The Multi-unit Abutment, Multi-unit Angled Abutment is intended to be surgically placed in the maxillary or mandibular arches for the purpose providing prosthetic support for dental restorations (Crown, bridges, and overdentures) in partially or fully edentulous individuals.
Device Story
Dental implant abutment system providing prosthetic support for restorations; includes straight and angled abutments, screws, healing caps, temporary cylinders, and CCM cylinders. Components connect to endosseous dental implants (Xpeed AnyRidge, AnyOne, AnyRidge Octa 1) to restore chewing function. Operated by dental professionals in clinical settings. Abutments and screws are fabricated from Ti-6Al-4V ELI; CCM cylinders use Co-Cr-Mo alloy with POM sleeves. Device components are non-sterile and require user-performed steam sterilization. Output is a stable platform for screw-retained dental prostheses. Benefits include restoration of chewing function and support for various prosthetic designs.
Clinical Evidence
No clinical data submitted. Substantial equivalence is supported by bench testing, including fatigue testing per ISO 14801 and biocompatibility evaluation per ISO 10993-1.
Technological Characteristics
Materials: Ti-6Al-4V ELI (ASTM F136-13), Co-Cr-Mo alloy, POM. Principle: Mechanical connection to endosseous implants via internal hex, octa, or conical interfaces. Form factor: Straight and angled abutments, various screws, and cylinders. Connectivity: None. Sterilization: Non-sterile, validated for user steam sterilization (ISO 17665-1, 2).
Indications for Use
Indicated for partially or fully edentulous individuals requiring prosthetic support for dental restorations (crowns, bridges, overdentures) in the maxillary or mandibular arches.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 20, 2021
MegaGen Implant Co. Ltd. % You Kim Chief Researcher DaeGyeong Regulatory Affairs Institute 32, Innovalley-ro Daegu, Dong-gu 41065 REPUBLIC OF KOREA
Re: K203808
Trade/Device Name: Multi-unit Abutment, Multi-unit Angled Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: April 9, 2021 Received: September 22, 2021
Dear You Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K203808
Device Name
Multi-unit Abutment, Multi-unit Angled Abutment
Indications for Use (Describe)
The Multi-unit Abutment, Multi-unit Angled Abutment is intended to the maxillary or mandbular arches for the purpose providing prosthetic support for dental restorations (Crown, bridges, and overdentures) in partially or fully edentulous individuals.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
==============================================================================================================================================================================
{3}------------------------------------------------
# 510(k) Summary for K203808
## Date: October 19, 2021
## 1. Applicant / Submitter
MegaGen Implant Co., Ltd. 45, Secheon-ro, 7-gil, Dasa-eup, Dalesong-gun, Daegu, Republic of Korea Tel: + 82-53-222-2828
## 2. Submission Correspondent
You Jung Kim DaeGyeong Regulatory affairs Institute 32, Innovalley-ro, Dong-gu, Daegu, 41065, Republic of Korea Tel: +82-53-247-2258 Fax: +82-53-247-2254 Email: ra3@dgri.co.kr
## 3. Device
- Multi-unit Abutment, Multi-unit Angled Abutment Trade Name:
- Endosseous Dental Implant Abutment Common Name:
- . Classification Name: Endosseous dental implant abutment
- Classification Product Code: NHA
- . Classification regulation: Class II, 21 CFR 872.3630
## 4. Predicate Device
- . Primary Predicate Device: K182448 – AnyRidge Octa 1 Implant System
#### . Reference Devices:
- K123988 AnyOne Internal Implant System K192401 - Straumann® Screw-Retained Abutments K201621 – Magicore II System K141457 – Dentium Implantium® and SuperLine® Abutments K110955 - AnyRidge Internal Implant System K171142 - Healing Cap Multi-Unit Titanium K052369 - ExFeel Dental Implant System
{4}------------------------------------------------
## 5. Description
The Multi-unit Abutment, Multi-unit Angled Abutment of a dental implant system to provide support for prosthetic restorations. It is attached to the implants and intended to be used in the upper or lower jaw and used for supporting tooth replacements to restore patient's chewing function. The Multi-unit Abutment, Multi-unit Angled Abutment consists of straight & angled abutments and prosthetic components in varying sizes for single & multiple unit screw retained restorations.
The Multi-unit Abutment, Multi-unit Angled Abutment is compatible with following MegaGen Implants cleared under:
| Manufacturer | Compatible Implant<br>System | Device Name | 510(k)<br>Number | Connection | Diameter (mm) |
|---------------------------------|----------------------------------------------|---------------------------------------|--------------------|------------------|--------------------------------------------------------|
| MegaGen<br>Implant Co.,<br>Ltd. | Xpeed AnyRidge<br>Internal Implant<br>System | Xpeed<br>AnyRidge<br>Internal Fixture | K123870<br>K140091 | Internal<br>Hex | 4.0, 4.4, 4.9, 5.4,<br>5.9, 6.4, 6.9, 7.4,<br>7.9, 8.4 |
| | AnyOne<br>Internal Implant<br>System | AnyOne<br>Internal Fixture | K123988 | Internal<br>Hex | 3.9, 4.3, 4.8, 5.3,<br>5.8, 6.3, 6.8, 7.3,<br>7.8, 8.3 |
| | AnyRidge Octa 1<br>Implant System | AnyRidge Octa<br>1 Fixture | K182448 | Internal<br>Octa | 3.6, 3.7, 4.0, 4.1,<br>4.4, 4.8, 5.0, 5.5 |
The Multi-unit Abutment, Multi-unit Angled Abutment is consisted of the following devices.
| | Device | | Content |
|-------------------|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. | Abutment | Multi-unit Abutment | Description The Multi-unit Abutment is used for fabricating screw-retained prosthesis. There are two types of Multi-unit Abutments, N type and S type, depending on the connection type and prosthetics compatibility. The N type (one-piece type) is screwed directly in to the endosseous dental implant by their lower threaded part, and the S type (two-piece type) is connected the fixture with Multi-unit Abutment Screw. |
| | | Material | Ti-6Al-4V ELI (ASTM F136-13) |
| | | Dimension (Diameter & Total Length) | $\varnothing$ 4.8 x 8.05, 8.75, 8.94, 9.10, 9.75, 9.80, 9.84, 10.75, 10.80, 10.84, 11.75, 11.80, 11.84, 12.80, 12.84mm<br>$\varnothing$ 5.0 x 6.2, 7.2, 8.2, 9.2, 10.2 mm |
| | | Post Heights | 1.8, 2.2 mm |
| | | Gingival Heights | 0.6, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0, 5.5 mm |
| | | Angulation | Straight |
| | | Compatible Implant System | Xpeed AnyRidge Internal Implant System<br>AnyOne Internal Implant System<br>AnyRidge Octa 1 Implant System |
| | | Multi-unit Angled Abutment | Description The Multi-unit Angled Abutment is used for fabricating screw-retained prosthesis and correcting the prosthetic angulation of implant. There are two types of Multi-unit Angled Abutments, N type and S type, depending on the prosthetics compatibility. The N type (two-piece type) is connected the fixture with Multi-unit Abutment Screw, and the S type (two-piece type) is connected the fixture with Abutment Screw or Multi Post Screw. |
| | | Material | Ti-6Al-4V ELI (ASTM F136-13) |
| | | Dimension (Diameter & Length) | $\varnothing$ 4.8 x 3.4, 3.9, 3.94, 4.35, 4.4, 4.9, 4.94, 5.35, 5.4, 5.85, 5.9, 5.94, 6.35, 6.4, 6.85, 6.9, 6.94, 7.35, 7.85, 8.85 mm<br>$\varnothing$ 5.0 x 4.5, 4.7, 5.5, 5.6, 5.7, 5.8, 6.5, 6.6, 6.7, 6.8, 7.5, 7.6, 7.7, |
| | | | 7.8, 8.6, 8.8 mm |
| | | Post Heights | 2.2, 3.9 mm |
| | | Gingival Heights | 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, mm |
| | | Angulation | 17°, 29°, 30° |
| | | Compatible<br>Implant System | Xpeed AnyRidge Internal Implant System<br>AnyOne Internal Implant System<br>AnyRidge Octa 1 Implant System |
| | Multi-unit<br>Abutment<br>Screw | Description | The Multi-unit Abutment Screw is used for connecting Multi-unit Abutment or Multi-unit Angled Abutment to the Fixture. |
| | | Used with | Multi-unit Abutment (S type) in Xpeed AnyRidge & AnyOne<br>Multi-unit Angled Abutment (N type) in Xpeed AnyRidge & AnyOne<br>& AnyRidge Octa 1 |
| | | Material | Ti-6Al-4V ELI (ASTM F136-13) |
| | | Dimension<br>(Diameter &<br>Total Length) | Ø 2.1 x 7.0 mm<br>Ø 2.2 x 4.4 mm<br>Ø 2.4 x 6.8 mm<br>Ø 2.95 x 10.8, 11.5, 11.8, 12.5, 12.8, 13.5, 13.8, 14.5, 15.5 mm |
| | | Compatible<br>Implant System | Xpeed AnyRidge Internal Implant System<br>AnyOne Internal Implant System<br>AnyRidge Octa 1 Implant System |
| | Abutment<br>Screw | Description | The Abutment Screw is used for connecting Multi-unit Angled Abutment to the Fixture. |
| | | Used with | Multi-unit Angled Abutment (S type) in AnyOne |
| | | Material | Ti-6Al-4V ELI (ASTM F136-13) |
| | | Dimension<br>(Diameter &<br>Total Length) | Ø 2.3 x 7.7 mm |
| | | Compatible<br>Implant System | AnyOne Internal Implant System |
| | Multi Post<br>Screw | Description | The Multi Post Screw is used for connecting Multi-unit Angled Abutment to the Fixture. |
| | | Used with | Multi-unit Angled Abutment (S type) in Xpeed AnyRidge |
| | | Material | Ti-6Al-4V ELI (ASTM F136-13) |
| | | Dimension<br>(Diameter &<br>Total Length) | Ø 2.1 x 7.0 mm |
| | | Compatible<br>Implant System | Xpeed AnyRidge Internal Implant System |
| 2. Components | Healing Cap | Description | The Healing Cap is used for protecting Multi-unit (Angled) Abutment and minimizing irritation to tongue and oral mucosa during period of gingival healing.<br>It is connected to the Abutment using Cylinder Screw. |
| | | Material | Ti-6Al-4V ELI (ASTM F136-13) |
| | | Dimension<br>(Diameter & Length) | Ø 4.9 x 4.20 mm<br>Ø 6.8 x 4.20 mm |
| | | Compatible<br>Implant System | Xpeed AnyRidge Internal Implant System<br>AnyOne Internal Implant System<br>AnyRidge Octa 1 Implant System |
| | Temporary<br>Cylinder | Description | The Temporary Cylinder is used in conjunction with Multi-unit (Angled) Abutment to provide support for provisional restoration. It is connected to the Abutment using Cylinder Screw. |
| | | Material | Ti-6Al-4V ELI (ASTM F136-13) |
| | | Dimension<br>(Diameter & Length) | Ø 4.8 x 12.0 mm |
| | Cuff Heights | 3.0 mm | |
| | Angulation | Straight | |
| | Compatible<br>Implant System | Xpeed AnyRidge Internal Implant System<br>AnyOne Internal Implant System<br>AnyRidge Octa 1 Implant System | |
| CCM<br>Cylinder | Description | The CCM Cylinder is used in conjunction with Multi-unit<br>(Angled) Abutment to provide support for screw type final<br>prosthesis by casting with non-precious metal alloy (Co-Cr-Mo<br>Alloy). It is connected to the Abutment using Cylinder Screw. | |
| | Material | Body: Co-Cr-Mo Alloy / Sleeve: POM | |
| | Dimension<br>(Diameter & Length) | $Ø$ 4.8 x 13.0, 15.0 mm | |
| | Post Height | 10.0, 13.0 mm | |
| | Cuff Heights | 2.0, 3.0 mm | |
| | Angulation | Straight | |
| | Compatible<br>Implant System | Xpeed AnyRidge Internal Implant System<br>AnyOne Internal Implant System<br>AnyRidge Octa 1 Implant System | |
| Cylinder<br>Screw | Description | The Cylinder Screw is used for connecting Multi-unit (Angled)<br>Abutment to Healing Cap, Temporary cylinder or CCM Cylinder. | |
| | Material | Ti-6Al-4V ELI (ASTM F136-13) | |
| | Dimension<br>(Diameter & Length) | $Ø$ 2.0 x 3.4 mm | |
| | Compatible<br>Implant System | Xpeed AnyRidge Internal Implant System<br>AnyOne Internal Implant System<br>AnyRidge Octa 1 Implant System | |
{5}------------------------------------------------
{6}------------------------------------------------
[Note 1] Some of devices described in this submission, Multi-unit Abutment and Multi-unit Abutment screw in AnyOne Internal Implant System had been FDA cleared with K123988, but it is being submitted to change their identifier without any modifications, and to add a new model.
[Note 2] Some of devices described in this submission, Multi-unit Abutment, Multi-unit Angled Abutment and Multi-unit Abutment screw in AnyRidge Octa 1 Implant System had been FDA cleared with K182448, but it is being submitted to change their identifier with modification of surface treatment (Machined →Anodizing), and to add a new model. The changes are explained not affecting substantial equivalence in the part of 'VII. Substantial Equivalence Comparison'.
{7}------------------------------------------------
## 6. Indication for use
The Multi-unit Abutment, Multi-unit Angled Abutment is intended to be surgically placed in the maxillary or mandibular arches for the purpose providing prosthetic support for dental restorations (Crown, bridges, and overdentures) in partially or fully edentulous individuals.
## 7. Basis for Substantial Equivalence
The Multi-unit Abutment, Multi-unit Angled Abutment is substantially equivalent to the predicate devices in terms of indication for use, technical characteristic and function. They are made of the same material and have similar design. The size range of the subject device slightly differ from the predicate device however it is very minor not affecting substantial equivalence.
The subject device is composed of the abutments and prosthetic components that are used for single & multiple unit screw retained restorations, while the predicate devices is addressed complete dental implant system including various fixtures and abutments.
For this reason, the indication for use statement appear to be different between subject and predicate devices but proposed indication is available to apply to predicate devices as well, since the submission device is a sub-set of devices included in the predicate devices.
Also, the point of those indications is ultimately identical in the way that all of devices in subject & predicate devices are intended to be provided the prosthetic support for dental restorations such as crown, bridges, and overdenture to restore patient's chewing function; and this is minor difference in wording not affecting the substantial equivalence of the subject device.
Therefore, the indication for use of subject device is substantially equivalent to the predicate devices.
The Indications for the subject device are identical except for the portion of K182448's indications that are specific to dental implant bodies, and this submission does not include any dental implant bodies, so the omission of this statement is acceptable.
In order to demonstrate the difference in design does not raise any new issues, the performance test on the subject and predicate device have been performed in consideration of the worst case according to 'ISO 14801' and 'Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutment' to figure out the physical property. The test result supports the substantial equivalence to the predicate device.
Based on the comparison charts below and test results provided in this submission, we conclude that the subject device is substantially equivalent to the predicate devices.
{8}------------------------------------------------
## Multi-unit Abutment
| | Subject Device | Predicate Device | Reference Device 1 | Reference Device 2 | |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) No. | K203808 | K182448 | K123988 | K192401 | |
| Device Name | Multi-unit Abutment<br>For Multi-unit Abutment,<br>Multi-unit Angled<br>Abutment | Multi-unit Abutment<br>For AnyRidge Octa 1 Implant<br>System | Multi-unit Abutment<br>For AnyOne Internal Implant<br>System | Straumann® Screw-<br>Retained Abutments | |
| Manufacturer | MegaGen Implant Co.,<br>Ltd. | MegaGen Implant Co.,<br>Ltd. | MegaGen Implant Co.,<br>Ltd. | Straumann USA, LLC | |
| Indications<br>for Use<br>Statement | The Multi-unit Abutment,<br>Multi-unit Angled<br>Abutment is intended to<br>be surgically placed in the<br>maxillary or mandibular<br>arches for the purpose<br>providing prosthetic<br>support for dental<br>restorations (Crown,<br>bridges, and overdentures)<br>in partially or fully<br>edentulous individuals. | The AnyRidge Octa 1<br>Implant System is intended<br>to be surgically placed in<br>the maxillary or<br>mandibular arches for the<br>purpose of providing<br>prosthetic support for<br>dental restorations (Crown,<br>bridges, and overdentures)<br>in partially or fully<br>edentulous individuals. It is<br>used to restore a patient's<br>chewing function in the<br>following situations and<br>with the clinical protocols:<br>-Delayed loading.<br>-Immediate loading when<br>good primary stability is<br>achieved and with<br>appropriate occlusal<br>loading. Larger implants<br>are dedicated for the molar<br>region. | The AnyOne Internal<br>Implant System is intended<br>to be surgically placed in<br>the maxillary or<br>mandibular molar areas for<br>the purpose providing<br>prosthetic support for<br>dental restorations (Crown,<br>bridges, and overdentures)<br>in partially or fully<br>edentulous individuals. It is<br>used to restore a patient's<br>chewing function. Smaller<br>implants (less than 6.0<br>mm) are dedicated for<br>immediate loading when<br>good primary stability is<br>achieved and with<br>appropriate occlusal<br>loading. Larger implants<br>are dedicated for the molar<br>region and are indicated<br>for delayed loading. | Prosthetic components<br>directly or indirectly<br>connected to the<br>endosseous dental implant<br>are intended for use as an<br>aid in prosthetic<br>rehabilitations. Temporary<br>components can be used<br>prior to the insertion of the<br>final components to<br>maintain, stabilize and<br>shape the soft tissue<br>during the healing phase;<br>they are to be placed out<br>of occlusion. Final<br>abutments may be placed<br>into occlusion for implants<br>with sufficient primary<br>stability or for implants<br>that are fully<br>osseointegrated.<br>Temporary Abutments<br>have a maximum duration<br>of usage of 180 days. | |
| Design | Image: Multi-unit Abutment | Image: Multi-unit Abutment | Image: Multi-unit Abutment | Image: Screw-Retained Abutments | |
| Diameter (Ø) | 4.8, 5.0 mm | 4.8 mm | 5.0 mm | 3.5, 4.6 mm | |
| Total Length | 6.2, 7.2, 8.05, 8.2,<br>8.75, 8.94, 9.1, 9.2,<br>9.75, 9.8, 9.84, 10.2,<br>10.75, 10.8, 10.84,<br>11.75, 11.8, 11.84, 12.8,<br>12.84 mm | 9.8, 10.8, 11.8, 12.8<br>mm | 6.2, 7.2, 8.2, 9.2, 10.2<br>mm | Not Known | |
| Post Height | 1.8, 2.2 mm | 2.2 mm | 1.8 mm | Not Known | |
| Gingival<br>Height | 0.6, 1.5, 2.0, 2.5, 3.0,<br>3.5, 4.0, 4.5, 5.0, 5.5<br>mm | 1.3, 2.3, 3.3, 4.3 mm | 1.5, 2.5, 3.5, 4.5, 5.5<br>mm…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.