K210611 · Siemens Medical Solutions USA, Inc. · LNH · Jul 1, 2021 · Radiology
Device Facts
Record ID
K210611
Device Name
MAGNETOM Free.Max
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LNH · Radiology
Decision Date
Jul 1, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
AI/ML
Indications for Use
The MAGNETOM MR system is indicated for use as a magnetic resonance diagnostic device (MRDD), which produces transverse, sagittal, coronal, and oblique cross sectional images that display the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images may also be produced. Depending on the region of interest, contrast agents may be used. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist in diagnosis.
Device Story
MAGNETOM Free.Max is an 80 cm bore, 0.55T superconducting MRI system. It acquires MR signals to produce cross-sectional images (transverse, sagittal, coronal, oblique) and derived physical parameters of head, body, and extremities. Operated by trained healthcare professionals in clinical settings. Features include myExam Companion (3D camera, cockpit, assist, autopilot) for workflow guidance and patient setup. Employs advanced software (syngo MR XA40A) for image reconstruction, denoising (Deep Resolve Gain), and sharpness enhancement (Deep Resolve Sharp). Output is interpreted by physicians to assist in clinical diagnosis. Benefits include improved patient access via 80cm bore and optimized workflow via automated positioning and guidance.
Clinical Evidence
Bench testing only. Included image quality assessments, SNR/uniformity measurements, heating measurements, and software verification/validation. A clinical study of 12 individuals was conducted to evaluate Peripheral Nerve Stimulation (PNS) effects. No diagnostic clinical trials were required; sample clinical images were provided to demonstrate performance.
Technological Characteristics
0.55T actively shielded superconducting magnet; 80 cm bore. Includes gradient coils, RF system, myExam 3D camera, and various coils (Body, Head/Neck, Spine, BioMatrix Contour). Connectivity via DICOM. Software syngo MR XA40A. Conforms to IEC 60601-1, IEC 60601-2-33, ISO 14971, IEC 62304, and NEMA MS standards.
Indications for Use
Indicated for patients requiring magnetic resonance imaging of the head, body, or extremities to assist in diagnosis. No specific age or gender contraindications provided.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
Siemens Medical Solutions USA, Inc. Andrew Turner Regulatory Affairs Specialist 40 Liberty Boulevard, Mailcode 65-1A Malvern, Pennsylvania 19355
Re: K210611
Trade/Device Name: MAGNETOM Free.Max Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH, MOS Dated: May 27, 2021 Received: June 1, 2021
Dear Andrew Turner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
July 1, 2021
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210611
Device Name MAGNETOM Free.Max
#### Indications for Use (Describe)
The MAGNETOM MR system is indicated for use as a magnetic resonance diagnostic device (MRDD), which produces transverse, sagittal, coronal, and oblique cross sectional images that display the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images may also be produced. Depending on the region of interest, contrast agents may be used.
These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist in diagnosis.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Section 5: 510(k) Summary
K210611
# Section 5 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act 1990 and 21 CFR § 807.92.
## 1. General Information
| Establishment: | Siemens Medical Solutions USA, Inc. |
|----------------|-------------------------------------|
| | 40 Liberty Boulevard |
| | Mail Code 65-1A |
| | Malvern, PA 19355, USA |
| | Registration Number: 2240869 |
#### June 30, 2021 Date Prepared:
Siemens Shenzhen Magnetic Resonance Ltd. Manufacturer: Siemens MRI Center, Gaoxin C. Ave., 2nd Hi-Tech Industrial Park 518057 Shenzhen PEOPLE'S REPUBLIC OF CHINA Registration Number: 3004754211
> Siemens Healthcare GmbH Henkestrasse 127 91052 Erlangen Germany Registration Number: 3002808157
#### 2. Contact Information
Andrew Turner, Requlatory Affairs Technical Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Mail Code 65-1A Malvern, PA 19355, USA Phone: (610) 850-5627 Fax: (610) 448-1787 E-mail: andrew.turner@siemens-healthineers.com
Siemens MR System: MAGNETOM Free.Max(0.55T) with Software Version syngo MR XA40A
{4}------------------------------------------------
#### Device Name and Classification 3.
| Device name: | MAGNETOM Free.Max |
|-----------------------|---------------------------------------------|
| Trade name: | MAGNETOM Free.Max |
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology |
| CFR Code: | 21 CFR § 892.1000 |
| Classification: | II |
| Product Code: | Primary: LNH<br>Secondary: MOS |
## 4. Legally Marketed Predicate Device
| Trade name: | MAGNETOM Sempra |
|-----------------------|---------------------------------------------|
| 510(k) Number: | K183221 |
| Clearance Date: | February 14, 2019 |
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology |
| CFR Code: | 21 CFR § 892.1000 |
| Classification: | II |
| Product Code: | Primary: LNH<br>Secondary: LNI, MOS |
## 5. Intended Use
The indications for use for the subject devices are within that of the predicate device:
The MAGNETOM MR system is indicated for use as a magnetic resonance diagnostic device (MRDD), which produces transverse, sagittal, coronal, and oblique cross sectional images that display the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images may also be produced. Depending on the region of interest, contrast agents may be used.
These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist in diagnosis.
#### 6. Device Description
The subject device, MAGNETOM Free.Max with software syngo MR XA40A, is an 80 cm bore Magnetic Resonance Imaging system with an actively shielded 0.55T superconducting magnet. Which is the first 0.55T MRI system for clinical use in the U.S.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a triangular shape.
MAGNETOM Free.Max is a MRI system that has equivalencies to the predicate device MAGNETOM Sempra with syngo MR XA12 (K183221, cleared on February 14, 2019). Whilst the subject device comes with significant new and modified hardware, the resultant customer clinical value is based on the predicate device MAGNETOM Sempra with syngo MR XA12 (K183221).
A high level summary of the new hardware comparing to the predicate device is provided below:
| Hardware New Hardware | - | Magnet with no Quench Pipe including<br>Autoramp up and Ramp down capability |
|-----------------------|---|---------------------------------------------------------------------------------------|
| | - | Gradient Coil and Gradient Power Amplifier |
| | - | System Cover |
| | - | Patient Table (fixed with/without vertical drive) |
| | - | RF system |
| | - | Select&GO Display (TPAN_2G) and Control<br>Panel (CPAN_2G) |
| | - | Patient Communication Unit (Intercom with<br>System Power On/Off Function integrated) |
| | - | Computer system |
| | - | myExam 3D Camera |
| New Coils | - | Body Coil |
| | - | Head/Neck Coil |
| | - | Spine Coil |
| | - | BioMatrix Contour L Coil |
| | - | Contour S Coil |
lole software syngo MR XA40A brings new/modified software features:
| Software | New Applications/<br>Software Features |
|----------|-------------------------------------------|
| | - Sequence Booster |
| | - SMS Averaging |
| | - SWI with 3D segmented EPI |
| | - Ramp sampling mode |
| | - Skewed SPAIR |
| | - Accelerated Shim (STEAM-based field map |
| | - Deep Resolve Gain |
| | - Deep Resolve Sharp |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.
The following new features for the subject device are migrated from the reference software version syngo MR XA30A (K202014, September 8, 2020):
- ı Motion correction for Haste with multiple averages
- myExam Companion, which consists of: New Software Platform/ Workflow
- myExam 3D camera ー
- myExam Cockpit (new name for Dot — Cockpit)
- myExam Assist (new name for Dot | Engines)
- myExam Autopilot.
- New Scanner User Interface: ।
- Manual positioning —
- Select&GO extension
- Guidance of general patient setup, and coil preparation
- । System functional safety test
Modified Software Software Platform/ Workflow
ı Improved System on/off/standby behavior
# 7. Substantial Equivalence
MAGNETOM Free.Max with software syngo MR XA40A is substantially equivalent to the predicate device:
| Predicate Device | FDA Clearance Number<br>and Date | Product<br>Code | Manufacturer |
|---------------------------------------|---------------------------------------|-----------------|----------------------------|
| MAGNETOM Sempra<br>with syngo MR XA12 | K183221,<br>cleared February 14, 2019 | LNH<br>LNI, MOS | Siemens Healthcare<br>GmbH |
The subject software syngo MR XA40A is modified based on the reference software XA30A; MAGNETOM C!, which is a low field device(0.35T) with permanent magnet is listed as a reference device as well:
| Reference Devices | FDA Clearance Number<br>and Date | Product<br>Code | Manufacturer |
|--------------------------------------|---------------------------------------|------------------|------------------------------------------------|
| MAGNETOM Aera with<br>syngo MR XA30A | K202014,<br>cleared September 8, 2020 | LNH,<br>LNI, MOS | Siemens AG / Siemens<br>Healthcare GmbH |
| MAGNETOM C! | K082331,<br>cleared October 1, 2008 | LNH,<br>MOS | Siemens Shenzhen<br>Magnetic Resonance<br>Ltd. |
{7}------------------------------------------------
#### Technological Characteristics 8.
The subject device. MAGNETOM Free.Max with software syngo MR XA40A. is substantially equivalent to the predicate device with regard to the operational environment, programming language, operating system and performance.
The subject device conforms to the standard for medical device software (IEC 62304) and other relevant IEC and NEMA standards.
There are some differences in technological characteristics between the subject device and predicate device, including different hardware and modified software. These differences have been tested and the conclusions from the non-clinical data suggests that the features bear an equivalent safety and performance profile to that of the predicate device.
# 9. Nonclinical Tests
| Performance Test | Tested Hardware or<br>Software | Source/Rationale for test |
|----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Sample clinical images | Coils, new and modified<br>software features, pulse<br>sequence types | Guidance for Submission of<br>Premarket Notifications for<br>Magnetic Resonance<br>Diagnostic Devices |
| Image quality assessments<br>by sample clinical images.<br>In some cases, a<br>comparison of the image<br>quality / quantitative data<br>was made. | - new / modified pulse<br>sequence types and<br>algorithms.<br>- comparison images<br>between the new / modified<br>features and the predicate<br>device features | Diagnostic Devices |
| Performance bench test | - SNR and image uniformity<br>measurements for coils<br>- heating measurements for<br>coils | |
| Software verification and<br>validation | mainly new and modified<br>software features | Guidance for the Content of<br>Premarket Submissions for<br>Software Contained in<br>Medical Devices |
The following performance testing was conducted on the subject devices.
The results from each set of tests demonstrate that the devices perform as intended and are thus substantially equivalent to the predicate device to which it has been compared.
## 10. Clinical Tests / Publications
In order to practically learn Peripheral Nerve Stimulation (PNS) effects of the subject system, a clinical study of 12 individuals were conducted (see Results of stimulation study in Appendix 3).
{8}------------------------------------------------
No additional clinical tests were conducted to support substantial equivalence for the subject device; however, as stated above, sample clinical images were provided. Clinical publications were referenced to provide information on the use of the following features and functions.
| Feature / Function | Clinical Publication |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SMS Averaging for TSE | [1] C. H. OH et al., 1984. Line-Integral Projection<br>Reconstruction (LPR) with Slice Encoding Techniques:<br>Multislice Regional Imaging in NMR Tomography, IEEE<br>Trans Med Imaging. 1984; 3(4):170-178 |
| Ramp sampling mode<br>for Beat | [2] Matt A.Bernstein, Kevin F.king, Xiaohong Joe Zhou,<br>Handbook Of MRI Pulse Sequences, 2004: P706-P712. |
| SWI with 3D segmented<br>EPI | [3] Zwanenburg JJ, et al. Fast high resolution whole brain<br>T2* weighted imaging using echo planar imaging at 7T.<br>Neuroimage. 2011; 56:1902-1907. |
| Accelerated Shim<br>(STEAM-based field<br>map) | [4] Liu W, et al. 3D Flow Compensated Interleaved EPI for a<br>Fast High-Resolution Susceptibility-Weighted Imaging at<br>1.5T. Proc. Intl. Soc. Mag. Reason. Med 2019; 27: 3326. |
| | [5] Nehrke, K., and Börnert, P. "DREAM—a novel approach<br>for robust, ultrafast, multislice B1 mapping." Magnetic<br>resonance in medicine 68.5 (2012): 1517-1526. |
| Skewed SPAIR | [6] Hwang, T. L., van Zijl, P. C., and Garwood, M. 1999.<br>Asymmetric adiabatic pulses for NH selection. J. Magn.<br>Reson. 138: 173-177. |
| | [7] Pfeuffer J. et al. "Zoomed Functional Imaging in the<br>Human Brain at 7 Tesla with Simultaneous High Spatial and<br>High Temporal Resolution” Neurolmage 17(1), 2002, p. 272-<br>286. |
| Deep Resolve Gain | [8] Kellman P. et al. Image Reconstruction in SNR Units: A<br>General Method for SNR Measurement. MRM 2005;<br>54:1439. Erratum in MRM 2007; 58:311. |
| | [9] Blu T. et al. The SURE-LET approach to image denoising<br>IEEE Transactions on Image Processing 16(11):2778-86 |
| Deep Resolve Sharp | [10] Yulun Zhang, et al. Residual Dense Network for Image<br>Super-Resolution, IEEE conference on computer vision and<br>pattern recognition. 2018 |
| | [11] Eirikur Agustsson, Radu Timofte. NTIRE 2017<br>Challenge on Single Image Super-Resolution: Dataset and<br>Study. CVPRW, 2017. 5 |
| | [12] Justin Johnson, et al. Perceptual Losses for Real-Time<br>Style Transfer and Super-Resolution. European conference<br>on computer vision. Springer, Cham, 2016 |
Siemens MR System: MAGNETOM Free.Max(0.55T) with Software Version syngo MR XA40A
{9}------------------------------------------------
## 11. Safety and Effectiveness
The device labeling contains instructions for use and any necessary cautions and warnings to ensure safe and effective use of the device.
Risk Management is ensured via a risk analysis in compliance with ISO 14971, to identify and provide mitigation of potential hazards early in the design cycle and continuously throughout the development of the product. Siemens adheres to recognized and established industry standards, such as the IEC 60601-1 series, to minimize electrical and mechanical hazards. Furthermore, the device is intended for healthcare professionals familiar with and responsible for the acquisition and post processing of magnetic resonance images.
MAGNETOM Free.Max with software syngo MR XA40A conform to the following FDA recognized and international IEC, ISO and NEMA standards:
| Recognition<br>Number | Product<br>Area | Title of Standard | Reference<br>Number and date | Standards<br>Development<br>Organization |
|-----------------------|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|------------------------------------------|
| 19-4 | General | Medical electrical equipment -<br>part 1: general requirements for<br>basic safety and essential<br>performance | ES60601-<br>1:2005/(R) 2012<br>and A1:2012 | AAMI / ANSI |
| 19-8 | General | Medical electrical equipment -<br>Part 1-2: General requirements<br>for basic safety and essential<br>performance - Collateral<br>Standard: Electromagnetic<br>disturbances - Requirements<br>and tests | 60601-1-2, Ed.<br>4.0:2014 | IEC |
| 12-295 | Radiology | Medical electrical equipment -<br>Part 2-33: Particular<br>requirements for the basic<br>safety and essential<br>performance of magnetic<br>resonance equipment for<br>medical diagnosis | 60601-2-33, Ed.<br>3.2 b:2015 | IEC |
| 5-40 | General | Medical devices - Application of<br>risk management to medical<br>devices | 14971, Ed. 2:2007 | ISO |
| 5-114 | General | Medical devices - Application of<br>usability engineering to medical<br>devices | 62366, Edition<br>1.0:2015 | AAMI<br>ANSI<br>IEC |
| 13-79 | Software | Medical device software -<br>Software life cycle processes | 62304:2006 +<br>A1:2015 | AAMI<br>ANSI<br>IEC |
| 12-232 | Radiology | Acoustic Noise Measurement<br>Procedure for Diagnosing<br>Magnetic Resonance Imaging<br>Devices | MS 4:2010 | NEMA |
### Traditional Premarket Notification 510(k)
{10}------------------------------------------------
## Section 5: 510(k) Summary
| 12-288 | Radiology | Characterization of Phased<br>Array Coils for Diagnostic<br>Magnetic Resonance Images<br>(MRI) | MS 9:2008 | NEMA |
|--------|------------------|-------------------------------------------------------------------------------------------------------------------------------------------|--------------------|---------------------|
| 12-300 | Radiology | Digital Imaging and<br>Communications in Medicine<br>(DICOM) Set 03/16/2012<br>Radiology | PS 3.1 - 3.20:2016 | NEMA |
| 2-258 | Biocompatibility | biological evaluation of medical<br>devices - part 1: evaluation and<br>testing within a risk<br>management process<br>(Biocompatibility) | 10993-1:2018 | AAMI<br>ANSI<br>ISO |
# 12. Conclusion as to Substantial Equivalence
MAGNETOM Free.Max with software syngo MR XA40A has the same basic technological characteristics as the predicate device, MAGNETOM Sempra with syngo MR XA12, with respect to the magnetic resonance features and functionalities. And the indications for use of the subject device falls within the intended use of the predicate device.
While there are some differences in technical features compared to the predicate device, the differences have been tested and the conclusions from all verification and validation data suggest that the features bear an equivalent safety and performance profile to that of the predicate device and reference devices.
Siemens believes that MAGNETOM Free.Max with software syngo MR XA40A is substantially equivalent to the currently marketed device MAGNETOM Sempra with software syngo MR XA12 (K183221, cleared on February 14, 2019).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.