K210550 · Ultradent Products, Inc. · EBZ · Mar 1, 2021 · Dental
Device Facts
Record ID
K210550
Device Name
VALO Grand Corded and Accessory Lenses
Applicant
Ultradent Products, Inc.
Product Code
EBZ · Dental
Decision Date
Mar 1, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6070
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
VALO Grand Corded: Source of illumination for curing photo-activated dental restorative materials and adhesives. VALO Accessory Lenses: The VALO Accessory Lenses are multiple-use accessory lenses intended to provide illumination to aid in visualization during oral procedures and augment the VALO family of curing lights, which are a source of illumination for curing photo-activated dental restorative materials and adhesives. VALO Accessory Lenses are not intended for complete cure of photo-activated materials and adhesives.
Device Story
VALO Grand Corded is a corded LED dental curing light; operates in broadband spectrum (385-515nm) to polymerize photo-activated dental materials. Accessory lenses (PointCure, ProxiCure, TransLume Green, Black Light, Interproximal, White Light) attach to the handpiece to provide specialized illumination for visualization, transillumination, or localized curing. Used by dentists/dental professionals in clinical settings. Handpiece powered by AC supply (100-240V). Output intensity varies by mode (Standard, High Power Plus, Xtra Power). Lenses augment visualization of oral structures, prostheses, or fluorescing resins. Clinical benefit includes improved procedural visualization and efficient material polymerization.
Clinical Evidence
Bench testing only. Testing included radiant exitance and emission spectrum per ISO 10650:2018, illuminance and color temperature testing, beam profile testing, duty cycle testing, and disinfection validation.
Technological Characteristics
Handpiece: Anodized aluminum. Accessory lenses: Delrin housing, Neodymium magnets, Polymethylpentene (TPX) molded lenses, glass filters, phosphor filters. Light source: LED (blue/violet). Power: AC wall-powered (9VDC, 500mA). Connectivity: Standalone. Standards: ISO 10650:2018, IEC 80601-2-60:2019, ISO 14971:2007, IEC 62366-1:2015, ISO 10993-1:2018.
Indications for Use
Indicated for dentists and dental professionals for the illumination and polymerization of photo-activated dental restorative materials and adhesives, and for visualization aid during oral procedures.
Regulatory Classification
Identification
An ultraviolet activator for polymerization is a device that produces ultraviolet radiation intended to polymerize (set) resinous dental pit and fissure sealants or restorative materials by transmission of light through a rod.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 1, 2021
Ultradent Products, Inc % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059
Re: K210550
Trade/Device Name: VALO Grand Corded and Accessory Lenses Regulation Number: 21 CFR 872.6070 Regulation Name: Ultraviolet activator for polymerization Regulatory Class: Class II Product Code: EBZ,EAQ, PEQ Dated: February 19, 2021 Received: February 25, 2021
Dear Dave Yungvirt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
## K210550
Device Name VALO Grand Corded and Accessory Lenses
Indications for Use (Describe) VALO Grand Corded: Source of illumination for curing photo-activated dental restorative materials and adhesives.
#### VALO Accessory Lenses:
The VALO Accessory Lenses are multiple-use accessory lenses intended to provide illumination to aid in visualization during oral procedures and augment the VALO family of curing lights, which are a source of illumination for curing photo-activated dental restorative materials and adhesives. VALO Accessory Lenses are not intended for complete cure of photo-activated materials and adhesives.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K210550
# Section 5: 510(k) Summary
This summary of substantial equivalence information is being submitted in accordance with the requirements of 21 CFR 807.92 for VALO™ Grand Corded and Accessory Lenses.
## I. Applicant's Name and Address
Ultradent Product, Inc. 505 West Ultradent Drive (10200 South) South Jordan, UT 84095
| Contact Person: | Mr. Adam Black |
|-----------------|----------------------------|
| Title: | Regulatory Affairs Manager |
| Telephone: | 801-553-4425 |
| Fax: | 801-553-4609 |
Date Summary Prepared: 12 February 2021
### II. Name of the Device
| Device: | Dental Curing Light |
|------------------------------|-----------------------------------------|
| Trade/Device Name: | VALO™ Grand Corded and Accessory Lenses |
| Review Panel: | Dental |
| Regulation Number: | 21 CFR 872.6070 |
| Device Class: | Class II |
| Classification Product Code: | EBZ |
| Subsequent Product Code: | EAQ |
### III. Device Description
### VALO™ Grand Corded:
With its broadband spectrum, VALO Grand Corded is designed to polymerize all light cured products in the wavelength range of 385-515nm per ISO 10650.
VALO has a medical grade, international power supply and is suitable for power outlets from 100 to 240 volts. The handpiece is designed to rest in a standard dental unit bracket or can be custom mounted using the bracket included with the kit.
### Accessory Lenses:
- . PointCure™ Lens
- o Augments the VALO curing light to polymerize composite through a translucent prosthetic.
- . ProxiCure™ Ball Lens
- o Augments the VALO curing light to polymerize composite and help shape contact area matrix of an interproximal restoration.
{4}------------------------------------------------
- TransLume™ Green Lens
- Augments the VALO curing light to aid in visualization by providing longer wavelength light to o transilluminate teeth and dental prostheses.
- Black Light Lens
- o Augments the VALO curing light to provide aid in visualization of fluorescing chemicals in dental resins.
- Interproximal Lens
- Augments the VALO curing light to aid in visualization of teeth and dental prostheses. o
- White Light Lens
- Augments the VALO curing light to provide a visual aid for accurate color/shade comparison or o whenever natural light is needed.
#### IV. Statement of Intended Use
#### VALO Grand Corded:
Source of illumination for curing photo-activated dental restorative materials and adhesives.
#### VALO Accessory Lenses:
The VALO Accessory Lenses are multiple-use accessory lenses intended to provide illumination to aid in visualization during oral procedures and augment the VALO family of curing lights, which are a source of illumination for curing photo-activated dental restorative materials and adhesives. VALO Accessory Lenses are not intended for complete cure of photo-activated materials and adhesives.
### V. Predicate Device
VALO Grand Corded and Accessory Lenses identified primary predicate: K190627 – VALO™ Grand Corded by Ultradent Products, Inc and secondary predicate: K101140 – FUSION-DOE by DentLight Inc.
### VII. Comparison of Technological Characteristics
#### Predicate technological comparison:
The technology, delivery, and intended use of VALO™ Grand Corded and Accessory Lenses are substantially equivalent to the identified predicate devices as outlined in Table 5-1:
{5}------------------------------------------------
## Table 5-1: VALO™ Grand Corded and Accessory Lenses substantial equivalence comparison
| Descriptive<br>Information | Device: VALO™ Grand Corded and<br>Accessory Lenses | Primary Predicate:<br>VALO™ Grand Corded<br>(K190627) | Secondary Predicate:<br>Dentlight Oral Exam Light Kit<br>("DOE") (K101140) |
|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code/<br>Classification | EBZ, EAQ - Class II | EBZ — Class II | EAQ — Class II |
| Indications for<br>Use | VALO Grand Corded:<br>The source of illumination for curing<br>photo-activated dental restorative<br>materials and adhesives.<br><br>Accessory Lenses: The VALO<br>Accessory Lenses are multiple-use<br>accessory lenses intended to provide<br>illumination to aid in visualization<br>during oral procedures and augment<br>the VALO family of curing lights, which<br>are a source of illumination for curing<br>photo-activated dental restorative<br>materials and adhesives. VALO<br>Accessory Lenses are not intended for<br>complete cure of photo-activated<br>materials and adhesives. | The source of illumination for<br>curing photo-activated dental<br>restorative materials and<br>adhesives. | Dentlight Oral Exam Light Kit is<br>indicated to be used by a<br>dentist and physician for<br>illumination to aid in<br>visualization during oral<br>procedures and an adjunct to<br>enhance the visualization for<br>oral examination of mucosal<br>abnormalities and oral lesions |
| Intended User | Dentist or dental professional | Dentist or dental professional | Dentist or dental<br>professional |
| Device Design:<br>Power Source | AC Power Supply:<br>Wall powered. | AC Power Supply:<br>Wall powered. | Rechargeable Li-Ion batteries |
| | Output: 9VDC, 500mA. Input: 100VAC -<br>240VAC with adapters for international<br>capability. Ratings: Medical Grade, (UL, CE, RoHS,<br>WEEE) Cord: 6 ft (1.8m), 2.5mm DC connector<br><br>Power On Button: Located on the<br>handle of the wand, back side and front<br>side<br>UL Approved | Output: 9VDC, 500mA. Input:<br>100VAC - 240VAC with adapters<br>for international capability.<br>Ratings: Medical Grade, (UL, CE, RoHS,<br>WEEE) Cord: 6 ft (1.8m), 2.5mm DC connector<br><br>Power On Button: Located on<br>the handle of the wand, back<br>side and front side<br>UL Approved | |
| Device Design:<br>Operational<br>Modes (Curing<br>EBZ) | VALO Grand Corded:<br>*Standard Power Mode: 1000 mW/cm2<br>*High Power Plus Mode: 1600mW/cm2<br>Xtra Power Mode: 3200mW/cm2<br>*as measured by Demetron<br><br>Accessory Lenses:<br>≥800 mW/cm2 (PointCure –<br>Recommened with High Power Mode)<br>(ProxiCure – Recommended with mode<br>suitable for material) | *Standard Power Mode: 1000<br>mW/cm2<br>*High Power Plus Mode:<br>1600mW/cm2<br>Xtra Power Mode: 3200mW/cm2<br>*as measured by Demetron | N/A |
| Device Design:<br>Operational<br>Modes<br>(Diagnostic EAQ) | Accessory Lenses:<br>≥25 mW/cm2, ≤420 nm wavelength<br>(Black Light Lens – Recommended with<br>Standard Curing mode)<br><br>≥1,000 lx luminescence, 5,000-6,000 K<br>color temperature, ≥85 CRI (White<br>Light Lens – Recommended with<br>Standard Curing mode)<br><br>≥15,000 lx luminescence, 500-570nm<br>peak wavelengths (TransLume Green<br>Lens – Recommended with Standard<br>Curing mode) | N/A | Violet or Blacklight: 395 – 420<br>nm peak wavelength, ≥25<br>mW/cm2* (Can be used in either<br>mode)<br><br>White light: ≥1,000 lx<br>luminescence*, 5,000 - 6,000 K<br>color temperature*, ≥65 CRI*<br>(Can be used in either mode)<br><br>*According to Manufacturer<br>marketing material |
| ≥15,000 lx luminescence (Interproximal Lens - Recommended with Standard Curing mode) | | | |
| Device Design: Light Source | LED light, blue and violet wavelengths<br>12mm head size | LED light, blue and violet wavelengths<br><br>12mm head size | High Power LED |
| Device Design:<br>Accessories | Barrier Sleeve VALO™,<br>Blue-Light Blocking Glasses (VALO™ Grand Corded)<br><br>PointCure Lens, ProxiCure Ball Lens,<br>Translume Green Lens, Black Light Lens,<br>Inter-proximal Lens, White Light Lens | Barrier Sleeve VALO™,<br>Blue Light Blocking Glasses | -Light blocking safety glasses<br>-Safety light blocking shields<br>-Full accessories for fluorescent<br>and transillumination oral exams<br>-Tacking Tip<br>-Curing Caps (3mm, 9mm, 13mm) |
| Composition of<br>Materials | VALO Grand Corded: Aluminum, anodized black<br><br>Accessory Lenses:<br>Lens housing - Delrin<br>Lens housing magnet – Neodymium<br>Molded lens – Polymethylpentene TPX<br>RT18 (PointCure, ProxiCure, TransLume Green, Interproximal)<br>Black Light Filter - Glass<br>White Light Filter – Type 2 phosphor | Aluminum, anodized black | Solid aircraft aluminum |
| Technical<br>Specifications:<br>Light Intensity | VALO Grand Corded: Standard: 900 mW/cm²<br>High: 1,500 mW/cm²<br>Xtra: 2,100 mW/cm²<br>As measured by traceable Gigahertz<br>spectrum analyzer<br><br>Accessory Lenses: | Standard: 900 mW/cm²<br>High: 1,500 mW/cm²<br>Xtra: 2,100 mW/cm²<br>As measured by traceable<br>Gigahertz spectrum analyzer | Curing Mode:<br>Pulse Mode: 2,000 mW/cm²<br>Plasma Mode: 4,000 mW/cm²<br><br>Violet Light Head: |
| | | | |
| | ≥800 mW/cm² (PointCure/ProxiCure)<br>≥25 mW/cm², ≤420 nm wavelength<br>(Black Light Lens)<br>≥1,000 lx luminescence, 5,000-6,000 K<br>color temperature, ≥85 CRI (White<br>Light Lens)<br>≥15,000 lx luminescence, 500-570nm<br>peak wavelengths (TransLume Green<br>Lens)<br>≥15,000 lx luminescence (Interproximal<br>Lens) | | ≥25 mW/cm²* (Can be used in<br>either mode)<br><br>White Light Head:<br>≥1,000 lx luminescence*<br><br>*According to Manufacturer<br>marketing material |
| Technical<br>Specifications:<br>Peak<br>Wavelength | VALO Grand Corded: Nominal values:<br>395-415nm and 440-480nm<br><br>Accessory Lenses:<br>All lenses match the curing lights peak<br>wavelengths except:<br>≤420 nm wavelength (Black Light Lens),<br>500-570nm peak wavelengths<br>(TransLume Green Lens) | Nominal values: 395-415nm and<br>440-480nm | Nominal values for curing light<br>source and White Light Head:<br>440-480nm<br><br>Violet Light Head: 395-420nm |
| Technical<br>Specifications:<br>Depth of Cure | 2mm | 2mm | 2mm |
| Recognized<br>Standards | ISO 10650:2018<br>IEC 80601-2-60:2019<br>ISO 14971:2007/(R)2010<br>IEC 62366-1:2015<br>ISO 10993-1:2018 | ISO 10650:2015<br>IEC 80601-2-60:2012<br>IEC 60601-1 Ed. 3.1<br>ISO 14971:2007/(R)2010<br>IEC 62366-1:2015 | ADA 48:2004<br>ISO 10650:2007<br>IEC 60601-1:2007<br>ISO 14971:2007 |
| Parameters of<br>Disinfection | VALO Grand Corded:<br>Acceptable Cleaners -<br>Lysol Brand III Disinfectant Spray<br>(Recommended)<br>Cavicide products (non-bleach) | Acceptable Cleaners:<br>Lysol Brand III Disinfectant Spray<br>(Recommended)<br>Cavicide products (non-bleach)<br>Isopropyl alcohol | Acceptable Cleaners:<br>Lysol Brand III Disinfectant Spray<br>(Recommended)<br>Cavicide products (non-bleach)<br>Isopropyl alcohol |
| Isopropyl alcohol | Ethyl alcohol-based cleaners | Ethyl alcohol-based cleaners | |
| Ethyl alcohol-based cleaners | Lysol Concentrate (alcohol- | Lysol Concentrate (alcohol- | |
| Lysol Concentrate (alcohol-based only) | based only) | based only) | |
| <b>Accessory Lenses:</b><br>Cleaners – Henry Schein General | | | |
| Purpose Cleaner or equivalent product | | | |
| Disinfectant - Cidex® OPA Solution or | | | |
| equivalent product | | | |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
As outlined in the comparison tables above, the VALO™ Grand Corded and Accessory Lenses are similar to the identified primary and secondary predicate devices with respect to intended use, power sources, light sources, principal of operation, intended user, and biocompatibility.
As this submission is to declare a significant change of introducing new accessories to a previously cleared device, the testing completed to demonstrate equivalence with the predicate included compatibility and functionality with the proposed new accessories. These tests show that the proposed VALO™ Accessory Lenses supplement the use of parent device of VALO™ Grand Corded. The proposed VALO™ Accessory Lenses supplement the VALO™ Grand Corded curing light as outlined above.
Device design validation and verification activities have been performed to FDA Guidance Document Dental Curing Lights – Premarket Notification [510(k)] and recognized standards and via internal testing protocols.
Non-clinical tests performed to establish substantial equivalence to identified predicate devices included radiant exitance and emission spectrum testing according to ISO 10650:2018, illuminance and color temperature testing, beam profile testing, duty cycle testing, and disinfection validation testing.
Conclusion: Based on these comparisons to the predicate device, we believe that VALO™ Grand Corded and Accessory Lenses are substantially equivalent to the predicate devices, in that they augment the use of the already cleared predicate, and do not raise new concerns of safety or efficacy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.