The Mpres stem is a cementless neck preserving stem designed for use in total or partial hip arthroplasty for primary or revision surgery. Total Hip Arthroplasty with the Mpres stem is indicated in the following cases: · Severely painful and/or disabled joint as a result of arthritis, rheumatoid polyarthritis or congenital hip dysplasia. · Avascular necrosis of the femoral head. - · Acute traumatic fracture of the femoral head. - Failure of previous hip surgery: o Conservative hip surgery, o Internal fixation, o Arthrodesis, o Hip resurfacing replacement. Partial hip arthroplasty with the Mpres stem is indicated in the following cases: - · Acute traumatic fracture of the femoral head. - · Avascular necrosis of the femoral head. · Primary pathology involving the femoral head but with a non-deformed acetabulum.
Device Story
Mpres Neck Preserving Stem is a cementless short femoral stem for total or partial hip arthroplasty; used in primary or revision surgery via minimally invasive techniques. Device replaces femoral head/neck; provides structural support for hip joint. Implanted by orthopedic surgeons in clinical/hospital settings. Surgeon selects appropriate size (12 sizes) and neck configuration (130° or 123° CCD angle) based on patient anatomy. Device integrates with various femoral heads (CoCr, Endo, MectaCer BIOLOX). Output is a stable prosthetic hip joint; restores joint function; reduces pain; improves mobility for patients with degenerative or traumatic hip conditions.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by bench testing, including design validation (sawbone/cadaveric), pull-off testing, range of motion (ROM) evaluation, neck and shaft fatigue testing, coating validation, and fretting corrosion analysis. Pyrogenicity was assessed via LAL and USP <151> testing.
Technological Characteristics
Cementless short femoral stem; material: Ti6Al7Nb titanium alloy (ISO 5832-11). Surface: Titanium plasma spray (ASTM F1580) and Hydroxyapatite coating (ASTM F1185). Available in 12 sizes; two neck configurations (130°/123° CCD). Sterilization: Gamma radiation. Mechanical fixation via press-fit.
Indications for Use
Indicated for patients requiring total or partial hip arthroplasty due to arthritis, rheumatoid polyarthritis, congenital hip dysplasia, avascular necrosis of the femoral head, acute traumatic fracture of the femoral head, or failure of previous hip surgeries (conservative, internal fixation, arthrodesis, or resurfacing).
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
September 22, 2021
Medacta International SA Chris Lussier Senior Director, Quality and Regulatory Medacta USA 3973 Delp Street Memphis, Tennessee 83118
Re: K210263
Trade/Device Name: Mpres Neck Preserving Stem Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, MEH Dated: August 16, 2021 Received: August 17, 2021
Dear Chris Lussier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Limin Sun, Ph.D. Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
## Indications for Use
510(k) Number (if known)
Device Name
Mpres Neck Preserving Stem
Indications for Use (Describe)
The Mpres stem is a cementless neck preserving stem designed for use in total or partial hip arthroplasty for primary or revision surgery.
Total Hip Arthroplasty with the Mpres stem is indicated in the following cases:
· Severely painful and/or disabled joint as a result of arthritis, rheumatoid polyarthritis or congenital hip dysplasia.
· Avascular necrosis of the femoral head.
- · Acute traumatic fracture of the femoral head.
- Failure of previous hip surgery:
o Conservative hip surgery,
o Internal fixation,
o Arthrodesis,
o Hip resurfacing replacement.
Partial hip arthroplasty with the Mpres stem is indicated in the following cases:
- · Acute traumatic fracture of the femoral head.
- · Avascular necrosis of the femoral head.
· Primary pathology involving the femoral head but with a non-deformed acetabulum.
Type of Use (Select one or both, as applicable)
| <div> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|----------------------------------------------------------------------------------------------------------------|
| <div> <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Senior Director of Quality and Regulatory, Medacta USA Date Prepared: September 21, 2021
#### II. Device
| Device Proprietary Name: | Mpres Neck Preserving Stem |
|--------------------------|-------------------------------------------------------------------------------------------------|
| Common or Usual Name: | Hip Prosthesis |
| Classification Name: | Hip joint metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis |
| Primary Product Code: | LZO |
| Secondary product code: | MEH |
| Regulation Number: | 21 CFR 888.3353 |
| Device Classification | II |
#### III. Predicate Device
Substantial equivalence is claimed to the following primary predicate device:
- > MiniHip, K083312, Corin USA
In addition the following reference devices are cited within the submission:
- > MiniHip, K111046, Corin USA
- > SMS Cementless Stem Extension, K201673, Medacta International SA
- > SMS Cementless Stem, K181693, Medacta International SA
- > MiniMAX, K170845, Medacta International SA
#### IV. Device Description
The Mpres Neck Preserving Stem is a cementless short femoral stem intended to be used in total or partial hip arthroplasty for primary or revision surgery with minimally invasive hip replacement techniques.
{4}------------------------------------------------
The Mpres Neck Preserving Stem is made of titanium alloy (Ti6Al7Nb) according to ISO 5832-11 and coated with Titanium plasma spray according to ASTM F1580 and Hydroxyapatite in compliance with ASTM F1185. It is available in 12 sizes (sizes 3-14) for each neck variation configuration: a standard 130° CCD angle or an high offset 123° CCD angle.
Identically to the reference devices (SMS Cementless Stem - K201673 and K181963), the Mpres Neck Preserving Stem can be combined with the CoCr ball head (K072857, K080885 and K103721), Endo Head (K111145), the MectaCer BIOLOX® forte (K073337), MectaCer BIOLOX® Delta Femoral Heads (K112115) or MectaCer BIOLOX® Option Heads (K131518).
#### V. Indications for Use
The Mpres stem is a cementless neck preserving stem designed for use in total or partial hip arthroplasty for primary or revision surgery.
Total Hip Arthroplasty with the Mpres stem is indicated in the following cases:
- Severely painful and/or disabled joint as a result of arthrosis, traumatic arthritis, ● rheumatoid polyarthritis or congenital hip dysplasia.
- Avascular necrosis of the femoral head.
- Acute traumatic fracture of the femoral head.
- Failure of previous hip surgery: ●
- o Conservative hip surgery,
- Internal fixation, O
- o Arthrodesis,
- Hip resurfacing replacement. O
Partial hip arthroplasty with the Mpres stem is indicated in the following cases:
- Acute traumatic fracture of the femoral head. ●
- . Avascular necrosis of the femoral head.
- Primary pathology involving the femoral head but with a non-deformed acetabulum.
#### VI. Comparison of Technological Characteristics
The Mpres Neck Preserving Stem implants and the predicate device (Corin MiniHip, K083312 and K111046) are similar in the following characteristics:
- shape and design; ●
- cementless use;
- CP-Ti and HA coating; ●
- mirror polishing surfaces;
- biocompatibility;
- device usage;
- . gamma radiation sterile; and
- packaging. ●
{5}------------------------------------------------
The Mores Neck Preserving Stem implants differ from the predicate device (Corin MiniHip. K083312 and K111046) as follow:
- sizes; .
- stem length:
- neck offset; and ●
- . materials.
## Discussion
The shape and design, the indications for use as well as the coating, device usage, biocompatibility, sterility and shelf life are similar to the predicate devices (Corin MiniHip, K083312 and K111046). Medacta International SA has not made any change to the materials and coating, device usage, biocompatibility, sterility, shelf life, and packaging of the subject devices respect to the reference devices (K201673, K1819693 and K170845).
The comparison of technological characteristics and performance data provided within this submission, shows that there are no new risks associated with the subject devices design, and supports the substantial equivalence of the Mpres Neck Preserving Stem implants to the identified predicate devices.
#### Performance Data VII.
Based on the risk analysis, design validation and characterization testing were conducted to written protocols. The following validation and tests were provided in support of the substantial equivalence determination:
### Non-Clinical Studies
- DESIGN VALIDATION ●
- Mpres Sawbone Design validation report O
- Plastic Stem Impactor, cadaveric evaluation O
- CHARACTERIZATION TESTING
- Pull-Off Test On CoCr Femoral Head according to Test Protocol IL 07.09.033 (per o ASTM F2009-00:2005, ISO 5832-9:2007 and ISO 5832-12:1996), EndoLab Test Report, No .: 167.090722.10.1309 (per ISO 7206-10:2003), CeramTec AG Test Reports 3128 (per ISO 5832-11 and ISO 7206-10) and 3300 (per ISO 7206-10).
- o Mpres evaluation of the ROM (per EN ISO 21535:2009).
- o Mpres Neck and Shaft Fatigue test according to Test Protocol IL 07.09.001 (per ISO 7206-6:2013 and ISO 7206-4:2010), Fatigue test worst case rationales (Neck worst case per ASTM F2068-15, ASTM F2996-13, ISO 5832-11 and ISO 7206-6:2013) (Shaft worst case per ASTM F2068-15, ISO 7206-4:2010, ASTM F2996-13 and ISO 5832-11), Endolab Test Report 167 201005 10 3284-part1-rev0 (per ISO 7206-6:2013), Accentus Test Report OTC357 Final Report Medacta and Straight Cantilever Test Report (per ISO 7206-4:2010).
{6}------------------------------------------------
- 0 Mpres Coating validation rationale (per ISO 5832-11).
- 0 Mpres femoral stems Fretting Corrosion Rationale (per ISO 5832-11 and ISO 21534).
- PYROGENICITY: .
- 0 Bacterial endotoxin test (LAL test) according to European Pharmacopoeia §2.6.14 (which is equivalent to USP chapter <85>)
- o Pyrogen test according to USP chapter <151> for pyrogenicity determination
- o The subject devices are not labeled as non-pyrogenic or pyrogen free.
### Clinical Studies:
- No clinical studies were conducted. ●
### VIII. Conclusion
It is demonstrated in the submission that the Mpres Neck Preserving Stem implants are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.