The source of illumination for curing photo-activated dental restorative materials and adhesives.
Device Story
VALO Grand Corded is a mains-powered, hand-held LED dental curing light. It emits light in the 395-480 nm range to polymerize photo-initiated dental resins and adhesives. Operated by dentists or dental professionals in a clinical setting, the device features three power modes (Standard, High Power Plus, Xtra Power) ranging from 1000 to 3200 mW/cm2. The device includes a 12mm light head, three buttons (two for power, one for mode selection), and is used with blue-light blocking glasses for eye protection. The clinician manually selects illumination time and power settings to cure restorative materials. The device benefits patients by providing a larger curing area compared to the predicate, facilitating efficient polymerization of dental composites.
Clinical Evidence
Bench testing only. Verification and validation activities were performed in accordance with FDA Guidance Document 'Dental Curing Lights – Premarket Notification [510(k)]' and recognized standards (including ADA) to confirm performance equivalence to the predicate.
Indicated for use by dentists or dental professionals as a source of illumination for curing photo-activated dental restorative materials and adhesives.
Regulatory Classification
Identification
An ultraviolet activator for polymerization is a device that produces ultraviolet radiation intended to polymerize (set) resinous dental pit and fissure sealants or restorative materials by transmission of light through a rod.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 29, 2019
Ultradent Products, Inc. Karen Kakunes Regulatory Affairs Manager 505 West Ultradent Drive (10200 South) South Jordan, Utah 84095
Re: K190627
Trade/Device Name: VALO Grand Corded Regulation Number: 21 CFR 872.6070 Regulation Name: Ultraviolet Activator For Polymerization Regulatory Class: Class II Product Code: EBZ Dated: March 8, 2019 Received: March 12, 2019
Dear Karen Kakunes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# VALO® Grand Corded Special 510(K)
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K190627
Device Name VALO Grand Corded
Indications for Use (Describe)
The source of illumination for curing photo-activated dental restorative materials and adhesives.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo in brown color. The logo is a stylized design that resembles the letters 'U' and 'E' connected together. The symbol '®' is located at the top right corner of the logo. Below the logo, the text 'K190627' is displayed.
# Special 510(k) Summary for VALO® Grand Corded
This summary of the Special 510(k) substantial equivalence information is being submitted in accordance with the requirements of 21 CFR 807.92 for VALO® Grand Corded.
#### l. Applicant's Name and Address
Ultradent Products, Inc. 505 West 10200 South South Jordan, UT 84095 FDA Establishment Registration #: 1718912
| Contact Person: | Ms. Karen Kakunes |
|-----------------|----------------------------|
| Title: | Regulatory Affairs Manager |
| Telephone: | 801-553-4366 |
| FAX: | 801-553-4609 |
Date Summary Prepared: 22 April 2019
#### II. Name of the Device
| Trade Name: | VALO® Grand Corded |
|------------------------------|------------------------------------------|
| Common Name: | Dental Curing Light |
| Device Classification: | Class II |
| Classification Product Code: | EBZ |
| Classification Name: | Ultraviolet Activator for Polymerization |
| Regulation No. | 21 CFR 872.6070 |
#### III. Device Description:
VALO® Grand Corded is a mains-powered, visible light activator for polymerization of dental resins of all photo-initiated dental materials. The VALO® Grand Corded version functions the same as the predicate, VALO® (K083647), as a professional hand-held, LED-based, visible light curing device. The modified device, VALO® Grand Corded, is manufactured from the same materials (anodized aluminum, rubber buttons, LED light source, printed circuit board, silicone adhesive, plastic), is used for the same indications, and has the same intended use as the predicate devices have three power output modes ranging from 800 - 2300 mW/cm2. Both the new device and predicate cure dental composite materials in the 395 -480 nm range using an LED light source. Differences from the predicate include an increased light head diameter to provide a larger curing area over a tooth, an additional LED activation ('power') button, and the middle power mode setting has been increased to make it more central between the Standard and Xtra Power modes. Additionally, VALO Grand Corded will include one pair of Blue Light Blocking Glasses, which will replace the Light Shield.
Image /page/3/Picture/11 description: The image shows the text '1 of 4' in black font. The text is positioned against a white background at the top, with a brown or orange color filling the lower portion of the image. The text indicates that this is the first image in a series of four.
505 West 10200 South
South Jordan, Utah 84095, USA
www.ultradent.com
800.552.5512
{4}------------------------------------------------
#### IV. Indications for Use:
The source of illumination for curing photo-activated dental restorative materials and adhesives.
#### V. Predicate Device:
The primary predicate device, VALO®, K083647, 23 Jan 2009 and reference device VALO Grand, K160551, 13 Jul 2016.
#### VI. ldentification of Risk Analysis Method:
Risk Analysis was performed on VALO® Grand Corded utilizing processes based on ISO 14971. Risks associated to patient safety and product efficacy for VALO® Grand Corded have been identified, assessed, and mitigated to a substantially equivalent level.
#### Comparison of Technological Characteristics VII.
VALO® Grand Corded and VALO® have similar technological characteristics as described in Table 5.1.
| Descriptive Information | Device: VALO® Grand Corded<br>dental curing light | Predicate: VALO® dental curing light<br>(K083647) |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The source of illumination for curing<br>photo-activated dental restorative<br>materials and adhesives. | The source of illumination for curing<br>photo-activated dental restorative<br>materials and adhesives. |
| Principle of Operation | Hand-held professional dental<br>curing light which emanates light<br>energy with peak wavelengths<br>between 395 - 480 nm range from<br>an LED light source which cures<br>photocurable, polymer resin-based<br>dental restorative materials | Hand-held professional dental curing<br>light which emanates light energy<br>with peak wavelengths between 395<br>- 480 nm range from an LED light<br>source which cures photocurable,<br>polymer resin-based dental<br>restorative materials |
| Intended User | Dentist or dental professional | Dentist or dental professional |
| Power source | Wall powered, 9VDC, medical grade<br>with adapters for international<br>capability<br>UL Approved | Wall powered, 9VDC, medical grade<br>with adapters for international<br>capability<br>UL Approved |
| Operational modes | *Standard Power Mode: 1000 mW/cm2 | *Standard Power Mode: 1000 mW/cm2 |
# Table 5.1: Substantial equivalence comparison
{5}------------------------------------------------
| Descriptive Information | Device: VALO® Grand Corded<br>dental curing light | Predicate: VALO® dental curing light<br>(K083647) |
|-----------------------------|------------------------------------------------------|-------------------------------------------------------------------|
| | *High Power Plus Mode: 1600mW/cm² | *High Power Mode: 1400mW/cm² |
| | Xtra Power Mode: 3200mW/cm² | Xtra Power Mode: 3200mW/cm² |
| | Device indicates illumination time<br>selection | Device indicates illumination time<br>selection |
| | Device indicates time and time selection | Device indicates time and time selection |
| | *as measured by Demetron | *as measured by Demetron |
| Light source | LED light, blue and violet wavelengths | LED light, blue and violet wavelengths |
| | 12mm head size | 10 mm head size |
| Accessories | Barrier Sleeve VALO®,<br>Blue-Light Blocking Glasses | Barrier Sleeve VALO®,<br>VALO® Light Shield |
| Composition of<br>Materials | Aluminum, anodized black | Aluminum, anodized black |
| Acceptable Cleaners | Acceptable Cleaners: | Anti-microbial surface disinfectant |
| | Lysol Brand III Disinfectant Spray<br>(Recommended) | DO NOT autoclave. |
| | Cavicide products (non-bleach) | DO NOT immerse in any kind of<br>ultrasonic batch or any liquids. |
| | Isopropyl alcohol | |
| | Ethyl alcohol-based cleaners | |
| | Lysol Concentrate (alcohol-based only) | |
| Usability/Ergonomics | 3 buttons – 2 power, 1 mode select | 2 buttons – 1 power, 1 mode select |
As outlined in the comparison tables above, VALO® Grand Corded is similar to the primary predicate, VALO® (K083647), in technology and identical in indications for use, power source, depth of cure, peak wavelength, and similar light source (LED). VALO® Grand Corded is used and cleaned in the same way by the same type of users. VALO Grand (K160551) is included as a reference device for the difference in head size between the subject device and the primary predicate.
The major functional changes between the submitted device, VALO® Grand Corded and the predicate device VALO® is in the light head diameter (12 mm vs 10 mm), an additional LED activation button, and the change in intensity of the High Power Plus (middle) mode setting (1600 mW/cm²). The VALO® Blue-Light Blocking Glasses were not included as part of the original VALO® submission. However, the original submission included a recommended light shield attachment as an accessory during use. VALO® Grand Corded is replacing the light shield supplied with the device to instead include Blue-Light Blocking Glasses. These differences do not impact safety and effectiveness based on validation and verification activities described below.
{6}------------------------------------------------
Device design validation and verification activities have been performed to FDA Guidance Document Dental Curing Lights – Premarket Notification [510(k)] and recognized standards, including ADA , where available, and via internal testing protocols. Results confirm that VALO® Grand Corded performs equivalently to the predicate VALO® with the modifications that have been made.
Based on these comparisons to already-cleared predicate device, we believe that VALO® Grand Corded is substantially equivalent to VALO®.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.