K211132 · Ariste Medical, LLC · FTL · Mar 18, 2022 · General, Plastic Surgery
Device Facts
Record ID
K211132
Device Name
ARISTE AB Mesh
Applicant
Ariste Medical, LLC
Product Code
FTL · General, Plastic Surgery
Decision Date
Mar 18, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ARISTE AB Mesh is a non-resorbable polymeric surgical mesh intended to reinforce soft tissue where weakness exists.
Device Story
ARISTE AB Mesh is a non-resorbable sterile prosthesis for soft tissue reconstruction; constructed of knitted polypropylene fibers. Mesh is coated with a thiolene polymer acting as a carrier for antimicrobial agents rifampin and minocycline (approx. 171μg/cm² each). Device is implanted by surgeons in clinical settings to reinforce tissue defects or repair hernias. Antimicrobial coating provides protection from microbial colonization during implantation. Device provides mechanical support for tissue deficiencies. No clinical data or clinical impact regarding the reduction of bacterial colonization has been demonstrated.
Clinical Evidence
No clinical data. Evidence consists of bench testing and GLP animal studies. Bench testing evaluated physical/functional characteristics (thickness, stiffness, tensile/burst/suture pullout strength, tear resistance) and drug elution/kinetics. In vitro studies showed >4 log antibacterial activity against MRSA, S. epidermidis, S. pyogenes, E. coli, A. baumannii, and E. aerogenes. Animal studies (rabbit abdominal wall defect model) confirmed clearance of MRSA and E. coli for 7 days.
Technological Characteristics
Non-resorbable polypropylene knitted mesh; coated with thiolene polymer containing rifampin and minocycline (171μg/cm² each). Dimensions: 0.49 mm thickness, 145 g/m2 density. Sterilized via E-Beam (ISO/AAMI 11137) to SAL 10^-6. Double sterile barrier packaging. Biocompatibility per ISO 10993-1:2018.
Indications for Use
Indicated for use as a soft tissue patch to reinforce soft tissue where weakness exists and for the surgical repair of damaged or ruptured soft tissue, including hernias and body wall defects. Contains rifampin and minocycline to help provide protection from microbial colonization during surgical implantation. Intended for single patient one-time use.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Predicate Devices
PIVIT A/B ST AND PIVIT A/B ST-R Surgical Mesh (K093524)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services-USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
March 18, 2022
Ariste Medical LLC % Brian Young VP Regulatory and Clinical Affairs Health Policy Associates Inc 690 Canton Street, Suite 302 Westwood, Massachusetts 02492
Re: K211132
Trade/Device Name: ARISTE AB Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTL Dated: May 5, 2021 Received: May 7, 2021
# Dear Brian Young:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Deborah Fellhauer, RN, BSN Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211132
Device Name ARISTE AB Mesh
#### Indications for Use (Describe)
ARISTE AB Mesh is intended for use as a soft tissue patch to reinforce soft tissue where weakness exists and for the surgical repair of damaged or ruptured soft tissue. Indications for use include the repair of hernias and/or body wall defects which require the use of reinforcing or bridging material to obtain the desired surgical outcome. The non-absorbable polymer coating on the mesh contains the antimicrobial agents, rifampin and minocycline to help provide protection from microbial colonization of the device during surqical implantation.
ARISTE AB Mesh is intended for single patient one-time use only.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# SECTION 5: 510(K) SUMMARY for K211132
The following information is provided in accordance with 21 CFR 807.92, for the Premarket 510(k) Summary:
## APPLICANT INFORMATION:
| Company Name: | ARISTE Medical, LLC |
|----------------------|-----------------------------|
| Company Address: | 9950 S 300 W |
| | Sandy UT 84070 |
| SUBMITTED BY: | |
| Company Name: | ARISTE Medical LLC |
| Contact Name: | Lisa K Jennings, PhD |
| Contact Address: | 140 Collins St |
| | Memphis, TN 38112 |
| Phone: | (901) 866-1700 |
| E-mail: | Ljennings@aristemedical.com |
| Date Prepared: | 3/16/2022 |
| DEVICE INFORMATION | |
| Trade Name: | ARISTE AB Mesh |
| CommonName: | Surgical Mesh |
| Classification Name: | Surgical Mesh |
| Regulation Number: | 21 CRF 878.3300 |
Regulatory Class: II Product Code: FTL
# PREDICATE DEVICE INFORMATION:
The ARISTE AB Mesh claims substantial equivalence to PIVIT A/B ST AND PIVIT A/B ST-R Surgical Mesh cleared by FDA under K093524. XenMatrix AB Surgical Graft (K133223) and Prolene™ Mesh (K962530) are identified as reference devices.
# DEVICE DESCRIPTION:
ARISTE AB Mesh is a non-resorbable sterile prosthesis designed to provide mechanical support for reconstruction of soft tissue deficiencies. The mesh is constructed of polypropylene fibers knitted together to form the mesh. The resulting structure is an implant which reinforces the tissue defect.
{4}------------------------------------------------
The mesh is coated with a thiolene polymer. The coating acts as a carrier for antibiotic agents rifampin and minocycline in equal concentrations of approximately 171μg/cm² to help provide protection from bacterial colonization of the device during surgical implantation. While the effectiveness of the rifampin and minocycline components has been evaluated via in vitro and in vivo testing, no clinical data nor clinical impact associated with this claimed reduction of bacterial colonization of the ARISTE AB mesh has been demonstrated.
# INTENDED USE:
ARISTE AB Mesh is a non-resorbable polymeric surgical mesh intended to reinforce soft tissue where weakness exists.
### INDICATIONS FOR USE:
ARISTE AB Mesh is intended for use as a soft tissue patch to reinforce soft tissue where weakness exists and for the surgical repair of damaged or ruptured soft tissue. Indications for use include the repair of hernias and/or body wall defects which require the use of reinforcing material to obtain the desired surgical outcome. The non-resorbable polymer coating on the mesh contains the antimicrobial agents, rifampin and minocycline to help provide protection from microbial colonization of the device during surgical implantation.
ARISTE AB Mesh is intended for single patient one-time use only.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:
ARISTE AB Mesh is similar in intended use and in technological characteristics and performance when compared to the cited predicate devices. ARISTE AB Mesh and the cited predicates are intended for use in the reconstruction and repair of soft tissue deficiencies where weakness exists such as hernia repair. In addition, these devices are similar in technological characteristics with regard to sterilization, packaging and labeling
Ariste AB Mesh is the same as the primary predicate PIVIT AB Mesh in the following respects: intended use; antimicrobial agents; mechanism of action; mesh material and method of construction; product handling; and method of implantation.
ARISTE AB Mesh differs from the primary predicate PIVIT AB Mesh in the following technological respects. Each technological difference is explained and supported with performance data.
- . Antimicrobial amounts
- The subject device contains equal amounts (171 ug/cm2) of rifampin O and minocycline whereas the predicate contains 86.11 ug/cm² of both rifampin and minocycline.
- o A 510(k) cleared device, Xenmatrix AB Surgical Graft, containing approximately the same antibiotic concentrations (180 ug/cm2) is used as a reference device for purposes of demonstrating safety and effectiveness of the antibiotic concentration used.
{5}------------------------------------------------
- In vitro and in vivo performance and safety testing demonstrating of O antimicrobial characteristics are provided.
- Polymer coating .
- The subject device utilizes a non-resorbable polymer coating as a O substrate for the antibiotics whereas the PIVIT® AB uses a resorbable polyarylate.
- O Bench and animal Performance testing is provided.
- . Physical characteristic
- Mesh thickness (0.49 mm Ariste vs. 0.51-0.53 mm Prolene), O weave characteristics (1.0 mm x 0.6 mm Ariste vs. 0.8-1.6 mm Prolene), and density (145 g/m2 Ariste vs 79.5-108 g/m2 Prolene) are similar to the predicate Prolene Mesh.
- Bench and animal performance testing demonstrating the device's O physical characteristics are provided.
# PERFORMANCE TEST SUMMARY
The ARISTE AB Mesh underwent an extensive safety and performance testing program, including bench and animal testing, to demonstrate that the device meets the requirements as defined in user specifications, and performs as intended. -
### Packaging, Sterilization and Shelf Life Testing
The ARISTE AB Mesh packaging system consists of a double sterile barrier. The ARISTE AB Mesh is E-Beam sterilized to an SAL of 1 X 10-6 using validated methods pursuant to ISO/AAMI 11137. A combination of real time and accelerated age life testing supports 6 months expiry dating.
### Biocompatibility testing
Biocompatibility testing was performed in accordance with ISO 10993-1:2018 and the "Use of International Standard ISO-10993-1, 'Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a risk management process," (June 16, 2016) in accordance with the device's tissue contact category and duration of implantation (permanent).
### Bench testing
In vitro testing was performed to assess the physical and performance characteristics of the ARISTE AB Mesh in accordance with FDA's "Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh'' (March 2, 1999) as well as FDA's "Premarket Notification [510(k)] Submissions for Medical Devices that Include Antimicrobial Agents" Antimicrobial Agents" (July 19, 2007). The in vitro testing included the following physical and performance characterization evaluations:
- a) Physical Characteristics:
{6}------------------------------------------------
- i. Device Thickness
- ii. Device Stiffness (ratio between strength and elongation)
- b) Functional Characteristics:
- i. Tensile strength (break strength)
- ii. Elongation
- Burst Strength iii.
- Suture Pullout Strength iv.
- Tear Resistance (single rip/tongue method) v.
Results indicated that the ARISTE AB Mesh meets or exceeds the minimum mechanical strength requirements.
### Drug Elution and Antibiotic Effectiveness
Drug content and impurities analytical testing of the antibacterial agents rifampin and minocycline as well as in vitro testing were performed on the device: including elution/extraction, time kill studies, drug content and impurity, and drug release kinetics.
In in vitro studies, ARISTE AB Mesh demonstrated greater than 4 Logs of antibacterial activity against a panel of gram positive and gram-negative organisms including methicillin-resistant Staphylococcus aureus (MRSA), Staphylococcus epidermidis, Streptococcus pyogenes, Escherichia coli, Acinetobacter baumannii, and Enterobacter aerogenes for the first 2 days following implantation.
### Animal Testing
The following GLP animal performance studies were performed to confirm safety and effectiveness in clinically relevant animal models.
- Tensile Strength Study in a Rabbit Abdominal Side Wall Defect Model "T-Peel Study"
- 7-Day Rabbit MRSA Inoculation Study
- 7-Day Rabbit MRSA and E. Coli Inoculation Study
In vivo animal inoculation confirmed clearance of clinically relevant organisms, Staphylococcus aureus (MRSA) and Escherichia coli for a period of 7 days.
# CONCLUSION
The ARISTE AB Mesh is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.