Deep Transcranial Magnetic Stimulation (DTMS) System
Applicant
Brainsway , Ltd.
Product Code
OBP · Neurology
Decision Date
Aug 17, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5805
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BrainsWay Deep TMS™ System is indicated for the treatment of depressive episodes and for decreasing anxiety symptoms for those who may exhibit comorbid anxiety symptoms in adult patients suffering from Major Depressive Disorder (MDD) and who failed to achieve satisfactory improvement from previous antidepressant medication treatment in the current episode.
Device Story
The BrainsWay Deep TMS System is a non-invasive neuromodulation device used in clinical settings to treat MDD and comorbid anxiety. It consists of a cart (neurostimulator, cooling system, positioning device) and a helmet containing an H1-Coil. The system delivers brief, high-intensity magnetic pulses to the brain; these pulses induce electric fields in deep brain structures, causing localized axonal depolarization and neuronal activation. The device is operated by trained personnel. By stimulating targeted brain regions, the system aims to alleviate depressive and anxiety symptoms in patients who have not responded to medication. Clinical efficacy is monitored by healthcare providers using standardized depression and anxiety scales (e.g., HDRS-21).
Clinical Evidence
Evidence includes 3 RCTs and 11 clinical studies (total N=573). Primary endpoint was change in HDRS-21 Anxiety-Somatization factor score. Meta-analysis of 3 RCTs showed a weighted pooled effect size (Cohen's d) of 0.55. Individual RCT effect sizes ranged from 0.34 to 0.905. Results demonstrate a clinically significant decrease in anxiety symptoms in MDD patients compared to sham or standard-of-care medication. Supportive meta-analyses reported pooled effect sizes of 1.45 (Kezior et al.) and -1.282 (Hung et al.).
Technological Characteristics
System comprises a cart-based neurostimulator, cooling system, and helmet with an H1-Coil. Operates via electromagnetic induction of electric fields in brain tissue. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 62304 (software), ISO 10993-5/10 (biocompatibility), and ISO 14971 (risk management).
Indications for Use
Indicated for adult patients with Major Depressive Disorder (MDD) who failed to achieve satisfactory improvement from previous antidepressant medication in the current episode, for the treatment of depressive episodes and reduction of comorbid anxiety symptoms.
Regulatory Classification
Identification
A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 17, 2021
Brainsway Ltd. Ahava Stein Regulatory Consultant A. Stein-Regulatort Affairs Consulting Ltd. 20 Hata'as Str., Suite 102 Kfar Saba 4442520 Israel
Re: K210201
Trade/Device Name: Deep Transcranial Magnetic Stimulation (DTMS) System Regulation Number: 21 CFR 882.5805 Regulation Name: Repetitive transcranial magnetic stimulation system Regulatory Class: Class II Product Code: OBP Dated: January 17, 2021 Received: January 25, 2021
Dear Ahava Stein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210201
Device Name Brainsway Deep TMS System
Indications for Use (Describe)
The BrainsWay Deep TMS™ System is indicated for the treatment of depressive episodes and for decreasing anxiety symptoms for those who may exhibit comorbid anxiety symptoms in adult patients suffering from Major Depressive Disorder (MDD) and who failed to achieve satisfactory improvement from previous antidepressant medication treatment in the current episode.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="text-decoration: line-through;">X</span> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;">□</span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
# BRAINSWAY DEEP TMS™ SYSTEM
## 510(k) Number K210201
### Applicant Name:
| Company Name: | BrainsWay Ltd |
|---------------|----------------------------------------|
| Address: | BrainsWay Ltd. |
| | 19 Hartom St. (Bynet Bldg) |
| | Har Hotzvim, Jerusalem, ISRAEL 9777518 |
| | Tel: +972-2-5813140 |
| | Fax: +972-2-5812517 |
| | E-mail: ahava@asteinrac.com |
### Contact Person:
| Official Correspondent: | Ahava Stein |
|-------------------------|---------------------------------------------------------|
| Company Name: | A. Stein - Regulatory Affairs Consulting Ltd. |
| Address: | 18 Hata'as Str., Suite 102 |
| | Kfar Saba 4442520 Israel |
| | Tel: +972-9-7670002 |
| | Fax: +972-9-7668534 |
| | E-mail: ahava@asteinrac.com |
| Date Prepared: | June 9, 2021 |
| Trade Name: | BrainsWay Deep TMS™ System |
| Classification Name: | CFR Classification section 882.5805; (Product Code OBP) |
| Classification: | Class II Medical Device |
### Predicate Device:
The subject device is a modification of the BrainsWay Deep TMS System ("predicate device") that was cleared in 510(k) document no. K122288 and the BrainsWay Deep TMS System cleared in K173540 ("reference device").
| Predicate | Device | Manufacturer | 510(k) No. |
|-----------|---------------------------|----------------|------------|
| Main | BrainsWay Deep TMS System | BrainsWay Ltd. | K122288 |
| Reference | BrainsWay Deep TMS System | BrainsWay Ltd. | K173540 |
{4}------------------------------------------------
### Device Description:
The BrainsWay Deep TMS System enables direct non-invasive activation of deep brain structures. Transcranial magnetic stimulation (TMS) is a non-invasive technique used to apply brief magnetic pulses to the brain. The pulses are administered by passing high currents through an electromagnetic coil placed adjacent to a patient's scalp. The pulses induce an electric field in the underlying brain tissue. When the induced field is above a certain threshold, and is directed in an appropriate orientation relative to the brain's neuronal pathways, localized axonal depolarizations are produced, thus activating neurons in the targeted brain structure.
The BrainsWay Deep TMS System is composed of the following main components:
- 1. Cart
- a) TMS Neurostimulator
- b) Cooling System
- c) Positioning Device
- 2. Helmet
- a) Aiming Apparatus (i.e., ruler/grid)
- b) Electromagnetic Coil (H1-Coil)
- c) Cap
## Intended Use/Indication for Use:
The BrainsWay Deep TMS™ System is indicated for the treatment of depressive episodes and for decreasing anxiety symptoms for those who may exhibit comorbid anxiety symptoms in adult patients suffering from Major Depressive Disorder (MDD) and who failed to achieve satisfactory improvement from previous antidepressant medication treatment in the current episode.
## Performance Standards:
BrainsWay Deep TMS System complies with the following FDA recognized consensus standards:
- EC 60601-1 Medical Electrical Equipment Part 1: General requirements for . basic safety and essential performance (Ed 3.1, 2005 + CORR.1:2006 + CORR.2:2007 + A1:2012 AND 2006 + AC:2010 + A1:2013)
- . IEC 60601-1-2 Medical Electrical Equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and test (Ed 4 2014)
- IEC 62304 Medical Devices Software life-cycle processes (2006 + A1:2015) .
- ISO 10993-5 Biological evaluation of medical devices Part 5: Tests for in vitro . cytotoxicity
- . ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity
- ISO 14971 Medical devices Application of risk management to medical devices .
{5}------------------------------------------------
#### Non-Clinical (Bench) Performance Data:
Tests were conducted on the BrainsWay Deep TMS System. The tests were performed according to the FDA Guidance Document Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation (rTMS) Systems. These tests included Output Waveform, Electrical Field Spatial Distribution and Magnetic Field Strength Gradient Testing. Additional performance testing included electrical and mechanical safety testing, electromagnetic compatibility testing and software validation testing in compliance with FDA guidelines for software validation and IEC 62304 standard requirements. The results of the performance tests demonstrated that the BrainsWay Deep TMS System is substantially equivalent to the predicate device.
### Animal Performance Data / Histology Data:
Not Applicable
#### Clinical Performance Data:
The clinical performance supporting the safety and effectiveness of the BrainsWay Deep TMS device for decreasing anxiety symptoms for those who may exhibit comorbid anxiety symptoms in adult patients suffering from Major Depressive Disorder (as described in the Indications for Use statement) was based on 3 Randomized Controlled Trials (RCTs) and supportive data from meta-analyses from 11 clinical studies (including the 3 RCTs and another 8 open label studies) using the BrainsWay Deep TMS device in 573 patients.
The Deep TMS vs. Sham between group (adjusted) effect size for the Multicenter MDD Study (Levkovitz 2015) was 0.34 and for the Kaster 2018 study was 0.36. The effect size for the RCT that used medication as SOC, Filipcic 2019, was 0.905. (See the table below). A meta-analysis of the three RCT studies, comparing the BrainsWay Deep TMS treatment to Sham or medication as standard of care treatment, demonstrated an overall, weighted pooled (between group, adjusted) effect size of 0.55 (Cohen's d), representing a medium effect size of the BrainsWay Deep TMS device for decreasing anxiety symptoms in this patient population, when compared to Sham or medication as standard of care.
The above effect sizes achieved with the BrainsWay Deep TMS device for decreasing anxiety symptoms, although based on the HDRS-21Anxiety-Somatization Factor score, may be compared to the effect size seen in the Multicenter MDD Study for all MDD subjects using the HDRS-21 score, where an effect size of 0.76 (adjusted difference between groups: -3.11±1.6, p value: 0.008) was reported. This effect size compares
{6}------------------------------------------------
favorably to the weighted effect sizes obtained in the above mentioned 3 RCTs which used Deep TMS, demonstrating a clinically significant decrease in anxiety symptoms in adult patients suffering from MDD (as described in the Indications for Use statement). The supportive data included two meta-analyses published by Kezior et al (including 6 of the 11 studies) and by Hung et al (including 8 of the 11 studies) reporting a pooled weighted effect size of 1.45 (Cohen's d) and -1.282 (Hedges' g), respectively.
In summary, the gamut of clinical data with the BrainsWay Deep TMS device demonstrates an effect size that is consistent, robust and similar to that reported forMDD subjects in the Multicenter MDD Study.. Hence, the results are clinically meaningful for decreasing anxiety symptoms for those who may exhibit comorbid anxiety symptoms in adult patients suffering from Major Depressive Disorder who failed to achieve satisfactory improvement from previous anti-depressant medication treatment in the current episode.
#### Substantial Equivalence:
The modified device has the same intended use as the BrainsWay Deep TMS Systems cleared in 510(k) K122288 and K173540. The subject device has an expanded indication for use, to include decreasing anxiety symptoms for those who may exhibit comorbid anxiety symptoms in patients suffering from Major Depressive Disorder who failed to achieve satisfactory improvement from previous anti-depressant medication treatment in the current episode.
Both the modified device and the cleared BrainsWay Deep TMS System are similar in terms of their intended prescription use only, suitable for adult population, indicated for anatomical sites according to indications for use and to be used in hospital or clinic settings.
All components of the subject BrainsWay Deep TMS System are identical to the previously cleared, predicate devices and have the same mechanism of operation and use the same underlying technology. The performance characteristics, including the Electrical and Magnetic Field Distribution testing are the same as the previously cleared BrainsWay Deep TMS Systems (K122288 and K173540). Supportive clinical data has been provided to demonstrate the safety and effectiveness of the subject BrainsWay Deep TMS Systems for the expanded indications for use.
#### Conclusions:
Consequently, it can be concluded that the modified BrainsWay Deep TMS System is substantially equivalent to the main predicate BrainsWay Deep TMS System, cleared
{7}------------------------------------------------
under 510(k) K122288 and the reference predicate BrainsWay TMS System, cleared under 510(k) K173540. Therefore, the modified BrainsWay Deep TMS System can be legally marketed in the USA.
{8}------------------------------------------------
| Study | Design | Population | Sample Size, Age | Tx Duration | Stimulation Protocol | Assessment | Effect size (Cohensc with lower / upper 95% limits) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|----------------------------------------|--------------------------------------------------------------------------------------------------|-------------|-------------------------------------------------------------|-------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Levkovitz 2015<br>Efficacy and safety of deep transcranial magnetic stimulation for major depression: a prospective multicenter randomized controlled trial<br>Analysis provided in 510(k) K210201 | RCT | Treatment Resistant Depression, DSM-IV | SS:<br>DTMS: 89<br>Sham:92<br><br>Age:<br>DTMS:<br>$45.1\pm11.7$<br>Sham:<br>$47.6\pm11.6$ | 4 weeks | 18Hz, 120% MT,<br>20 mins, 1980 pulses/session, 20 sessions | Change in HDRS-21 A/S Factor Score from baseline at 5 weeks | 0.34 (0.009,0.667)<br>HDRS A-S (PP):<br>DTMS: Pre: $7.55\pm1.92$<br>Post: $4.25\pm2.88$<br>Diff: $-3.18\pm2.79$<br>Adj Diff: -2.93+0.31(SE)<br>Sham: Pre: $6.83\pm1.94$<br>Post: $4.62\pm2.7$<br>Diff: $-2.05\pm2.50$<br>Adj Diff:-2.03+0.32(SE)<br>Adjusted Diff btwn groups: -0.90±0.41<br>P value: 0.0306 |
| Kaster 2018<br>Efficacy, tolerability, and cognitive effects of deep transcranial magnetic stimulation for late-life depression: a prospective randomized controlled trial | RCT | Treatment Resistant Depression, DSM-IV | SS:<br>DTMS: 25<br>Sham: 27<br><br>Age: DTMS:<br>$65.0 \pm 5.5$<br>Sham: $65.4 \pm 5.5$ | 4 weeks | 18Hz, 120% MT,<br>61 mins, 6012 pulses/session, 20 sessions | Change in HDRS-21 A/S Factor Score from baseline at 5 weeks | 0.36 (-0.23,0.95)<br>HDRS A-S (ITT):<br>DTMS: Pre: $6.48\pm1.98$<br>Post: $3.65\pm2.70$<br>Diff: $-2.75\pm3.04$<br>Adj Diff:-2.9+0.60(SE)<br>Sham: Pre: $6.85\pm1.51$<br>Post: $4.85\pm2.78$<br>Diff:-2.0+2.62<br>Adj Diff:-1.9+0.52(SE)<br>Adjusted Diff btwn groups: -1.0±0.80(SE)<br>P value: 0.2171 |
| Filipcic 2019<br>Efficacy of repetitive transcranial magnetic stimulation using a figure-8-coil or an H1-Coil in treatment of major depressive disorder; A randomized clinical trial | RCT | Treatment Resistant Depression, DSM-IV | SS:<br>DTMS: 65<br>Med as SOC:<br>72<br><br>Age:<br>DTMS:<br>$50 + 2.7$<br>MedSOC:<br>$53 + 2.2$ | 4 weeks | 18Hz, 120% MT,<br>20 mins, 1980 pulses/session, 20 sessions | Change in HDRS-21 A/S Factor Score from baseline at 5 weeks | 0.90 (0.55,1.26)<br>(DH1 vs DSOC)<br>HDRS A-S (PP):<br>DTMS: Pre: $5.35\pm2.20$<br>Post: $2.08\pm1.97$<br>Diff: $-3.28\pm2.39$<br>Adj Diff:-3.41+0.27(SE)<br>SOC:<br>Pre: $5.79\pm2.71$<br>Post: $4.69\pm2.46$<br>Diff: $-1.10\pm2.89$<br>Adj Diff:-0.96+0.26(SE)<br>Adjusted Diff btwn groups: -2.46±0.37(SE)<br>P value: <0.0001 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.