K082331 · Siemens Medical Solutions USA, Inc. · LNH · Oct 1, 2008 · Radiology
Device Facts
Record ID
K082331
Device Name
MAGNETOM C!
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LNH · Radiology
Decision Date
Oct 1, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The MAGNETOM C! is indicated for use as magnetic resonance diagnostic devices (MRDD) that produce transverse, sagittal, coronal and oblique cross sectional images, and that display the internal structure and/or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images when interpreted by a trained physician yield information that may assist in diagnosis. The MAGNETOM C! may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in room display and MR safe biopsy needles. Disclaimer: Utility of contrast enhanced breast MRI for the detection of breast masses has not been documented and is thus not indicated.
Device Story
MAGNETOM C! is an open, whole-body MRI scanner; produces transverse, sagittal, coronal, and oblique cross-sectional images of head, body, or extremities. Input: magnetic resonance signals; output: diagnostic images for physician interpretation. Used in clinical settings for diagnostic imaging and interventional procedures (e.g., biopsies) with MR-compatible tools. Modifications include updates to RF infrastructure, signal unit, filter plate, gradient amplifier, physiological measurement unit, measurement/reconstruction system, magnet temperature unit, patient table control, power distribution, external field interference shielding, and cooling system. System provides high-resolution imaging and reduced scan times; assists physicians in clinical diagnosis.
Clinical Evidence
Bench testing only. Performance evaluated against NEMA standards for signal-to-noise ratio, image uniformity, geometric distortion, high-contrast spatial resolution, and slice thickness. No clinical data provided.
Technological Characteristics
Open, whole-body MRI scanner. Modifications include RF infrastructure, RF signal unit, gradient amplifier, and integrated cooling system. Conforms to NEMA standards for performance measurement and IEC standards for safety. Software updated to support hardware modifications.
Indications for Use
Indicated for patients requiring diagnostic MR imaging of head, body, or extremities; supports interventional procedures using MR-compatible devices. Not indicated for contrast-enhanced breast MRI for mass detection.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
### 510(k) Summary 5
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92.
# l. General Information
| Establishment | Siemens Medical Solutions. Inc.<br>51 Valley Stream Parkway<br>Malvern. PA 19355 |
|---------------|----------------------------------------------------------------------------------|
|---------------|----------------------------------------------------------------------------------|
Registration Number 2240869
- Manufacturer Siemens Mindit Magnetic Resonance Ltd. Siemens MRI Center, Gaoxin C. Ave. 2nd Hi-Tech Industrial Park, ShenZhen 518057, PR. China
- Registration Number 3004754211
Contact Person Ms. Judy Campbell Technical Specialist. Regulatory Submissions 51 Valley Stream Parkway Malvern. PA 19355 Phone: (610)448-4918 Fax: (610) 448-1787
| Device Name | |
|----------------------|--------------------------------------|
| Trade Name: | MAGNETOM C! System |
| Classification Name: | Magnetic Resonance Diagnostic Device |
| CFR Code: | 21 CFR § 892.1000 |
| Classification: | Class II |
# Performance Standards
None established under Section 514 the Food, Drug and Cosmetic Act.
{1}------------------------------------------------
# II. Safety and Effectiveness Information Supporting Substantial Equivalence.
#### Intended Use
The MAGNETOM C! is indicated for use as magnetic resonance diagnostic devices (MRDD) that produce transverse, sagittal, coronal and oblique cross sectional images, and that display the internal structure and/or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images when interpreted by a trained physician yield information that may assist in diagnosis.
The MAGNETOM C! may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in room display and MR safe biopsy needles.
Disclaimer: Utility of contrast enhanced breast MRI for the detection of breast masses has not been documented and is thus not indicated.
### Device Description
The MAGNETOM C! system is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times, which was described in premarket notification K043030 which received FDA clearance on December, 10, 2004. Siemens intends to modify the RF Infra Structure, RF Signal Unit, Filter Plate, Gradient Amplifier, Physiological Measurement Unit, Measurement and Reconstruction System, Magnet Temperature Unit, Patient Table Control, Power Distribution System, External Field Interference, Integrated Cooling System and software update for the existing MAGNETOM C! Magnetic Resonance System.
#### Substantial Equivalence
Siemens believes that, within the meaning of the Safe Medical Device Act of 1990, the MAGNETOM C!, which is configured with modified hardware and software, is substantially equivalent to the following cleared medical devices, which offers the same applications and handling:
| Predicate Device Name | FDA Clearance<br>Number | FDA Clearance<br>Date |
|----------------------------------|-------------------------|-----------------------|
| Siemens MAGNETOM 1.5T<br>ESSENZA | K071925 | 08/14/2007 |
| Siemens MAGETOM 0.35 T C ! | K043030 | 12/10/2004 |
{2}------------------------------------------------
### General Safety and Effectiveness Concerns
Operation of the MAGNETOM C! System is substantially equivalent to the commercially available MAGNETOM C! (K043030) System and 1.5 T ESSENZA System (071925). Below are the parameters specified by the FDA guidance document for MR Diagnostic Devices that will be evaluated based on the modifications:
Performance Levels
- · Signal to Noise
- · Image Uniformity
- · Geometric Distortion
- · High Contrast Spatial Resolution
- · Slice Thickness
Safety Levels
### None
The MAGNETOM C! will conform to the FDA recognized NEMA Standards for the measurement of performance and safety parameters and the international IEC standard for safety issues with Magnetic Resonance Imaging Devices This will assure that the performance of this device can be considered safe and effective with respect to the currently available MAGNETOM C!( K043030) and MAGNETOM ESSENZA (K071925).
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and care. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", indicating the department's name and national affiliation.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 01 2008
Ms. Judith Campbell Regulatory Technical Specialist Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway MALVERN PA 19355
Re: K082331
Trade/Device Name: MAGNETOM C! Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Codc: LNH and MOS Dated: August 11, 2008 Received: August 14, 2008
Dear Ms. Campbell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device. can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
hoque An Whang
Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use Statement 4
510(k) Number (if known)
Device Name: MAGNETOM C!
# Indications for Use:
The MAGNETOM C! is indicated for use as maqnetic resonance diagnostic devices (MRDD) that produce transverse, sagittal, coronal and oblique cross sectional images, and that display the internal structure and/or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images when interpreted by a trained physician yield information that may assist in diagnosis.
The MAGNETOM C! may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in room display and MR safe biopsy needles
Disclaimer: Utility of contrast enhanced breast MRI for the detection of breast masses has not been documented and is thus not indicated.
(please do not write below this line- continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation
Prescription Use ✔
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Over-The-Counter Use
tion Use
OR
(Division Division of Reproductive, Abdominal and Radiological Devices 510(k) Number __
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.