The Pulse Oximeter is intended for measuring the functional oxygen saturation and pulse rate (PR) through a patient's finger. It is applicable for spot-checking SpO2 and pulse rate (PR) of adult and pediatric patients in homes and clinics.
Device Story
Portable fingertip pulse oximeter; uses photoelectric oxyhemoglobin inspection technology; emits red and near-infrared light through a clamping finger-type sensor; measures light absorption based on Lambert-Beer Law; calculates SpO2 and pulse rate; powered by internal AAA batteries; used by clinicians or patients in homes and hospitals for spot-checking; provides real-time SpO2 and PR values on a display; assists healthcare providers in assessing patient oxygenation status; benefits patients through non-invasive, rapid monitoring.
Clinical Evidence
Clinical validation performed per ISO 80601-2-61 on 12 healthy adult volunteers (5 male, 7 female) with varying skin pigmentations. 278 data points collected across 70%-100% SpO2 range. Subject device compared against laboratory CO-Oximeter. Results: Arms < ±2% SpO2; Bland-Altman analysis confirmed agreement within 95% limits.
Technological Characteristics
Materials: ABS enclosure, silicone clip. Sensing: Photoelectric plethysmography (Red 660nm, IR 905nm). Power: 2x AAA 1.5V alkaline batteries. Connectivity: Standalone. Safety: Type BF applied part. Standards: ISO 10993-1, ISO 80601-2-61, IEC 60601-1, IEC 60601-1-11.
Indications for Use
Indicated for adult and pediatric patients in home and hospital environments (including clinical use in internist/surgery, anesthesia, intensive care, etc.) for non-invasive spot-checking of functional oxygen saturation (SpO2) and pulse rate (PR).
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 23, 2021
Shenzhen Hexin ZONDAN Medical Equipment Co., Ltd. John Liu RA Manager Floor 14, Block D, Dianlian Technology Building, the Crossing between South Circle Road and South Fuli Road, Shenzhen, Guangdong 518106 China
Re: K203854
Trade/Device Name: Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: March 23, 2021 Received: March 23, 2021
Dear John Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K203854
Device Name Pulse Oximeter
Indications for Use (Describe)
The Pulse Oximeter is non-invasive device intended for spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The portable fingertip device is indicated for adult and pediatric patients in home and hospital environments (including clinical use in internist/ surgery, anesthesia, intensive care, etc).
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:10pt">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image contains a logo for ZONDAN. The logo features a green graphic of two overlapping cards with a white heart and heartbeat symbol on them. To the right of the graphic is the company name "ZONDAN" in bold, black letters, and below that is the tagline "Core Technology Care for Life" in a smaller, gray font.
# 510(k) Summary
This summary of 510(k) information is submitted as required by requirements of 21 CFR 807.92.
### 1. Submitter's Identification
| Applicant: | Shenzhen Hexin ZONDAN Medical Equipment Co., Ltd. |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Floor 14, Block D, Dianlian Technology Building, the<br>Crossing between South Circle Road and South Fuli Road,<br>Guangming District, 518106 Shenzhen, Guangdong, China |
| Contact person: | Li Feng<br>General Manager<br>Tel: +86-755-26865970<br>E-mail: 13602607257@163.com |
| Prepared by: | Mr. John Liu<br>Regulatory Manager<br>E-mail: liuming@zondan.com |
| Date prepared: | Dec. 21, 2020 |
## 2. Subject device information
Device Common Name: Oximeter Device Trade Name: Pulse Oximeter Model: A2, A3, A4, A5 Regulation Number: 21 CFR 870.2700 Regulation Name: Oxiemter Regulatory Class: II Product Code: DQA
## 3. Predicate device information
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for Zondan. The logo consists of a green square with a white heart and an EKG line inside of it. To the right of the square is the word "ZONDAN" in large, bold, black letters. Below the word "ZONDAN" is the phrase "Core Technology Care for Life" in smaller, gray letters.
510(k) Number: K163135 Trade/ Device Name: Fingertip Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Manufacturer: Shenzhen Fitfaith Technology Co., Ltd.
#### 4. Device description
The Pulse Oximeter is intended for measuring the functional oxygen saturation and pulse rate (PR) through a patient's finger. It is applicable for spot-checking SpO2 and pulse rate (PR) of adult and pediatric patients in homes and clinics.
The Pulse Oximeter is as a stand-alone device, which is powered by an internal battery. The device is for multi-use. The device is not for life supporting or life sustaining, not for implant. The device is not sterile, the device is not a reprocessed single-use device, and the device does not contain drug or biological products.
### 5. Intended Use
The Pulse Oximeter is intended for measuring the functional oxygen saturation and pulse rate (PR) through a patient's finger. It is applicable for spot-checking SpO2 and pulse rate (PR) of adult and pediatric patients in homes and clinics.
#### 6. Indication for Use
The Pulse Oximeter is non-invasive device intended for spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The portable fingertip device is indicated for adult and pediatric patients in home and hospital environments (including clinical use in internist/ surgery, anesthesia, intensive care, etc).
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for ZONDAN. The logo consists of a green graphic on the left, which includes a white heart with an EKG line inside of it. To the right of the graphic is the company name, "ZONDAN" in bold, black letters. Below the company name is the tagline "Core Technology Care for Life" in a smaller, gray font.
# 7. Substantial Equivalence Comparison
| Characteristics | Subject device | Predicate device<br>(K163135) | Remark |
|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Device name | Pulse Oximeter | Fingertip Pulse<br>Oximeter | NA |
| Model | A2, A3, A4, A5 | A300 | NA |
| Manufacturer | Shenzhen Hexin<br>ZONDAN Medical<br>Equipment Co.,Ltd. | Shenzhen Fitfaith<br>Technology Co., Ltd. | NA |
| Classification | Class II | Class II | Substantially<br>Equivalent. |
| Product code | DQA | DQA | Substantially<br>Equivalent. |
| Prescription or<br>OTC | Prescription | Prescription | Substantially<br>Equivalent. |
| Intended use | The Pulse Oximeter is<br>intended for measuring<br>the functional oxygen<br>saturation and pulse<br>rate (PR) through a<br>patient's finger. It is<br>applicable for<br>spot-checking SpO2<br>and pulse rate (PR) of<br>adult and pediatric<br>patients in homes and<br>clinics. | The Fingertip Pulse<br>Oximeter is intended<br>for measuring the<br>functional oxygen<br>saturation and pulse<br>rate (PR) through a<br>patient's finger. It is<br>applicable for<br>spot-checking SpO2<br>and pulse rate (PR) of<br>adult and pediatric<br>patients in homes and<br>clinics. | Substantially<br>Equivalent. |
| Indications for | The Pulse Oximeter is | The Fingertip Pulse | Substantially |
| use | non-invasive device<br>intended for<br>spot-checking of<br>functional oxygen<br>saturation of arterial<br>hemoglobin (SpO2) and<br>pulse rate (PR). The<br>portable fingertip<br>device is indicated for<br>adult and pediatric<br>patients in home and<br>hospital environments<br>(including clinical use<br>in internist/ surgery,<br>anesthesia, intensive<br>care, etc). | Oximeter is<br>non-invasive device<br>intended for<br>spot-checking of<br>functional oxygen<br>saturation of arterial<br>hemoglobin (SpO2)<br>and pulse rate (PR).<br>The portable fingertip<br>device is indicated for<br>adult and pediatric<br>patients in home and<br>hospital environments<br>(including clinical use<br>in internist/ surgery,<br>anesthesia, intensive<br>care, etc). | Equivalent. |
| Design principle | A mathematical formula is established making<br>use of Lambert Beer Law according to Spectrum<br>Absorption Characteristics of Reductive<br>hemoglobin (RHb) and Oxyhemoglobin (HbO2)<br>in red and near-infrared zones. Operation<br>principle of the instrument: Photoelectric<br>Oxyhemoglobin Inspection Technology is<br>adopted in accordance with Capacity Pulse<br>Scanning and Recording Technology, so that two<br>beams of different wavelength of lights can be<br>focused onto a human nail tip through a<br>clamping finger-type sensor A measured signal | | Substantially<br>Equivalent. |
| Environment of<br>use | Home and Hospital | Home and Hospital | Substantially<br>Equivalent. |
| Patient type | Adult and Pediatric | Adult and Pediatric | Substantially<br>Equivalent. |
| Intended<br>application site | Fingertip | Fingertip | Substantially<br>Equivalent. |
| Type, Degree of<br>protection<br>against electric<br>shock | Internally powered<br>equipment;<br>Type BF applied part. | Internally powered<br>equipment;<br>Type BF applied part. | Substantially<br>Equivalent. |
| Power supply | 2*AAA 1.5V alkaline<br>battery | 2*AAA 1.5V alkaline<br>battery | Substantially<br>Equivalent. |
| Work current | <40mA | <40mA | Substantially<br>Equivalent. |
| Light<br>specification | Wavelength:<br>RED: 660nm±5nm,<br>IR: 905nm±5nm | Wavelength:<br>RED: 660nm±3nm,<br>IR: 905nm±5nm | Substantially<br>Equivalent. |
| | Maximum optical<br>power:<br>RED: 1.5mW,<br>IR: 1.2mW | Maximum optical<br>power:<br>RED: 1.5mW,<br>IR: 1.2mW | Substantially<br>Equivalent. |
| SpO2 display<br>range | 0%~100% | 0%~100% | Substantially<br>Equivalent. |
| SpO2<br>measurement | 70%-100%, accuracy:<br>±2%;<br><70%, no definition. | 70%-100%, accuracy:<br>±2%;<br><70%, no definition. | Substantially<br>Equivalent. |
| accuracy | | | |
| SpO2 resolution | 1% | 1% | Substantially Equivalent. |
| Pulse rate display range | 30bpm~240bpm | 25bpm~250bpm | Different Note 01 |
| Pulse rate measurement accuracy | $\pm$ 1bpm | $\pm$ 2bpm | Different Note 02 |
| Pulse rate resolution | 1bpm | 1bpm | Substantially Equivalent. |
| Operation environment | Temperature:<br>5°C ~ 40°C<br>Humidity:<br>30% ~ 80%<br>(non-condensing)<br>Atmospheric pressure:<br>70kPa ~ 106kPa | Temperature:<br>5°C ~ 40°C<br>Humidity:<br>15% ~ 85%<br>(non-condensing) | Different Note 03 |
| Storage & Transport environment | Temperature:<br>-20°C ~ +55°C<br>Humidity:<br>10% ~ 93%<br>(non-condensing)<br>Atmospheric pressure:<br>70kPa ~ 106kPa | Temperature:<br>-20°C ~ +55°C<br>Humidity:<br>10% ~ 95%<br>(non-condensing) | |
| Contacting material | ABS (Acrylonitrile butadiene styrene copolymers) for enclosure, silicone for clip | ABS (Acrylonitrile butadiene styrene copolymers) for enclosure, silicone for clip | Substantially Equivalent. |
| Contact<br>duration | <24h | <24h | Substantially<br>Equivalent. |
| Contact type | Skin surface-contacting | Skin<br>surface-contacting | Substantially<br>Equivalent. |
| Biocompatibility of patient<br>contact parts | Comply with ISO<br>10993-1 | Comply with ISO<br>10993-1 | Substantially<br>Equivalent. |
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for ZONDAN. The logo consists of a green icon with a white heart and an EKG line inside, followed by the word "ZONDAN" in large, bold, black letters. Below the word "ZONDAN" is the tagline "Core Technology Care for Life" in a smaller, gray font.
{7}------------------------------------------------
Image /page/7/Picture/2 description: The image shows the logo for ZONDAN. The logo features a green graphic on the left, which includes a heart with an EKG line inside of it. To the right of the graphic is the company name, "ZONDAN" in large, bold, black letters. Below the company name is the tagline "Core Technology Care for Life" in a smaller, gray font.
{8}------------------------------------------------
Image /page/8/Picture/2 description: The image shows the logo for ZONDAN. The logo consists of a green graphic with a white heart and heartbeat line inside of it, and the text "ZONDAN" in bold black letters. Below the company name is the text "Core Technology Care for Life" in a smaller, lighter font.
{9}------------------------------------------------
Image /page/9/Picture/2 description: The image shows the logo for Zondan. The logo features a green icon with a white heart and a heartbeat line inside. Next to the icon is the word "ZONDAN" in bold, black letters. Below the word is the tagline "Core Technology Care for Life" in a smaller, gray font.
#### Note 01:
The pulse rate display range of the subject device is different from the predicate device. But the subject device complies with the claimed display range according to bench test report.
#### Note 02:
The pulse rate measurement accuracy of the subject device is different from the predicate device. But the subject device complies with the claimed measurement accuracy according to bench test report.
#### Note 03:
The operation and storage & transport environment of subject device are different from predicate device. But the subject device has been verified all the full claimed range according to the IEC 60601-1, IEC 60601-1-11 and ISO 80601-2-61 standards.
Based on the aforementioned substantial equivalence comparison table, the subject device is as same as predicate device. Only Pulse rate display range (refer to Note 01), Pulse rate measurement accuracy (refer to Note 02), Operation and Storage & Transport environment (refer to Note 03) are different which had been validated. However, the differences would not raise any safety and effectiveness issue based on tests in this submission.
Thus, the subject device is substantially equivalent to the predicate device which is
{10}------------------------------------------------
Image /page/10/Picture/2 description: The image shows the logo for Zondan. The logo features a green square with a white heart and heartbeat line inside. To the right of the square is the word "ZONDAN" in large, bold, black letters. Below the word "ZONDAN" is the phrase "Core Technology Care for Life" in a smaller, gray font.
legally marketed in US.
## 8. Brief discussions of the non-clinical tests
The following tests were conducted to demonstrate substantial equivalence to the predicate device:
| Test Standard | Description | Result |
|----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| ANSI AAMI<br>ES60601-1:2005/(R)2012<br>and A1:2012 | Medical electrical equipment - Part 1:<br>General requirements for basic safety and<br>essential performance (IEC 60601-1:2005,<br>MOD) | Pass |
| IEC 60601-1-2 Edition<br>4.0 2014-02 | Medical electrical equipment - Part 1-2:<br>General requirements for basic safety and<br>essential performance - Collateral Standard:<br>Electromagnetic disturbances -<br>Requirements and tests | Pass |
| IEC 60601-1-11 Edition<br>2.0 2015-01 | Medical electrical equipment - Part 1-11:<br>General requirements for basic safety and<br>essential performance - Collateral Standard:<br>Requirements for medical electrical<br>equipment and medical electrical systems<br>used in the home healthcare environment | Pass |
| ISO 80601-2-61 Second<br>edition 2017-12<br>(Corrected version<br>2018-02) | Medical electrical equipment - Part 2-61:<br>Particular requirements for basic safety and<br>essential performance of pulse oximeter<br>equipment | Pass |
| ISO 10993-1 Fourth<br>edition 2009-10-15 | Biological evaluation of medical devices -<br>Part 1: Evaluation and testing within a risk<br>management process [Including: Technical<br>Corrigendum 1 (2010)] | Pass |
| Pulse Oximeter_Traditional 510k Submission | | |
| ISO 10993-5 Third<br>edition 2009-06-01 | Biological evaluation of medical devices -<br>Part 5: Tests for in vitro cytotoxicity | Pass |
| ISO 10993-10 Third<br>Edition 2010-08-01 | Biological evaluation of medical devices -<br>Part 10: Tests for irritation and skin<br>sensitization | Pass |
| ISO 14971 Second<br>edition 2007-03-01 | Medical devices - Application of risk<br>management to medical devices | Pass |
{11}------------------------------------------------
It is concluded that the subject device is in compliance with the requirements of the aforementioned tests.
#### 9. Brief discussions of the clinical tests
The functional oxygen saturation (SpO2) measurement has been validated in accordance with ISO 80601-2-61. The clinical testing was completed on healthy, non-smoking, a total of 12 healthy adult volunteers (5 men and 7 women) with light to dark skin pigmentations in the range of 70% to 100% against a laboratory CO-Oximeter. The subjects include 10 with light skin, and 2 with dark skin pigmentation. Total 278 data points were sampled for analysis. The measured arterial hemoglobin saturation value (SpO2) of the proposed device was compared with arterial hemoglobin oxygen (SaO2) value, determined from blood samples with a CO-oximeter. The accuracy of the device is in comparison with the CO-oximeter samples measured over the SpO2 range 70%-100%.
Accuracy data was calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-2-61, the SpO2 accuracy result showed that the root-mean-square (Arms) value of the Pulse Oximeter is less than ±2% SpO2 over the declared range of 70% to 100%; and the Agreement between Methods of Measurement with Multiple Observations per each subject was analyzed with the Bland and Altman statistics, the analysis demonstrated that the vast majority of data is within ±95% limit of agreement.
{12}------------------------------------------------
Image /page/12/Picture/2 description: The image contains a logo for ZONDAN, a company that appears to be in the healthcare industry. The logo features a green icon resembling a medical chart with a heart and heartbeat symbol inside. Next to the icon is the company name, "ZONDAN," in bold, black letters, followed by the tagline "Core Technology Care for Life" in a smaller, lighter font.
### 10.Software information
The software level of concern for the subject device is MODERATE. According to FDA Guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff", the software validation documentation summarizes the required information for a MODERATE Level of Concern device.
### 11. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in in this premarket notification, Shenzhen Hexin ZONDAN Medical Equipment Co., Ltd. concludes that the subject device is as safe and as effective, and thus substantially equivalent, to the predicate device, A300 Fingertip Pulse Oximeter manufactured by Shenzhen Fitfaith Technology Co., Ltd..
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.