K210405 · Cadillac Products, Inc. · QPC · Dec 15, 2021 · General, Plastic Surgery
Device Facts
Record ID
K210405
Device Name
ProTEC-USA EZDoff Gown
Applicant
Cadillac Products, Inc.
Product Code
QPC · General, Plastic Surgery
Decision Date
Dec 15, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
ProTEC-USA EZDoff Gown is intended to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material. The ProTEC-USA EZDoff Gown is a single use, disposable gown provided non-sterile. The back of the gown is open and non-protective. The gown is not intended for use in the operating room.
Device Story
Single-use, non-sterile, disposable isolation gown; constructed from 41 gsm blue polyolefin (polyethylene) film; features open back, thumb loops, back perforation for removal, and waist ties. Used in healthcare settings by personnel to provide barrier protection against microorganisms, body fluids, and particulates. Not intended for operating room use. Provides Level 3 liquid barrier performance per AAMI PB70:2012. Device is donned by healthcare personnel; provides physical barrier to reduce cross-contamination risk.
Clinical Evidence
No clinical testing was performed. Evidence is based on bench testing, including flammability (16 CFR 1610), hydrostatic pressure (AATCC 127), water impact (AATCC 42), breaking strength (ASTM D5034), tearing strength (ASTM D5587), linting (ISO 9073-10), and air permeability (ASTM C522). Biocompatibility testing (cytotoxicity, sensitization, irritation) performed per ISO 10993 standards.
Technological Characteristics
Material: 41 gsm blue polyolefin (polyethylene) film. Design: Open back, thumb loops, waist ties, heat-sealed sleeves. Barrier: AAMI PB70:2012 Level 3. Flammability: 16 CFR 1610 Class 1. Biocompatibility: ISO 10993-1, -5, -10. Sterilization: Non-sterile.
Indications for Use
Indicated for health care patients and personnel to protect against transfer of microorganisms, body fluids, and particulate material. Not for use in operating rooms.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
December 15, 2021
Cadillac Products, Inc. % Dennis Gucciardo Partner Morgan Lewis & Bockius LLP 1111 Pennsylvania Ave Washington, District of Columbia 20004-2541
Re: K210405
Trade/Device Name: ProTEC-USA EZDoff Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: OPC Dated: November 15, 2021 Received: November 15, 2021
Dear Dennis Gucciardo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
- 79 -
510(k) Number (if known) K210405
Device Name ProTEC-USA EZDoff Gown
Indications for Use (Describe)
ProTEC-USA EZDoff Gown is intended to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material. The ProTEC-USA EZDoff Gown is a single use, disposable gown provided non-sterile. The back of the gown is open and non-protective. The gown is not intended for use in the operating room.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------------|
| | Prescription Use (Part 21 CFR 801 Subpart D) |
| | x Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K210405
# 510(k) SUMMARY
# Cadillac Product Inc.'s ProTEC USA EZDoff Gown
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared:
Cadillac Products, Inc. 5800 Crooks Road Suite 100 Troy, MI 48098
Phone: (248) 813-8200 Fax: (248) 813-8282
Contact Person: Andrew Stone Date Summary Prepared: December 11, 2021
# Application Correspondent
Dennis C. Gucciardo, Partner
Morgan, Lewis & Bockius, LLP 1111 Pennsylvania Ave. NW Washington, DC 20004-2541
Phone: (202) 739-5278
#### Name of Device and Name:
ProTEC-USA EZDoff Gown
#### Common or Usual Name:
Singe Use/Non-Sterile Open Back Protective Gowns
#### Classification Name:
Class II, 21 CFR § 878.4040; Product Code QPC
#### Predicate Device
ValueCare®Open Back Protective Gown (K160337)
#### Intended Use / Indications for Use
ProTEC-USA EZDoff Gown is intended to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material. The ProTEC-USA EZDoff Gown is a single use, disposable gown provided non-sterile. The back of the gown is open and nonprotective. The gown is not intended for use in the operating room.
{4}------------------------------------------------
# Device Description
The ProTEC-USA EZDoff Gown is constructed of 41 gsm blue polyolefin (Polyethylene) film offered in one universal size. The body of the gown is provided with belt ties that are constructed of the same 41 gsm blue polyolefin (Polyethylene) film. The sleeves of the gown are sealed using a heat-sealing method.
### Comparison of Technological Characteristics with the Predicate Device
The technological characteristics of the ProTEC-USA EZDoff Gown are nearly identical to the cleared ValueCare®Open Back Protective Gown (K160337). Both devices are made from extruded plastic film and include sleeves that are sealed with a heat treatment. A summary of the technologies and characteristics between the predicate device and the ProTEC-USA EZDoff Gown are provided in the table below.
| | Predicate Device:<br>ValueCare® Open Back<br>Protective Gown (K160337) | Proposed Device: ProTEC-<br>USA EZDoff Gown<br>(K210405) | Comparison |
|---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| Intended Use /<br>Indications for<br>Use | These gowns are intendedto<br>protect health care patients<br>and health care personnel<br>from the transfer<br>of<br>microorganisms, body fluids<br>and particulate material. The<br>back of the gown is open and<br>non- protective. They are not<br>intended for use in the<br>operating room. | ProTEC-USA EZDoff Gown<br>is intended to protect health<br>care patients and health<br>care personnel from the<br>transfer of microorganisms,<br>body fluids and particulate<br>material. The ProTEC-USA<br>EZDoff Gown is a single<br>use, disposable gown<br>provided non-sterile. The<br>back of the gown is open<br>and non-protective. The<br>gown is not intended for use<br>in the operating room. | Same |
| Design | Design includes open back,<br>thumb<br>loops,<br>back<br>perforation for easy removal<br>and waist tie | Design includes open back,<br>thumb<br>loops,<br>back<br>perforation for easy removal<br>and waist tie | Same |
| Material<br>Composition | Made from extruded plastic<br>film | Made from extruded plastic<br>film<br>(Polyolefin<br>(Polyethylene) film) | Same |
| Sterility | Non-sterile | Non-sterile | Same |
| Use | Single use; disposable | Single use; disposable | Same |
| Color | Blue | Blue | Same |
| Labeling Claims | These gowns are intendedto<br>protect health care patients<br>and health care personnel<br>from the transfer<br>of<br>microorganisms, body fluids<br>and particulate material. The<br>back of the gown is open and<br>non- protective. They are not<br>intended for use in the<br>operating room.<br>Non-Sterile | These gowns are intended to<br>protect health care patients<br>and health care personnel<br>from the transfer<br>of<br>microorganisms, body fluids<br>and particulate material. The<br>back of the gown is open and<br>non- protective. They are not<br>intended for use in the<br>operating room.<br>Non-Sterile | Same |
| | - Not Made with<br>Natural Rubber<br>Latex<br>Meets ANSI/AAMI<br>PB70: 2012 Level 3<br>Open Back Non-<br>protective<br>SingleUse,<br>Disposable | - Not Made with<br>Natural Rubber<br>Latex<br>Meets ANSI/AAMI<br>PB70: 2012 Level 3<br>Open Back Non-<br>protective<br>Single Use,<br>Disposable | |
| Flammability<br>16 CFR 1610 | Meets Class 1 "normal<br>flammability " in accordance<br>to 16 CFR Part 1610 | Meets Class 1 "normal<br>flammability " in accordance<br>to 16 CFR Part 1610 | Same |
| Liquid Barrier<br>Performance<br>Classification<br>Properties | Device was tested in<br>accordance with AAMI<br>PB70:2012 and meets Level<br>3 requirements for an<br>isolation gown. | Device was tested in<br>accordance with AAMI<br>PB70:2012 and meets Level<br>3 requirements for an<br>isolation gown. | Same |
| Biocompatibility | Pass<br>ISO 10993-1 | Pass<br>ISO 10993-1 | Same |
| Tearing<br>Strength /<br>Resistance | Not Reported | 8N (Machine Direction)<br>22N (Transverse Direction) | Not reported for<br>comparison |
| Lint | Not Reported | Log10<4 | Not reported for<br>comparison |
| Air Permeability | Not Reported | 0 ft³/min/ft² | Not reported for<br>comparison |
| Tensile Strength | Not Reported | 52N (Machine Direction)<br>41N (Transverse Direction) | Not reported for<br>comparison |
| Sterilization<br>Modality | None<br>(Non-Sterile) | None<br>(Non-Sterile) | Same |
{5}------------------------------------------------
# Summary of Non-Clinical Testing Performed on the Proposed Device
| Test Method | Purpose | Acceptance Criteria | Results |
|----------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Flammability | Test performed in<br>accordance with<br>(CPSC), 16 CFR<br>Part 1610 | Average burn time ≥<br>3.5s | Pass<br>Meets Class 1<br>Flammability. |
| Hydrostatic Pressure | Test performed in<br>accordance with AAMI<br>PB70:2012 utilizing<br>three (3)<br>nonconsecutive lots. | AATCC 127 ≥ 50 cm<br>AQL of 4.0 | Pass<br>Meets Level 3 liquid<br>barrier requirements. |
| Water Impact | Test performed in<br>accordance with AAMI<br>PB70:2012 utilizing<br>three (3)<br>nonconsecutive lots. | AATCC 42 ≤ 1.0g<br>AQL of 4.0 | Pass<br>Meets Level 3 liquid<br>barrier requirements. |
| Breaking strength | Test performed in<br>accordance with<br>ASTM D5034<br>utilizing three (3)<br>nonconsecutive<br>lots. | The standard does not<br>include an acceptance<br>criteria. | The test results<br>reported values were<br>52N (Machine<br>Direction), 41N<br>(Transverse<br>Direction). |
| Tearing strength | Test performed in<br>accordance with<br>ASTM D5587 | The standard does not<br>include an acceptance<br>criteria. | The test results<br>reported<br>values were |
{6}------------------------------------------------
| | utilizing three<br>(3)<br>nonconsecutive<br>lots. | | 8N (Machine<br>Direction), 22N<br>(Transverse<br>Direction). |
|------------------|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------|--------------------------------------------------------------|
| Linting | Test performed in<br>accordance with ISO<br>9073-10 utilizing three<br>(3) nonconsecutive lots. | The standard does not<br>include an acceptance<br>criteria. | The test results<br>reported value was<br>$Log_{10}<4$ . |
| Air Permeability | Test performed in<br>accordance with ASTM<br>C522. | The standard does not<br>include an acceptance<br>criteria. | The test results<br>reported value was 0 ft3/min/ft2 |
# Biocompatibility Testing Performed on the Proposed Device
| Test Method | Purpose | Acceptance Criteria | Results |
|---------------|----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|---------|
| Cytotoxicity | Test performed in<br>accordance with ISO<br>10993-<br>5:2009. | Under the conditions<br>of the studies ISO<br>10993-<br>5:2009, the proposed<br>device is non-<br>cytotoxic. | Pass |
| Sensitization | Test performed in<br>accordance with ISO<br>10993-<br>10:2010. | Under the conditions<br>of the studies: ISO<br>10993-<br>10:2010, the proposed<br>device is non-<br>sensitizing | Pass |
| Irritation | Test performed in<br>accordance with ISO<br>10993-<br>10:2010. | Under the conditions<br>of the studies: ISO<br>10993-<br>10:2010, the proposed<br>device is non-irritating. | Pass |
The EZDoff Gown was tested and conformed to the following standards and requirements:
| Standard<br>Number | Standards<br>Organization | Standards Title | Version | Date |
|------------------------------|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|---------|------------|
| ASTM F2407 | American Society<br>for Testing &<br>Materials | Standard Specification for<br>Surgical Gowns Intended for Use<br>in Healthcare Facilities | 2013 | May 2013 |
| AAMI<br>PB70:2012 | American<br>National Standard | Liquid Barrier Performance<br>Classification of Protective<br>Apparel and Drapes Intended for<br>use in Health Care Facilities | 2012 | 6/21/2012 |
| (CPSC), 16 CFR<br>Part 1610 | Consumer<br>Product Safety<br>Commission | Standard for Flammability of<br>Clothing Textiles | 2008 | 3/25/2008 |
| ISO 10993-<br>1:2009/(R)2013 | International<br>Organization of<br>Standardization | Biological Evaluation of Medical<br>Devices - Part 1: Evaluation and<br>testing with a risk management<br>process | 2013 | 10/15/2013 |
| ISO 10993-<br>5:2009 | International<br>Organization of<br>Standardization | Biological Evaluation of Medical<br>Devices - Part 5: Tests for In-<br>Vitro Cytotoxicity | 2009 | 6/1/2009 |
| ISO 10993-<br>10:2010 | International<br>Organization of<br>Standardization | Biological Evaluation of Medical<br>Devices - Part 10: Tests for | 2010 | 8/1/2010 |
{7}------------------------------------------------
| | | irritation and delayed-type of<br>hypersensitivity | | |
|---------------|------------------------------------------------|------------------------------------------------------------------------------------|------|-----------|
| ASTM D5587-15 | American Society<br>for Testing &<br>Materials | Standard Test Method for<br>Tearing Strength of Fabrics by<br>Trapezoid Procedure | 2019 | 7/1/2019 |
| ASTM D5034-09 | American Society<br>for Testing &<br>Materials | Standard Test Method for<br>Breaking Strength and<br>Elongation of Textile Fabrics | 2017 | 7/15/2019 |
| ASTM C522 | American Society<br>for Testing &<br>Materials | Standard Test Method for Airflow<br>Resistance of Acoustical<br>Materials | 2016 | 4/1/2016 |
# Summary of Clinical Testing
No clinical testing was performed.
# Conclusion:
The conclusions drawn from the nonclinical testing demonstrate that the ProTEC-USA EZDoff Gown is as safe, as effective, and performs as well as or better than the legally marketed device predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.