K201484 · Koncept Innovators, LLC · LRK · Nov 23, 2020 · Dental
Device Facts
Record ID
K201484
Device Name
Somnos Anti-Snoring Mouth Guard
Applicant
Koncept Innovators, LLC
Product Code
LRK · Dental
Decision Date
Nov 23, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
SOMNOS ANTI-SNORING MOUTH GUARD is an over-the-counter mouthguard intended for use in adult persons 18 years of age or older as an aid for the reduction of snoring.
Device Story
Somnos Anti-Snoring Mouth Guard is a self-fitting, adjustable mandibular advancement device (MAD) for home use by adults to reduce snoring. The device consists of two prefabricated, single-piece trays made of thermoplastic resin (ethylene-vinyl-acetate and polycarbonate) that are customized via a 'boil and bite' method. When worn during sleep, the device repositions the mandible anteriorly (up to 6 mm) to increase pharyngeal space, improving air exchange. The user self-fits the device and adjusts the jaw position for comfort. The device is supplied with a spatula for handling during molding and a storage case. It is non-sterile and reusable by a single individual.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including functional performance (hardness, water sorption, solubility, flexural strength, fracture toughness, optical microscopy) per ISO 20795-2 and biocompatibility testing (cytotoxicity, skin sensitization, oral mucosal irritation) per ISO 10993.
Technological Characteristics
Mandibular advancement device; materials: Ethylene-Vinyl-Acetate (moldable) and Polycarbonate (rigid); spatula/case: Polypropylene. Manufacturing: injection molding. Biocompatibility: ISO 10993-1. Functional testing: ISO 20795-2. Dimensions: adjustable anterior mandibular repositioning up to 6 mm. Non-sterile, reusable, single-patient use.
Indications for Use
Indicated for adult persons 18 years of age or older as an aid for the reduction of snoring. Not indicated for obstructive sleep apnea.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo includes the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 23, 2020
Koncept Innovators, LLC. % Maureen Garner President New World Regulatory Solutions, Inc. 11700 W. Charleston Blvd. Suite 170-390 Las Vegas, Nevada 89135
Re: K201484
Trade/Device Name: Somnos Anti-Snoring Mouth Guard Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK, LQZ Dated: October 15, 2020 Received: October 19, 2020
Dear Maureen Garner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael E. Adjodha -S
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201484
Device Name SOMNOS ANTI-SNORING MOUTH GUARD
Indications for Use (Describe)
SOMNOS ANTI-SNORING MOUTH GUARD is an over-the-counter mouthguard intended for use in adult persons 18 years of age or older as an aid for the reduction of snoring.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Koncept Innovators, Inc. K201484
SOMNOS ANTI-SNORING MOUTH GUARD
# Section 5: 510(k) Summary
[Compliance with 21 CFR 807.92]
## Submission Sponsor
Koncept Innovators, Inc. 1619 Executive Ave Myrtle Beach, SC 29577
| Contact: | Sylvan Newby |
|----------|----------------------|
| Title: | President |
| Phone: | 843-945-4031 |
| Email: | sylvan@glidegear.com |
### Submission Correspondent
New World Regulatory Solutions, Inc. 11700 W. Charleston Blvd Suite 170-390 Las Vegas, NV 89135
| Contact: | Maureen Garner |
|----------|-------------------|
| Title: | President |
| Phone: | 732-779-7422 |
| Email: | NWRSinc@gmail.com |
# Date Prepared
October 28, 2020
#### Type of 510(k) Submission
Traditional
#### Device Identification
| Trade /Proprietary Name: | SOMNOS ANTI-SNORING MOUTH GUARD |
|--------------------------|---------------------------------|
| Common Name: | Device, Anti-snoring |
| Classification: | II |
| FDA Product Code: | LRK |
| Review Panel: | Dental |
# Legally Marketed Predicate Device Primary Predicate Device:
Snore Rx, K170825
This predicate has not been subject to a design-related recall.
{4}------------------------------------------------
# SOMNOS ANTI-SNORING MOUTH GUARD
# Device Description
SOMNOS ANTI-SNORING MOUTH GUARD is comprised of a mouthguard, a spatula and a storage case. The mouthguard is a self-fitting adjustable Mandibular Advancement Device (MAD) removable intraoral anti-snoring device used to reduce nighttime snoring in adult persons 18 years of age or older. SOMNOS ANTI-SNORING MOUTH GUARD consists of two customized prefabricated single-piece trays that conform to the patient's dentition in the upper and lower jaws. The device functions as a mandibular anterior repositioner, which acts to increase the user's pharyngeal space, resulting in the improvement of the ability to exchange air during sleep. SOMNOS ANTI-SNORING MOUTH GUARD, while in situ. moves the lower jaw forward and helps reduce the likelihood of snoring. It is intended to be worn while sleeping. The supplied spatula is used to separate the SOMNOS ANTI-SNORING MOUTH GUARD. SOMNOS ANTI-SNORING MOUTH GUARD is conformable to wear, and it can be self-fitted by the 'boil and bite' method (when heated the device conforms to person's teeth) and allows slight adjustment of the anterior positioning of the jaw to the person's comfort.
SOMNOS ANTI-SNORING MOUTH GUARD is constructed of two (2) preformed and fixed single piece trays composed of Thermoplastic resin Propylene Elastomer (TPE) material, specifically Ethylene-Vinyl-Acetate for the moldable part and Polycarbonate for the rigid part of the device. The spatula and storage case are composed of 100% Polypropylene. The materials in Somnos are commonly used in other 510(k) cleared dental mouth guards for bruxism and antisnoring. The three components (mouthguard, spatula and storage case) are manufactured by injection molding. They are non-flavored and have no color additives.
SOMNOS ANTI-SNORING MOUTH GUARD is provided in one model/size, is reusable by a single individual who is at least eighteen (18) years of age and is supplied as non-sterile.
Each box of SOMNOS ANTI-SNORING MOUTH GUARD is supplied with the following items:
- . (1) Anti-snoring Mouthguard
- . (1) Spatula
- (1) Storage case ●
- (1) Instructions for Use ●
# Indications for Use
SOMNOS ANTI-SNORING MOUTH GUARD is an over-the-counter mouth guard intended for use on adult persons 18 years of age or older as an aid for the reduction of snoring.
# Summary of Substantial Equivalence
The intended, indications, and directions for use of SOMNOS ANTI-SNORING MOUTH GUARD are equivalent to those of the predicate device, SnoreRx. The design, manufacturing, and principal of operation are similar to those of the predicate device and do not raise any new issues concerning safety or effectiveness.
# Summary of Technological Characteristics
The tables below compare the technological characteristics of SOMNOS ANTI-SNORING MOUTH GUARD and the primary predicate, SnoreRx.
{5}------------------------------------------------
# SOMNOS ANTI-SNORING MOUTH GUARD
| | Subject Device | Primary Predicate Device |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name /<br>Brand Name | SOMNOS ANTI-SNORING MOUTH<br>GUARD | SnoreRx |
| 510(k)# | K201484 | K170825 |
| Indications for Use | SOMNOS ANTI-SNORING MOUTH<br>GUARD is intended for use in adult<br>persons 18 years of age or older as an aid<br>for the reduction of snoring. | SnoreRx is intended for use on adult<br>patients 18 years of age or older as an aid<br>for the reduction of snoring. |
| Device Description | Mandibular repositioning device that<br>advances the lower jaw to increase<br>pharyngeal space.<br><br>Adjustment of the relative position of the<br>trays by the use of latch adjustment to<br>obtain a fixed advancement. | Mandibular repositioning device that<br>advances the lower jaw to increase<br>pharyngeal space.<br><br>Adjustment of the relative position of the<br>trays by the use of latch adjustment to<br>obtain a fixed advancement. |
| Patient Population | Adult persons 18 years of age or older | Adult patients 18 years of age or older |
| Duration of Use | Single patient, multi-use | Single patient, multi-use |
| Desirable<br>Characteristics | Home use, heat sensitive/moldable<br>position [Boil and Bite Method (Heat &<br>Bite Self-Fit)], adjustable jaw<br>advancement | Home use, heat sensitive/moldable<br>position [Boil and Bite Method (Heat &<br>Bite Self-Fit)], adjustable jaw<br>advancement |
| Rx or OTC | OTC | OTC |
| Specifications<br>-Mechanical | -Repositions mandible-anteriorly up to 6<br>mm | -Repositions mandible-anteriorly up to 6<br>mm |
| Method of<br>Manufacturing | Injection Molding | Injection Molding |
| Biocompatibility | Biocompatible Materials Used – ISO<br>10993 | Biocompatible Materials Used – ISO<br>10993 |
| Specifications:<br>-Physical<br>-Use | -Custom-fitted intraoral device<br>-Reusable | -Custom-fitted intraoral device<br>-Reusable |
| Sterility | Non-Sterile | Non-Sterile |
| Materials | -Ethylene Vinyl Acetate copolymer<br>-Polycarbonate resin<br>No Flavor: No Color Additives | -Ethylene Vinyl Acetate copolymer<br>-Polycarbonate resin<br>No Flavor: No Color Additives |
## Similarities Between Subject and Predicate Devices
# Differences Between Subject and Predicate Devices
| Trade Name /<br>Brand Name | Subject Device | Primary Predicate Device |
|------------------------------------|--------------------------------------------------------|------------------------------------------------------------|
| SOMNOS ANTI-SNORING MOUTH<br>GUARD | SnoreRx | |
| Accessories<br>Provided | "Spatula" for handling devices during<br>molding steps | "Fitting Tray" for handling device during<br>molding steps |
| Total allowed<br>molding attempts | 1 | 3 |
| Replacement<br>recommended | 6 months | 8 to 12 months |
{6}------------------------------------------------
#### Koncept Innovators, Inc.
### SOMNOS ANTI-SNORING MOUTH GUARD
In accordance with section 513(i)(1)(A) of the FDCA, a device is substantially equivalent when it has the same intended use and same or similar technological characteristics as a legally marketed predicate device. As demonstrated in this traditional 510(k), any difference between the subject device and the cited predicate are minor and do not raise questions of safety or effectiveness. It is on this basis that SOMNOS ANTI-SNORING MOUTH GUARD is substantially equivalent to the cited predicate device.
# Summary of Performance Data
There were no clinical tests performed or provided in this submission.
The following functional performance tests were conducted on SOMNOS ANTI-SNORING MOUTH GUARD in accordance with FDA Recognized Consensus Standard, ISO 20795-2 Dentistry - Base polymers - Part 2: Orthodontic base polymers:
- Hardness, Water Sorption and Solubility Testing; ●
- Spatula & Mouth guard Boiling/Cooling Water Testing;
- Flexural Strength & Flexural Modulus; ●
- Fracture Toughness; and
- Optical Microscopy (OM).
A biocompatibility assessment was conducted using SOMNOS ANTI-SNORING MOUTH GUARD as a final finished device in accordance with International Standard ISO 10993-1:2009 Biological evaluation of medical devices – Part 1 Evaluation and testing within a risk management process and FDA's Guidance Document. Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," June 16, 2016.
The following biocompatibility performance tests were conducted on SOMNOS ANTI-SNORING MOUTH GUARD:
- In Vitro Cytotoxicity Test using ISO 10993-5:2009 Test Method - MTT Method MEM with 10%FBS extract
- Skin Sensitization Test using ISO 10993-10:2010 Test Method Guinea Pig Maximization ● Test 0.9% Sodium Chloride Injection Extract
- Skin Sensitization Test using ISO 10993-10:2010 Test Method Guinea Pig Maximization ● Test Sesame Oil Extract
- Oral Mucosal Irritation Test using ISO 10993-10:2010 Test Method - Hamster 0.9% Sodium Chloride Extract
- Oral Mucosal Irritation Test using ISO 10993-10:2010 Test Method Hamster Sesame ● Oil Extract
{7}------------------------------------------------
#### SOMNOS ANTI-SNORING MOUTH GUARD
Koncept Innovators conducted a risk analysis on SOMNOS ANTI-SNORING MOUTH GUARD in accordance with ISO 14971. taking into account the issues raised in the FDA Guidance Document, "Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea -Guidance for Industry and FDA." With respect to perceivable conditions in which the device would be subjected to a worst-case environmental of human error scenario, Koncept Innovators believes the outcomes of these risks are considered acceptable within the context of ISO 14971, and that all potential risks have been mitigated to the lowest form. The identified potential risks have been addressed through device design or with communication with the user through the instructions for use. SOMNOS ANTI-SNORING MOUTH GUARD uses a STOP BANG questionnaire on the labeling to assess the risks of sleep apnea and contains appropriate warnings and labeling for the device to be used without a prescription. SOMNOS ANTI-SNORING MOUTH GUARD is not indicated for obstructive sleep apnea.
SOMNOS ANTI-SNORING MOUTH GUARD relied on biocompatibility and functional performance testing as the basis for non-clinical data. The non-clinical testing performed demonstrated that the SOMNOS ANTI-SNORING MOUTH GUARD is substantially equivalent to the predicate device. SnoreRx.
# Statement of Substantial Equivalence
SOMNOS ANTI-SNORING MOUTH GUARD has the same intended use and indications for use as the predicate, SnoreRx. Any minor differences in the technological features of the subject device when compared to the predicate device have been successfully evaluated through safety and performance testing and other verification and validation activities such that the information submitted to the FDA demonstrates that the subject device, when compared to the predicate device. does not raise any new questions of safety and effectiveness. SOMNOS ANTI-SNORING MOUTH GUARD, as designed by and manufactured for Koncept Innovators, Inc. has been determined to be substantially equivalent to the predicate device, SnoreRx.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.