K203804 · Global Med Systems SL · GEX · Feb 23, 2021 · General, Plastic Surgery
Device Facts
Record ID
K203804
Device Name
Global Med Systems Milesman Compact Laser
Applicant
Global Med Systems SL
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 23, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Global Med Systems Milesman Compact Laser is intended for use in dermatological and general surgical procedures including hair removal and permanent hair reduction. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen. It is intended for use on all skin types (Fitzpatrick skin types 1-VI), including tanned skin.
Device Story
Device is a diode laser system (810 nm) for hair removal and permanent hair reduction. System comprises a main console, handpiece, umbilical, and footswitch. Operator uses a touch screen on the console to configure settings; laser firing is controlled via footswitch or handpiece trigger. Laser energy is absorbed by melanin in hair follicles, converting to heat to damage follicles and inhibit growth. Handpiece face features active cooling via circulating coolant fluid. Used in clinical settings by trained professionals. Output is pulsed laser energy; healthcare provider uses device to perform hair reduction procedures, potentially benefiting patients through long-term hair regrowth reduction.
Clinical Evidence
No clinical data. Bench testing only. Evidence includes electrical safety (IEC 60601-1, IEC 60601-2-22), EMC (IEC 60601-1-2), biocompatibility (cytotoxicity, dermal irritation, skin sensitization), and software verification/validation.
Technological Characteristics
810 nm diode laser; 10x10 mm spot size; 1-10 Hz repetition rate; 12-30 J/cm² fluence; 5-150 ms pulse duration; 120 W max average power; 2000 W peak power. Features active contact cooling. Non-sterile, cleanable. Touch screen interface. Software-controlled.
Indications for Use
Indicated for hair removal and permanent hair reduction in patients of all skin types (Fitzpatrick I-VI), including tanned skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 23, 2021
Global Med Systems SL % Kevin Morningstar Senior Consultant Morningstar Consulting Group 20319 E Costilla Ave Centennial, Colorado 80016
Re: K203804
Trade/Device Name: Global Med Systems Milesman Compact Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: December 21, 2020 Received: December 28, 2020
Dear Kevin Morningstar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
K203804
Device Name
Global Med Systems Milesman Compact Laser
Indications for Use (Describe)
The Global Med Systems Milesman Compact Laser is intended for use in dermatological and general surgical procedures including hair removal and permanent hair reduction. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen. It is intended for use on all skin types (Fitzpatrick skin types 1-VI), including tanned skin.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Milesman Compact logo. The logo consists of a blue square with a white "M" inside, followed by the word "milesman" in blue. Below the word "milesman" is a green rectangle with the word "compact" in white.
# Compact 510(k) Summary
I. SUBMITTER: Fernando Diéz, CEO Global Med Systems S.L. AV / Padre Isla 127. Bajo, 24008 Leon Spain Phone: +34 619 118 489 Fax: N/A fdiez@milesman.com
Contact Person: Kevin Morningstar, Regulatory Consultant (510k Preparer and FDA correspondent) Morningstar Consulting Group LLC 20319 E Costilla Ave Centennial, CO 80016 (720) 940-8271 Fax: N/A kevin.morningstar@gmail.com
Date Prepared: February 14, 2021
II. DEVICE
Name of Device: Global Med Systems Milesman Compact Laser Common or Usual Name: Powered Laser Surgical Instrument Classification Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Requlation Number: 21 CFR 878.4810 Regulatory Class: Class II Product Code: GEX
III. PREDICATE DEVICE
K073300 Milesman Premium Diode Laser System by Milesman S.L.
This predicate has not been subject to a design-related recall.
Reference Device: K140009 Alma Soprano™ Laser by Alma Lasers Ltd.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Milesman Compact. The logo consists of a blue square with a white "M" inside, followed by the word "milesman" in blue. Below the word "milesman" is a green rectangle with the word "compact" in white.
## IV. DEVICE DESCRIPTION
The Global Med Systems Milesman Compact Laser delivers pulsed laser energy at 810 nm to the surface of the skin using a handpiece which contains the laser diodes. The Milesman Compact consists of the following components:
- 1. The main console unit
- 2. Handpiece
- 3. Umbilical (between main console and handpiece)
- 4. Footswitch
The system is operated using the touch screen on top of the main console unit. Firing of the laser is controlled with either the footswitch or the trigger on the handpiece.
During hair removal, the laser emits a wavelength of light that is absorbed by the pigment (melanin) in the hair. The light energy is converted to heat, which damages the tube-shaped sacs within the skin (hair follicles) that produce the hair. This damage inhibits or delays future hair growth.
The face of the hand piece which comes into contact with skin is actively cooled with a circulating coolant fluid.
## V. INDICATIONS FOR USE
The Global Med Systems Milesman Compact Laser is intended for use in dermatological and general surgical procedures including hair removal and permanent hair reduction. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen. It is intended for use on all skin types (Fitzpatrick skin types 1-VI), including tanned skin.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Milesman Compact. The logo consists of a blue square with a white "M" inside, followed by the word "milesman" in blue. Below the word "milesman" is the word "compact" in white on a green background.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE AND REFERENCE DEVICES
| | Alma Soprano* | Milesman Premium | Milesman Compact | Comparison Remarks |
|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | (Reference) - K140009 | (Predicate) - K073300 | (Subject) - K203804 | (subject to predicate/reference) |
| Trade/Proprietary<br>Name | The Modified Alma Lasers XL™<br>Family of Multi-Application and<br>Multi- Technology Platforms<br>[SopranoXL SopranoXLi and<br>SopranoICE] | Milesman Premium Pulsed Diode<br>Array Laser System | Milesman Compact Diode<br>Laser System | Similar (predicate and subject device are<br>made by the same company; Compact is a<br>smaller lighter version of the Premium,<br>made possible by using the latest available<br>laser diode technology) |
| Generic/Common<br>Name | Hair removal laser | Hair removal laser | Hair removal laser | Identical |
| Classification<br>Name | Laser Instrument, Surgical,<br>Powered | Laser Instrument, Surgical,<br>Powered | Laser Instrument, Surgical,<br>Powered | Identical |
| Regulation Name | Laser surgical instrument for use<br>in general, and plastic surgery<br>and in dermatology | Laser surgical instrument for use<br>in general, and plastic surgery<br>and in dermatology | Laser surgical instrument for<br>use in general, and plastic<br>surgery and in dermatology | Identical |
| Regulatory Class | Class II | Class II | Class II | Identical |
| Product Code | GEX | GEX | GEX | Identical |
| Regulation Number | 21CFR 878.4810 | 21CFR 878.4810 | 21CFR 878.4810 | Identical |
| Review Panel | General and Plastic Surgery | General and Plastic Surgery | General and Plastic Surgery | Identical |
| Indications for use | The Hair Removal (HR) and<br>Super Hair Removal (SHR)<br>Mode is intended for permanent<br>reduction in hair regrowth,<br>defined as a long term, stable<br>reduction in the number of hairs<br>regrowing when measured at 6,<br>9 and 12 months after the<br>completion of a treatment<br>regimen.<br>The treatment of benign vascular<br>lesions (The laser blanch (LB)<br>mode).<br>Use on all skin types (Fitzpatrick<br>I-VI) including tanned skin (HR,<br>SHR and LB modes) | The Milesman Premium Pulsed<br>Diode Array Laser System is<br>intended for use in general and<br>plastic surgery and dermatology<br>procedures for the treatment of<br>vascular lesions, such as<br>angiomas, hemangiomas,<br>telangiectasia and other benign<br>vascular lesions and treatment<br>for pseudofolliculitis barbae. The<br>Milesman Premium is also used<br>for removal of unwanted hair,<br>permanent hair reduction and the<br>treatment of benign pigmented<br>lesions and leg veins in all skin<br>types (Fitzpatrick I-VI), including<br>tanned skin. | The Global Med Systems<br>Milesman Compact Laser is<br>intended for use in<br>dermatological and general<br>surgical procedures including<br>hair removal and permanent<br>hair reduction. Permanent hair<br>reduction is defined as the<br>long-term, stable reduction in<br>the number of hairs regrowing<br>when measured at 6, 9, and 12<br>months after the completion of<br>a treatment regimen. It is<br>intended for use on all skin<br>types (Fitzpatrick skin types 1-<br>VI), including tanned skin. | Similar but narrower in scope (subject<br>device indication does not mention vascular<br>lesions, benign pigmented lesions or leg<br>veins as the predicate does; subject device<br>indication includes the same "permanent"<br>definition that the reference device does<br>whereas predicate does not; predicate<br>mentions specific medical conditions<br>whereas subject and reference devices do<br>not) |
| Light/Laser Source | Diode | Diode | Diode | Identical |
| | Alma Soprano* | Milesman Premium | Milesman Compact | Comparison Remarks |
| | (Reference) - K140009 | (Predicate) - K073300 | (Subject) - K203804 | (subject to predicate/reference) |
| Wavelength | 810 nm | 810 nm (nominal) | 810 nm | Identical |
| Repetition rate | .5-3 Hz ("HR" or "hair removal"<br>mode)<br>5-10 Hz ("SHR" or "super hair<br>removal" mode) | 1-3 Hz | 1-10 Hz | Similar (subject device is more comparable<br>to the reference device than the predicate<br>when combining predicate's "HR" mode and<br>"SHR" mode) |
| Fluence -<br>Energy Density | 2-120 J/cm² ("HR" mode)<br>2-20 J/cm² ("SHR" mode) | 10 - 100 J/cm² | 12-30 J/cm² | Similar (subject device has a narrower range<br>of values than either the predicate or<br>reference (in "HR" mode) device) |
| Maximum fluence<br>utilized per<br>treatment protocols | 40 J/cm² | 40 J/cm² | 30 J/cm² | Similar (subject device has a lower<br>maximum value than either the predicate or<br>reference device) |
| Pulse Duration | 3.3-200 ms | 5 - 400 ms | 5 to 150 ms | Similar (Reference device provides the<br>greatest range of values; subject device<br>provides the narrowest range) |
| Spot Size | 12 x 10 mm | 10 x 10 mm | 10 x 10 mm | Identical (subject device compared to<br>predicate; reference device has larger spot<br>size) |
| Maximum average<br>power | 100 W | 120 W | 120 W | Similar (subject device and predicate are<br>identical but both are 20% greater than the<br>reference device) |
| Peak Power | Unknown | 2000 W | 2000 W | Identical (subject device compared to<br>predicate) |
| Tissue Cooling | Contact | Contact | Contact | Identical |
| How Supplied | Non-sterile - cleanable | Non-sterile - cleanable | Non-sterile - cleanable | Identical |
| Exposure Indicator | Audible & visual indicator | Audible & visual indicator | Audible & visual indicator | Identical |
| User Interface | Touch screen | Touch Screen | Touch Screen | Identical |
| Laser Dimension | 53 cm x 57 cm x 120 cm | 40 cm x 50 cm x 36 cm | 35 cm x 42 cm x 23 cm | Different (the subject device is somewhat<br>smaller than the predicate; both are smaller<br>than the reference device) |
| Weight | 50 kg. / 110 lbs. | 45 kg. / 120.64 lbs. | 9 kg. / 19 lbs. | Different (the subject device is considerably<br>lighter than both the predicate and reference<br>devices) |
{6}------------------------------------------------
## Global Med Systems Milesman Compact Laser 510(k) K203804
Image /page/6/Picture/1 description: The image shows the logo for Milesman Compact. The logo consists of a blue square with a white "M" inside, followed by the word "milesman" in blue. Below the word "milesman" is the word "compact" in white, set against a green background.
* 810 nm diode laser version of the Soprano™ with 1.2 cm² laser spot size, in either HR (hair removal) mode
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Milesman Compact. The logo consists of a blue square with a white "M" inside, followed by the word "milesman" in blue. Below the word "milesman" is a green rectangle with the word "compact" in white.
### VII. PERFORMANCE DATA
The following performance data on the current design were provided in support of the substantial equivalence determination.
IEC 60825-1 Edition 3.0 2014-05 Safety of laser products – Part 1: Equipment classification and requirements
IEC 60601-2-22 Medical electrical equipment - Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
### Biocompatibility testing
Compact Biocompatibility report for in vitro cytoxicity Compact Biocompatibility report for Dermal Irritation Compact Biocompatibility report for Skin Sensizitation
### Electrical safety and electromagnetic compatibility (EMC)
IEC 60601-1:2005 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
Note: A gap analysis was conducted to demonstrate compliance with the FDArecognized standard for electrical safety (AAMI/ANSI ES60601-1:2005(R) 2012 and A1:2012) because testing to the non-FDA-recognized standard IEC 60601-1:2005 was conducted.
## Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.
Software verification and validation testing was performed at both the unit level and integrated level using written protocols. The written protocols include a description of the acceptance criteria, which in most cases is taken as defined by the SRS (Software Requirements Specification). Test results are documented in test reports. Verification testing exercised software functionality by comparing actual outputs to those expected in response to pre-established inputs. Validation testing was done at the system level by operating the device in simulated use scenarios. A traceability matrix was used to account for each software requirement in the SRS to ensure testing thoroughness. For test failures, the software under test was analyzed and rewritten and the software test repeated. All software successfully passed verification and validation testing.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Milesman Compact. The logo consists of a blue square with a white "M" inside, followed by the word "milesman" in blue. Below the word "milesman" is a green rectangle with the word "compact" in white.
Animal Study An animal study was not conducted.
Clinical Studies Clinical testing was not conducted.
## VIII. CONCLUSIONS
The non-clinical data support the safety of the device and design verification and validation demonstrate that the Global Med Systems Milesman Compact device performs as intended in the specified use conditions. The Global Med Systems Milesman Compact device performs comparably to the predicate device (and reference device) that was marketed for the same intended use and shares the same or similar indications for use, similar design features, and functional features with, and thus is as safe, as effective, and performs as well as the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.