visor2 is used to aid 3D navigation for positioning of Transcranial Magnetic Stimulation (TMS) coils over preselected brain regions based on data from MRI measurements or scalp landmarks and optional EMG recording in clinical procedures. The visor2 system is compatible with the following commercially available TMS coils of the following TMS companies: - Mag & More: PMD70-aCool Double coil Figure-8 (510575) - MagStim Company Ltd.: Double 70mm Remote Coil (3190-00 rev2), Double 70mm Air Film Coil (3910 00); - MagVenture A/S: Double 75mm (C-B60), Double Cooled 75mm (Cool-B65), Double 95mm (D-B80), Angulated Double Cooled 97mm (Cool-B70 rev2), Butterfly figure-8 (C-B70) - Neuronetics Inc.: NeuroStar coil The visor2 system must be operated by a trained physician and must not be used during surgical operations.
Device Story
3D navigation system for TMS coil positioning; inputs include MRI data, scalp landmarks, and optional EMG signals; uses optical stereotactic tracking (infrared camera and tracking tools) to register patient head and coil in 3D space; provides visual guidance on All-in-One touch display; operated by trained physicians in clinical/outpatient settings; aids accurate coil placement relative to anatomical/functional targets; benefits patient by ensuring precise stimulation delivery; does not perform stimulation itself.
Clinical Evidence
Bench testing only. No clinical data required. Performance validated via electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and system accuracy testing. Overall system accuracy demonstrated <4mm. EMG acquisition accuracy validated to 10% of parameters. Biocompatibility confirmed for patient-contacting pointer tool.
Technological Characteristics
Cart-based system with All-in-One PC; optical tracking via NDI infrared camera; tracking tools made of Polyurethane (PUR); connectivity via DICOM for MRI/fMRI import; optional EMG amplifier (eego EE-410); software-controlled navigation; electrical safety per IEC 60601-1; EMC per IEC 60601-1-2.
Indications for Use
Indicated for 3D navigation to aid positioning of TMS coils over preselected brain regions using MRI data or scalp landmarks, with optional EMG recording, in clinical procedures. For use by trained physicians; not for surgical use.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
HORIZON TMS Therapy System with Navigation (K183376)
Nexstim Navigational Brain Stimulation (NBS) System 4 (K112881)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 8, 2022
Eemagine Medical Imaging Solutions GmbH % Gary Syring Principal Consultant Quality & Regulatory Associates, LLC 800 Levanger Lane Stoughton, Wisconsin 53589
Re: K210109
Trade/Device Name: visor2 system Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW, GWF, IKN, OBP Dated: March 9, 2022 Received: March 11, 2022
Dear Gary Syring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known) K210109
Device Name visor2 system
Indications for Use (Describe)
visor2 is used to aid 3D navigation for positioning of Transcranial Magnetic Stimulation (TMS) coils over preselected brain regions based on data from MRI measurements or scalp landmarks and optional EMG recording in clinical procedures. The visor2 system is compatible with the following commercially available TMS coils of the following TMS companies:
- · Mag & More: PMD70-aCool Double coil Figure-8 (510575)
- MagStim Company Ltd.: Double 70mm Remote Coil (3190-00 rev2), Double 70mm Air Film Coil (3910 00);
- MagVenture A/S: Double 75mm (C-B60), Double Cooled 75mm (Cool-B65), Double 95mm (D-B80), Angulated Double Cooled 97mm (Cool-B70 rev2), Butterfly figure-8 (C-B70)
- · Neuronetics Inc .: NeuroStar coil
The visor2 system must be operated by a trained physician and must not be used during surgical operations.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <div> <span> <svg height="12" width="12"> <rect fill="black" height="12" width="12" x="0" y="0"></rect> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
RAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for eemagine medical imaging solutions. The word "eemagine" is written in gray, with the words "medical imaging solutions" written in a smaller font below it. To the right of the word "eemagine" are two blue circles with the letter "e" inside them. A blue line connects the two circles, resembling an EKG readout.
eemagine Medical Imaging Solutions GmbH, Gubener Str. 47, D-10243 Berlin
| 510(k) Summary | Date |
|---------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| This summary is provided to support the 510(k) pre-market notification for the visor2<br>system. | Berlin |
| Company Name: eemagine | Reference (eemagine):<br>e.g. contract ID |
| Company Contact: Frank Zanow, Chief Executive Officer<br>eemagine Medical Imaging Solutions GmbH | |
| Date Summary Prepared: April 8, 2022 | eemagine GmbH<br>Gubener Str.47<br>D-10243 Berlin<br>Tel. +49 30 29048404<br>Fax. +49 30 29048405<br>www.eemagine.com |
| Trade Name: visor2 system | |
| Common Name: Accessory to Repetitive Transcranial Magnetic Stimulation | |
| Classification Name: Repetitive Transcranial Magnetic Stimulation System<br>21 CFR 882.4560, Product Code: HAW, Class II | |
| Primary Predicate: K191422, Neural Navigator<br>Manufacturer: Soterix Medical, Inc., New York, NY 10001 | CEO/Geschäftsführer:<br>Ralf Hauffe, Ph.D. |
| Reference Devices: K183376, HORIZON TMS Therapy System with Navigation<br>Manufacturer: Magstim Company Ltd., Whitland, GB Sa34 0hr | Registration: HRB 7128<br>VAT ID: DE201608185 |
| K112881, Nexstim Navigational Brain Stimulation (NBS)<br>System 4, Nexstim NBS System 4 with NEXSPEECH®<br>Manufacturer: NEXSTIM OY, Helsinki, FI-00510 | Commerzbank Berlin<br>BIC Code CoBaDEBB120<br>IBAN |
DE73120400000458888500
# Product Description
The visor2 system is a 3D navigation system, supporting external/surface positioning of labeling identified commercially available Transcranial Magnetic Stimulation (TMS) coils based on the anatomical and functional areas for which use of the coil has been FDA cleared and as identified in cleared labeling for the coil. The visor2 system optionally records and processes electrophysiological signals to guide the user in the TMS coil placement. The visor2 system is compatible with TMS stimulators identified in the labeling that meet IEC 60601-1 and IEC 60601-1-2 electrical safety and EMC standards, a compatible optical tracking system, and further compatible accessories. Optionally, a compatible amplifier identified in the labeling may be applied for measuring electrophysiological signals. The visor2 system must be operated by a trained physician, and must not be used during surgical operations.
The visor2 system consists of a cart, an All-in-One medical computer with touch interface, a mounting plate for an optional amplifier applied for EMG recording and a Tracking camera with mounting arm. There are two fundamental visor2 system variants: single or dual display. the second configuration is available with an additional touch display added next to the All-in-One computer. Both variants are identical with regard to hardware features.
{4}------------------------------------------------
# Indications for Use
visor2 is used to aid 3D navigation for positioning of Transcranial Magnetic Stimulation (TMS) coils over preselected brain regions based on data from MRI measurements or scalp landmarks and optional EMG recording in clinical procedures. The visor2 system is compatible with the following commercially available TMS coils of the following TMS companies:
- · Mag & More: PMD70-aCool Double coil Figure-8 (510575)
- MagStim Company Ltd.: Double 70mm Remote Coil (3190-00 rev2), Double 70mm Air Film Coil (3910 00);
- MagVenture A/S: Double 75mm (C-B60), Double Cooled 75mm (Cool-B65), Double 95mm (D-B80), Angulated Double Cooled 97mm (Cool-B70 rev2), Butterfly figure-8 (C-B70)
- · Neuronetics Inc.: NeuroStar coil
The visor2 system must be operated by a trained physician and must not be used during surgical operations.
| Substantial Equivalence Technical Characteristics | | | | | |
|----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Subject Device<br>visor2 system | Primary<br>Predicate<br>Neural<br>Navigator<br>(K191422) | Reference<br>Device<br>HORIZON TMS<br>Therapy System<br>with Navigation<br>(K183376) | Reference<br>Device<br>Nexstim<br>Navigational<br>Brain<br>Stimulation<br>(NBS) System 4,<br>Nexstim NBS<br>System 4 with<br>NEXSPEECH®<br>(K112881) | Discussion of comparison |
| Product Code,<br>Classification | HAW<br>(Subsequent:<br>GWF, IKN, OBP)<br>21 CFR<br>882.4560,<br>Class II | HAW<br>(Subsequent:<br>GWF, IKN, OBP)<br>21 CFR<br>882.4560,<br>Class II | OBP, HAW, IKN<br>21 CFR<br>882.5805,<br>Class II | GWF, HAW,<br>IKN<br>21 CFR<br>882.1870,<br>Class II | Classification of Subject Device,<br>Primary Predicate and Reference<br>Device matches.<br>The Subject Device is an<br>accessory to TMS therapy<br>systems, providing location for<br>the TMS coil. |
| Intended use | As an accessory<br>to TMS systems,<br>provide<br>navigation,<br>location support<br>for placement of<br>the TMS coil on<br>the skull in<br>relation to a prior<br>acquired MRI<br>image. | A<br>neuronavigation<br>system indicated<br>for accurate<br>positioning of the<br>treatment coil of<br>the CloudTMS<br>Therapy System<br>with respect to<br>target brain<br>regions based on<br>data obtained<br>from MRI<br>measurements. | A repetitive<br>transcranial<br>magnetic<br>stimulation<br>system that<br>delivers<br>transcranial<br>repetitive pulsed<br>magnetic fields of<br>sufficient<br>magnitude to<br>induce neural<br>action potentials<br>in the prefrontal<br>cortex to treat the<br>symptoms of<br>major depressive | Non-invasive<br>mapping of the<br>primary motor<br>cortex of the<br>brain to its<br>cortical gyrus.<br>The NBS System<br>provides<br>information that<br>may be used in<br>the assessment of<br>the primary motor<br>cortex for pre-<br>procedural<br>planning. | The intended use of Subject<br>Device, Primary Predicate and<br>both Reference Devices match<br>fully with regards to all aspects<br>related to effectiveness and safety<br>(navigation of a TMS coil relative<br>to MRI and/or scalp<br>landmarks/targets).<br>As accessory to the TMS system,<br>the device under review provides<br>TMS coil location support.<br>The device under review provides<br>a locating feature to TMS<br>systems. The device under review<br>is an optional accessory to |
| Substantial Equivalence Technical Characteristics | | | | | |
| Feature | Subject Device<br>visor2 system | Primary<br>Predicate<br>Neural<br>Navigator<br>(K191422) | Reference<br>Device<br>HORIZON TMS<br>Therapy System<br>with Navigation<br>(K183376) | Reference<br>Device<br>Nexstim<br>Navigational<br>Brain<br>Stimulation<br>(NBS) System 4,<br>Nexstim NBS<br>System 4 with<br>NEXSPEECH®<br>(K112881) | Discussion of comparison |
| Indications<br>for Use | visor2 is used to<br>aid 3D navigation<br>for positioning of<br>Transcranial<br>Magnetic<br>Stimulation<br>(TMS) coils over<br>preselected brain<br>regions based on<br>data from MRI<br>measurements or<br>scalp landmarks<br>and optional<br>EMG recording<br>in or clinical<br>procedures. The<br>visor2 system is<br>compatible with<br>the following<br>commercially<br>available TMS<br>coils of the<br>following TMS<br>companies:<br>Mag & More:<br>PMD70-aCool<br>Double coil<br>MagStim<br>Company Ltd.:<br>Double 70mm<br>Remote Coil<br>(3190-00 rev2),<br>Double 70mm<br>Air Film<br>(3910-00)<br>MagVenture A/S:<br>Double 75mm<br>(C-B60), Double<br>Cooled 75mm<br>(Cool-B65), | The Neural<br>Navigator is a<br>neuronavigation<br>system indicated<br>for accurate<br>positioning of the<br>treatment coil of<br>the CloudTMS<br>Therapy System<br>with respect to<br>target brain<br>regions based on<br>data obtained<br>from MRI<br>measurements.<br>Specifically, the<br>Neural Navigator<br>is indicated for<br>use with the<br>following<br>CloudTMS<br>Therapy System<br>coils<br>manufactured by<br>Neurosoft Ltd:<br>AFEC-02-100<br>and<br>AFEC-02-100-C. | HORIZON® is<br>indicated for the<br>treatment of<br>Major Depressive<br>Disorder in adult<br>patients who have<br>failed to achieve<br>satisfactory<br>improvement<br>from prior<br>antidepressant<br>medication in the<br>current episode. | The Nexstim<br>Navigated Brain<br>Stimulation<br>System (NBS<br>System) 4 is<br>indicated for non-<br>invasive mapping<br>of the primary<br>motor cortex of<br>the brain to its<br>cortical gyrus.<br>The NBS System<br>4 provides<br>information that<br>may be used in<br>the assessment of<br>the primary motor<br>cortex for<br>preprocedural<br>planning.<br>Nexstim<br>NEXSPEECH,<br>when used<br>together with the<br>Nexstim NBS<br>System 4, is<br>indicated for<br>noninvasive<br>localization of<br>cortical areas that<br>do not contain<br>essential speech<br>function. Nexstim<br>NEXSPEECH<br>provides<br>information that<br>may be used in<br>pre-surgical<br>planning in<br>patients<br>undergoing brain<br>surgery.<br>Intraoperatively,<br>the localization<br>information<br>provided by | The indication of use of Subject<br>Device, Primary Predicate and<br>both Reference Devices match<br>fully with regard to all aspects<br>related to effectiveness and safety<br>(navigation of a TMS coil relative<br>to MRI and/or scalp<br>landmarks/targets). The Subject<br>Device is more flexible with<br>regard to supported coils than any<br>of the Reference and Predicate<br>system.<br>As accessory to the TMS system,<br>the device under review provides<br>TMS coil location support.<br>The visor2 system under review is<br>an optional accessory to support<br>locating/navigating TMS coil<br>feature to existing TMS systems<br>for their intended use.<br>Note:<br>The reference device K183376<br>HORIZON TMS Therapy System<br>with Navigation includes a<br>StimGuide feature to support<br>TMS coil location/navigation.<br>The StimGuide feature is a sub<br>licensed variation of the visor2<br>system, was reviewed in the<br>Horizon TMS Therapy System<br>with Navigation, K183376, to<br>support TMS coil location. |
| Substantial Equivalence Technical Characteristics | | | | | |
| Feature | Subject Device<br>visor2 system | Primary<br>Predicate<br>Neural<br>Navigator<br>(K191422) | Reference<br>Device<br>HORIZON TMS<br>Therapy System<br>with Navigation<br>(K183376) | Reference<br>Device<br>Nexstim<br>Navigational<br>Brain<br>Stimulation<br>(NBS) System 4,<br>Nexstim NBS<br>System 4 with<br>NEXSPEECH®<br>(K112881) | Discussion of comparison |
| | Double 95mm<br>(D-B80),<br>Angulated<br>Double Cooled<br>97mm<br>(Cool-B70 rev2),<br>Butterfly figure-8<br>(C-B70)<br>Neuronetics Inc.:<br>NeuroStar coil<br>The visor2<br>system must be<br>operated by a<br>trained physician<br>and must not be<br>used during<br>surgical<br>operations. | | | NEXSPEECH® is<br>intended to be<br>verified by direct<br>cortical<br>stimulation.<br>The Nexstim<br>NBS System 4<br>and NBS System<br>4 with<br>NEXSPEECH®<br>are not intended<br>to be used during<br>a surgical<br>procedure. The<br>Nexstim NBS<br>System 4 and<br>NBS System 4<br>with<br>NEXSPEECH®<br>are intended to be<br>used by trained<br>clinical<br>professionals. | |
| TMS Coil<br>Position<br>Principle | Based on<br>anatomy (MRI<br>picture) and<br>indirect targeting<br>of treatment<br>target through<br>measured<br>distance and<br>direction using<br>optical<br>stereotactic<br>navigation. | Based on<br>Anatomy (MRI<br>picture) and<br>electro-magnetic<br>field navigation. | Indirect targeting<br>of treatment<br>target through<br>measured<br>distance and<br>direction using<br>optical<br>stereotactic<br>navigation and<br>anatomy. | Based on<br>Anatomy (MRI<br>picture) and<br>optical<br>stereotactic<br>navigation. | All devices support the<br>location/navigation of the TMS<br>coil. The reference device<br>K183376 HORIZON TMS<br>Therapy System with Navigation<br>includes a StimGuide feature to<br>support TMS coil<br>location/navigation uses the same<br>methodology as the visor2 system<br>under review.<br>The StimGuide feature is a sub<br>licensed variation of the visor2<br>system.<br>Subject Device, Primary Predicate<br>and both Reference Devices all<br>use the same or equivalent<br>position principle registering the<br>physical head/scalp and a coil via<br>tracker tools tracked by a stereo<br>infrared camera 3D tracking<br>system or in case of the Primary<br>Predicate Device electro-magnetic<br>field tracking. Electro-magnetic |
| Substantial Equivalence Technical Characteristics | | | | | |
| Feature | Subject Device<br>visor2 system | Primary<br>Predicate<br>Neural<br>Navigator<br>(K191422) | Reference<br>Device<br>HORIZON TMS<br>Therapy System<br>with Navigation<br>(K183376) | Reference<br>Device<br>Nexstim<br>Navigational<br>Brain<br>Stimulation<br>(NBS) System 4,<br>Nexstim NBS<br>System 4 with<br>NEXSPEECH®<br>(K112881) | Discussion of comparison |
| Fundamental | Infrared locators<br>and 3D tracking | Electro-magnetic<br>field tracking | Infrared locators<br>and 3D tracking | Infrared locators<br>and camera | field tracking and optical field<br>tracking are both relying on<br>identifying the position of tools in<br>3D space. Optical tracking is in<br>general more precise as it relies<br>on shorter wavelength than<br>electrical tracking. All devices<br>define targets either via<br>landmarks in a co-registered MRI<br>image (same principle in all<br>systems using nasion-ear<br>landmarks) or directly on the<br>scalp. Further, all devices define a<br>point that represents the centre<br>position of the coil for navigation.<br>This is either referred to as a<br>geometrical centre of the coil<br>defining the maximum of the field<br>simply by geometry of the coil or<br>referring to the electrical field<br>maximum by design for any<br>treatment coil including tilted<br>designs. This point is at the<br>geometric centre of the coil and as<br>close as possible to the scalp<br>surface. The display of any<br>modelled induced electrical field<br>in the head tissue does not impact<br>the accuracy of the stimulus<br>delivered relative to the target but<br>only illustrates the field<br>distribution of a TMS stimulus<br>induced in head tissue. The<br>Primary Predicate device as well<br>as the Subject Device and the<br>Horizon Reference device rely on<br>geometry directly for the defining<br>the stimulation centre point while<br>reference K112881 calculates an<br>E-field for this purpose. With<br>regard to effectiveness and safety<br>all devices follow the same<br>working principle and perform<br>with identical effectiveness. The<br>Subject Device does not impose<br>any new or additional safety risks.<br>Comparison and discussion for |
| Substantial Equivalence Technical Characteristics | | | | | |
| Feature | Subject Device<br>visor2 system | Primary<br>Predicate<br>Neural<br>Navigator<br>(K191422) | Reference<br>Device<br>HORIZON TMS<br>Therapy System<br>with Navigation<br>(K183376) | Reference<br>Device<br>Nexstim<br>Navigational<br>Brain<br>Stimulation<br>(NBS) System 4,<br>Nexstim NBS<br>System 4 with<br>NEXSPEECH®<br>(K112881)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.