C-Qual M™ Breastboard; Monarch™ Overhead Arm Positioner
K211001 · Medtec, Inc. Dba Civco Medical Solutions and Civco · IYE · May 20, 2021 · Radiology
Device Facts
Record ID
K211001
Device Name
C-Qual M™ Breastboard; Monarch™ Overhead Arm Positioner
Applicant
Medtec, Inc. Dba Civco Medical Solutions and Civco
Product Code
IYE · Radiology
Decision Date
May 20, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
The device is indicated to aid in supporting and positioning adult and adolescent patients undergoing radiation therapy of the breast and chest region including electron, photon, and proton treatments. Device is also used to position the patient during image acquisition to support treatment planning including in Computed Tomography (CT) and Magnetic Resonance (MR) imaging systems. The device is not intended for use with patients under 12 years of age.
Device Story
Patient positioning system for radiation therapy and imaging; consists of base, back support, bottom stop, and overhead arm positioner. Supports patient in supine position; allows angular adjustments (5° to 25°) for back support; provides arm support and head cushioning. Used in healthcare facilities by clinical staff. Facilitates patient immobilization during CT and MR imaging for treatment planning. Benefits include stable, reproducible patient positioning for radiation delivery and image acquisition. Device is reusable and non-sterile.
Clinical Evidence
No clinical data. Bench testing only. MR safety and compatibility testing performed per ASTM F2052-15, F2213-06, and F2119-07; device demonstrated safety at 1.5T and 3.0T. Biocompatibility testing performed per ISO 10993-5:2009 and ISO 10993-10:2010 for limited contact duration (<24 hours).
Technological Characteristics
Mechanical patient positioning assembly; includes base, back support, bottom stop, and overhead arm positioner. Materials evaluated for biocompatibility (ISO 10993). MR safety tested per ASTM F2052-15, F2213-06, F2119-07. Non-sterile, reusable. No electronic or software components.
Indications for Use
Indicated for adult and adolescent patients (12+ years) requiring support and positioning during radiation therapy (electron, photon, proton) or image acquisition (CT, MR) for the breast and chest region.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
Proton Positioning and Immobilization Devices (Breast Devices) (K180021)
Submission Summary (Full Text)
{0}------------------------------------------------
May 20, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
MEDTEC, Inc. dba CIVCO Medical Solutions and CIVCO Radiotherapy % Ms. Alena Newgren Regulatory Manager 1401 8th Street SE ORANGE CITY IA 51041
### Re: K211001
Trade/Device Name: C-Qual M" Breastboard; Monarch™ Overhead Arm Positioner Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: IYE, LNH Dated: March 31, 2021 Received: April 2, 2021
Dear Ms. Newgren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K211001
#### Device Name
C-Qual M Breastboard with Monarch Overhead Arm Positioner
Indications for Use (Describe)
The device is indicated to aid in supporting adult and adolescent patients undergoing radiation therapy of the breast and chest region including electron, photon treatments. Device is also used to position the patient during image acquisition to support treatment planning including in Computed Tomography (CT) and Magnetic Resonance (MR) imaging systems. The device is not intended for use with patients under 12 years of age.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the CIVCO Radiotherapy logo. The logo is blue and features the word "CIVCO" in a stylized font, with a blue circle above the "V". Below the word "CIVCO" is the word "Radiotherapy" in a smaller, sans-serif font. The logo is simple and modern, and the blue color gives it a professional and trustworthy look.
# 510(k) Summary
K211001
## A. Submitter Information
| | Submitter Name & Address: MEDTEC, Inc. d/b/a "CIVCO Medical Solutions" and "CIVCO<br>Radiotherapy"<br>1401 8th St. SE<br>Orange City, IA 51041 |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Alena Krishna, Regulatory Affairs Manager<br>Telephone: 319-248-6650<br>Alena.Newgren@civcort.com |
| Date Summary Prepared: | March 31, 2021 |
| Trade Names: | C-Qual M™ Breastboard; Monarch™ Overhead Arm Positioner |
| Common Names: | Breastboard; Overhead Arm Positioner |
| Classification Names &<br>Numbers: | Medical charged-particle radiation therapy system (892.5050)<br>System, Nuclear Magnetic Resonance Imaging (892.1000) |
| Device Class: | Class II |
| Review Panels: | Radiology |
| Product Codes: | IYE, LNH |
### B. Predicate Devices
The Breastboard and Overhead Arm Positioner are substantially equivalent to the following predicate device:
| Predicate Device | Manufacturer |
|----------------------------------------------------------------------------|--------------------|
| K180021: Proton Positioning and Immobilization Devices<br>(Breast Devices) | CIVCO Radiotherapy |
The purpose of this 510(k) is to add MR to the indications for use for the MTM410.
## C. Device Descriptions
The Breastboard and Overhead Arm Positioner are used to position the patient during external beam radiation therapy. The Breastboard is comprised of a Base, Back Support, and Bottom Stop. The Base serves as the foundation of the assembly and provides a mounting surface for the Back Support, Angle Supports, and the Bottom Stop. The inferior end of the Back Support
## www.civcort.com
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the CIVCO Radiotherapy logo. The logo is blue and features the word "CIVCO" in a stylized font, with a blue circle above the "V". Below the word "CIVCO" is the word "Radiotherapy" in a smaller, sans-serif font. The logo is simple and modern, and the blue color gives it a professional and trustworthy feel.
is mounted to the Base while the superior end of the Back Support is raised/lowered as needed for patient setup using Low Angle Supports. The Bottom Stop is located near the inferior end of the Base and prevents the patient from sliding down the Board.
The Low Angle Support is used to position the Back Support at 5°, 7.5°, 10°, 12.5°, and 15° anqles. Features on the High Angle Support receive the cutouts on the Low Angle Support and hold the anqular position of the Support.
The High Angle Support is used to position the Back Support at 17.5°, 20°, 22.5°, and 25° angles. Features on the Base receive the cutouts on the High Angle Support and hold the angular position of the Support.
A Lock provides a method of locking the Back Support to the Base at 0° to facilitate easier handling and storage of the Breastboard.
The Overhead Arm Positioner provides support for the patient's arms above the head and is attached to the Back Support via a locking assembly. The Overhead Arm Positioner adjusts longitudinally on the Back Support to accommodate a variety of patient sizes. A Hand Grip is mounted on the Overhead Arm Positioner and the location of the Hand Grip is adjustable longitudinally.
A Head Support rests on a recessed area of the Overhead Arm Positioner and provides a cushion for positioning of the head of the patient. The Head Support can also be mounted directly on the Breastboard. A Thermoplastic Frame may also be mounted to the Overhead Arm Positioner to enable attachment of a thermoplastic mask for additional patient support.
The Breastboard and Overhead Arm Positioner can also be used for supporting and elevating patients undergoing treatment who cannot lay down flat on the table due to their habitus.
The Breastboard and Overhead Arm Positioner are reusable devices that are provided nonsterile. The devices are used in a healthcare facility/hospital. The following model is included in this submission:
| Part No. | Device Name |
|----------|---------------------------------------------------------------|
| MTM410 | C-Qual MRTM Breastboard and MonarchTM Overhead Arm Positioner |
### D. Indications for Use/Intended Use Statements
Indications for Use/Intended Use: The device is indicated to aid in supporting and positioning adult and adolescent patients undergoing radiation therapy of the breast and chest region including electron, photon, and proton treatments. Device is also used to position the patient during image acquisition to support treatment planning including in Computed Tomography
### www.civcort.com
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the CIVCO Radiotherapy logo. The logo is blue and features the word "CIVCO" in a stylized font. A blue circle is placed above the V. Below the word "CIVCO" is the word "Radiotherapy" in a smaller font.
(CT) and Magnetic Resonance (MR) imaging systems. The device is not intended for use with patients under 12 years of age.
### E. Comparison of Technological Characteristics
There are no technological differences between the predicate and proposed device. Both the proposed and predicate Breastboard are comprised of a Base, Back Support, and Bottom Stop. Both devices consist of High Angle and Low Angle Supports that allow elevation from 5 to 25° in 2.5 degree increments. The Overhead Arm Positioner adjusts longitudinally on the Back Support via a locking assembly.
### F. Non-Clinical Testing
To add MR indications to the Breastboard and Overhead Arm Positioner (MTM410), the device was tested for MR safety and compatibility in accordance with ASTM Standard F2052-15. Testing was also completed using ASTM Standards F2213-06 and F2119-07 as guidance. The device passed the acceptance criteria for magnetically induced torque and magnetically induced displacement force and demonstrate that the device is safe for use in field strengths of 1.5 T and 3.0 T. Image artifact was observed, with the worst case artifact occurring near the location of the locking knob on the Breastboard. Information regarding size of this artifact has been included in the Instructions for Use. In addition to MR safety testing, other nonclinical testing demonstrated that the differences in technological characteristics did not raise any new issues of safety or effectiveness.
Biocompatibility testing was completed for patient-contacting materials in accordance with ISO 10993-5:2009 and ISO 10993-10:2010. The Breastboard and Overhead Arm Positioner is intended for limited contact duration (<24 hours) for surface devices (skin).
### G. Conclusion
This premarket submission for the Breastboard and Overhead Arm Positioner has demonstrated substantial equivalence as defined and understood in the Federal Food. Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.
### www.civcort.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.