Snore Bandit is intended for use by adult patients (18 years or older) as an aid to reduce snoring.
Device Story
Snore Bandit is an intraoral anti-snoring appliance. User heats device in hot water ('boil-and-bite' technique) to soften material, then bites down to mold it to teeth and gums while holding jaw in forward position. Once cooled, device maintains shape. During sleep, device advances and holds lower jaw forward, enlarging and stiffening airway to reduce snoring. Device features multi-layered breathing space design to maximize airflow and prevent lip blockage. Used by patients at home without prescription. Output is physical mandibular repositioning. Benefits include reduced snoring. Labeling includes STOP-BANG questionnaire to help users identify potential undiagnosed obstructive sleep apnea.
Clinical Evidence
No clinical testing was performed in association with this submission.
Technological Characteristics
Intraoral mandibular advancement device. Material: Ethylene Vinyl Acetate (EVA) copolymer (USP Class VI). Manufacturing: Molding process. Design: Multi-layered breathing space. Non-sterile. Biocompatibility: Tested per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (sensitization, irritation).
Indications for Use
Indicated for adult patients (18 years or older) as an aid to reduce snoring.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 1, 2021
SnoreBandit LLC % Colette Cozean Regulatory Consultant The EyeDeas Company 21581 Midcrest Dr. Lake Forest, California 92630
Re: K210910
Trade/Device Name: Snore Bandit Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK Dated: May 24, 2021 Received: May 26, 2021
Dear Colette Cozean:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA
has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210910
Device Name Snore Bandit Anti-Snoring Appliance
| Indications for Use (Describe) |
|-----------------------------------------------------------------------------------------------------|
| Snore Bandit is intended for use by adult patients (18 years or older) as an aid to reduce snoring. |
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
5. 510(k) Summary
K210910
# 510(k) Summary
| Applicant: | Snore Bandit LLC<br>8176 Merlewood Ave<br>Las Vegas, NV 89117<br>Phone: 702-283-3557 |
|------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Contact person: | Colette Cozean, Ph,D.<br>21581 Midcrest Dr.<br>Lake Forest, CA 92603 USA<br>949-855-2885<br>colettecozean@gmail.com |
| Date Prepared: | March 24, 2021 |
| Proprietary Name: | Snore Bandit ™ Anti-Snoring Appliance |
| Common name: | Intraoral Device for Snoring |
| Classification Name: | Device, anti-snoring<br>(Class II, 21 CFR 872.5570, Product Code LRK) |
| Primary Predicate Device: | Sleep Tight Mouthpiece STZ (K191618)-OTC Reference |
| Reference Predicate Devices: | Pure Sleep (K190058) – Over the Counter Reference |
## Description of the Device
The Snore Bandit Anti-Snoring Appliance advances and holds the lower jaw forward during sleep which enlarges and stiffens the airway to alleviate or reduce snoring. The Snore Bandit uses the common 'boil-and bite' technique familiar to athletes who heat and mold their mouth guard to fit their teeth. In this process, the user softens the mouthpiece in hot water according to detailed directions provided with packaging. The user bites into the softened plastic and compresses the softened material to their teeth and gums while holding the jaw slightly forward until the material cools and stiffens. After cooling, the device maintains its shape and is used to hold the jaw open and forward during sleep to alleviate or reduce snoring. The product is non-sterile and provided in a sealed box with instructions.
{4}------------------------------------------------
## Indications for Use
Snore Bandit is intended for use by adults (18 years or older) as an aid to reduce snoring.
## A. Technological Characteristics
The appliance is equivalent in design and functionality to the primary predicate, Sleep Tight Mouthpiece (STZ) device (K191618). Both devices work by advancing the mandible and allows the patient to breathe through their mouth and nose utilizing space between the upper and lower teeth. Both are custom fit for comfort to the individual user by a standard 'boil and bite' method. Both use an ethylene vinyl acetate material. Both are manufactured via a molding process. Both devices are provided non-sterile. The main difference between the subject device and the primary predicate device is a multi-layered breathing space design rather than a singlelayer breathing space design in the primary predicate device. This opens the airway and allows maximum airflow and prevents the patient's lips from blocking the breathing ports.
| | Subject | OTC Primary<br>Predicate | OTC Reference<br>Predicate | OTC Reference<br>Predicate |
|----------------------------------------|------------------------------------------------------------------------------------------------------|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Name | Snore Bandit | STZ | PureSleep | ZQuiet |
| Classification<br>name | Device, anti-snoring | Device, anti-snoring | Intraoral<br>Device for<br>Snoring and<br>Obstructive<br>Sleep Apnea | Device, anti-snoring |
| 510(k) Number | K210910 | K191618 | K190058 | K180124 |
| Class | II | II | II | II |
| Product Code | LRK | LRK | LRK | LRK |
| 21 CFR | 872.5570 | 872.5570 | 872.5570 | 872.5570 |
| Technology &<br>Mechanism of<br>Action | Mandibular<br>advancement<br>design to<br>increase<br>pharyngeal<br>space to<br>alleviate<br>snoring | Mandibular<br>advancement | Mandibular<br>repositioning<br>device (MRD)<br>that advances<br>the lower jaw<br>to increase<br>pharyngeal<br>space and<br>alleviate<br>snoring | Mandibular<br>advancement |
### Comparison to Predicate Devices
{5}------------------------------------------------
| Indications for<br>use | "Snore Bandit<br>is intended<br>for use on<br>adults 18<br>years of age<br>or older as an<br>aid to reduce<br>snoring" | "Sleep Tight<br>Mouthpiece<br>STZ is<br>intended for<br>use on adult<br>patients 18<br>years of age<br>or older as an<br>aid to reduce<br>snoring" | Adults | "ZQuiet is intended<br>as an aid in the<br>reduction of snoring<br>for adults at least 18<br>years old." |
|------------------------|------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Prescription<br>Status | OTC | OTC | OTC | OTC |
| Provided Sterile | NO | NO | NO | NO |
| Materials | Ethylene Vinyl<br>Acetate (EVA)<br>Copolymer,<br>no colorant | Thermoplastic<br>Elastomer<br>(Ethyl Vinyl<br>Acetate) | Polypropylene<br>homopolymer<br>and Ethylene<br>Vinyl Acetate<br>(EVA) | Thermoplastic<br>Elastomer with a<br>colorant |
# B. Technology - Mechanism of Action
The subject device and primary predicate device (K191618) are identical in the mechanism of action. They both work by utilizing an intraoral mouthpiece to advance the lower mandible and create space between the upper and lower teeth (breathing spaces) giving the patient the ability to breathe through their mouth and nose.
The mechanism of action for the other two OTC reference predicate devices, the PureSleep and ZQuiet, is solely advancing the mandible. With an intraoral device that alters the mandible position, mandibular changes can lead to pain or discomfort in the temporomandibular joint (TMJ). The subject device has no new risks regarding mandibular advancement.
Utilizing the same mechanism of action as the predicate devices (primary and reference), the subject device raises no new concerns regarding the effectiveness or safety as compared to the other predicate devices.
{6}------------------------------------------------
# C. Indication for Use
The subject device uses language for the indication for use in substance to previously approved OTC devices for the reduction of snoring. The subject device is "intended for use by adults (18 years or older) as an aid to reduce snoring."
This is almost identical to the indications for use for the primary predicate devices, and the two reference devices, the Sleep Tight Mouthpiece (STZ) is intended for use by adult patients 18 years of age as an aid to reduce snoring, PureSleep is an intraoral mandibular repositioning device used on adults as an aid during sleep to reduce snoring and ZQuiet, whose indications is "intended as an aid in the reduction of snoring for adults at least 18 years old" respectively.
As the indications for use are identical, there are no new concerns regarding safety as compared to predicate devices (primary and reference).
## D. Prescription Status
The subject device is intended to be sold over-the-counter. All of the predicate devices (primary and reference), STZ, PureSleep, and ZQuiet have clearance for over-the-counter use. The subject labeling specifically states that the device does not treat obstructive sleep apnea and the STOP-BANG questionnaire is included in the labeling to help mitigate the risk of undiagnosed OSA among potential Snore Bandit users.
Substantial equivalent labeling and instructions for use to that of the over-the-counter predicate devices, there are no new concerns regarding effectiveness or safety as compared to predicate devices (primary and reference).
## E. Labeling
The subject appliance has similar labeling as the OTC reference predicate devices (primary and reference), Sleep Tight Mouthpiece (STZ), PureSleep, and ZQuiet. It utilizes a validated questionnaire to access the risk of sleep apnea and contains the appropriate warnings and labeling for the device to be used without requiring a prescription.
# F. Technology - Materials
To support this application, the sponsor has included Biocompatibility testing results demonstrating the subject device is biocompatible under ISO 10993, including cytotoxicity, sensitization, and irritation testing. The device was shown to meet the requirements of the ISO 10993 guidelines for each of these tests. Other biocompatibility tests recommended for permanent mucosal contact were not conducted because all materials used in the device are certified USP Class VI and used in many other currently marketed medical devices, including another mouthguard called Somnos Anti-Snoring Mouth Guard (K201484) made by the same manufacturer that makes Snore Bandit.
{7}------------------------------------------------
Because the same materials and manufacturing processes are used in the primary predicate device, there are no new concerns regarding effectiveness or safety as compared to the predicate devices.
## G. Biocompatibility
The materials and manufacturing process are similar to the predicate devices (primary and reference). One hundred percent of the Snore Bandit is made of Elvax. There is no added colorant in Snore Bandit.
Testing results have been include cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), and irritation (ISO 10993-10). As all the materials used in the device are certified USP Class VI and used in many other currently marketed medical devices, no additional biocompatibility testing was conducted.
## Clinical Testing
No clinical testing was performed in association with this submission.
## Conclusions
The results of the comparison of design, materials, intended use, labeling, and technological characteristics demonstrate the subject device is substantially equivalent in safety and efficacy to the legally marketed predicate devices, both the primary and reference predicates. Therefore, the sponsor concludes the proposed Snore Bandit device is substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.