K133223 · C.R. Bard, Inc. · PIJ · Sep 19, 2014 · General, Plastic Surgery
Device Facts
Record ID
K133223
Device Name
XENMATRIX AB SURGICAL GRAFT
Applicant
C.R. Bard, Inc.
Product Code
PIJ · General, Plastic Surgery
Decision Date
Sep 19, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The XenMatrix™ AB Surgical Graft is intended for implantation to reinforce soft tissue where weakness exists and for surgical repair of damaged or ruptured soft tissue, including: abdominal plastic and reconstructive surgery; muscle flap reinforcement; hernia repair including abdominal, inguinal, femoral, diaphragmatic, scrotal, umbilical, and incisional hernias. The Rifampin and Minocycline coating has been shown in preclinical in vitro and in vivo testing to reduce or inhibit microbial colonization on the device. The claim of reduction of bacterial colonization of the device has not been established with human clinical data, nor has a clinical impact associated with this claim been demonstrated.
Device Story
XenMatrix™ AB Surgical Graft is an acellular, sterile, non-pyrogenic porcine dermal matrix used for soft tissue reconstruction. Device is supplied dry and requires hydration in sterile saline (5-40 minutes) prior to implantation. Surfaces are coated with a bioresorbable L-tyrosine succinate polymer containing antimicrobial agents Rifampin and Minocycline (180 µg/cm² each) to inhibit microbial colonization. Used by surgeons in clinical settings for hernia repair and soft tissue reinforcement. Antimicrobial coating is intended to protect the graft from bacterial colonization; however, clinical impact of this reduction has not been established in humans. Device provides mechanical support for tissue healing; antimicrobial agents elute during hydration and post-implantation.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing (thickness, stiffness, burst strength, suture pullout, tear resistance) and in vivo porcine/rabbit studies. Porcine study (28-day) evaluated mechanical strength, tissue in-growth, and histology. Rabbit infection models demonstrated reduced bacterial colonization compared to predicate. All testing confirms substantial equivalence to predicates; clinical impact of antimicrobial colonization reduction not established.
Technological Characteristics
Acellular porcine dermal matrix; 1.5-2.3mm thickness. Coated with bioresorbable L-tyrosine succinate polymer containing Rifampin and Minocycline (180 µg/cm²). Biocompatibility per ISO 10993. Hydration required prior to use. Non-electronic, mechanical implant.
Indications for Use
Indicated for patients requiring soft tissue reinforcement or repair of damaged/ruptured soft tissue, including abdominal plastic/reconstructive surgery, muscle flap reinforcement, and various hernia repairs (abdominal, inguinal, femoral, diaphragmatic, scrotal, umbilical, incisional).
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Ventrio™ Light Hernia Patch with TRM Antimicrobial Coating (K113229)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of a stylized eagle or bird with three human profiles incorporated into its design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 19, 2014
Davol Incorporated Mr. Tony John, MS Regulatory Affairs Specialist 100 Crossings Boulevard Warwick, Rhode Island 02886
Re: K133223
Trade/Device Name: XenMatrix™ AB Surgical Graft Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: PIJ, FTM, OXH Dated: August 6, 2014 Received: August 7, 2014
Dear Mr. John:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### INDICATION FOR USE STATEMENT
510(k) Number (if known): K133223
Device Name: XenMatrix™ AB Surgical Graft
Intended for implantation to reinforce soft tissue where weakness exists and for surgical repair of damaged or ruptured soft tissue, including: abdominal plastic and reconstructive surgery; muscle flap reinforcement; hernia repair including abdominal, inguinal, femoral, diaphragmatic, scrotal, umbilical, and incisional hernias. The Rifampin and Minocycline coating has been shown in preclinical in vitro and in vivo testing to reduce or inhibit microbial colonization on the device. The claim of bacterial colonization of the device has not been established with human clinical data, nor has a clinical impact associated with this claim been demonstrated.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Davol Inc. Subsidiary of C.R. Bard, Inc. 100 Crossings Blvd. Warwick, RI 02886 (401) 463-7000
Image /page/3/Picture/1 description: The image contains the logo for BARD Davol Inc. The logo is green and features the word "BARD" in a stylized font, with "Davol INC." written below it in a smaller font. To the right of the logo is a vertical line, followed by the words "TECHNOLOGY", "TECHNIQUE", "TRAINING", and "TRUST" stacked vertically in a smaller font.
#### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) Summary is provided per the requirements of section 807.92(c).
## Submitter Information:
| Company Name: | Davol Inc., Subsidiary of C. R. Bard, Inc. |
|------------------------|--------------------------------------------|
| Company Address: | 100 Crossings Boulevard |
| | Warwick, RI 02886 |
| Telephone: | (401) 825-8692 |
| Fax: | (401) 825-8765 |
| Submitter's Name: | Tony John, MS |
| | Regulatory Affairs Specialist |
| Date Summary Prepared: | August 25, 2014 |
## Device Identification:
| Trade Name: | XenMatrix™ AB Surgical Graft |
|----------------------|---------------------------------------|
| Common/Usual Name: | Surgical Mesh |
| Classification Name: | Mesh, Surgical<br>Antimicrobial Agent |
| Device Class: | Class II |
| Regulation Number: | 21 CFR § 878.3300 |
| Product Code: | FTM |
#### Predicate Device Names:
- Porcine Dermal Matrix Surgical Mesh, K081272 (Brennen Medical, LLC), ● FDA cleared on 31 July, 2008 - Currently marketed by Davol as XenMatrix™ Surgical Graft
- SURGISIS® Biodesign Tissue Graft, K073391 (Cook Biotech Incorporated), . FDA cleared on 21 March, 2008
- Ventrio™ Light Hernia Patch with TRM Antimicrobial Coating, K113229 . (Davol, Inc.), FDA cleared on 20 July, 2012
#### Device Description:
The XenMatrix™ AB Surgical Graft is an acellular, sterile, non-pyrogenic porcine dermal matrix for use in the reconstruction of soft tissue deficiencies. The graft is packed dry and must be hydrated in sterile saline prior to use. The thickness of the device is 1.5 to 2.3mm.
## PREMARKET NOTIFICATION FOR XENMATRIX™ AB SURGICAL GRAFT
#### SECTION 7
{4}------------------------------------------------
Product sizes when hydrated are:
- . 6x6cm
- 6 x 10 cm ●
- 6 x 16 cm
- 8 x 8 cm ●
- 10 x 10 cm
- 10 x 15 cm ●
- 15 x 20 cm ●
- . 19 x 28 cm
- 19 x 35 cm ●
- 10 x 20 cm ●
- 20 x 20 cm
- 20 x 25 cm
- 10 x 28 cm .
- 15 x 25 cm
The XenMatrix™ AB Surgical Graft surfaces are coated with an antimicrobial coating, which is comprised of a bioresorbable L-tyrosine succinate polymer and antimicrobial agents Rifampin and Minocycline at 180 µg/cm each. The coating is shaded orange in color due to the color of the antimicrobial agents. The bioresorbable L - Tyrosine succinate polymer is essentially absorbed in 12 months based on in vitro studies.
The antimicrobial coating present on this device is intended to protect the graft from bacterial colonization. In preclinical studies the coating has been shown to reduce or inhibit microbial colonization of the device (please see performance data, section 5., below). The claim of reduction of bacterial colonization of the device has not been established with human clinical data, nor has a clinical impact associated with this claim been demonstrated.
# Intended Use:
The XenMatrix™ AB Surgical Graft is intended for implantation to reinforce soft tissue where weakness exists and for surgical repair of damaged or ruptured soft tissue, including: abdominal plastic and reconstructive surgery; muscle flap reinforcement; hernia repair including abdominal, inguinal, femoral, diaphragmatic, scrotal, umbilical, and incisional hernias. The Rifampin and Minocycline coating has been shown in preclinical in vitro and in vivo testing to reduce or inhibit microbial colonization on the device. The claim of reduction of bacterial colonization of the device has not been established with human clinical data, nor has a clinical impact associated with this claim been demonstrated.
# Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Use:
The XenMatrix™ AB Surgical Graft is similar in intended use and in technological characteristics and performance when compared to the cited predicate devices. XenMatrix™ AB and the cited predicates are intended for use in the reconstruction and repair of soft tissue deficiencies where weakness exists such as hernia repair. In addition, these devices are similar in technological characteristics with regard to sterilization, packaging and labeling.
The XenMatrix™ AB Surgical Graft and the predicate Porcine Dermal Matrix Surgical Mesh (K081272, marketed by Davol as XenMatrix™) are based on the same mesh matrix. The XenMatrix™ AB device contains a higher concentration, 180 µg/cm² (each drug) of the same antibacterial drug coating as the Ventrio Light Hernia Patch with TRM
> PREMARKET NOTIFICATION FOR XENMATRIX™ AB SURGICAL GRAFT SECTION 7
{5}------------------------------------------------
Antimicrobial Coating (K113229). The Ventrio Light Hernia Patch device is a nonresorbable, synthetic surgical mesh containing 115 ug/cm- per Rifampin and Minocycline (each drug). Preparation, i.e., hydration, rinsing, of the XenMatrix™ AB device prior to use elutes Rifampin and Minocycline to equivalent concentrations as the comparator predicate surgical mesh. The instructions for use cite the specific hydration steps, i.e., submersion of the device in room temperature sterile water for at least 5 minutes but not exceeding 40 minutes, prior to surgical implantation.
The device differs from the SURGISIS Biodesign Tissue Graft (K073391) and Ventrio Light Hernia Patch with TRM Antimicrobial Coating (K113229) in material substrate.
Both XenMatrix™ AB and SURGISIS are based on porcine derived material. SURGISIS does not have any antimicrobial coating.
Where technological differences exist between the device and the predicate devices, performance testing indicates that the devices are substantially equivalent.
# Performance Data:
The XenMatrix™ AB surgical mesh was evaluated in the following assessments:
- 1. Biocompatibility testing
Biocompatibility testing in accordance to the current ISO 10993 series was conducted on the finished device and the results indicate that the device is biocompatible per these standards.
- 2. Bench testing
Bench testing in accordance with the FDA Guidance for the Preperation of a Premarket Notification Application for a Surgical Mesh was conducted comparing the XenMatrix™ AB device with the cited predicates. The testing included the following:
- a. Physical Characteristics:
- i. Device Thickness
- ii. Device (Flexural) Stiffness
- b. Functional Characteristics:
- i. Burst Strength
- ii. Suture Pullout Strength
- iii. Tear Resistance
Results demonstrate that the physical and functional characteristics of XenMatrix™ AB Surgical Mesh are substantially equivalent to the predicate comparator devices.
## 3. In vivo strength determinations
The device was implanted in a 28 day, porcine evaluation and tested for strength characteristics in comparison to the cited predicate device at Time zero (To) and Day 28 (D28). This study assessed the following characteristics of the implanted mesh:
## PREMARKET NOTIFICATION FOR XENMATRIX™ AB SURGICAL GRAFT SECTION 7
{6}------------------------------------------------
- a. Mechanical Testing
- i. Tensile Testing
- ii. Tissue in-growth Testing
- iii. Device Burst Testing
- b. Percent Area Contracture
- c. Peritoneal Tissue Attachments
- d. Histology
Results demonstrate that the in vivo performance of XenMatrix™ AB Surgical Mesh is substantially equivalent to the predicate device.
# 4. Drug Content and Impurities of the Antimicrobial Agents Rifampin and Minocycline
Analytical and in vitro testing was also performed on the device and included speed to kill, kinetic drug release (KDR), drug content and impurity, and polymer degradation testing. The test results demonstrated substantial equivalence to the predicate comparator surgical mesh with respect to these parameters.
- 5. Animal Testing
An in vivo porcine implantation study was performed to investigate the device's mechanical strength and the host inflammatory response to the device over a 28 day duration. At 28 days, the XenMatrix™ AB surgical mesh had greater tissueingrowth/T-Peel Force values than the surgical mesh predicate. .The mechanical strength values (i.e., Ultimate Load/Burst Force, Peak Tensile Strength) of the graft alone were lower than the surgical mesh predicate .
In addition, two in vivo dorsum-implanted rabbit infection model studies were performed. Devices were inoculated with bacteria at implantation, and at 7 days, post-implantation, bacterial colonization quantifications were conducted. At that time point the antimicrobial coating on the XenMatrix™ AB was observed to prevent bacterial colonization of the device in comparison to the surgical mesh predicate device.
The relevance of these studies to human clinical performance outcomes has not been demonstrated.
The correlation of these studies has not been demonstrated to be predictive of positive human clinical outcomes.
- 6. Human clinical data None.
The claim of reduction of colonization has not been established with human clinical data, nor has a clinical impact associated with this claim been demonstrated.
## PREMARKET NOTIFICATION FOR XENMATRIX™ AB SURGICAL GRAFT SECTION 7
{7}------------------------------------------------
# Conclusion:
Results of the testing performed in support of this submission demonstrate that the XenMatrix™ AB Surgical Graft is substantially equivalent to currently marketed predicate devices. The claim of reduction of bacterial colonization of the device has not been established with human clinical data, nor has a clinical impact associated with this claim been demonstrated.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.