The Brainsway Deep TMS System is indicated for the treatment of depressive episodes in adult patients suffering from Major Depressive Disorder who failed to achieve satisfactory improvement from previous anti-depressant medication treatment in the current episode.
Device Story
Brainsway Deep TMS (DTMS) System is a non-invasive neurostimulation device for treating Major Depressive Disorder. It uses an electromagnetic H1 coil to deliver brief, high-current magnetic pulses to deep brain structures, inducing electric fields that modulate neuronal activity. The system comprises an H1 coil, neurostimulator, cooling system, positioning device, and cart. Operated by clinicians in a clinical setting, the device delivers repetitive pulse trains at 18 Hz. Treatment involves determining the patient's motor threshold, positioning the coil over the prefrontal cortex, and administering pulses. The output affects clinical decision-making by providing a non-pharmacological intervention for treatment-resistant depression. Benefits include potential symptom relief for patients who have failed antidepressant therapy. The system is designed to reach deeper brain regions than standard figure-8 coils.
Clinical Evidence
Prospective, double-blind, randomized, multi-center trial (n=181 in PP analysis set). Primary endpoint: change in HDRS-21 scores. Results: DTMS group showed mean change of -6.39 vs -3.28 for sham (p=0.0080). Response rates: 38.4% (DTMS) vs 21.4% (sham) (p=0.0138). Remission rates: 32.6% (DTMS) vs 14.6% (sham) (p=0.0051). Efficacy maintained at 16 weeks. Adverse events included application site pain/discomfort and jaw pain; one device-related seizure reported.
Technological Characteristics
System uses electromagnetic H1 coil (air-core) for deep brain stimulation. Operates at 18 Hz, 120% of motor threshold. Compliance with IEC 60601-1 (electrical/mechanical safety) and IEC 60601-1-2 (EMC). Includes active cooling system for coil. Software-controlled pulse delivery. Form factor: mobile cart with positioning arm.
Indications for Use
Indicated for adult patients with Major Depressive Disorder (MDD) who failed to achieve satisfactory improvement from previous antidepressant medication in the current episode. Contraindicated in patients with metal implants in or around the head, history of seizure, or those who failed to respond to previous ECT treatment.
Regulatory Classification
Identification
A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.
{0}------------------------------------------------
### 510(K) SUMMARY
# BRAINSWAY DEEP TMS SYSTEM
## 510(k) Number K122288
# Applicant Name:
| Company Name: | |
|---------------|--|
| Address: | |
Brainsway Ltd. 19 Hartum St. (Bynet Bldg) Har Hotzvim, Jerusalem, ISRAEL 91451 Tel: +972-2-5813140 Fax: +972-2-5812517 E-mail: hila@Brainsway.com
### Contact Person:
| Official Correspondent: | Ahava Stein |
|-------------------------|---------------------------------------------------------|
| Company Name: | A. Stein - Regulatory Affairs Consulting Ltd. |
| Address: | 20 Hata'as Str., Suite 102<br>Kfar Saba 44425<br>Israel |
| Tel: | +972-9-7670002 |
| Fax: | +972-9-7668534 |
| E-mail: | ahava@asteinrac.com |
| Date Prepared: | July 10, 2012 |
| Trade Name: | Repetitive Transcranial Magnetic Stimulator |
Classification Name: CFR Classification section 882.5805 (Product code OBP)
Classification: Class II Medical Device
Predicate Device:
The Brainsway Deep TMS System is substantially equivalent to the following predicate device:
| Manufacturer | Device | 510(k) No. | Product<br>Code |
|--------------|--------------------------------------------|------------|-----------------|
| Neuronetics | NEUROSTAR TMS SYSTEM | K061053 | OBP |
| Neuronetics | NEUROSTAR TMS THERAPY<br>SYSTEM, MODEL 1.1 | K083538 | OBP |
{1}------------------------------------------------
### Device Description:
The Brainsway Deep TMS (DTMS) System is intended for the treatment of depressive episodes in patients suffering from MDD. Transcranial magnetic stimulation (TMS) is a non-invasive technique used to apply brief magnetic pulses to the brain. The pulses are administered by passing high currents through an electromagnetic coil placed adjacent to a patient's scalp. The pulses induce an electric field in the underlying brain tissue. When the induced field is above a certain threshold, and is directed in an appropriate orientation relative the brain's neuronal pathways, localized axonal depolarizations are produced, thus activating neurons in the targeted brain structure.
The electromagnetic radiation emitted is at a low frequency of the order of 1-10 kHz. The only interaction with the human body is caused by the briefly changing magnetic field, and its effect may be neuronal activation.
Many neurological and psychiatric disorders are associated with abnormal neuronal activity patterns in deep brain regions. These regions cannot be affected directly, but only indirectly, through secondary processes involving cortical structures, which are directly activated by TMS and then affect the deeper structures.
The Brainsway DTMS System's unique design enables to produce directed electromagnetic fields that can induce excitation or inhibition of neurons deep inside the brain non-invasively.
The Brainsway Deep TMS (DTMS) System is composed of the following main components:
- An Electromagnetic Coil (H1 Coil) 1.
- 2. A TMS Neurostimulator
- 3. A Cooling System
- A Positioning Device 4.
- 5. A cart
### Intended Use/Indication for Use:
The Brainsway Deep TMS System is indicated for the treatment of depressive episodes in adult patients suffering from Major Depressive Disorder who failed to achieve satisfactory improvement from previous anti-depressant medication treatment in the current episode.
### Performance Standards:
The Brainsway Deep TMS System has been tested and complies with the following voluntary recognized standards:
- Electrical & Mechanical Safety testing according to IEC 60601-1 .
- . Electromagnetic Compatibility testing according to IEC 60601-1-2
- ETSI EN 301 489-1 V1.8.1 Electromagnetic Compatibility and Radio Spectrum . Matters (ERM); Electromagnetic Compatibility (EMC) Standard for Radio Equipment and Services; Part 1: Common Technical Requirements (2008) Immunity
{2}------------------------------------------------
### Non-Clinical Performance Data:
The non-clinical performance testing with the Brainsway Deep TMS System included testing of the magnetic field characteristics of the system, as required by the FDA Guidance document for TMS systems. Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation (rTMS) Systems, as follows:
### Output Waveform
The output waveform produced by the H-coil was measured in a human head phantom model. The probe was located in the head model in a position analogous to the left lateral prefrontal cortex, where the coil's induced field is maximal according to the electric field distribution mapping. The electric field pulse outputs and waveforms were measured at predefined stimulator outputs.
#### . Electric Field Spatial Distribution
The field distribution produced by the H-coil was measured in a human head phantom model. The probe was moved in three directions inside the phantom model using a displacement system with 1 mm resolution, and the field distribution of the H1-coil was measured in the whole head model volume in 1cm resolution. Axial and coronal field maps were produced. The field maps were superimposed on anatomical T1-weighted MRI coronal slices, to show the induced field in each anatomical brain region.
### Magnetic Field Strength Gradient
The magnetic field strength was also measured in a human head phantom model. The magnetic field strength at the H-coil surface and at certain points at various distances from the coil was measured. The probe was located in the head model at a distance of 0 and 2 cm from the coil edge, in a position analogous to the left lateral prefrontal cortex. where the coil's induced field is maximal according to the electric field distribution mapping. The stimulator was operated at 50% and at 100% output and the magnetic field output waveform was measured along three orthogonal axes: X- Anteriorposterior, Y- Lateral-medial, and Z- Superior-inferior. These results demonstrate the ability of the H-coil to stimulate deeper brain structures without overstimulation of superficial regions.
### Cooling System Testing
The cooling system tests included measurement of the air flow, suction pressure, compression pressure and currents measured at different voltage levels. The temperature of the cooling system was measured over an extended period of time to test the life expectancy of the system. Life expectancy testing demonstrated that the cooling system thermostat temperature remains consistent during a continuous operation of >180 days.
The performance tests, along with the clinical study validation demonstrate that the Brainsway DTMS System may be safely and effectively used for treatment of Major Depressive Disorder.
{3}------------------------------------------------
### Clinical Performance Data:
The safety and effectiveness of the Brainsway Deep TMS System for treatment of Major Depressive Disorder was demonstrated in a prospective, double blind, randomized, controlled, multi-center trial. The study was conducted at 20 study sites in the United States (13 sites), Israel (4 sites), Germany (2 sites) and Canada (1 site), During the initial treatment phase. TMS sessions were performed daily for 4 weeks and during the maintenance phase, subjects were treated twice a week for another 12 weeks.
The study results were summarized for three patient analysis sets, the ITT analysis set (n=229, 108 DTMS and 121 Sham), the mITT analysis set (n=212, 101 DTMS and 111 Sham) and the PP analysis set (n=181, 89 DTMS and 92 Sham). The most appropriate analysis set for the purpose of assessing the effectiveness of the Brainsway Deep TMS System is the PP analysis set. The reason for this is that the ITT analysis set and the mITT analysis set include 31 subjects (12 DTMS and 19 Sham) who did not receive the adequate DTMS treatment regimen. Furthermore, the ITT analysis set also includes subjects who did not meet the inclusion/exclusion criteria for the study. The IFU contains clear instructions under the Warnings and Precautions section regarding the appropriate subjects who may be treated with the Brainsway Deep TMS System and clear instructions how to administer the treatment so that subjects will receive the appropriate stimulation intensity during the treatment. The efficacy results for the PP analysis set reflect the intended patient population and therefore are most relevant and presented below. The PP analysis set includes 181 subjects (89 DTMS and 92 Sham) between the ages of 22 and 68, diagnosed with Major Depressive Disorder, with an episode of less than 7 years. with an HDRS-21 score of greater than 20, did not receive benefit from 1 to 4 antidepressant treatments or were intolerant to at least 2 antidepressant treatments, and met the safety screening questionnaire for TMS. These subjects did not have MDD with psychotic features, another primary psychotic disorder or significant neurological disorder, were not at risk of seizure or suicide, did not fail to respond to previous ECT treatment and did not have any metal implants in or around the head or any other implants which could be affected by TMS.
The model estimated mean change from baseline in HDRS-21 scores in the Deep TMS group was -6.39 points across 5 weeks compared to only -3.28 points in the sham group in the PP analysis set. This difference of 3.11 points was statistically significant (p=0.0080). This change was found almost statistically significant in the mITT analysis i set (p=0.0578), but was not found statistically significant in the ITT analysis set. The response rate was significantly better in the Deep TMS group (38.4% and 37.0%) compared to the sham group (21.4% and 22.8%) (p=0.0138 and p=0.0310) in the PP analysis set and the mITT analysis set, respectively. The remission rate was significantly better in the Deep TMS group (32.6% and 30.4%) compared to the sham group (14.6% and 15.8%) (p=0.0051 and p=0.0158) in the PP analysis set and the mITT analysis set, respectively. The response and remission rates were not found statistically significant in the ITT analysis set.
The efficacy of the Deep TMS Treatment was maintained at 16 weeks, based on a statistically significant change from baseline in HDRS-21 scores (p=0.0259) in the PP analysis set. This was not shown in the mITT or ITT analysis sets. The efficacy of the Deep TMS Treatment was further maintained at 16 weeks based on a significantly better response rate in the Deep TMS group compared to the sham group (p=0.0086 and
{4}------------------------------------------------
p=0.0276), in the PP analysis set and the mITT analysis set. This was not shown in the ITT analysis set. The efficacy of the Deep TMS Treatment, based on statistically significant differences in the CGI-S scores, was demonstrated at 5 weeks and also maintained at 16 weeks in the PP analysis set and the mITT analysis set. The efficacy of the Deep TMS Treatment, based on statistically significant differences in the CGI-I, PGI and GAF scores, was demonstrated at 16 weeks in the PP analysis set and the mITT analysis set. These scores were also statistically significantly lower in the Deep TMS treatment group at 5 weeks in the PP analysis set. These scores were not found to be statistically significantly different in the ITT analysis set. No significant differences were found in the SF-36 scores, however all the SF-36 Quality of Life parameters were better in the Deep TMS group than in the sham group in all the analysis sets.
The results of the cognitive tests (MMSE, BSRT and AMI-SF) demonstrated that the Deep TMS treatment does not have a negative cognitive effect on MDD subjects in all the analysis sets.
The following table summarizes the main adverse events reported in the study for all the study subjects.
| | Deep TMS Treatment<br>(N=111 Subjects) | | Sham Treatment<br>(N-122 Subjects) | | |
|------------------------------------------|----------------------------------------|-----------|------------------------------------|-----------|---------|
| Anticipated Event | No of Subjects | Incidence | No of Subjects | Incidence | p-value |
| Eye Pain | 2 | 1.9% | 4 | 3.3% | 0.6864 |
| Dental Pain | 3 | 2.8% | 2 | 1.7% | 0.6687 |
| Pain in Jaw | 11 | 10.2% | 1 | 0.8% | 0.0017 |
| Application Site Discomfort | 21 | 19.4% | 5 | 4.1% | 0.0003 |
| Application Site Pain | 27 | 25.0% | 1 | 0.8% | <.0001 |
| Muscle Twitching | 7 | 6.5% | 2 | 1.7% | 0.0878 |
| Headache | 51 | 47.2% | 44 | 36.4% | 0.1079 |
| Anxiety | 6 | 5.6% | 9 | 7.4% | 0.6042 |
| Insomnia | 8 | 7.4% | 9 | 7.4% | 1.0000 |
| Upper Respiratory Tract<br>Infection Nos | 9 | 8.3% | 7 | 5.8% | 0.6049 |
| Allergy | 6 | 5.6% | 3 | 2.5% | 0.3128 |
Table 1: Anticipated Events
The adverse events reported with the Deep TMS System are typical to TMS treatments and were reported with similar or lower incidence rates compared to these events reported with other TMS devices.
There was one serious adverse reported in the study that was device related. The event was reported in a subject who experienced a seizure, following drinking at least a half a bottle of wine the night before the treatment. This SAE is considered device related, albeit with the caveat that withdrawal from alcohol forms a strong hazard for application of TMS due to the resulting significant seizure threshold lowering potential.
In summary, the safety and efficacy results of the Multicenter Deep TMS clinical study demonstrated the safety and effectiveness of the Brainsway Deep TMS System for treatment of Major Depressive Disorder.
,
{5}------------------------------------------------
### Substantial Equivalence:
The Brainsway DTMS System is substantially equivalent to the NeuroStar TMS Therapy® System (manufactured by Neuronetics Ltd., and the subject of 510(k) document nos. K061053 and K083538). The Brainsway DTMS System is similar to the NeuroStar TMS Therapy® System (K061053 and K083538) regarding intended use and basic technological characteristics.
The Brainsway DTMS System and the NeuroStar TMS Therapy® System are both intended for the treatment of adult patients with Major Depressive Disorder.
The design of the Brainsway DTMS System is similar to the design of the NeuroStar TMS Therapy® System, as both systems are based on applying transcranial magnetic stimulation by means of repetitive pulse trains at a predetermined frequency. Both systems use the same mechanism of action, i.e., an electromechanical instrument that produces and delivers brief duration, rapidly alternating (pulsed) magnetic fields to induce electrical currents in localized regions of the prefrontal cortex.
Transcranial magnetic stimulation is enabled in the Brainsway DTMS System and in the NeuroStar TMS Therapy® System, as both devices have the same system components, consisting of a mobile cart or console, an electromagnetic coil, a positioning arm, a TMS stimulator and software. The basic operational procedure is the same in both the Brainsway DTMS System and the NeuroStar TMS Therapy® System consisting of system setup, patient preparation, determination of patient's motor threshold, coil positioning and administration of treatment at predefined treatment stimulation parameters. The main differences between the Brainsway DTMS System and the NeuroStar TMS Therapy® System are the design of the electromagnetic coil enabling activation of deep brain regions with the Brainsway DTMS coil and the pulse sequence parameters administered during treatment, as described in the table below:
{6}------------------------------------------------
| Device Treatment Parameters | Brainsway DTMS Treatment | NeuroStar TMS Therapy |
|----------------------------------|-----------------------------------|------------------------------|
| & Specifications | | |
| Magnetic Field Intensity: | 120% of the MT | 120% of the MT |
| Frequency: | 18 Hz | 10 Hz |
| Train duration: | 2 sec | 4 sec |
| Inter-train interval: | 20 sec | 26 sec |
| Number of trains: | રેર | 7 ર |
| Magnetic Pulses per Session: | 1980 | 3000 |
| Treatment Session Duration: | 20.2 min | 37.5 min |
| Sessions per Week: | 5 | 5 |
| Treatment Schedule: | 5 daily sessions for 4 weeks | 5 daily sessions for 6 weeks |
| | Bi-weekly sessions for another 12 | |
| | weeks (optional maintenance | |
| | treatments) | |
| Area of brain to be stimulated: | Prefrontal Cortex | Prefrontal Cortex |
| Applicator: | | |
| Configuration | Biphasic H-Coil | Biphasic Figure 8 Coil |
| Core material | Air | Ferromagnetic |
| Output Stimulation Parameters: | | |
| Amplitude in Standard | | |
| Motor Threshold (SMT) units | 0.6 - 1.4 | 0.22 - 1.6 |
| Pulse width (usec) | 370 | 185 |
| Frequency range (Hz) | 0.02 - 30 | 0.1 - 30 |
| Pulse train duration range (sec) | 1 - 20 | 1 - 20 |
| Inter-train interval range (sec) | 10 - 60 | 10-60 |
| Maximum trains per session | ~140 | ~140 |
| Maximum # of pulses per | 5000 | 5000 |
| session (cumulative exposure) | | |
Treatment is administered in each system according to a standard pulse sequence that has been demonstrated to be effective during the clinical trials performed with the system.
The basic software capabilities related to treatment administration are the same in the Brainsway DTMS System and in the NeuroStar TMS Therapy® System. This is enabled in the Brainsway DTMS System using the commercial TMS stimulator software. The NeuroStar TMS Therapy® System has its own system software that provides capabilities beyond setting the treatment administration parameters.
Both the Brainsway DTMS System and the NeuroStar TMS Therapy® System meet the same electrical and mechanical safety standards (IEC 60601-1) and the same. EMC standards (IEC 60601-1-2).
### Conclusions:
In summary, the indications for use, anatomical sites and treatment environment for the Brainsway DTMS System and for the NeuroStar TMS Therapy® System are similar. Furthermore, the technological characteristics of the Brainsway DTMS System and the NeuroStar TMS Therapy® System, including basic design, mechanism of action, components, specifications, treatment procedure, etc. are the same. The minor differences in the system components and software do not affect the treatment procedure or outcome. The differences in the coil and the treatment stimulation
{7}------------------------------------------------
parameters have been demonstrated in a clinical study as safe and effective and therefore, do not raise new questions of safety and effectiveness.
As aforementioned, the Brainsway DTMS System has been validated in performance testing, including a clinical study. The complete clinical study report is provided in Section 20 (Clinical) of this 510(k) submission. The clinical performance data supports the safety and effectiveness of the Brainsway DTMS System. Consequently, the Brainsway DTMS System is substantially equivalent to the previously cleared NeuroStar TMS Therapy® System (K061053 and K083538).
{8}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/8/Picture/12 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three stripes forming its wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Public Health Service
January 7, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Brainsway, Ltd c/o Ahava Stein A. Stein Regulatory Affairs Consulting 20 Hata'as St. (POB 124) 44425 Kfar Saba ISRAEL
Re: K122288
Trade/Device Name: Brainsway Deep TMS System Regulation Number: 21 CFR 882.5805 Regulation Name: Repetitive Transcranial Magnetic Stimulator for Treatment of Major Depressive Disorder Regulatory Class: Class II Product Code: OBP Dated: October 29, 2012 Received: November 1, 2012
Dear Ms. Stein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{9}------------------------------------------------
Page 2 – Ms. Ahava Stein
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Victor Krauthamer -S
Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
# INDICATIONS FOR USE
510(k) Number (if known): K122288
Device Name:
Brainsway Deep TMS System
Indications For Use:
The Brainsway Deep TMS System is indicated for the treatment of depressive episodes in adult patients suffering from Major Depressive Disorder who failed to achieve m addit patients buttoring from previous anti-depressant medication treatment in the current episode.
ブ Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Brian D. Pullin -S |
|----------------------------------------------------------------|
| (Division Sign Off) |
| Division of Neurological and Physical Medicine Devices (DNPMD) |
| 510(k) Number | K122288 |
|---------------|---------|
|---------------|---------|
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.