Campylobacter Quik Chek

K191456 · Techlab, Inc. · LQP · Jun 20, 2019 · Microbiology

Device Facts

Record IDK191456
Device NameCampylobacter Quik Chek
ApplicantTechlab, Inc.
Product CodeLQP · Microbiology
Decision DateJun 20, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 866.3110
Device ClassClass 1

Indications for Use

The CAMPYLOBACTER QUIK CHEK test is a rapid membrane enzyme-linked immunosorbent assay for the qualitative detection of a Campylobacter-specific antigen in human fecal specimens. The CAMPYLOBACTER QUIK CHEK test is designed to detect C. jejuni, C. coli, C. lari, and C. upsaliensis from patients with signs and symptoms of gastroenteritis. The test is intended for use with preserved fecal specimens in transport media and unpreserved fecal specimens. Test results should be considered in conjunction with clinical findings and patient history.

Device Story

The CAMPYLOBACTER QUIK CHEK is a diagnostic device used for the detection of Campylobacter species. It functions as a serological reagent test. The device is intended for use in clinical laboratory settings by trained personnel to assist in the diagnosis of Campylobacter infections. The test provides qualitative results to healthcare providers, aiding in clinical decision-making regarding patient treatment and management.

Clinical Evidence

Bench testing only. Analytical validation performed using clinical isolates of C. lari and C. upsaliensis spiked into fecal matrix at 2x LoD. Study included 135 data points per species tested over three days by multiple technicians. Results showed 100% positive percent agreement (95% CI: 97.2%, 100%) for both species across unpreserved, Cary Blair, and C&S media. Interfering substances and prozone effect studies confirmed no impact on assay performance.

Technological Characteristics

Serological reagent test for Campylobacter species detection. Classified as Class I, Product Code LQP under 21 CFR 866.3110.

Indications for Use

Indicated for qualitative detection of Campylobacter-specific antigen (C. jejuni, C. coli, C. lari, and C. upsaliensis) in human fecal specimens from patients with signs and symptoms of gastroenteritis.

Regulatory Classification

Identification

Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identify Campylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases. Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.

Submission Summary (Full Text)

{0} Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov # SPECIAL 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ## I Background Information: A 510(k) Number K191456 B Applicant Techlab, Inc. C Proprietary and Established Names CAMPYLOBACTER QUIK CHEK D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | LQP | Class I | 21 CFR 866.3110 - Campylobacter fetus Serological Reagents | MI – Microbiology | ## II Review Summary: This 510(k) submission contains information/data on modifications made to the submitter's own CLASS I device requiring 510(k). The following items are present and acceptable: 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. CAMPYLOBACTER QUIK CHEK 510(k) number: K173217 2. Submitter's statement that the INDICATIONS FOR USE/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). K191456 - Page 1 of 5 {1} 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for an update to the Indications for Use to include detection of additional Campylobacter species. 4. Comparison Information (i.e., similarities and differences) to the submitter's legally marketed predicate device including, labeling, intended use, and physical characteristics. | Device & Predicate Device(s): | CAMPYLOBACTER QUIK CHEK K191456 (Device) | CAMPYLOBACTER QUIK CHEK K173217 (Predicate) | | --- | --- | --- | | Similarities | | | | Intended Use/Indications for Use | The CAMPYLOBACTER QUIK CHEK test is a rapid membrane enzyme-linked immunosorbent assay for the qualitative detection of a Campylobacter-specific antigen in human fecal specimens. The CAMPYLOBACTER QUIK CHEK test is designed to detect C. jejuni, C. coli, C. lari, and C. upsaliensis from patients with signs and symptoms of gastroenteritis. The test is intended for use with preserved fecal specimens in transport media and unpreserved fecal specimens. Test results should be considered in conjunction with clinical findings and patient history. | The CAMPYLOBACTER QUIK CHEK test is a rapid membrane enzyme-linked immunosorbent assay for the qualitative detection of a Campylobacter-specific antigen in human fecal specimens. The CAMPYLOBACTER QUIK CHEK test is designed to detect C. jejuni and C. coli from patients with signs and symptoms of gastroenteritis. The test is intended for use with preserved fecal specimens in transport media and unpreserved fecal specimens. Test results should be considered in conjunction with clinical findings and patient history. | | Measured Analyte | Campylobacter-specific antigen | Same | | Target Population | Persons suspected of having Campylobacter infection | Same | | Type of Test | Qualitative | Same | | Sample Matrix | Human Fecal Specimen | Same | | Controls | Positive and negative control included in the kit. Internal Control line | Same | | Storage | Refrigerated (2-8°C) | Same | | Format | Single Use Membrane Cassette - 25 tests | Same | | Specimen Type | Unpreserved fecal specimens | Same | K191456 - Page 2 of 5 {2} K191456 - Page 3 of 5 | | and fecal specimens in Cary-Blair and C&S Transport Media | | | --- | --- | --- | | Antibody Format | Monoclonal/Polyclonal | Same | | Time to Result | < 30 minutes | Same | | Differences | | | | Species Detected | C. jejuni, C. coli, C. lari, and C. upsaliensis | C. jejuni and C. coli | 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis. A formal risk assessment was conducted for the CAMPYLOBACTER QUIK CHEK assay. This assessment included Failure Mode Effects and Analysis (FMEA) for the assay design. This analysis identified false negative and false positive results i.e., true Campylobacter infection not identified or false Campylobacter infection identified, as the worst-case risks to the patient. It was concluded based on FMEA and verification testing that the risks associated with detection of C. lari and C. upsaliensis are mitigated and therefore CAMPYLOBACTER QUIK CHEK is safe and effective for its intended use. b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied. The following studies were performed to demonstrate risk mitigation: **Limit of Detection** The limit of detection (LoD) for the modified CAMPYLOBACTER QUIK CHEK test was determined visually using whole organisms of both C. lari and C. upsaliensis spiked into unpreserved and preserved media, i.e., raw stool fecal matrix and Cary Blair and C&amp;S media, respectively. The concentrations were calculated in colony forming units per milliliter (CFU/mL) and their equivalent in CFU/test by factoring in the dilutions and the final volume used in the assay. The LoD is the concentrations of organisms that yields a positive result 95% of the time and a negative result 5% of the time. See the LoDs in Table 1 listed by organism and by stool (unpreserved and preserved). Table 1. LoD for C. lari and C. upsaliensis by Visual Reading | C. lari | | C. upsaliensis | | | --- | --- | --- | --- | | CFU/mL | CFU/test | CFU/mL | CFU/test | | Fecal Matrix | | | | | 1.23 X 10^{6} | 18,636 | 2.68 X 10^{6} | 40,606 | | Cary Blair | | | | | 3.54 X 10^{6} | 55,313 | 2.43 X 10^{6} | 37,969 | | C&S | | | | | 2.27 X 10^{6} | 35,469 | 5.04 X 10^{6} | 78,750 | {3} # Inclusivity Study The sponsor tested the reactivity of Campylobacter isolates from Centre National de Reference des Campylobacters et Helicobacters - Centre Hospitalier Universitaire de Bordeaux by spiking each isolate listed in Table 2 into negative fecal matrix at $2 - 3\mathrm{x}$ LoD and testing with the modified CAMPYLOBACTER QUIK CHEK. All results were positive. Table 2. Strains Evaluated for Reactivity with Modified CAMPYLOBACTER QUIK CHEK | C. lari | C. upsaliensis | | --- | --- | | 2013/0823H | 2016/0385 | | 2014/2772 | 2016/1931 | | 2015/0519 | 2016/1950 | | 2015/0814 | 2016/2697 | | 2015/1582 | 2016/2826 | | 2015/1657 | 2017/0349 | | 2015/2189 | 2017/0506H | | 2015/2983 | 2017/2584 | | 2016/0235 | 2018/0319H | | 2016/1130H | 2018/1669 | # Interfering Substances Study The modified CAMPYLOBACTER QUIK CHEK test was evaluated for interfering substances with the substances and concentrations listed in Table 3. None of the substances where shown to interfere with the performance of the modified CAMPYLOBACTER QUIK CHEK test. Table 3. Interfering Substances | Barium sulfate (5% w/v) | Benzalkonium Chloride (1% w/v) | | --- | --- | | Ciprofloxacin (0.25% w/v) | Ethanol (1% w/v) | | Hog gastric mucin (3.5% w/v) | Human blood (40% v/v) | | Hydrocortisone (1% w/v) | Imodium (5% v/v) | | Kaopectate (5% v/v) | Leukocytes (0.05% w/v) | | Maalox Advanced (5% v/v) | Mesalazine (10% w/v) | | Metronidazole (0.25% w/v) | Mineral Oil (10% w/v) | | Mylanta (4.2 mg/mL) | Naproxen Sodium (5% w/v) | | Nonoxynol-9 (1% w/v) | Nystatin (1% w/v) | | Palmitic Acid/Fecal Fat (40% w/v) | Pepto-Bismol (5% v/v) | | Phenylephrine (1% w/v) | Polyethylene glycol 3350 (10% w/v) | | Prilosec OTC (5 μg/mL) | Sennosides (1% w/v) | | Simethicone (10% w/v) | Stearic Acid/Fecal Fat (40% w/v) | | Tagamet (5 μg/mL) | TUMS (50 μg/mL) | | Human Urine (5% v/v) | Vancomycin (0.25% w/v) | K191456 - Page 4 of 5 {4} # Prozone Effect Study The purpose of this study was to demonstrate that a high concentration of Campylobacter (antigen) does not interfere with a positive reaction in the modified CAMPYLOBACTER QUIK CHEK test. High samples were prepared by spiking a negative fecal pool with $1.6 \times 10^{8}$ whole organisms of C. lari and $1.28 \times 10^{8}$ whole organisms of C. upsaliensis. A total of five different dilutions of C. lari and C. upsaliensis were prepared and tested. Testing was performed in triplicate according to the Package Insert instructions. The results demonstrated that there was no overall hook affect, and that elevated levels of whole organism did not affect the detection of Campylobacter. # Analytical Validation Study The performance of the modified CAMPYLOBACTER QUIK CHEK test was determined by testing a panel of five C. lari and five C. upsaliensis clinical isolates. All samples were prepared by spiking negative, pooled fecal matrix with whole organism at 2x LoD. Performance was assessed for unpreserved stool and stool preserved in Cary Blair and C&amp;S transport media. Due to availability of limited number of representative isolates, each isolate was tested once a day over a three-day period by multiple technicians at a single internal site for a total of 135 data points. The results of this study demonstrated positive percent agreement (PPA) with expected results of $100\%$ for C. lari and $100\%$ for C. upsaliensis. Results are illustrated in Table 4. Table 4. Analytical Validation Study - Performance of the Modified CAMPYLOBACTER QUIK CHEK | Species | Percent Agreement | 95% CI | | --- | --- | --- | | C. lari | 100% (135/135) | 97.2%, 100% | | C. upsaliensis | 100% (135/135) | 97.2%, 100% | The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the specific modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared predicate device. K191456 - Page 5 of 5
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...