The Fingertip Pulse Oximeter is non-invasive device intended for spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The portable fingertip device is indicated for adult and pediatric patients in home and hospital environments (including clinical use in internist/ surgery, anesthesia, intensive care, etc).
Device Story
Fingertip pulse oximeter; portable, battery-powered device. Inputs: red (660nm) and infrared (905nm) light beams transmitted through fingertip. Principle: Photoelectric Oxyhemoglobin Inspection Technology; Capacity Pulse Scanning and Recording; Lambert-Beer Law applied to differential light absorption of oxyhemoglobin and reductive hemoglobin. Output: SpO2 and pulse rate displayed on device screen. Used in homes and clinical environments (surgery, anesthesia, ICU) by patients or clinicians for spot-checking. Provides real-time physiological data to assist in monitoring patient oxygenation status; aids in identifying potential hypoxemia or pulse rate abnormalities.
Clinical Evidence
Controlled desaturation study conducted on 12 healthy, non-smoking adults (ages 21-32). 294 data sets collected (288 valid). Subject device SpO2 compared against CO-oximeter (control) across 70-100% range. Results met ISO 80601-2-61 criteria for accuracy and precision. Bland-Altman analysis showed majority of data within 95% limits of agreement. No adverse effects reported.
Technological Characteristics
Materials: ABS enclosure, silica gel clip. Sensing: Photoelectric sensor (660nm red, 905nm IR). Power: 2x AAA 1.5V alkaline batteries. Form factor: Portable fingertip clip. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993-5/10, ISO 80601-2-61.
Indications for Use
Indicated for adult and pediatric patients for non-invasive spot-checking of functional arterial hemoglobin oxygen saturation (SpO2) and pulse rate (PR) in home and hospital settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 19, 2017
Shenzhen Fitfaith Technology Co.,ltd % Migo. Yang Consultant Shenzhen Joyantech Consulting Co., Ltd. 1122#, Internation Mayor Communication Center, Baishizhong Rd 55#, Nanshan District, Shenzhen, 518000 CN
Re: K163135
Trade/Device Name: Fingertip Pulse Oximeter (Model: A300, A310, M110, M120, M130, M150, M160, M170, M230) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: July 15, 2017 Received: August 7, 2017
Dear Migo. Yang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Tara A. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K163135
#### Device Name
Fingertip Pulse Oximeter (Model: A300, A310, M100, M110,M120,M130,M150,M170,M230)
Indications for Use (Describe)
The Fingertip Pulse Oximeter is non-invasive device intended for spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The portable fingertip device is indicated for adult and pediatric patients in home and hospital environments (including clinical use in internist/ surgery, anesthesia, intensive care, etc).
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 10pt;">☑</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> </div> | <div> <span> <span style="font-size: 10pt;">☐</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) information is submitted as required by requirements of SMDA and 21 CFR §807.92.
## 1 Administrative Information
| Submission Date | Aug 18,2016 |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>information | Submitter's Name: Shenzhen Fitfaith Technology Co.,Ltd.<br>Address: Area B, Floor 9, Building D1, Tangwei Industrial Park<br>Donglong Road, Guangming New District, ShenZhen,<br>Guangdong, China.<br>Contact person: Yuan.Junfeng<br>TEL: +86-0755-29544253-803<br>FAX: +86-0755-29544253-801<br>E-Mail: 160666@qq.com |
| Submission<br>Correspondent | Contact person: Miss Migo. Yang<br>E-Mail: migo@cefda.com<br>Shenzhen Joyantech Consulting Co., Ltd.<br>1122#, International Mayor Communication Center,<br>Baishizhong Road 55#, Nanshan District, Shenzhen,<br>Guangdong, China. |
| Image: Logo | Contact person: Mr. Field.Fu<br>E-Mail: cefda13485@163.com<br>Shenzhen Joyantech Consulting Co., Ltd.<br>1122#, International Mayor Communication Center,<br>Baishizhong Road 55#, Nanshan District, Shenzhen,<br>Guangdong, China |
| Establishment<br>registration number | NA |
## 2 Device Information
| Type of 510(k)<br>submission: | Traditional |
|-------------------------------|-------------------------------------------------------------------------------------------|
| Trade Name: | Fingertip Pulse Oximeter (Model: A300, A310, M100,<br>M110,M120,M130,M150,M160,M170,M230) |
| Classification<br>name: | Oximeter |
| Review Panel: | Anesthesiology |
| Product Code: | DQA |
| Device Class: | II |
| Regulation | 870.2700 |
| Number: | K163135 |
{4}------------------------------------------------
#### 3 Predicate Device Information
Shenzhen Jumper Medical Equipment Co., Ltd. Sponsor: JPD-500A Fingertip Pulse Oximeter Device: K140582 510(K) Number:
#### 4 Device Description
The Fingertip Pulse Oximeteris intended for spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of adult and pediatric patients in the home and hospital.
The fingertip pulse oximeter features a small size, low power consumption, a convenient operation, and portability.It is only necessary for a patient to put one of his/her fingers into the fingertip clips for measurement.
Principle of the fingertip pulse oximeter as follows:
A mathematical formula is established making use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin (RHb) and Oxyhemoglobin (HbO2) in red and infrared zones.
Operation principle of the instrument:
Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm red and 905nm infrared light) can be focused onto a human nail tip through a clamping finger-type sensor. Relevant data is shown on the Oximeter's display through electronic circuits and a microprocessor.
#### 5 Intended Use
The subject device is intended for measuring the functional oxygen saturation and pulse rate (PR) through a patient's finger. It is applicable for spot-checking SpO2 and pulse rate (PR) of adult and pediatric patients in homes and clinics.
#### 6 Indications for Use
The Fingertip Pulse Oximeter is non-invasive device intended for spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The portable fingertip device is indicated for adult and pediatric patients in home and hospital environments (including clinical use in internist/ surgery. anesthesia, intensive care, etc).
{5}------------------------------------------------
#### 7 Contraindications
- High-frequency electrosurgical
- Placement of a sensor on an extremity with a blood pressure cuff arterial catheter, or intravascular line
- The patient has hypotension severe vasoconstriction severe anemia or hypothermia.
- The patient is in cardiac arrest or is in shock.
- Fingernail polish or false fingernails may cause inaccurate SpO2 readings.
{6}------------------------------------------------
## 8 SE Comparison
| Characteristics | Subject device | Predicate device<br>(k140582) | Remark | |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|------------------|
| Device name | Fingertip Pulse Oximeter<br>A300,A310,M100,M110,M<br>120,M130,M150,M160,M1<br>70,M230 | Fingertip Pulse Oximeter | / | |
| Model | Note:<br>A300 and M110 are totally<br>identical model except for<br>model name based on<br>customer's requirement;<br>A310 and M100 are totally<br>identical model except for<br>model name based on<br>customer's requirement; | JPD-500A | / | |
| Manufacturer | Shenzhen Fitfaith<br>Technology Co.,Ltd. | Shenzhen Jumper<br>Medical Equipment<br>Co.,Ltd | / | |
| Intended patient<br>population | Adult & Pediatric | Adult & Pediatric | same | |
| Intended application<br>site | Fingertip | | same | |
| Intended<br>Environments | home & hospital | | same | |
| Prescription & OTC | Prescription | | same | |
| Intended use | The subject device is intended for measuring the<br>functional oxygen saturation and pulse rate through<br>patient's finger. It is applicable for spot-checking SpO2<br>and pulse rate of adult and pediatric patients in homes<br>and clinics. | | same | |
| Indications for Use | The Fingertip Pulse<br>Oximeter is non-invasive<br>device intended for spot-<br>checking of functional<br>oxygen saturation of<br>arterial hemoglobin<br>(SpO2) and pulse rate<br>(PR). The portable<br>fingertip device is<br>indicated for adult and<br>pediatric patients in home<br>and hospital environments<br>(including clinical use in<br>internist/ surgery,<br>anesthesia, intensive<br>care, etc). | The JPD-500A Fingertip<br>Pulse Oximeter is non-<br>invasive device intended<br>for spot-checking of<br>functional oxygen<br>saturation of arterial<br>hemoglobin (Spo2) and<br>pulse rate. The portable<br>fingertip device is<br>indicated for adult and<br>pediatric patients in home<br>and hospital<br>environments (including<br>clinical use in internist/<br>surgery, anesthesia,<br>intensive care, etc). | Similar(Note01) | |
| Characteristics | | Subject device | Predicate device<br>(k140582) | Remark |
| Working Principle | | A mathematical formula is established making use of<br>Lambert Beer Law according to Spectrum Absorption<br>Characteristics of Reductive hemoglobin (RHb) and<br>Oxyhemoglobin (HbO2) in red and near-infrared<br>zones. Operation principle of the instrument:<br>Photoelectric Oxyhemoglobin Inspection Technology is<br>adopted in accordance with Capacity Pulse<br>Scanning and Recording Technology, so that two<br>beams of different wavelength of lights can be focused<br>onto a human nail tip through a clamping finger-type<br>sensor. A measured signal obtained by a<br>photosensitive element, will be shown on the<br>Oximeter's display through process in electronic<br>circuits and microprocessor. | | same |
| Light<br>Specification<br>on | wavelength | 660nm±3nm for red light,<br>905nm±5nm for infrared light (IR) | | same |
| | maximum<br>optical<br>power | 1.5 mW for red light (660nm)<br>1.2 mW for IR (905nm) | | same |
| Contact Material | | ABS for enclosure, silica gel for clip | | same |
| Internal Power supply | | 2*AAA 1.5v alkaline battery | | same |
| Working Current | | Less than 40mA(Normal) | 35mA | Similar(Note02) |
| Resolution | SpO2 | 1% | | same |
| | Pulse<br>rate | 1 bpm | | same |
| Measurement<br>Range | SpO2 | 0%~100% | | same |
| | Pulse<br>rate | 25~250 bpm | | same |
| Measurement<br>Accuracy | SpO2 | ±2%: (70%~100%)<br>Unspecified : (0%~69%) | ±2%: (70%~100%)<br>no definition(0%~69%) | same |
| | Pulse<br>rate | ±2 bpm | ±2 bpm | same |
| | | | | |
| Operating<br>Environment | Temperature | 5°C~40°C | | Similar (Note03) |
| | Humidity | 15%~85% non-<br>condensing | 15%~80% non-<br>condensing | |
| Storage &<br>Transport<br>Environment | Temperature | -20°C~+55°C | -10°C~+50°C | Similar(Note04) |
| | Humidity | 10%~95% | 10%~93% | |
### Table 1. Substantial Equivalence Comparison
{7}------------------------------------------------
Note01:
The trade name is different, but SE determination does not depend on trade name.
Note02:
The working currents of subject device and predicate device are similar. The subject device met the requirements of IEC 60601-1.
{8}------------------------------------------------
Note03:
The humidity under operating environment of subject device and predicate device are similar. The subject device met the requirements of ISO 80601-2-61. Note04:
The Storage & Transport environment of subject device and predicate device are similar. The subject device met the requirements of ISO 80601-2-61.
The subject device is the same as the predicate device in intended use, intended patient population, intended application site, working principle, contact material, internal power supply, display range, resolution. Only their appearance, indication for use (Note01), working current (Note02), operation environment (Note03), storage transport environment (Note04), are a little bit different. However, the minor differences do not raise different questions of safety and effectiveness.
#### 9 Brief discussions of tests
#### 9.1 Brief discussions of the nonclinical tests
The subject device conforms to the following standards:
- ゃ IEC 60601-1:2005+A1:2012: Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance;
- IEC 60601-1-2:2014 Medical Electrical Equipment Part 1-2: General ゃ Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests;
- や IEC 60601-1-11: Requirements for Medical Electrical (ME) Equipment and ME Systems used in the Home Healthcare Environment;
- ゃ ISO 10993-5:2009 Biological evaluation of medical devices – Part 5: Tests for In Vitro cytotoxicity;
- � ISO 10993-10:2010 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization;
- ゃ ISO 80601-2-61:2011 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment;
- ゃ Cleaning and low level disinfection validation.
#### 9.2 Brief discussions of clinical tests
Clinical testing according to ISO 14155:2011 and ISO 80601-2-61: 2011 has also been performed on the device.
Clinical hypoxia accuracy testing (controlled desaturation study) was conducted during induced hypoxia studies on healthy, non-smoking, light-to-dark-skinned subjects in an independent research laboratory. The measured arterial
{9}------------------------------------------------
hemoglobin saturation value (SpO2) of the subject device was compared with arterial hemoglobin oxygen (SaO2) value, determined from blood samples with a CO-oximeter (control device). The accuracy of the subject device is in comparison with the control device over the SpO2 range of 70~100%. Data was calculated and analyzed using the mean bias (B), root-mean-square (Ams), PRECISION (standard deviation of the residuals (Sres) for all subjects, per ISO 80601-2-61, the result showed that the error is far less than the scope specified in the ISO 80601-2-61;
Besides, the Agreement between methods of measurement with multiple observations for both all subjects pooled and individual test subjects were analyzed with the Bland and Altman statistics, the analysis demonstrated that the vast majority of data is within ±95% limit of agreement, the data points beyond or below this scope were regarded as outliers. The outliers only occurred occasionally and after being analyzed, it was determined that the outliers do not raise performance concerns regarding the accuracy and precision of the device.
During the clinical study, 12 subjects were enrolled, who are healthy, nonsmoking, competent adults, between twenty-one and thirty-two (21-32) years of age, and they were provided EC (Ethics Committee)-approved informed consent as documented on an informed consent form. No case was lost in this trial. The trial completed 12 cases, on which 294 data sets were collected, of which 288 were valid. 6 data sets which less than 70% were excluded. The result met the criteria specified in the ISO 80601-2-61; In addition, there were no reported adverse effects during these investigations.
#### 10 Other information (such as required by FDA guidance)
No other information.
#### 11 Conclusions
Based on the clinical and non-clinical testing performed, the results demonstrate that the subject device fingertip pulse oximeter is substantially equivalent to the predicate device JPD-500A Fingertip Pulse Oximeter manufactured by Shenzhen Jumper Medical Equipment Co., Ltd.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.