HORIZON® is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
Device Story
Computerized, electromechanical TMS system; delivers non-invasive, pulsed magnetic fields to induce electrical currents in cerebral cortex. Used for Major Depressive Disorder treatment. System components: mainframe, user interface (touchscreen), stimulating coils, StimGuide navigation, and arm. Operated by licensed physician in clinics/hospitals. StimGuide uses optical stereotactic navigation to assist coil positioning over target area (DLPFC). Motor threshold (MT) determination via visual observation of muscle response or optional EMG. Treatment delivered via repetitive pulse trains at predetermined frequency/intensity. Provides non-invasive cortical stimulation; helps clinicians target treatment area precisely; potentially benefits patients by providing alternative therapy for treatment-resistant depression.
Clinical Evidence
Bench testing only. Electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and biocompatibility (ISO 10993-1) testing performed. Software verification and validation completed. Magnetic field characteristics previously cleared in K180907. No new clinical data presented; equivalence supported by predicate performance and non-clinical testing.
Indicated for treatment of Major Depressive Disorder in adults who failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
Regulatory Classification
Identification
A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.
Predicate Devices
Mag Vita TMS Therapy System w/Theta Burst Stimulation (K173620)
Nexstim Navigated Brain Therapy (NBT) System 2 (K171902)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 3, 2019
Magstim Company Ltd. Tom Campbell Regulatory Affairs Manager Spring Gardens Whitland. Carmarthenshire SA34 0HR, United Kingdom
Re: K183376
Trade/Device Name: HORIZON TMS Therapy System with Navigation Regulation Number: 21 CFR 882.5805 Regulation Name: Repetitive Transcranial Magnetic Stimulation System Regulatory Class: Class II Product Code: OBP, HAW, IKN Dated: February 28, 2019 Received: March 4, 2019
Dear Tom Campbell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Actinclude requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Pamela D. Scott-S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183376
Device Name
HORIZON® TMS Therapy System with Navigation
Indications for Use (Describe)
HORIZON® is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "magstim" in a bold, sans-serif font. The word is a dark blue color, and there is a light blue circle above the "i" in "magstim". There is a registered trademark symbol to the right of the word.
# 510(k) SUMMARY Magstim's HORIZON® TMS Therapy System with Navigation
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Magstim Company Limited Spring Gardens, Whitland, Carmarthenshire SA34 0HR, United Kingdom
Phone: +44 (0) 1994 240798 Facsimile: +44 (0) 1994 240061
Contact Person: Tom Campbell
Date Prepared: April 3, 2019
### Name of Device
HORIZON® TMS Therapy System with Navigation
### Common or Usual Name
Repetitive Transcranial Magnetic Stimulation (rTMS) System
### Classification
Repetitive Transcranial Magnetic Stimulation (rTMS) System
21 C.F.R. § 882.5805, Class II, product code OBP
Additional Product Codes: HAW, IKN
### Predicate Devices
Maq Vita TMS Therapy System w/Theta Burst Stimulation, Tonica Elektronik A/S. (K173620) (Primary Predicate) HORIZON® TMS Therapy System, The Magstim Company Limited. (K180907) (Secondary Predicate) Nexstim Navigated Brain Therapy (NBT) System 2, Nexstim Plc. (K171902) (Reference Predicate)
### Device Description
The HORIZON® TMS Therapy System with Navigation (HORIZON®) is a computerized, electromechanical medical device that produces and delivers non-invasive, magnetic stimulation using brief duration rapidly alternating, or pulsed, magnetic fields to induce electrical currents directed at spatially discrete regions of the cerebral cortex. This method of cortical stimulation by application of brief magnetic pulses to the head is known as Transcranial Magnetic Stimulation.
{4}------------------------------------------------
The HORIZON® is a non-invasive tool for the stimulation of cortical neurons for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from antidepressant medication in the current episode. The HORIZON® is used for patient treatment by prescription only under the supervision of a licensed physician. It can be used in both inpatient and outpatient settings, including physicians' offices, clinics, and hospitals.
HORIZON® is an integrated system consisting of a combination of hardware, software, and accessories. Its technological characteristics are described in further detail below.
# Intended Use / Indications for Use
HORIZON® is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
# Technological Characteristics
HORIZON® is comprised of the following components:
- 1. HORIZON®TMS Therapy System
- a. HORIZON® User Interface:
- b. HORIZON® Mainframe;
- c. HORIZON® Power Supply:
- d. HORIZON® MT Remote Coil:
- e. HORIZON® E-z Cool Coil W/StimGuide Tracking Reference ('E-z Cool Coil (SG)');
- f. HORIZON® E-z Cart and Arm.
- 2. StimGuide
- 3. Accessories
- a. Interconnecting Cables;
- b. Footswitch:
- c. Pointer Tool:
- d. Head Reference.
The operator controls HORIZON® via the HORIZON® User Interface, using a graphic LCD panel with touchscreen technology. The operator instructions, given through the HORIZON® User Interface, direct the HORIZON® Mainframe in charging and discharging the device's high voltage pulse capacitor. The discharge is delivered to the patient via the stimulating coil. Motor threshold level can be determined using either the HORIZON® MT Remote Coil or the HORIZON® E-z Cool Coil (SG). Treatment is delivered to the treatment area via the HORIZON® E-z Cool Coil (SG), which is positioned above the treatment area. Positioning, and fixation, of the coil over the treatment area is accomplished using StimGuide and the HORIZON® E-z Arm. The HORIZON® Power Supply provides power to charge the high voltage pulse capacitor in the HORIZON® Mainframe.
Software documentation for a "moderate" level of concern has been provided.
# Non-Clinical Testing
Electrical safety and electromagnetic compatibility ("EMC") testing were conducted on HORIZON® to demonstrate that the device is compliant with IEC 60601-1 (Ed. 3.1.) and IEC
{5}------------------------------------------------
60601-1-2 (Ed. 4.). Environmental testing and acoustic output measurements has been conducted during IEC 60601-1 (Ed. 3.1) testing to demonstrate safety and performance.
Biocompatibility evaluation has demonstrated that patient-contacting components of the HORIZON® TMS Therapy System meet the requirements of ISO 10993-1 (Ed. 4.) standards. In addition, acoustic output measurements have been conducted during IEC 60601-1 (Ed. 3.1) testing to demonstrate safety and performance.
The software verification and validation testing further demonstrated that the software performs as intended and in accordance with specifications, with exception of identified unresolved anomalies.
The potential risks of HORIZON® have been identified and evaluated in compliance with ISO14971, and the risks were determined to be acceptable, or have been addressed with risk control measures.
As required by FDA's Guidance Document titled "Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation (rTMS) Systems", non-clinical magnetic field characteristic testing the HORIZON® TMS Therapy System was cleared by the FDA earlier in K180907. This testing has not been repeated as the magnetic field characteristics of the E-z Cool Coil (SG) are unchanged from the HORIZON® E-z Cool Coil.
# Substantial Equivalence
HORIZON® is substantially equivalent to its predicate devices, the HORIZON® TMS Therapy System (K180907) and the Mag Vita TMS Therapy System w/Theta Burst Stimulation (K173620).
HORIZON® and the secondary predicate device (K180907) have identical intended use and indications for use, equivalent principles of operation, as well as the same key technological characteristics.
The technological difference between HORIZON® and that cleared earlier in K180907, includes the addition of StimGuide simplifies repeatable coil positioning using optical stereotactic navigation and introduces the option to use EMG for MT determination.
The design of HORIZON® is substantially equivalent to its predicate devices, as all systems are based on applying transcranial magnetic stimulation by means of repetitive pulse trains at a predetermined frequency. All systems use the same mechanism of action, i.e., an electromechanical instrument that produces and delivers brief duration. rapidly alternating (pulsed) maqnetic fields to induce electrical currents in localized regions of the prefrontal cortex.
Transcranial magnetic stimulation is enabled in HORIZON® and its predicate devices, as all have the same key system components, consisting of electromagnetic coils, a coil holding mechanism, a TMS stimulator and software.
The principles of operation of HORIZON® are similar to that cleared earlier in K180907, the differences being:
{6}------------------------------------------------
#### Alternative method to determine and register treatment related landmarks
The operation procedure for HORIZON® differs from that of the secondary predicate device (K180907) however the use of optical stereotactic coil positioning is supported by the reference predicate device, Nexstim Navigated Brain Therapy (NBT) System 2 (K171902). Combining optical stereotactic navigation with non-invasive TMS and simultaneous EMG. StimGuide is capable of locating areas of the brain capable of evoking muscle response when stimulated and locating the target area for depression therapy.
Non-clinical performance testing demonstrates that HORIZON® is both equivalent and as safe and effective as the secondary predicate device.
#### Additional method to determine Motor Threshold (MT) using electromyography (EMG)
The operation procedure for HORIZON® differs from that of the secondary predicate device (K180907) however is supported by the reference predicate device, Nexstim Navigated Brain Therapy (NBT) System 2 (K171902). Whereas, the secondary predicate device uses a visual observation of contractions of the Abductor Pollicis Brevis (APB) to determine resting Motor Threshold, HORIZON® introduces the option to use EMG in addition to visual observations of the APB response.
The clinical evidence to support the change to use EMG in addition to a visual APB response is supported by FDA clearance of reference predicate device, the Nexstim Navigated Brain Therapy (NBT) System 2 (K171902).
#### Modification to the labelling to include a recommendation for an additional treatment protocol known as iTBS.
This treatment protocol, is identical to the treatment protocol recommended by the primary predicate device, Maq Vita TMS Therapy System w/Theta Burst Stimulation (K173620).
Non-clinical performance testing has demonstrated that the change to the HORIZON® TMS Therapy System labelling to add a recommendation for an additional treatment protocol known as iTBS, raises no new issues of safety or effectiveness.
The similarities and minor differences between HORIZON®, HORIZON® TMS Therapy System (Secondary Predicate). Mag Vita TMS Therapy System w/Theta Burst Stimulation (Primary Predicate) and Nexstim Navigated Brain Therapy (NBT) System 2 (Reference Predicate) are described in Table 1.
HORIZON® meets the same electrical and mechanical safety standards (IEC 60601-1) and the same EMC standards (IEC 60601-1-2).
{7}------------------------------------------------
### Conclusions
In summary, the intended use and indications for use for HORIZON® and the secondary predicate device, the HORIZON® TMS Therapy System (K180907) are identical.
Furthermore, HORIZON® shares the same key technological characteristics, as the secondary predicate device as both have the same key system components, consisting of electromagnetic coils, a coil holding mechanism, a TMS stimulator and software. The difference includes the addition of StimGuide for navigated coil positioning.
The principles of operation differ from that of the secondary predicate device (K180907). The differences include the addition of StimGuide which simplifies coil positioning using optical stereotactic navigation, the option to use EMG for MT determination and a modification to the labeling to include a recommendation for an additional treatment protocol known as iTBS. These changes are supported by the clearance of the primary predicate device (K173620) and the reference predicate device (K171902).
Non-clinical performance testing demonstrates that HORIZON® is both equivalent and as safe and effective as the secondary predicate device, the HORIZON® TMS Therapy System (K180907).
Thus, HORIZON® is considered substantially equivalent to its predicate devices.
{8}------------------------------------------------
# Table 1: Substantial Equivalence Summary
| Criteria | HORIZON® TMS Therapy<br>System with Navigation<br>(Subject of this submission) | Mag Vita TMS Therapy<br>System w/Theta Burst<br>Stimulation<br>(K173620)<br>(Primary Predicate) | HORIZON® TMS Therapy System<br>(K180907)<br>(Secondary Predicate) | Nexstim Navigated<br>Brain Therapy (NBT)<br>System 2<br>(K171902)<br>(Reference Predicate) | | |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Manufacturer | Magstim Company Limited | Tonica Elektronik A/S | Magstim Company Limited | Nexstim Plc | | |
| Device Name | HORIZON® TMS Therapy<br>System with Navigation | Mag Vita TMS Therapy<br>System w/Theta Burst<br>Stimulation | HORIZON® TMS Therapy System | Nexstim Navigated Brain<br>Therapy (NBT) System 2 | | |
| Clearance date | 08/14/2018 | 08/14/2018 | 08/03/2018 | 11/10/2018 | | |
| 510(k) number | K173620 | K173620 | K180907 | K171902 | | |
| Device code | OBP | OBP | OBP | OBP, GWF, HAW, IKN | | |
| Intended Use/<br>Indications for Use | The HORIZON® TMS Therapy<br>System is indicated for the<br>treatment of Major Depressive<br>Disorder in adult patients who<br>have failed to achieve<br>satisfactory improvement from<br>prior antidepressant medication<br>in the current episode. | The Mag Vita TMS Therapy<br>System w/Theta Burst Stimulation<br>is indicated for the treatment of<br>Major Depressive Disorder in adult<br>patients who have failed to receive<br>satisfactory improvement from<br>prior antidepressant medication in<br>the current episode. | The HORIZON® TMS Therapy System is<br>indicated for the treatment of Major Depressive<br>Disorder in adult patients who have failed to<br>achieve satisfactory improvement from prior<br>antidepressant medication in the current<br>episode. | Nexstim Navigated Brain<br>Therapy (NBT) System 2 is<br>indicated for the treatment of<br>Major Depressive Disorder in<br>adult patients who have failed<br>to receive satisfactory<br>improvement from prior<br>antidepressant medication in<br>the current episode. | | |
| Standard Treatment Protocol | | | | | | |
| Magnetic Field<br>Intensity | 120% of the MT | N/A | 120% of the MT | 120% of the MT | | |
| Stimulus Frequency | 10 Hz | N/A | 10 Hz | 10 Hz | | |
| Stimulus Train<br>duration | 4 sec | N/A | 4 sec | 4 sec | | |
| Inter-train interval | 11-26 sec | N/A | 11-26 sec | 26 sec | | |
| Number of trains | 75 | N/A | 75 | 75 | | |
| Magnetic Pulses per<br>Session | 3000 | N/A | 3000 | 3000 | | |
| Treatment Session<br>Duration | 18.8 min-37.5 min | N/A | 18.8 min-37.5 min | 37.5 min | | |
| Sessions/week | 5 | N/A | 5 | 5 | | |
| Treatment Schedule | 5 daily sessions for 6 weeks | N/A | 5 daily sessions for 6 weeks | 5 daily sessions for 6<br>weeks | | |
| Area of brain to be<br>stimulated | Left Dorsolateral Prefrontal<br>Cortex | N/A | Left Dorsolateral Prefrontal Cortex | Left Dorsolateral<br>Prefrontal Cortex | | |
| iTBS Treatment Protocol | | | | | | |
| Stimulation Intensity | 120% of the MT | 120% of the MT | N/A | N/A | | |
| Repetition Rate | 50 Hz (5 pulses per sec) | 50 Hz (5 pulses per sec) | N/A | N/A | | |
| Train Duration | 2 sec | 2 sec | N/A | N/A | | |
| Inter-train Interval | 8 sec | 8 sec | N/A | N/A | | |
| Burst Pulses | 3 | 3 | N/A | N/A | | |
| Bursts | 200 | 200 | N/A | N/A | | |
| Inter Pulse interval | 20 msec | 20 msec | N/A | N/A | | |
| Number of trains | 20 | 20 | N/A | N/A | | |
| Number of Pulses per<br>Session | 600 | 600 | N/A | N/A | | |
| Treatment Session<br>Duration | 3.09 min | 3.09 min | N/A | N/A | | |
| Sessions/week | 5 | 5 | N/A | N/A | | |
| Treatment Schedule | 5 daily sessions for 6 weeks | 5 daily sessions for 6 weeks | N/A | N/A | | |
{9}------------------------------------------------
{10}------------------------------------------------
| Area of brain to be<br>stimulated | Left Dorsolateral Prefrontal<br>Cortex | | Left Dorsolateral Prefrontal<br>Cortex | N/A | | N/A | |
|------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | HORIZON®<br>MT Remote<br>Coil | HORIZON®<br>E-z Cool Coil<br>(SG) | Cool-B70 Coil | HORIZON®<br>MT Remote<br>Coil | HORIZON®<br>E-z Cool Coil | HORIZON®<br>AFC | Nexstim Stimulation Coil |
| Waveform | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic |
| Core Material | Air | Air | Air | Air | Air | Air | Air |
| Shape | Figure-of-<br>eight | Figure-of-<br>eight | Figure-of-eight | Figure-of-<br>eight | Figure-of-<br>eight | Figure-of-<br>eight | Figure-of-eight |
| Coil Winding<br>Configuration<br>Inner Diameter (ID)<br>Outer Diameter (OD)<br>Winding Height (H)<br>Number of Windings (NW) | ID: 55mm<br>OD: 92mm<br>H: 6mm<br>NW: 2x9 | ID: 37.5mm<br>OD: 97mm<br>H: 11mm<br>NW: 2x(3x23) | ID: 23mm<br>OD: 96mm<br>H: 12mm<br>NW: 2x11 | ID: 55mm<br>OD: 92mm<br>H: 6mm<br>NW: 2x9 | ID: 37.5mm<br>OD: 97mm<br>H: 11mm<br>NW:<br>2x(3x23) | ID: 43mm<br>OD: 92mm<br>H: 11mm<br>NW:<br>2x(3x19) | ID: unknown<br>OD: unknown<br>H: unknown<br>NW: 2x10 |
| Pulse Width | 330μs | 340μs | 280μs | 330μs | 340μs | 300μs | 230μs |
| Amplitude in SMT<br>units<br>(Standard Motor Threshold) | 0.28 - 1.9 | 0.28 - 1.9 | 0 - 1.7 | 0.28 - 1.9 | 0.28 - 1.9 | 0.28 - 1.9 | 0 - 2.5 |
| Frequency range (Hz)<br>at 100% | 1 - 20 | 1 - 20 | 0.1 - 30 | 1 - 20 | 1 - 20 | 1 - 20 | 0.1-50 Hz |
| Coil Position Principle | Indirect targeting of treatment<br>target through measured<br>distance and direction<br>(5.5cm) from MT Hotspot<br>using optical stereotactic<br>navigation. Measure derived<br>from statistical distance of<br>DLPFC from MT hotspot. | Indirect targeting of treatment<br>target through measured<br>distance and direction<br>(5.5cm) from MT Hotspot<br>using optical stereotactic<br>navigation. Measure derived<br>from statistical distance of<br>DLPFC from MT hotspot. | Indirect targeting of treatment<br>target through measured<br>distance and direction (5.5cm)<br>from MT Hotspot. Measure<br>derived from statistical<br>distance of DLPFC from MT<br>hotspot. | Indirect targeting of treatment target<br>through measured distance and direction<br>(5.5cm) from MT Hotspot. Measure<br>derived from statistical distance of DLPFC<br>from MT hotspot. | Indirect targeting of treatment target<br>through measured distance and direction<br>(5.5cm) from MT Hotspot. Measure<br>derived from statistical distance of DLPFC<br>from MT hotspot. | Indirect targeting of treatment target<br>through measured distance and direction<br>(5.5cm) from MT Hotspot. Measure<br>derived from statistical distance of DLPFC<br>from MT hotspot. | Individual patient direct<br>targeting of anatomical<br>treatment location<br>(DLPFC).<br>Placing of E-field<br>maximum location on 3D<br>model built from patients<br>individual MRI. |
{11}------------------------------------------------
| MT Response<br>Detection | Option 1. EMG provides<br>quantitative data based on<br>which user defines MT.<br>Option 2. Visual qualitative<br>monitoring for APB response | Visual qualitative monitoring<br>for APB response | Visual qualitative monitoring for APB<br>response | EMG provides quantitative<br>data based on which user<br>defines MT. |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|---------------------------------------------------|----------------------------------------------------------------------|
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|---------------------------------------------------|----------------------------------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.