K073300 · Milesman S.L. · GEX · Dec 4, 2007 · General, Plastic Surgery
Device Facts
Record ID
K073300
Device Name
MILESMAN PREMIUM
Applicant
Milesman S.L.
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 4, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Milesman Premium Pulsed Diode Array Laser System is intended for use in general and plastic surgery and dermatology procedures for the treatment of vascular lesions, such as angiomas, hemanqiomas, telangjectasia and other benign vascular lesions and treatment for pseudofolliculitis barbae. The Milesman Premium is also used for removal of unwanted hair, permanent hair reduction and the treatment of benign pigmented lesions and leg veins in all skin types (Fitzpatrick I-VI), including tanned skin.
Device Story
Milesman Premium Pulsed Diode Array Laser System delivers 800 nm pulsed laser light for photothermolysis of soft tissue. System comprises console, umbilical hose, and handpiece. User selects parameters via console touch screen. Used in general/plastic surgery and dermatology clinics by professionals. Device utilizes compression-based refrigeration and a diode array (2 matrices of 10 bars) to deliver energy for hair reduction and lesion treatment. Output allows clinicians to perform targeted dermatological procedures; benefits include efficient treatment of vascular/pigmented lesions and hair reduction.
Clinical Evidence
No clinical data; bench testing only. Substantial equivalence established via comparison of technological characteristics and intended use to predicate device.
Technological Characteristics
800 nm pulsed diode laser system. Features compression-based refrigeration and diode array (2 matrices of 10 bars). Console-based with handpiece delivery. Class II device.
Indications for Use
Indicated for patients requiring treatment of vascular lesions (angiomas, hemangiomas, telangiectasia, benign vascular lesions), pseudofolliculitis barbae, benign pigmented lesions, leg veins, and unwanted hair removal/permanent hair reduction. Applicable to all skin types (Fitzpatrick I-VI), including tanned skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
LightSheer™ Pulsed Diode Array Laser System (K003614)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(K) SUMMARY of SAFETY and EFFECTIVENESS
6073300
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990 and 21 CFR §807.92.
The assigned 5l0(k) number is:
### 1. Submitter's Identification:
Milesman, S.L. Max Planck 545, P.I. Roces, 33211 Gijon, Asturias, Spain
DEC 0 4 2007
Date Summary Prepared: August 15, 2007
#### 2. Name of the Device:
| Trade Name: | Milesman Premium Pulsed Diode Array Laser System |
|----------------------|--------------------------------------------------|
| Classification Name: | Laser Instrument, Surgical, Powered |
| | 21CFR Part 878.4810 |
| | Product Code: GEX |
| | Review Panel: General & Plastic Surgery |
#### 3. Predicate Device Information:
K#003614, LightSheer™ Pulsed Diode Array Laser System
#### 4. Device Description:
The Milseman Premium Pulsed Diode Array Laser System delivers pulsed laser light at 800 nm wavelength. The device consists of three interconnected sections: the system console, the handle hose (umbilical cord) to the handpiece, and the handpiece. The laser pulsed energy is delivered through the handpiece. A user is able to select specific laser parameters from the touch screen on top of the console.
{1}------------------------------------------------
# 5. Intended Use:
The Milesman Premium Pulsed Diode Array Laser System is intended for use in general and plastic surgery and dermatology procedures for the treatment of vascular lesions, such as angiomas, hemanqiomas, telangjectasia and other benign vascular lesions and treatment for pseudofolliculitis barbae.
The Milesman Premium is also used for removal of unwanted hair, permanent hair reduction and the treatment of benign pigmented lesions and leg veins in all skin types (Fitzpatrick I-VI), including tanned skin.
# 6. Comparison to Predicate Device:
The Milesman Premium and the LightSheer ET are very similar or identical in terms of the device structure and its technology. Both systems are medical laser devices that deliver intense pulsed light at 800nm and are designed for effective photothermolysis treatment of soft tissue in general and plastic surgery, i.e. benign pigmented lesions, vascular lesions, leg veins, and removal of unwanted hair and permanent hair reduction.
The main differences between the Milesman Premium and LightSheer™ devices are in the refrigeration systems and in the number of diode lasers. The Milesman Premium and predicate LightSheer™ ET use different methods for obtaining the refrigeration with Milesman Premium refrigeration based on refrigeration by compression which is more efficient than thermoelectric refrigeration used by the predicate device. The diode laser array of Milesman Premium is formed by 2 matrixes of 10 the bars. The LightSheer™ ET diode array is formed by 2 matrixes of 35 bars. Due to a different method of bars assembly, the Milesman Premium is able to deliver the same amount of energy as LightSheer™ while using fewer diode lasers.
The above features of Milesman Premium do not introduce any new questions regarding the safety and effectiveness of the device.
# 7. Performance Data:
The Milseman Premium Pulsed Diode Array Laser System and the predicate LightSheer™ Pulsed Diode Array Laser System have very similar specifications and indications for use and as such, the performance data are not required.
#### 8. Discussion of Clinical Tests Performed:
Non-Applicable
{2}------------------------------------------------
# 9. Conclusion:
The Milseman Premium Pulsed Diode Array Laser System has the same intended use and similar characteristics as the LightSheer™ Pulsed Diode Array Laser System. Moreover, documentation supplied in this submission demonstrates that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, the Milesman Premium System is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, represented by three curved lines that suggest feathers or wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 0 4 2007
Milesman, S.L. % Intertek Testing Services Mr. Daniel W. Lehtonen 2307 East Aurora Road, Unit B7 Twinsburg, Ohio 44087
Re: K073300
Trade/Device Name: Milesman Premium Pulsed Diode Array Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general, and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: November 21, 2007 Received: November 23, 2007
Dear Mr. Lehtonen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Daniel W. Lehtonen
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Mulhausen
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE
Page _ 1______________________________________________________________________________________________________________________________________________________________________
510(k) Number (if known): _ k_() 7 3 3 0 0
| Device Name | Milesman Premium Pulsed Diode Array Laser System |
|-------------|--------------------------------------------------|
|-------------|--------------------------------------------------|
The Milesman Premium Pulsed Diode Array Laser System is intended for use in general and plastic surgery and dermatology procedures for the treatment of vascular lesions, such as angiomas, hemangiomas, telangiectasia and other benign vascular lesions and treatment for pseudofolliculitis barbae.
The Milesman Premium is also used for removal of unwanted hair, permanent hair reduction and the treatment of benign pigmented lesions and leg veins in all skin types (Fitzpatrick I-VI), including tanned skin.
Prescription Use X Over-The Counter Use OR (Per 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
일부노동========================================================================================================================================================================= Concurrence of CDRH, Office of Device Evaluation ODE)
Evaluation (SDE)
(Division Sigand Neurological Devices
510(k) Number k, 073300
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.