Safecare Open Back Protective Gowns and ValueCare Open Back Protective Gowns
K160337 · Bam Corporation Limited · QPC · Jul 7, 2017 · General, Plastic Surgery
Device Facts
Record ID
K160337
Device Name
Safecare Open Back Protective Gowns and ValueCare Open Back Protective Gowns
Applicant
Bam Corporation Limited
Product Code
QPC · General, Plastic Surgery
Decision Date
Jul 7, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
These gowns are intended to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material. The back of the gown is open and non-protective . They are not intended for use in the operating room.
Device Story
Single-use, non-sterile, open-back protective gowns; manufactured from extruded plastic film. Features include thumb loops for sleeve retention and back perforations for rapid removal if contaminated. Used by healthcare personnel to provide barrier protection against body fluids and particulates. Not intended for operating room use. Provides Level 3 barrier protection per ANSI/AAMI PB70:2012.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-10:2010 for irritation/sensitization, ISO 10993-5:2009 for cytotoxicity), flammability (16 CFR Part 1610 Class 1), and fluid barrier performance (ANSI/AAMI PB70:2012 Level 3, AATCC 127:2014, AATCC 42:2013).
Technological Characteristics
Extruded plastic film; open-back design with thumb loops and back perforations; non-sterile; single-use. Meets ANSI/AAMI PB70:2012 Level 3 barrier requirements. Not made with natural rubber latex.
Indications for Use
Indicated for healthcare patients and personnel to provide barrier protection against microorganisms, body fluids, and particulate material. Not for use in the operating room.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
Poly-Med Disposable Personal Protective Gown Apron (K925077)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 8, 2021
BAM Corporation Limited Mary Mejaes Official Correspondent No. 138 Shatin Rural Committee Road, Shatin, N.t. Unit 1706, Tower 2, Grand Central Plaza Hong Kong, N.T. China
Re: K160337
Trade/Device Name: Safecare Open Back Protective Gowns ValueCare Open Back Protective Gowns Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: QPC
Dear Mary Mejaes:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated July 7, 2017. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Dr. Ryan Ortega, Acting Assistant Director, Office of Surgical and Infection Control Devices, at 240-402-2303 or Ryan.ortega@fda.hhs.gov.
Sincerely,
# Bifeng Qian -S
For Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 7, 2017
BAM Corporation Limited Mary Mejaes Official Correspondent No. 138 Shatin Rural Committee Road, Shatin, N.T. Unit 1706, Tower 2, Grand Central Plaza Hong Kong, CN
Re: K160337
Trade/Device Name: SafeCare Open Back Protective Gowns ValueCare Open Back Protective Gowns Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYC Dated: June 20, 2017 Received: June 27, 2017
Dear Mary Mejaes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark S. Fellman -S
for
Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K160337
Device Name ValueCare® Open Back Protective Gowns
Indications for Use (Describe)
These gowns are intended to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material. The back of the gown is open and non-protective . They are not intended for use in the operating room.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K160337
Device Name SafeCare® Open Back Protective Gowns
Indications for Use (Describe)
These gowns are intended to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material. The back of the gown is open and non-protective . They are not intended for use in the operating room.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k)Summary
# K160337
| Regulatory Affairs Contact: | Ms. Mary Mejaes<br>BAM Corporation Ltd.<br>Unit P, 12/F, Kings Wing Plaza 1<br>No. 3 On Kwan Street<br>Shek Mun, Shatin, N.T.<br>Hong Kong |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone: | (011) 852-3695-5272 |
| Fax: | (011) 852-2314-1317 |
| Email: | mary@bam-corp.com |
| Date Summary Prepared: | July 3, 2017 |
| Common Name: | Single Use/Non-Sterile Open Back Protective Gowns |
| Classification Name: | Surgical Isolation Gowns,<br>Class II, 21 CFR 878: 4040, Product Code : FYC |
| Proprietary Name: | SafeCare® Open Back Protective Gowns<br>ValueCare® Open Back Protective Gowns |
| Device Description: | SafeCare® Open Back Protective Gowns (Models: 46969-097X,<br>46969097XB, 46969097X-10, 46969-098 & 46969-098D) and ValueCare®<br>Open Back (Model: 4550-25X) Protective Gowns are open back gowns<br>made from extruded plastic film.<br><br>The SafeCare® and ValueCare® Open Back Protective gowns provide barrier<br>protection to the wearer from body fluids and particulate materials and<br>have passed barrier testing according to ANSI/AAMI PB70:2012 Level 3<br>protection or equivalent. |
| | The significant design points of the gowns are:<br>1. Made from extruded plastic film to provide an effective barrier meeting<br>the requirements for penetration testing as per ANSI/AAMI PB70: 2012<br>Level 3 according to AATCC 42:2013 and AATCC 127:2014 test methods.<br>2. Backless to provide comfort<br>3. Thumb loops at the wrist which help to hold gown sleeves down<br>4. Back has perforations so the gowns can be removed without taking<br>over the head in case they are contaminated |
| Predicate Device: | Poly-Med Disposable Personal Protective Gown Apron, 510(k) K925077 |
| | 510(k)Summary<br>Page 1 out of 4 |
{6}------------------------------------------------
## Indications for Use: These gowns are intended to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material. The back of the gown is open and non-protective. They are not intended for use in the operating room.
| Aspect | Subject Device:<br>ValueCare® Open Back<br>Protective Gowns<br>510K No: K160337 | Subject Device:<br>SafeCare® Open Back<br>Protective Gowns<br>510K No: K160337 | Predicate Device:<br>PolyMed Disposable<br>Personal Protective Gown<br>Apron<br>510K No: K925077 |
|-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Design | Design includes open back, thumb loops, back<br>perforation for easy<br>removal and waist tie | Design includes open back, thumb loops, back<br>perforation for easy<br>removal and waist tie | Design includes open back, thumb loops, back<br>perforation for easy<br>removal and waist tie |
| Material | Made from extruded<br>plastic film. | Made from extruded<br>plastic film. | Made from extruded<br>plastic film. |
| Material Color | Blue | Blue or Yellow | Blue, Red, Pink, White,<br>Yellow |
| Indications for Use | These gowns are intended<br>to protect health care<br>patients and health care<br>personnel from the<br>transfer of micro-<br>organisms, body fluids and<br>particulate material. The<br>back of the gown is open<br>and non- protective. They<br>are not intended for use in<br>the operating room. | These gowns are intended<br>to protect health care<br>patients and health care<br>personnel from the<br>transfer of micro-<br>organisms, body fluids and<br>particulate material. The<br>back of the gown is open<br>and non- protective. They<br>are not intended for use in<br>the operating room. | Designed for use in<br>healthcare to provide<br>liquid and aerosol<br>resistance. |
| Labeling Claims | These gowns are intended<br>to protect health care<br>patients and health care<br>personnel from the<br>transfer of micro-<br>organisms, body fluids and<br>particulate material. The<br>back of the gown is open<br>and non- protective. They<br>are not intended for use in<br>the operating room.<br>* Non-Sterile | These gowns are intended<br>to protect health care<br>patients and health care<br>personnel from the<br>transfer of micro-<br>organisms, body fluids and<br>particulate material. The<br>back of the gown is open<br>and non- protective. They<br>are not intended for use in<br>the operating room.<br>* Non-Sterile | In compliance with OSHA<br>Bloodborne Pathogen<br>Standard, 29, CFR Part<br>1910.1030.<br><br>Provides liquid and<br>aerosol blood resistance. |
| | * Not Made with<br>Natural Rubber Latex<br>* Meets ANSI/AAMI<br>PB70: 2012 Level 3<br>* Open Back Non-<br>protective<br>* Single Use, Disposable | * Not Made with<br>Natural Rubber Latex<br>* Meets ANSI/AAMI<br>PB70: 2012 Level 3<br>* Open Back Non-<br>protective<br>* Single Use, Disposable | Not provided |
| Biocompatibility -<br>Sensitization and<br>Irritation | Under the condition of the<br>studies:<br>ISO 10993-10:2010, the<br>device is non-irritating and<br>non- sensitizing | Under the condition of the<br>studies:<br>ISO 10993-10:2010, the<br>device is non-irritating and<br>non- sensitizing | Not provided |
| Biocompatibility-<br>Cytotoxicity | Under the condition of the<br>studies:<br>ISO10993-5:2009,<br>The test device extract did<br>not show potential toxicity<br>to L-929 cells | Under the condition of the<br>studies:<br>ISO 10993-5:2009,<br>The test device extract did<br>not show potential toxicity<br>to L-929 cells | Not provided |
| Barrier Testing | ANSI/AAMI PB70: 2012<br>Level 3:<br>AATCC 127:2014<br>Hydrostatic Pressure Test<br>AATCC 42:2013 Impact<br>Penetration Test | ANSI/AAMI PB70: 2012<br>Level 3:<br>AATCC 127:2014<br>Hydrostatic Pressure Test<br>AATCC 42:2013 Impact<br>Penetration Test | Not provided |
| Flammability | Meets Class 1 "normal<br>flammability " in<br>accordance to 16 CFR Part<br>1610 | Meets Class 1 "normal<br>flammability " in<br>accordance to 16 CFR Part<br>1610 | Not provided |
| Sterility | Provided non-sterile | Provided Non-sterile | Provided Non-sterile |
# Comparison of Subject and Predicate Devices:
{7}------------------------------------------------
# Summary of the technological characteristics of the device compared to the Predicate Device:
- The PolyMed Disposable Personal Protective Gown Apron 510(k) K925077 and the SafeCare® Open Back Protective Gowns and ValueCare® Open Back Protective Gowns 510(k) K160337 are both made from extruded Plastic Film material.
- The construction and design features which include an open back, thumb loops, heat sealed sleeve seams and perforated back for easy removal are found in both gowns.
- Both the subject and predicate device are being promoted as suitable to be worn by health care professionals as barrier protection.
- Neither gown is intended to be used for surgical procedures and both are provided non-sterile.
{8}------------------------------------------------
- The SafeCare® and ValueCare® Open Back Protective Gowns are substantially equivalent to the ● PolyMed Disposable Personal Protective Gown Apron in that they both provide the following characteristics: fluid barrier and tensile strength.
| Test Performed | Test Standard | Results |
|------------------------------|--------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dermal Irritation Test | ISO 10993-10:2010 Skin<br>Irritation | Under the conditions of the studies: ISO 10993-<br>10:2010, the subject device is non-irritating. |
| Dermal Sensitization<br>Test | ISO 10993-10:2010 Skin<br>Sensitization | Under the conditions of the studies: ISO 10993-<br>10:2010, the subject device is non-sensitizing. |
| Cytotoxicity Test | ISO 10993-5:2009 In Vitro<br>Cytotoxicity | Under the conditions of the studies: ISO 10993-<br>5:2009, the SafeCare® Open Back Protective<br>Gowns and ValueCare® Open Back Protective<br>Gowns extract did not show potential toxicity<br>to L-929 cells. |
| Flammability Test | 16 CFR Part 1610 | Subject Device meets Class 1 requirements per<br>NFPA 16 CFR Part 1610 |
| Fluid Penetration Test | ANSI/AAMI PB70:2012<br>AATCC 127:2014<br>Hydrostatic Pressure Test<br>and AATCC 42:2013 Impact<br>Penetration test | SafeCare® Open Back Protective Gowns and<br>ValueCare® Open Back Protective Gowns have<br>been successfully tested and meet the Level 3<br>requirements of the ANSI/AAMI PB70:2012<br>Liquid Barrier classifications. |
## Summary of Testing:
# Conclusion:
Based on the results of biocompatibility testing, physical performance testing, and intended use the subject devices are as safe and as effective and perform as well as the predicate device. The SafeCare® and ValueCare® Open Back Protective Gowns are substantially equivalent to the predicate device cleared under K925077.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.