K203821 · Dong Nai Garment Corporation (Donagamex) · QPC · Dec 17, 2021 · General, Plastic Surgery
Device Facts
Record ID
K203821
Device Name
Donagamex Blue Performance Protective Gown
Applicant
Dong Nai Garment Corporation (Donagamex)
Product Code
QPC · General, Plastic Surgery
Decision Date
Dec 17, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
Donagamex protective gowns are non-sterile, single use surgical apparel intended to be worn by healthcare personnel to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter. The Donagamex protective gowns meet the requirements of AAMI Level 3 barrier protection for a protective gown per ANSI/AAMI PB70: 2012 Liquid Barrier Performance and Classification of protective apparel and drapes intended for use in healthcare facilities (AAMI PB70). The Donagamex protective gowns are sold non-sterile, single use. They are not intended for use in the operating room.
Device Story
Non-sterile, single-use protective gown; worn by healthcare personnel to provide AAMI Level 3 barrier protection against microorganisms, body fluids, and particulate matter. Constructed from non-woven polypropylene spunbond fabric with polyethylene laminate film. Features elastic cuffs and belt tie. Not intended for operating room use. Provides physical barrier to reduce cross-contamination between patient and healthcare worker.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing including AAMI PB70 liquid barrier performance, physical/mechanical strength (ASTM D5034, D5587, D751, D3776), flammability (16 CFR 1610), lint generation (ISO 9073-10), and biocompatibility (ISO 10993-5, ISO 10993-10).
Technological Characteristics
Non-woven polypropylene spunbond fabric with polyethylene laminate film. AAMI Level 3 barrier protection per ANSI/AAMI PB70:2012. Tested per ASTM D5034 (breaking strength), D5587 (tearing strength), D751 (seam strength), D3776 (mass per unit area). Biocompatibility per ISO 10993-5 and ISO 10993-10. Non-sterile, single-use, disposable. No natural rubber latex.
Indications for Use
Indicated for healthcare personnel to protect patients and workers from microorganisms, body fluids, and particulate matter. Not for use in the operating room.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 10, 2022
Dong Nai Garment Corporation (Donagamex) % Bari Steinberg Official Correspondent for Dong Nai Garment Corporation (Donagamex) mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, New York 11021
Re: K203821
Trade/Device Name: Donagamex Blue Performance Protective Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: OPC
Dear Bari Steinberg:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated December 17, 2021. Specifically, FDA is updating this SE Letter to inlude the finalized trade name of the device as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Dr. Clarance Murray, OHT4: Office of Surgical and Infection Control Devices, 301-796-0870, Clarence.Murray@fda.hhs.gov.
Sincerely,
Liqun Zhao -S
For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 17, 2021
Dong Nai Garment Corporation (Donagamex) Bari Steinberg Official Correspondent for Dong Nai Garment Corporation (Donagamex) mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, New York 11021
Re: K203821
Trade/Device Name: Donagamex Blue Performance Surgical Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: OPC Dated: August 29, 2021 Received: September 7, 2021
Dear Bari Steinberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ligun Zhao -S
For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203821
Device Name
Donagamex Blue Performance Protective Gown
#### Indications for Use (Describe)
Donagamex protective gowns are non-sterile, single use surgical apparel intended to be worn by healthcare persomel to help protect both the patient and the healthcare worker of microorganisms, body fluids, and particulate matter.
The Donagamex protective gowns meet the requirements of AAMI Level 3 barrier protection for a protective gown per ANSI/A AMI PB70: 2012 Liquid Barrier Performance and Classification of protective apparel and drapes intended for use in healthcare facilities (AAMI PB70).
The Donagamex protective gowns are sold non-sterile, single use. They are not intended for use in the operating room.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY
The assigned 510(k) number is: K203821
Date Summary Prepared: December 17, 2021
#### 1. Submitter's Identification:
#### Dong Nai Garment Corporation
Road 2, Bien Hoa Industrial Zone 1 An Binh Ward, Bien Hoa City Dong Nai Province, Vietnam Tel.: +84 0253 836147 Fax: +84 0253 836151
Official Correspondent: Bari Steinberg, mdi Consultants Inc. E-mail: bari@mdiconsultants.com
#### 2. Device Name: Donagamex Blue Performance Protective Gown
#### 3. Regulatory Information:
Regulation Number: 21 CFR 878.4040 Device Class: Class II Regulation Name: Surgical Apparel Product Code: QPC Common Name: Gown, Non-Sterile, Non-Isolation
#### 4. Predicate Device Information:
510K Number: K160337 Device Name: ValueCare Open Back Protective Gown
#### 5. Intended Use:
Donagamex protective gowns are non-sterile, single use surgical apparel intended to be worn by healthcare personnel to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter.
The Donagamex protective gowns meet the requirements of AAMI Level 3 barrier protection for a protective gown per ANSI/AAMI PB70: 2012 Liquid Barrier Performance and Classification of protective apparel and drapes intended for use in healthcare facilities (AAMI PB70).
The Donagamex protective gowns are sold non-sterile, single use. They are not intended for use in the operating room.
{5}------------------------------------------------
#### 6. Device Description:
The Donagamex Blue Performance Protective Gown is a Class II medical device under the FDA product code QPC, General & Plastic Surgery Panel, and Regulation 21 CFR 878.4040. The device description of the Donagamex Blue Performance Protective Gown is in accordance with the Guidance on Premarket Notification [510(k)] Submissions for Surgical Gowns and Surgical Drapes, issued on August 1, 1993. The Donagamex Blue Performance Protective Gowns are non-woven, blue gowns, available in various sizes and have no areas of reinforcement. They are made from a layer of polyethylene laminated over spun-bonded polypropylene and have been tested according to AAMI PB70:2012 and meet AAMI Level 3 barrier level protection for a protective gown. The Donagamex Blue Performance Protective Gown is a single use, disposable medical device that will be sold non-sterile.
#### 7. Device and Predicate Device Technical Characteristics:
| COMPARISON<br>CRITERIA | SUBJECT DEVICE<br>(K203821) | PREDICATE DEVICE<br>(K160337) | COMPARISON<br>RESULT |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Device Name | Donagamex Blue<br>Performance Protective<br>Gown<br>(AAMI Barrier Level 3) | ValueCare Open Back<br>Protective Gowns<br>(AAMI Barrier Level 3) | N/A |
| Manufacturer | Dong Nai Garment<br>Corporation (Donagamex) | ValueCare | N/A |
| Classification | Surgical Apparel<br>21 CFR Part 878.4040<br>Product Code: QPC<br>Class II | Surgical Apparel<br>21 CFR Part 878.4040<br>Product Code: QPC<br>Class II | Same |
| Indications for Use | Donagamex Blue<br>Performance Protective<br>Gowns are non-sterile,<br>single use surgical apparel<br>intended to be worn by<br>healthcare professionals to<br>help protect both the<br>patient and the health care<br>personnel from the<br>transfer of microorganisms,<br>body fluids, and particulate<br>matter.<br>The Donagamex Protective<br>gown meets the<br>requirements of AAMI<br>Level 3 barrier protection<br>for a protective gown per | These gowns are intended<br>to protect health care<br>patients and health care<br>personnel from the transfer<br>of micro- organisms, body<br>fluids and particulate<br>material. The back of the<br>gown is open and non-<br>protective. They are not<br>intended for use in the<br>operating room. | Similar |
| | | | |
| ANSI/AAMI PB70: 2012<br>Liquid Barrier Performance<br>and Classification of<br>protective apparel and<br>drapes intended for use in<br>healthcare facilities (AAMI<br>PB70).<br><br>The Donagamex protective<br>gowns are sold non-sterile,<br>single use. They are not<br>intended for use in the<br>operating room. | | | |
| Sterile or Non-Sterile | Non-Sterile | Non-sterile | Same |
| Reusable or<br>Disposable (Single<br>Use) | Single use | Single use | Same |
| Design | Elastic at the cuff for<br>keeping the sleeves in<br>place on the wearer,<br>belt tie | Design includes open back,<br>thumb loops, back<br>perforation for easy removal<br>and waist tie | Similar |
| Material<br>Composition | Nonwoven polypropylene<br>spunbond fabric<br>with polyethylene laminate<br>film (SF) | Made from extruded plastic<br>film | Similar |
| Size | Medium to XX-Large | Not reported in 510k<br>Summary | N/A |
| Color | Blue | Blue | Same |
| Sterilization<br>Method (if<br>applicable) | None, non-sterile gowns | None, non-sterile gowns | Similar |
| Packaging | Bulk for non-sterile gowns | Not reported in 510k<br>Summary | N/A |
| Natural rubber<br>Latex | Not made with natural<br>rubber latex | Not made with Natural<br>Rubber Latex | Same |
| <b>PERFORMANCE TEST RESULTS :</b> | | | |
| AATCC 42 Water<br>Resistance: Impact<br>Penetration (g) | Meets AAMI PB70 Barrier<br>Classification for Level 3 | Meets AAMI PB70 Barrier<br>Classification for Level 3 in<br>critical zones | Same |
| AATCC 127 Water<br>Resistance:<br>Hydrostatic<br>Pressure (cm) | Meets AAMI PB70 Barrier<br>Classification for Level 3 | Meets AAMI PB70 Barrier<br>Classification for Level 3 in<br>critical zones | Same |
| Liquid Barrier<br>Performance<br>Classification | Device was tested in<br>accordance with AAMI PB<br>70:2012 and meets AAMI<br>Level 3 barrier protection<br>requirements for a<br>protective gown. Testing<br>was performed using 3<br>nonconsecutive lots and 32<br>samples per lot in each<br>critical zone area. The<br>critical zone areas tested<br>were the chest, back and<br>sleeve seam. | Device meets the barrier<br>protection requirements of<br>AAMI Level 3 per<br>ANSI/AAMI PB70:2012, but<br>has an open back which is<br>non-protective. | Similar |
| ASTM - D3776<br>Standard Test<br>Methods for Mass<br>Per Unit Area<br>(Weight) of Fabric | Pass | Not reported in 510k<br>Summary | N/A |
| ASTM - D5034<br>Standard Test<br>Method for<br>Breaking Strength<br>and Elongation of<br>Textile Fabrics<br>(Grab Test) | Pass | Not reported in 510k<br>Summary | N/A |
| ASTM - D5587<br>Standard Test<br>Method for<br>Tearing Strength of<br>Fabrics by<br>Trapezoid<br>Procedure | Pass | Not reported in 510k<br>Summary | N/A |
| ASTM - D751<br>Standard Test<br>Methods for<br>Coated Fabrics<br>(Seam Strength) | Pass | Not reported in 510k<br>Summary | N/A |
| Flammability Test<br>Method Standard<br>for Flammability of<br>Clothing Textiles<br>(16 CFR Part 1610) | Meets Class I Flammability<br>per CPSC, Part 1610 | Meets Class I Flammability<br>per CPSC, Part 1610 | Same |
| Textiles ISO 9073-<br>10<br>Test Method for<br>Nonwovens Part<br>10: Lint & other | Pass | Not reported in 510k<br>Summary | N/A |
| Particles<br>Generation in the<br>Dry State | | | |
| ISO 10993 Part-1<br>Biological<br>Evaluation of<br>Medical Devices.<br>Biocompatibility<br>tests included<br>cytotoxicity,<br>sensitization and<br>irritation. | Non-cytotoxic, non-<br>sensitizing, non-irritating | Non-cytotoxic, non-<br>sensitizing, non-irritating | Same |
#### Table 1: Side by Side Comparison
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
Both gowns are disposable, single use, non-reinforced gowns intended to help protect from the transfer of body fluids and particulate matter in settings other than the operating room. They are similar in color, design and are both offered as non-sterile.
The primary difference is that the ValueCare Open Back Protective Gown provides liquid barrier protection in front of gown only (critical zones), while the Donagamex Blue Performance Protective Gowns provide AAMI Level 3 in front and back of gown.
The above differences in design and AAMI liquid protection level raise no new issues of safety and effectiveness since both gowns are designed, tested, and labeled in compliance with the applicable AAMI PB70: 2012 liquid barrier requirements.
#### 8. Summary of Clinical Testing:
Clinical performance for non-sterile, disposable protective gowns is not applicable for this product. Numerous predicated devices (gowns) exist and have been used extensively for a number of years that are made from the same or similar materials.
#### 9. Summary of Non-Clinical Performance Testing:
The information in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to devices already in commercial distribution. Equivalence is demonstrated through intended use, materials, design test methods including physical, mechanical, liquid barrier and biocompatibility testing. The Donagamex Blue Performance Protective Gown was found to be acceptable for its intended use in each of these applicable industry recognized standards:
AATCC 42:2017 Standard Test Method for Water Resistance: Impact Penetration
AATCC 127:2018 Standard Test Method for Water Resistance: Hydrostatic Pressure
ASTM - D5034:2017 Standard Test Method for Breaking Strength and Elongation of Textile Fabrics (Grab Test)
ASTM - D5587:2019 Standard Test Method for Tearing Strength of Fabrics by Trapezoid Procedure
{9}------------------------------------------------
ASTM - D751:2019 Standard Test Methods for Coated Fabrics
ASTM - D3776:2020 Standard Test Methods for Mass Per Unit Area (Weight) of Fabric
16 CFR Part 1610 - Flammability Test Method Standard for Flammability of Clothing Textiles
ISO 9073-10 Textiles - Test Method for Nonwovens Part 10: Lint & other Particles Generation in the Dry State
AMMI /ANSI / ISO 10993-5:2009 – In Vitro Cytotoxicity using Agar Overlay test (USP, ISO)
AAMI / ANSI / ISO 10993-10:2010 – In Vitro Irritation reactivity using Primary Skin Irritation Test
AAMI / ANSI / ISO 10993-10:2010 – In Vitro Sensitization using the Buehler Method Test
#### Table 2: Non-Clinical Tests
| Test Method | Purpose | Acceptance Criteria | Results |
|---------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|------------------------------|---------|
| AATCC 42:2017<br>Standard Test Method<br>for Water Resistance:<br>Impact Penetration | Determine resistance of<br>gown to the<br>penetration of water by<br>impact | Level 3: ≤ 1.0 gm | Pass |
| AATCC 127:2018<br>Standard Test Method<br>for Water Resistance:<br>Hydrostatic Pressure | Determine resistance of<br>gown to the<br>penetration of water<br>under hydrostatic<br>pressure | Level 3: ≥ 50 cm | Pass |
| ASTM - D5034:2017<br>Standard Test Method<br>for Breaking Strength<br>and Elongation of<br>Textile Fabrics (Grab<br>Test) | Determine the breaking<br>strength of gown | More or equals to<br>7 lbs | Pass |
| ASTM - D5587:2019<br>Standard Test Method<br>for Tearing Strength of<br>Fabrics by Trapezoid<br>Procedure | Determine the tearing<br>strength of gown | More or equals to<br>2.3 lbs | Pass |
| ASTM - D751:2019<br>Standard Test Methods<br>for Coated Fabrics | Determine if any failure<br>occurs in gown seams | More or equals to<br>7 lbs | Pass |
| ASTM - D3776:2020<br>Standard Test Methods<br>for Mass Per Unit Area<br>(Weight) of Fabric | Determine the fabric<br>weight | N/A | N/A |
{10}------------------------------------------------
| Flammability Test<br>Method Standard for<br>Flammability of<br>Clothing Textiles (16<br>CFR Part 1610) | To test the clothing<br>textile flammability | Class I Normal<br>Flammability Result | Pass |
|-----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|---------------------------------------|------|
| ISO 9073-10<br>Textiles - Test Method<br>for Nonwovens Part 10:<br>Lint & other Particles<br>Generation in the Dry<br>State | Characterize the size<br>and quantity of the lint<br>generation | N/A | N/A |
### Table 3: Biocompatibility Tests
| Test Method | Purpose | Acceptance Criteria | Results |
|----------------------------------------------|--------------------------------------------------|----------------------------------------|---------------------------------------------------|
| In vitro Cytotoxicity<br>ISO 10993-5:2009 | To determine if device<br>extract is cytotoxic | The device must be non-<br>cytotoxic | Under the study<br>conditions non-<br>cytotoxic |
| Primary Skin Irritation<br>ISO 10993-10:2010 | To determine if device is<br>a skin irritant | The device must be a<br>non-irritant | Under the study<br>conditions not an<br>irritant |
| Dermal Sensitization<br>ISO 10993-10:2010 | To determine if device is<br>a dermal sensitizer | The device must be a<br>non-sensitizer | Under the study<br>conditions not a<br>sensitizer |
#### 10. Conclusions:
Based on the non-clinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, K160337.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.