K210976 · Meridian Bioscience, Inc. · LQP · Dec 23, 2021 · Microbiology
Device Facts
Record ID
K210976
Device Name
Curian Campy
Applicant
Meridian Bioscience, Inc.
Product Code
LQP · Microbiology
Decision Date
Dec 23, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3110
Device Class
Class 1
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K210976 · Dec 23, 2021
Curian Campy
Meridian Bioscience, Inc.
Archived clinical stool specimens
Archived clinical specimens were used to supplement the prospective clinical study to provide additional performance data (sensitivity and specificity) for the device.
Archived specimens; Retrospective clinical data
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Archived Specimens Study; Retrospective analysis of archived clinical specimens
Patients with signs and symptoms of gastroenteritis; Sample Size: 290 specimens (29 positive, 261 negative); Number of Sites: 5
Culture and speciation
Clinical sensitivity and specificity
Indications for Use
Curian Campy, for use with the Curian Analyzer, is a rapid, qualitative fluorescent immunoassay for the detection of a Campylobacter-specific antigen in human fecal specimens. Curian Campy is intended to detect C. jejuni, C. coli, C. upsaliensis, and C. lari in human stool from patients with signs and symptoms of gastroenteritis. The test is intended for use with unpreserved fecal specimens or preserved fecal specimens in transport media. Test results are to be used in conjunction with information available from the patient clinical evaluation and other diagnostic procedures. Curian Campy is intended to aid in the diagnosis of Campylobacter infection.
Device Story
Curian Campy is a rapid, qualitative fluorescent immunoassay for detecting Campylobacter-specific antigens in human stool. It utilizes the Curian Analyzer (K192817) to process samples and automate result interpretation. The device is used in clinical settings by healthcare professionals to aid in diagnosing Campylobacter infection in patients with gastroenteritis. Stool samples are collected via transfer pipette or a specific collection brush for non-pipettable specimens. The assay detects C. jejuni, C. coli, C. upsaliensis, and C. lari. The analyzer provides objective, automated results, reducing the subjectivity associated with visual interpretation of traditional lateral flow or ELISA tests. Results are intended to be used alongside clinical evaluation and other diagnostic procedures to inform patient management.
Clinical Evidence
Clinical performance was evaluated via a prospective study (n=1,474) and an archived study (n=290) against culture and speciation. In the prospective study, sensitivity was 85.7% (95% CI: 65.4-95.0%) and specificity was 98.1% (95% CI: 97.2-98.7%). In the archived study, sensitivity was 96.6% (95% CI: 82.8-99.4%) and specificity was 98.1% (95% CI: 95.6-99.2%). A contrived study (n=210) showed 100% PPA and 100% NPA. Analytical performance included precision, LoD determination, cross-reactivity, and interference testing.
Technological Characteristics
Lateral flow fluorescent immunoassay; utilizes monoclonal/polyclonal antibodies. Single-use cassette format. Requires Curian Analyzer for signal detection and automated interpretation. Compatible with unpreserved stool or stool in Cary-Blair/C&S transport media. Refrigerated storage (2-8 °C).
Indications for Use
Indicated for patients with signs and symptoms of gastroenteritis to detect Campylobacter-specific antigens (C. jejuni, C. coli, C. upsaliensis, and C. lari) in human stool specimens (unpreserved or preserved in transport media).
Regulatory Classification
Identification
Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identify Campylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases. Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K210976
B Applicant
Meridian Bioscience, Inc.
C Proprietary and Established Names
Curian Campy
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LQP | Class I, reserved | 21 CFR 866.3110 - Campylobacter fetus Serological Reagents | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain a substantial equivalence determination for the detection of Campylobacter jejuni, Campylobacter coli, Campylobacter lari, and Campylobacter upsaliensis antigens in human stool.
B Measurand:
Campylobacter jejuni, Campylobacter coli, Campylobacter lari, and Campylobacter upsaliensis antigens.
C Type of Test:
Lateral flow fluorescent immunoassay (FIA)
## III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
Curian Campy, for use with the Curian Analyzer, is a rapid, qualitative fluorescent immunoassay for the detection of a Campylobacter-specific antigen in human fecal specimens. Curian Campy is intended to detect C. jejuni, C. coli, C. upsaliensis, and C. lari in human stool from patients with signs and symptoms of gastroenteritis. The test is intended for use with unpreserved fecal specimens
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
or preserved fecal specimens in transport media. Test results are to be used in conjunction with information available from the patient clinical evaluation and other diagnostic procedures. Curian Campy is intended to aid in the diagnosis of Campylobacter infection.
## C Special Conditions for Use Statement(s):
- Rx - For Prescription Use Only
- The Curian Campy was evaluated using only fresh fecal samples and fecal samples stored in Cary Blair media or C&S media. The performance of fecal samples stored in other transport media (e.g., formalin, polyvinyl alcohol) has not been evaluated and therefore, should not be used.
- No data exists on the effects of colonic washes, barium enemas, laxatives, or bowel preparations on the performance of the Curian Campy. All of these procedures can result in extensive dilution or the presence of additives that may affect test performance.
- The fluorescence signal obtained with this assay is invisible to the unaided eye. The test results can only be obtained with the propose use of the Curian Analyzer.
## D Special Instrument Requirements:
The Curian Campy assay can only be run on the Curian Analyzer.
## IV Device/System Characteristics:
### A Device Description:
Curian Campy, for use with the Curian Analyzer, is a rapid, qualitative, fluorescent immunoassay for the detection of Campylobacter spp. (C. jejuni/C. coli/C. upsaliensis/C. lari) in human stool, where stool may be either unpreserved or preserved in C&S or Cary-Blair transport media. The Curian Campy assay consists of a test strip enclosed in a plastic frame (test card), positive control reagent, and an assay specific Aioprep sample preparation device. Curian Campy is a lateral flow-based immunoassay for the direct detection of Campylobacter antigens in human stool. Curian Campy utilizes monoclonal antibodies, specific for an antigen common to C. jejuni, C. coli, C. lari and C. upsaliensis, as the capture and detector antibodies.
The Aioprep is pre-filled with blue tinted Sample Diluent and contains a filter and a sample metering device. A sample of the patient's stool specimen is transferred from the collection container to the Aioprep, using either the included transfer pipette (for preserved or unpreserved, liquid/semi-solid specimens) or the assay specific sample collection brush (for preserved or unpreserved, formed/solid specimens) which is available separately as an accessory (not required for use). When using the transfer pipette, the sample is added directly into the Sample Diluent within the Aioprep. When using the sample collection brush, the brush with sample is inserted through the metering device to remove excess stool and then is pushed directly into the Sample Diluent within the Aioprep sample preparation device.
The diluted sample is mixed and dispensed dropwise into the sample port of the Curian Campy test card. If present, the Campylobacter antigen binds to the monoclonal detector antibody conjugated to fluorescent particles, forming a complex. As the sample moves through the test strip, the anti-Campylobacter capture antibody, bound to the assay membrane at the test position of the strip, binds the complex and yields a test line. When antigen is not present, a complex is not formed, and a test line will not occur. As the sample continues to move further up the test strip, the polyclonal capture antibody, bound to the assay membrane at the control position of the strip, binds the conjugated antibody and yields a control line. A line at the control position of the test strip should be present
K210976 - Page 2 of 13
{2}
each time a sample or external control is tested. If the control line is not generated, adequate sample flow has not occurred, and the Curian Analyzer will consider the test invalid.
The Curian Analyzer can be run in either "Incubate and Analyze" mode or "Analyze Now" mode. If operating the Curian Analyzer in the "Incubate and Analyze" mode, the test card is inserted into the Curian Analyzer, after which all further steps are automated. If operating the analyzer in "Analyze Now" mode, the test card is incubated at 19-27°C on the benchtop for 20 minutes before insertion into the Curian Analyzer. If a patient specimen cannot be tested immediately/as soon as possible after collection, unpreserved and preserved specimens (in Meridian Para-Pak C&S media) may be held for up to 7-days at 2-8°C. Alternatively, unpreserved or preserved specimens (in Meridian Para-Pak C&S media) may be frozen for up to 59 days at ≤-20°C prior to testing.
B Principle of Operation:
See Section IV.A above.
V Substantial Equivalence Information:
A Predicate Device Name(s):
CAMPYLOBACTER QUIK CHEK
B Predicate 510(k) Number(s):
K191456
C Comparison with Predicate(s):
| Similarities | | |
| --- | --- | --- |
| Item | DEVICE
Curian Campy
K210976 | PREDICATE
CAMPYLOBACTER QUIK CHEK
K191456 |
| Indications For Use | Curian Campy, for use with the Curian Analyzer, is a rapid, qualitative fluorescent immunoassay for the detection of a Campylobacter-specific antigen in human fecal specimens. Curian Campy is intended to detect C. jejuni, C. coli, C. upsaliensis, and C. lari in human stool from patients with signs and symptoms of gastroenteritis. The test is intended for use with unpreserved fecal specimens or preserved fecal specimens in transport media. Test results are to be used in conjunction with information available from the patient clinical evaluation and other diagnostic procedures. Curian Campy is intended to aid in the diagnosis of Campylobacter infection. | The CAMPYLOBACTER QUIK CHEK test is a rapid membrane enzyme-linked immunosorbent assay for the qualitative detection of a Campylobacter-specific antigen in human fecal specimens. The CAMPYLOBACTER QUIK CHEK test is designed to detect C. jejuni, C. coli, C. lari, and C. upsaliensis from patients with signs and symptoms of gastroenteritis. The test is intended for use with preserved fecal specimens in transport media and unpreserved fecal specimens. Test results should be considered in conjunction with clinical findings and patient history. |
K210976 - Page 3 of 13
{3}
| Classification | Class I | Same |
| --- | --- | --- |
| Product Code | LQP | Same |
| Regulation | 21 CFR 866.3110 | Same |
| Measured Analyte | Campylobacter-specific antigen | Same |
| Antibody Format | Monoclonal/Polyclonal | Same |
| Type of Test | Qualitative | Same |
| Sample Matrix | Human fecal specimen | Same |
| Specimen Type | Unpreserved fecal specimens and fecal specimens in Cary-Blair and C&S transport media | Same |
| Controls | Positive and negative control included in the kit. Internal control line | Same |
| Species Detected | C. jejuni, C. coli, C. lari, and C. upsaliensis | Same |
| Read Result Time | < 30 minutes | Same |
| Format | Single use cassette | Same |
| Kit Storage | Refrigerated (2-8°C) | Same |
| Differences | | |
| Item | DEVICE
Curian Campy
K210976 | PREDICATE
CAMPYLOBACTER QUIK CHEK
K191456 |
| Technology | Lateral flow fluorescent immunoassay (FIA) | Membrane Enzyme Linked Immunosorbent Assay (ELISA) |
| Result Interpretation Method | Result interpretation automated by Curian Analyzer | Visually Read |
## VI Standards/Guidance Documents Referenced:
- BS EN ISO 23640:2015 In vitro diagnostic medical devices - Evaluation of stability of in vitro diagnostic reagents
- ISO 14971 Third edition. 2019-12 Medical devices – Application of risk management to medical devices
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
All analytical studies were conducted in “Analyze Now” mode. The primary difference between these modes is that in “Analyze Now” mode, the operator is responsible for the incubation time; in the “Incubate and Analyze” mode, the instrument will wait to read the strip until the specified development time has passed. A validation study was conducted to demonstrate that the “Analyze Now” mode does not impact the results of the analytical studies in the conditions tested. See section VII.A.3. “Analyze Now” vs. “Incubate and Analyze” Read Mode Study for more information and the results of this study.
K210976 - Page 4 of 13
{4}
1. Precision/Reproducibility:
Reproducibility of the Curian Campy assay was assessed across three testing sites, one of which was internal. Two sample panels, one consisting of unpreserved (raw) stool and one consisting of preserved stool (in C&S transport media) were evaluated. Panels were created by spiking *C. jejuni* (strain ATCC #BAA-1234) whole organism into unpreserved and preserved pooled stool negative matrix at the following concentrations: negative, high negative (just below $C_5$), low positive ($\sim C_{95}$), and moderate positive (3x higher than the $C_{95}$). Ten blinded sample panels per stool matrix (i.e., unpreserved or preserved), each consisting of 16 samples per panel (1 negative, 5 high negatives, 5 low positives, and 5 moderate positives) were provided to each site for a total of 160 unpreserved stool samples and 160 preserved stool samples per site. Testing was conducted at each site twice a day over 5 non-consecutive days by two operators per day/site and included three different kit lots (2 lots per site). Positive and negative controls were run with each panel of masked samples. The results for unpreserved and preserved stool are shown in Table 1.
Table 1. Reproducibility Study Results
| Unpreserved Stool | | | | | |
| --- | --- | --- | --- | --- | --- |
| Panel Member | Percent Agreement with Expected Results | | | | |
| | Site 1 | Site 2 | Site 3 | Total | 95% CI |
| Moderate Positive | 100% (50/50) | 100% (50/50) | 100% (50/50) | 100% (150/150) | 97.5, 100% |
| Low Positive | 100% (50/50) | 82% (41/50) | 66% (33/50) | 82.7% (124/150) | 75.8, 87.9% |
| High Negative | 96% (48/50) | 100% (50/50) | 100% (50/50) | 98.7% (148/150) | 95.3, 99.6% |
| Negative | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (30/30) | 88.7, 100% |
| Preserved (C&S) Stool | | | | | |
| Panel Member | Site 1 | Site 2 | Site 3 | Total | 95% CI |
| Moderate Positive | 100% (50/50) | 100% (50/50) | 100% (50/50) | 100% (150/150) | 97.5, 100% |
| Low Positive | 100% (50/50) | 100% (50/50) | 100% (50/50) | 100% (150/150) | 97.5, 100% |
| High Negative | 90% (45/50) | 98.0% (48/49)¹ | 98% (49/50) | 95.3% (142/149) | 90.6, 97.7% |
| Negative | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (30/30) | 88.7, 100% |
¹ A single replicate was not performed because the Curian Analyzer inadvertently shut down and there was insufficient volume for a repeat test.
For the unpreserved low positive panel member, a low Percent Agreement (PA) with expected results was observed (82.7%). Upon review of the study information, it was determined that the PA for the unpreserved low positive sample was due to operator under-sampling with the Curian Campy transfer pipette. To mitigate the risk presented by under-sampling, the package insert instructions were revised to provide additional clarity regarding unpreserved specimen handling/processing instructions.
2. Linearity:
Not applicable
K210976 - Page 5 of 13
{5}
# 3. Analytical Specificity/Interference:
# Cross-Reactivity and Microbial Interference
The Curian Campy assay was evaluated for cross-reactivity with the bacteria, fungi, and viruses listed in Table 2 below. Bacteria and fungi were tested at $\geq 1.2\mathrm{x}10^{7}$ CFU/mL while all viruses, except Norovirus, were tested at $\geq 1.0\mathrm{x}10^{5}$ TCID $_{50}$ /mL. For Norovirus, five positive preserved clinical stool specimens were evaluated for cross-reactivity. Norovirus cross-reactivity testing was not performed with unpreserved stool. Testing for all other microorganisms was performed with both unpreserved and preserved (in C&S transport media) stool matrices. Each non-target microorganism was tested in triplicate.
For microbial interference testing, C. jejuni (strain ATCC #BAA-1234) whole organism was spiked at 2x LoD in both unpreserved and preserved (C&S) stool matrix containing non-target microorganisms (see Table 2). For Norovirus, unpreserved stool matrix was not evaluated; rather, the preserved clinical stool specimens described above were tested in the presence of C. jejuni at 2x LoD. Testing was performed in triplicate.
Table 2. Microorganisms Evaluated for Cross-Reactivity
| Bacteria/Fungi | |
| --- | --- |
| Acinetobacter baumannii | Klebsiella pneumoniae |
| Aeromonas hydrophila | Lactobacillus acidophilus |
| Bacillus cereus | Lactococcus lactis |
| Bacillus subtilis | Listeria monocytogenes |
| Bacteroides fragilis | Peptostreptococcus anaerobius |
| Campylobacter concisus | Plesiomonas shigelloides |
| Campylobacter fetus | Porphyromonas asaccharolytica |
| Campylobacter helveticus | Prevotella melaninogenica |
| Campylobacter hyointestinalis | Proteus vulgaris |
| Candida albicans | Pseudomonas aeruginosa |
| Citrobacter freundii | Pseudomonas fluorescens |
| Clostridium bifermentans | Salmonella enterica subs. enterica serovar Hilversum |
| Clostridiodes difficile | Salmonella enterica subs. enterica serovar Typhimurium |
| Clostridium perfringens | Salmonella enterica subs. enterica serovar Minnesota |
| Edwardsiella tarda | Serratia marcescens |
| Enterobacter cloacae | Shigella dysenteriae |
| Enterococcus faecalis | Shigella flexneri |
| Escherichia coli | Shigella sonnei |
| Escherichia coli EIEC | Staphylococcus aureus |
| Escherichia coli EPEC | Staphylococcus aureus (Cowan's) |
| Escherichia coli ETEC | Streptococcus agalactiae |
| Escherichia coli O157:H7 (non-toxigenic) | Staphylococcus epidermidis |
| Escherichia coli O157:H7 (toxigenic) | Steptococcus dysgalactiae subsp. Equisimilis |
K210976 - Page 6 of 13
{6}
K210976 - Page 7 of 13
| Bacteria/Fungi | |
| --- | --- |
| Escherichia fergusonii | Vibrio parahaemolyticus |
| Escherichia hermannii | Yersinia enterolitica |
| Helicobacter pylori | |
| Viruses | |
| Adenovirus Type 1 | Echovirus 11 |
| Adenovirus Type 2 | Echovirus 18 |
| Adenovirus Type 3 | Enterovirus 68 |
| Adenovirus Type 5 | Enterovirus 69 |
| Adenovirus Type 40 | Enterovirus 70 |
| Adenovirus Type 41 | Enterovirus 71 |
| Coxsackie virus B2 | Human coronavirus OC43 |
| Coxsackie virus B3 | Human Rotavirus |
| Coxsackie virus B4 | Norovirus¹ |
| Coxsackie virus B5 | Paraechovirus 1 (formerly Echovirus 22) |
| Echovirus 9 | |
¹ Five Norovirus positive preserved stool clinical specimens were evaluated.
C. helveticus (strain ATCC 51209) was cross-reactive at concentrations greater than $3.75 \times 10^{6}$ CFU/mL for unpreserved stool and $7.50 \times 10^{6}$ CFU/mL for preserved stool. None of the other microorganisms evaluated in this study were cross-reactive with the Curian Campy assay.
## Interfering Substances Study
The Curian Campy test was evaluated for interfering substances with the substances and concentrations listed in Table 3. Testing was performed with both unpreserved and preserved (C&S) stool matrices. None of the substances were shown to interfere with the performance of the Curian Campy test.
Table 3. Interfering Substance Panel
| Barium sulfate (5% w/v) | Mylanta (4.2 mg/mL) |
| --- | --- |
| Benzalkonium chloride (moist towelettes) (1% w/v) | Naproxen sodium (5% w/v) |
| Ciprofloxacin (0.25% w/v) | Nonoxynol-9 (1% w/v) |
| Ethanol (1% w/v) | Nystatin (1% w/v) |
| Hog gastric mucin (3.5% w/v) | Palmitic acid (fecal fat) (40% w/v) |
| Human whole blood (40% v/v) | Pepto-Bismol (5% v/v) |
| Human hemoglobin (10% w/v) | Phenylephrine (1% w/v) |
| Human urine (5% v/v) | Polyethylene glycol 3350 (10% w/v) |
| Hydrocortisone (1% w/v) | Prilosec OTC (5 ug/mL) |
| Imodium AD (5% v/v) | Sennosides (1% w/v) |
| Kaopectate (5% v/v) | Simethicone (10% w/v) |
| Leukocytes (0.05% v/v) | Stearic acid (fecal fat) (40% w/v) |
| Mesalazine (10% w/v) | Tagament (5 ug/mL) |
| Metronidazole (0.25% w/v) | TUMS (50 ug/mL) |
| Mineral Oil (10% w/v) | Vancomycin (0.25% w/v) |
{7}
K210976 - Page 8 of 13
# Prozone Study
The purpose of this study was to demonstrate that a high concentration of Campylobacter (antigen) does not interfere with a positive reaction in the Curian Campy test. High positive samples were prepared by spiking a negative fecal pool with $1.73 \times 10^{8}$ whole organisms of C. jejuni (strain ATCC #BAA-1234). Both unpreserved and preserved (C&S) samples were tested. A total of 7 different dilutions were prepared and evaluated for both unpreserved and preserved stool matrices. Testing was performed according to the Package Insert instructions in replicates of five for each dilution. The results demonstrated that were no overall hook affect, and that elevated levels of whole organism did not affect the detection of Campylobacter.
# "Analyze Now" and "Incubate and Analyze" Read Mode Study
To demonstrate that performance of the Curian Campy is not influenced by operating the device in "Analyze Now" mode under the conditions used for analytical studies, a comparison study was conducted with contrived positive samples prepared by spiking a negative fecal pool with C. jejuni (strain ATCC BAA-1234) at 2-3x LoD. Negative samples and positive clinical specimens were also included in the study. Each sample was tested on the Curian analyzer in "Analyze Now" mode and "Incubate and Analyze" mode.
There was $100\%$ agreement between the expected and observed results for all positive and negative samples. These results demonstrate that there is no notable difference in performance between "Analyze Now" and "Incubate and Analyze" modes in the analytical testing setting. These study results are acceptable and support performing analytical testing solely in "Analyze now" mode.
## 4. Assay Reportable Range:
Not applicable
## 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
### Specimen Stability and Freeze/Thaw Stability
The effect of specimen storage on antigen stability was evaluated for both unpreserved and preserved stool samples. For this analysis, a sample panel consisting of 2 clinical negative specimens (1 unpreserved and 1 preserved) and 4 contrived positive samples (2 unpreserved and 2 preserved) were evaluated. Contrived positive samples were generated by spiking C. jejuni (strain ATCC BAA-1234) whole organism at 2x LoD. After baseline testing was performed, stool samples were stored under the conditions outlined in Table 4 and tested according to the schedule described in the same table.
| Specimen Stability | | |
| --- | --- | --- |
| Stool Matrix | Storage Temperature | Time Interval |
| Preserved Stool | Room temperature (19-27°C) | 0, 49, 73, and 97-hours |
| Unpreserved and Preserved Stool^{1} | Refrigerated (2-8°C) | 0, 49, 97, and 192-hours |
| Unpreserved and Preserved Stool^{2} | Frozen (-20°C and -80°C) | 0, 30, 45, and 60-days |
| Freeze/Thaw Stability | | |
| Stool Matrix | Storage Temperature | # of Freeze/Thaw Cycles Evaluated |
| Unpreserved stool and Preserved Stool^{2} | Frozen (-20°C and -80°C) | 4 and 6 cycles |
{8}
K210976 - Page 9 of 13
$^{1}$ Only Meridian Para-Pak C&S medium was evaluated for refrigerated specimen stability. Other C&S and Cary Blair transport media containing patient stool should be stored according to the recommendations outlined in the respective transport media package insert.
$^{2}$ Only Meridian Para-Pak C&S medium was evaluated for frozen specimen stability and freeze/thaw stability. Other C&S and Cary Blair transport media containing patient stool should be stored according to the recommendations outlined in the respective transport media package insert.
Positive samples remained positive and negative samples remained negative throughout the study. Based on these results, the recommended storage conditions and timeframes are as follows:
## Unpreserved stool
- Up to 168-hours (7-days) refrigerated (2-8°C)
- Up to 59-days frozen (-20°C or -80°C)
- Up to 5 freeze/thaw cycles when stored frozen (-20°C or -80°C)
## Preserved stool
- Up to 96-hours (4-days) at room temperature (19-27°C)
- Up to 168-hours (7-days) refrigerated (2-8°C) for stool in Meridian Par-Pak C&S medium
- Up to 59-days frozen (-20°C or -80°C) for stool in Meridian Para-Pak C&S medium
- Up to 5 freeze/thaw cycles when stored frozen (-20°C or -80°C) in Meridian Para-Pak C&S medium
## Diluted Specimen Hold Time Study
This study was performed to determine storage conditions the Curian Campy Aioprep Sample Preparation Device containing stool, prior to testing with the Curian Campy assay. For this study, the sample panel evaluated during the specimen stability study was analyzed. After baseline testing was performed, samples were stored at room temperature (19-27°C) or refrigerated (2-8°C). For samples stored at room temperature, testing was performed at 5, 9, and 25-hours following storage, while refrigerated samples were tested at 25-hours post-storage. All positive samples were positive, and all negative samples were negative. These results support that stool specimens in the Aioprep may be stored up to 8-hours at room temperature or up to 24-hours refrigerated prior to testing with the Curian Campy assay.
## 6. Detection Limit:
### Limit of Detection Study
The limit of detection (LoD) for the Curian Campy test was determined using whole organisms of C. jejuni, C. coli, C. lari, and C. upsaliensis spiked into unpreserved (raw) stool and preserved stool in Cary Blair and C&S transport media. Three lots of the Curian Campy assay were evaluated in the study. The concentrations were calculated in colony forming units per milliliter (CFU/mL) and their equivalent in CFU/test by factoring in the dilutions and the final volume used in the assay. The LoD is the microorganism concentration that yields a positive result 95% of the time and a negative result 5% of the time. The LoD for each Campylobacter species and stool type are presented in Table 5.
{9}
Table 5. LoD Study Results
| C. jejuni (ATCC #BAA-1234) | | C. coli (ATCC #43484) | | C. upsaliensis (ATCC #49816) | | C. lari (ATCC #43675) | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| CFU/mL | CFU/test | CFU/mL | CFU/test | CFU/mL | CFU/test | CFU/mL | CFU/test |
| Unpreserved Stool Matrix | | | | | | | |
| 4.00x105 | 1818 | 3.00x106 | 13636 | 1.62x106 | 7386 | 5.00x106 | 22727 |
| Preserved (C&S) Stool Matrix | | | | | | | |
| 7.25x105 | 2266 | 1.57x107 | 49063 | 1.18x106 | 3681 | 1.16x107 | 36250 |
| Preserved (Cary Blair) Stool Matrix | | | | | | | |
| 7.25x105 | 2266 | 1.57x107 | 49063 | 2.36x106 | 7375 | 1.16x107 | 36250 |
# Inclusivity Study
The reactivity of the Curian Campy was established using the strains listed in Table 6.
For C. upsaliensis and C. lari reactivity testing, a subset of all isolates were clinical isolates obtained from patients with symptoms of gastroenteritis that were isolated at the Centre National de Reference des Campylobacters et Helicobacters – Universite de Bordeaux between 2013 and 2018. For reactivity testing, each strain was spiked into negative fecal matrix or preserved stool (C&S) at 2-4x the LoD of the corresponding species reference strain used to determine the LoD in Table 5 above, and then tested with the Curian Campy assay. Any strain which did not produce $100\%$ positive results at 2-4x LoD was tested at sequentially higher concentrations until $100\%$ positive results were obtained. All C. jejuni and C. coli strains were detected at 2-4x LoD in both unpreserved and preserved stool. All C. upsaliensis and C. lari strains were detected at 2-4x LoD in unpreserved and preserved stool, except for those denoted in Table 7, where a higher concentration relative to the LoD of the original reference strain was required. The reason for the elevated reactivity levels is unknown but may reflect biological variation among the strains.
Table 6. Strains Evaluated for Reactivity with the Curian Campy Assay
| C. jejuni | C. coli | C. upsaliensis | C. lari |
| --- | --- | --- | --- |
| Zeptometrix Z086 | ATCC 43482 | ATCC BAA-1059 | ATCC BAA-1060 |
| ATCC 33292 | ATCC 43476 | ATCC 49815 | ATCC 35222 |
| ATCC 49350 | ATCC 43478 | 2017/0506H | ATCC 35223 |
| ATCC 43442 | ATCC 43485 | 2016/1950 | 2013/0823H |
| ATCC 33560 | ATCC BAA-1061 | ATCC 43954 | 2015/0814 |
| | | ATCC 43953 | 2015/2983 |
| | | 2016/2697 | 2016/0235 |
| | | 2018/0319H | 2015/0519 |
| | | 2016/1931 | 2015/1582 |
| | | ATCC 700558 | 2015/2189 |
Table 7. Campylobacter Strains Detected at Elevated Concentrations
| Species | Strain | Reactivity Level in Unpreserved Stool | Reactivity Level in Stool Preserved in C&S |
| --- | --- | --- | --- |
| C. upsaliensis | 2016/1950 | 4x | 8x |
| | ATCC 43953 | 2x | 8x |
| | 2016/2697 | 56x | 78x |
| | 2018/0319H | 40x | 55x |
| | ATCC 700558 | 24x | 33x |
K210976 - Page 10 of 13
{10}
| C. lari | 2013/0823H | 8x | 4x |
| --- | --- | --- | --- |
| | 2015/0814 | 8x | 4x |
| | 2016/0235 | 8x | 10x |
| | ATCC 35222 | 8x | 10x |
| | ATCC 35223 | 24x | 10x |
7. Assay Cut-Off:
To determine the preliminary cut-off the Curian Campy assay read on the Curian Analyzer, a panel of clinical negative and contrived positive (near LoD) stool samples were evaluated. The initial cut-off was validated during the prospective clinical study by testing a population of 1464 specimens, of which 18 were positive and 1425 were negative by the Curian Campy assay. The results of the cut-off study were acceptable.
B Comparison Studies:
1. Method Comparison with Predicate Device:
Not applicable
2. Matrix Comparison:
Not applicable
C Clinical Studies:
1. Clinical Sensitivity:
Prospectively Collected Specimens Study
The performance of the Curian Campy test was assessed with specimens prospectively collected from patients with signs and symptoms of gastroenteritis. Specimens were collected from five independent clinical sites. The sensitivity and specificity for the Curian Campy test was calculated by comparing directly to culture and speciation results performed at the sites. Prospective testing consisted of 1474 stool specimens tested in clinical microbiology laboratories at Medical College of Wisconsin, Medfusion, Cincinnati Children's Hospital Medical Center, Indiana University, and Tricore Reference Laboratories. Both "Analyze Now" and "Incubate and Analyze" modes were used for specimen testing. The vast majority of stool specimens were pipettable and were sampled using the transfer pipette provided in the Curian Campy kit. A small subset were non-pipettable and thus were sampled with the Curian Campy Stool Collection Brush (sold separately by Meridian).
The ages of patients ranged from less than 1 year to 99 years with 15.7% of the specimens coming from patients that were ≤18 years of age. Of the 1474 patients tested, 60.8% were female and 39.2% were male. No difference in test performance was observed based on patient age or gender. Table 8 shows a summary of the clinical performance of the Curian Campy for all sites, combined.
K210976 - Page 11 of 13
{11}
Table 8. Clinical Performance of the Curian Campy with Prospective Specimens
| | Culture and Speciation | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| Curian Campy | Positive | 18 | 28** | 46 |
| | Negative | 3* | 1425 | 1428 |
| | Total | 21 | 1453 | 1474 |
| | | | | |
| Sensitivity | | 85.7% (18/21), 95% CI: (65.4, 95.0%) | | |
| Specificity | | 98.1% (1425/1453), 95% CI: (97.2, 98.7%) | | |
*The Standard of Care (SoC) testing of two of the three false negative specimens by a high sensitivity, FDA-cleared nucleic acid amplification (NAAT) assay showed that one of the specimens produced a negative Campylobacter result, whereas one of the specimens produced a positive Campylobacter result; the third specimen was not subjected to NAAT testing as part of the SoC.
**Of the 28 false positive specimens, 10 were subjected to SoC testing by a high sensitivity, FDA-cleared NAAT assay. Two of these 10 specimens produced a positive Campylobacter result, whereas eight produced a negative Campylobacter result. Eighteen specimens were not subjected to NAAT testing as part of the SoC.
The results of the study show that the Curian Campy exhibited a sensitivity of $85.7\%$ and a specificity of $98.1\%$ compared to culture and speciation. The sensitivity point estimate and lower bound of the two-sided $95\%$ confidence interval for the sensitivity point estimate were slightly lower than expectations for this type of assay. However, the sensitivity results were deemed acceptable without the need to conduct additional clinical specimen testing for the following reasons: (1) Challenges with performing prospective clinical studies during the COVID-19 pandemic, (2) One of the three specimens with a false negative Curian Campy result was negative by a high sensitivity FDA-cleared nucleic acid amplification (NAAT) assay. This result suggests that the Curian Campy sensitivity is likely higher than the $85.7\%$ -point estimate and, (3) The prospective study was supplemented with archived positive specimens (see section VII.C.Archived Study, below).
# Archived Specimens Study
Additional testing with the Curian Campy was performed at five clinical sites using 290 archived specimens, of which 29 were positive and 261 were negative by Campylobacter spp. culture and speciation. The patient ages ranged from $< 1$ year to 93 years. Compared to culture and speciation, the sensitivity and specificity point estimates for the Curian Campy were $96.6\%$ and $98.1\%$ , respectively. The results of the archived study are illustrated in Table 9. These results are acceptable.
Table 9. Clinical Performance of the Curian Campy with Archived Specimens
| | Culture and Speciation | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| Curian Campy | Positive | 28 | 5 | 33 |
| | Negative | 1 | 256 | 257 |
| | Total | 29 | 261 | 290 |
| | | | | |
| Sensitivity | | 96.6% (28/29), 95% CI: (82.8, 99.4%) | | |
| Specificity | | 98.1% (256/261), 95% CI: (95.6, 99.2%) | | |
# Contrived Study
The performance of the Curian Campy was further evaluated by testing a panel consisting of $C$ coli, $C$ lari, and $C$ upsaliensis isolates. All isolates, which were obtained from humans, were procured from ATCC. Samples were prepared by spiking negative, pooled fecal matrix with
K210976 - Page 12 of 13
{12}
whole organism at 2x and 8x the LoD of the reference strain, which was used to confirm the LoD. Both unpreserved stool and preserved stool (in C&S transport media) were evaluated during the study. For each Campylobacter species, 25 samples were prepared in both unpreserved stool and preserved stool, for a total of 50 positive samples/species. Sixty negatives were included for blinding and randomization purposes. A subset of the panel was provided to each of the five clinical testing sites for evaluation. Each site was provided both positive and negative samples. Testing was performed over a period of several days at each site. The results of this study demonstrated 100% positive percent agreement (PPA) with expected results for all C. coli, C. lari, and C. upsaliensis isolates, regardless of stool matrix. These results are acceptable.
2. Clinical Specificity:
See section VII.C.1.Prospective Study above.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Brush Bridging Study
The Curian Campy assay is for use with unpreserved human stool specimens and human stool specimens preserved in Cary-Blair and C&S transport media. Most specimens are easily sampled using the transfer pipette provided in the kit; however, some unpreserved stool specimens are non-pipettable and require use of a specific brush (i.e., Curian Campy Stool Collection Brush, sold separately by Meridian) to adequately collect the sample for analysis in the Curian Campy assay. Fifty-three non-pipettable stool specimens (5 Campylobacter positives and 48 negatives) were processed with the brush during the prospective and archived clinical studies, combined.
To further support use of the brush with the Curian Campy assay, an analytical bridging study was performed that evaluated a panel consisting of contrived positive and negative samples collected/processed with the brush. Contrived positive samples were generated by spiking C. jejuni (strain ATCC #BAA-1234) into negative, non-pipettable stool samples. Ten samples were prepared at 3x LoD and 25 samples were prepared at 5x LoD. Twenty-five negative samples were also tested. Three non-expert operators tested each sample at one internal site. All positive samples gave the expected positive results, and all negative samples were negative. These results indicate that non-pipettable stool specimens can be collected with the brush prior to testing with the Curian Campy assay.
D. Clinical Cut-Off:
Not applicable
E. Expected Values/Reference Range:
Not applicable
VIII. Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K210976 - Page 13 of 13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.