44 or 64 matrix pins |
| RF Energy Penetration<br>depth | Up to 400 µm | Up to 400 µm |
| Disposable Treatment Tip<br>Lifetime | Up to 400 pulses | Up to 400 pulses |
| Rated Voltage | 275 Vrms | 275 Vrms |
| Mode of operation | During treatment, the multi-<br>electrodepin array is placed on the<br>dry skin's surface. The RF current<br>flows betweenthe rows of the<br>pins, having the highest impact at<br>the electrode-skin contact points<br>where it creates spots of<br>demarcated ablation and<br>resurfacing of the skin (stratum<br>corneum, the deeper epidermis<br>and superficial dermis) to achieve<br>wrinkle | During treatment, the multi-<br>electrodepin array is placed on the<br>dry skin's surface. The RF current<br>flows betweenthe rows of the<br>pins, having the highest impact at<br>the electrode-skin contact points<br>where it creates spots of<br>demarcated ablation and<br>resurfacing of the skin (stratum<br>corneum, the deeper epidermis<br>and superficial dermis) to achieve<br>wrinkle |
## Table 8-3: Comparison of parameters for Profound Matrix System's Sublative RF applicator and eTwo System's Sublative applicator
{10}------------------------------------------------
| Specifications | Profound Matrix<br>Sublime<br>(Subject Device) | eTwo Skin Treatment System<br>Sublime (Refirme)<br>K141507 (Reference Device<br>#1) |
|-----------------------|------------------------------------------------|-------------------------------------------------------------------------------------|
| Energy Type | RF and IR | RF and IR |
| Applicator Dimensions | 20 x 20 cm / 7.8 x 7.8" | 20 x 20 cm / 7.8 x 7.8" |
| | [W x H] | [W x H] |
| Applicator Weight | 1.0 kg. / 2.2 lbs. | 1.0 kg. / 2.2 lbs. |
| Frequency | 1 MHz | 1 MHz |
| RF Power | 200 W (≤138.9 W/cm2) | 200 W (≤138.9 W/cm2) |
| | | |
| Treatment Area | 8 x 12 mm | 8 x 12 mm |
| Pulse Duration | Up to 200 ms | Up to 200 ms |
| Total RF Energy | Hz Mode: 50-200 J/cm3 | Hz Mode: 50-200 J/cm3 |
| | Hz Mode: 50-100 J/cm3 | Hz Mode…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.