K192222 · Poonglim Pharmatech, Inc. · FMI · May 12, 2020 · General Hospital
Device Facts
Record ID
K192222
Device Name
EZ-Inject Single use Needle
Applicant
Poonglim Pharmatech, Inc.
Product Code
FMI · General Hospital
Decision Date
May 12, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is intended for use to inject fluids into or withdraw fluids from parts of the body below the surface of the skin.
Device Story
The EZ-Inject Single Use Needle is a sterile, single-use hypodermic needle consisting of a beveled metal (SUS304) tube joined to a polypropylene (PP) female connector (hub) and protected by a PP needle cap. It is available in various gauges (18G-34G) and lengths (4mm-40mm). The device is used by healthcare professionals to inject or withdraw fluids from below the skin surface. It is supplied sterile via E.O. gas sterilization. The device functions as a standard hypodermic needle; the clinician attaches the needle to a compatible syringe via the hub, inserts the needle into the patient, and performs the fluid transfer. The device provides a physical conduit for fluid delivery or aspiration, facilitating medication administration or sample collection.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including biocompatibility (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, pyrogenicity, hemolysis), mechanical testing (elasticity, flexural strength, pullout, stiffness, resistance to breakage/corrosion), and sterility/LAL testing per ISO and USP standards.
Technological Characteristics
Materials: SUS304 stainless steel cannula, polypropylene (PP) hub and protector, epoxy adhesive. Design: Single lumen, beveled tip. Dimensions: 18G-34G, 4mm-40mm lengths. Connectivity: Standard Luer connector (ISO 80369). Sterilization: E.O. gas. Standards: ISO 7864, ISO 9626, ISO 80369-7, ISO 6009, ISO 10993, USP <85>, USP <788>.
Indications for Use
Indicated for the injection of fluids into or withdrawal of fluids from parts of the body below the surface of the skin.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 12, 2020
POONGLIM Pharmatech Inc. % Peter Chung President Plus Global 300 Atwood Pittsburgh, Pennsylvania 15213
Re: K192222
Trade/Device Name: EZ-Inject Single Use Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: March 27, 2020 Received: April 13, 2020
Dear Peter Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K192222
Device Name EZ-Inject Single use Needle
Indications for Use (Describe)
This device is intended for use to inject fluids into or withdraw fluids from parts of the surface of the skin
Type of Use (Select one or both, as applicable) |X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
### [as required by 807.92(c)]
### 1. Applicant
- Company: POONGLIM Pharmatech Inc. 1)
- 2) Address: 21, Jayumuyeok 1-gil, Gunsan-si, Jeollabuk-do, Korea
- Tel : 82-63-451-8141 3)
- 4) Fax : 82-63-451-8145
- 5) Prepared date : July 11, 2019
- Contact person : Peter Chung, 412-512-8802 6)
- Contact person address : 300 Atwood Street, Pittsburgh, PA, 15213, USA 7)
- 8) Submission date: August 11, 2017
- ರು) Prior related submission : K172483
#### Device Information 2.
- 1) Trade name : EZ-Inject Single use Needle
- Common name : Single-use needle, hypodermic needle 2)
- 3) Classification name : Needle, Hypodermic, Single Lumen
- 4) Product code : FMI
- 5) Regulation number : 880.5570
- ୧) Class of device : Class II
- 7) Panel : General hospital
- 8) Model/type Name : 159 model codes including Y18-13
- 9) List of models : Y (53 models), I (53 models), L (53 models)
| | 18-13 | 18-25 | 18-30 | 18-40 | 19-13 | 19-25 | 19-30 | 19-40 |
|---|-------|-------|-------|-------|-------|-------|-------|-------|
| Y | 21-13 | 21-25 | 21-30 | 21-40 | 22-13 | 22-25 | 22-30 | 22-40 |
| | 23-13 | 23-16 | 23-25 | 23-30 | 25-13 | 25-16 | 25-25 | 25-30 |
| | 26-13 | 26-16 | 26-25 | 27-13 | 27-16 | 27-25 | 29-08 | 29-13 |
| | 29-16 | 30-04 | 30-08 | 30-13 | 30-16 | 31-04 | 31-06 | 31-08 |
| | 31-13 | 32-04 | 32-06 | 32-08 | 32-13 | 33-04 | 33-06 | 33-08 |
| | 33-13 | 34-04 | 34-06 | 34-08 | 34-13 | | | |
| | | | | | | | | |
| | 18-13 | 18-25 | 18-30 | 18-40 | 19-13 | 19-25 | 19-30 | 19-40 |
| I | 21-13 | 21-25 | 21-30 | 21-40 | 22-13 | 22-25 | 22-30 | 22-40 |
| | 23-13 | 23-16 | 23-25 | 23-30 | 25-13 | 25-16 | 25-25 | 25-30 |
| | 26-13 | 26-16 | 26-25 | 27-13 | 27-16 | 27-25 | 29-08 | 29-13 |
| | 29-16 | 30-04 | 30-08 | 30-13 | 30-16 | 31-04 | 31-06 | 31-08 |
| | 31-13 | 32-04 | 32-06 | 32-08 | 32-13 | 33-04 | 33-06 | 33-08 |
| | 33-13 | 34-04 | 34-06 | 34-08 | 34-13 | | | |
| | | | | | | | | |
| | 18-13 | 18-25 | 18-30 | 18-40 | 19-13 | 19-25 | 19-30 | 19-40 |
| L | 21-13 | 21-25 | 21-30 | 21-40 | 22-13 | 22-25 | 22-30 | 22-40 |
| | 23-13 | 23-16 | 23-25 | 23-30 | 25-13 | 25-16 | 25-25 | 25-30 |
| | 26-13 | 26-16 | 26-25 | 27-13 | 27-16 | 27-25 | 29-08 | 29-13 |
| | 29-16 | 30-04 | 30-08 | 30-13 | 30-16 | 31-04 | 31-06 | 31-08 |
| | 31-13 | 32-04 | 32-06 | 32-08 | 32-13 | 33-04 | 33-06 | 33-08 |
| | 33-13 | 34-04 | 34-06 | 34-08 | 34-13 | | | |
{4}------------------------------------------------
## K192222
- 3. The legally marketed device to which we are claiming equivalence K172483, KOPAC Sterile Needle
### 4. Device description
The device consists of a metal tube that is beveled at one end and at the other end join to a female connector (hub). The needle cap covers intended to provide physical protection to the needle tube. This product is packed by blister paper and sterilized by E.O. gas.
Needle gauge : 18, 19, 21, 22, 23, 25, 26, 27, 29, 30, 31, 32, 33, 34G Needle length : 4, 6, 8, 13, 16, 25, 30, 40 mm Wall thickness RW (Regular-Wall) : 18G, 19G, 21G, 22G, 23G, 25G, 26G, 27G, 29G, 30G, 31G, 32G, 33G, 34G TW (Thin Wall) : 18G, 19G, 21G, 22G, 23G, 25G, 26G, 27G, 29G, 30G, 31G, 32G, 33G, 34G
ETW (Extra Thin Wall) : 18G, 19G, 21G, 23G, 29G, 30G, 31G, 32G, 33G, 34G
UTW (Ultra Thin Wall) : 19G, 21G, 29G, 30G, 31G, 32G
### 5. Intended Use:
This device is intended for use to inject fluids into or withdraw fluids from parts of the body below the surface of the skin.
#### 6. Performance data:
- Bench test were performed. Bench testing included biocompatibility, mechanical testing 1) including EO residues. The tests demonstrated that the device performs in a substantially equivalent manner to the predicate device. The following bench testing is performed to demonstrate the functionality is substantially equivalent.
| Requirement – Test (ISO 7864) | Testing report no. | Result |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|--------|
| Visual check | -PQC-90527-01 (1) (Needle inner and outer surfaces)<br>-PQC-200217-01 (Cover inner and outer surfaces) | Pass |
| Dimension | -PQC-90527-01 (1) (Outer diameter of needle tube, Length of the needle tube, needle hub and hub hole)<br>-PQC-200217-01 (Cover) | Pass |
| Elasticity | -PQC-90527-01 (1) (Needle) | Pass |
| Flexural strength | -PQC-90527-01 (1) (Needle)<br>- PQC-200217-01 (Cover strength) | Pass |
| Pullout | -PQC-90527-01 (1) (Needle)<br>-PQC-200217-02 (Hub/needle bond strength) | Pass |
| Cleanliness | -PQC-200302-01 | Pass |
| Limits for acidity or alkalinity | -PQC-200302-01 | Pass |
| Limits for extractable metals | -PQC-200302-01 | Pass |
| Needle hub (Conical fitting, Colour of hub) | -Refer to KTL test report for relevant test (20-008504-01-1 (Conical fitting))<br>-PQC-200302-01 | Pass |
| Needle tube (Tolerances on length, Lubricant) | -PQC-200302-01 | Pass |
| Needle point | -PQC-200302-01 | Pass |
| Bond between hub and needle tube | -Refer to Poonglim Pharmatech Inc. test report for relevant test (PQC-200217-02 (Hub/needle bond strength)) | Pass |
| Patency of lumen | -PQC-200302-01 | Pass |
| Tolerances on length | -PQC-200508-07 | Pass |
| Requirement – Test (ISO 9626) | Testing report no. | Result |
| Stiffness | | Pass |
| Resistance to breakage | -PQC-90821-01 (1) | Pass |
| Resistance to corrosion | | Pass |
| Cleanliness | | Pass |
{5}------------------------------------------------
# K192222
| Limits for acidity or alkalinity | | Pass |
|-----------------------------------|-----------------------|--------|
| Stiffness | -PQC-200302-02 | Pass |
| Resistance to breakage | | Pass |
| Resistance to corrosion | | Pass |
| Requirement – Test (USP 41, <85>) | Testing report no. | Result |
| LAL test (Bacterial Endotoxin) | -T2018-14332 | Pass |
| Requirement – Test (ISO 80369) | Testing report no. | Result |
| Luer connector performance | -20-008504-01-1 | Pass |
| Dimension | -PQC-200225-01 (Luer) | Pass |
#### Biocompatibility 2)
Category: External Communicating Device Contact: Blood path, indirect Contact duration: A-limited (≤24h)
| # | Test item | Test method / Test criteria | Test result |
|---|-----------------------------------|------------------------------------------------------------------------------------------------------------|-------------|
| 1 | Cytotoxicity | ISO 10993-5 Tests for in vitro cytotoxicity | Pass |
| 2 | Skin Sensitization<br>Test | ISO 10993-10 irritation and skin sensitization | Pass |
| 3 | Intracutaneous<br>Reactivity Test | ISO 10993-10 Test for irritation and skin sensitization, maximization<br>test for delayed hypersensitivity | Pass |
| 4 | Acute Systemic<br>Toxicity Test | ISO 10993-11 Test for systemic toxicity - Acute Systemic Toxicity | Pass |
| 5 | Pyrogen Test | ISO 10993-11 Tests for systemic toxicity, Annex(F) Information on<br>material-mediated pyrogens. | Pass |
| 6 | Hemolysis Test | ISO 10993-4 Selection of tests for interactions with blood | Pass |
| 7 | Particulate matter<br>injections | USP <788> Particulate Matter in Injection | Pass |
#### 3) Sterility and LAL test
| # | Test item | Test standard | Test result |
|---|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| 1 | LAL test | USP39 <85>, Bacterial Endotoxins Test (Unit : EU/Device) | Pass |
| 2 | E.O sterilization<br>validation | According to ISO 11135:2014<br>E.O 30%, CO2 70%<br>Temperature : 55°C<br>Exposure time : 5 hours | Pass |
| 3 | Sterility test | According to ISO 11737-2 | Pass |
| 4 | E.O Residual test | Under the conditions of ISO 10993-7:2008, Ethylene oxide sterilization<br>residuals, the test articles should meet the test requirements. | Pass |
The performance tests demonstrated that this device is performs in a substantially equivalent manner to the predicate device.
{6}------------------------------------------------
#### 7. Comparison Table
| Manufacturer | POONGLIM Pharmatech Inc. | POONGLIM Pharmatech Inc. | Remark |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| 510(K) No. | K192222 | K172483 | |
| Intended use | This device is intended for use to<br>inject fluids into or withdraw from<br>parts of the body below the surface<br>of the skin | This device is intended for use to<br>inject fluids into or withdraw from<br>parts of the body below the<br>surface of the skin | Same |
| Hub of<br>needle | Polypropylene (PP) | Polypropylene (PP) | Same |
| Protector | Polypropylene (PP) | Polypropylene (PP) | |
| Cannula | SUS304 | SUS304 | |
| Adhesive | Epoxy | Epoxy | |
| Length | 4, 6, 8, 13, 16, 25, 30, 40mm | 13, 16, 25, 40mm | See discussion below |
| Gauge | 18, 19, 21, 22, 23, 25, 26, 27, 29, 30,<br>31, 32, 33, 34G | 30G | See discussion below |
| Tip<br>configuration | Bevel | Bevel | Per 7864:2016<br>Section 4.11 |
| Cover<br>dimension | Y type : 49 mm, 63 mm<br>I type : 51 mm, 65 mm<br>L type : 61 mm, 75 mm | Y type : 49 mm, 63 mm<br>I type : 51 mm, 65 mm | Similar |
| Cover<br>strength | Y type (63 mm) : 2.0 N<br>I type (65 mm) : 3.5 N<br>L type (65 mm) : 2.5 N | Y type (63 mm) : 2.0 N<br>I type (65 mm) : 3.5 N | Similar |
| Hub/needle<br>bond<br>strength | 18 G-40 mm : 137.7 N<br>19 G-40 mm : 177.8 N<br>21 G-25 mm : 82.0 N<br>22 G-25 mm : 114.0 N<br>23 G-25 mm : 144.3 N<br>25 G-13 mm : 91.1 N<br>26 G-13mm : 74.0 N<br>27 G-13 mm : 82.0 N<br>29 G-13 mm : 71.3 N<br>30 G-13 mm : 67.6 N<br>31 G-4 mm : 45.0 N<br>32 G-4 mm : 43.5 N<br>33 G-8 mm : 22.6 N<br>34 G-8 mm : 29.0 N | 30 G-13 mm : 67.5 N | Similar<br>Although we<br>compared only 30 G,<br>the rest of the gauges<br>meet the<br>requirements of the<br>ISO 7864. |
Length: The additional lengths meet the requirements of ISO 7864:2016, ISO 9626:2016, ISO 80369-7:2016 and ISO 6009:2016. The 4mm needle has a minimum specification of 3.5mm to successfully inject fluids or withdrawal fluids below the surface of the skin.
Gauge: The additional gauge sizes meet the requirements of ISO 7864:2016, ISO 80369-7:2016 and ISO 6009:2016.
### 8. Conclusion
The device has completed testing to show that the device meets its intended use and demonstrates substantial equivalence to the predicate device, K172483. Therefore, it is concluded that the subject device, EZ-Inject Single use Needle, is substantially equivalent to the legally marketed predicate device, K172483.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.