K203665 · Pollogen, Ltd. · PAY · Feb 25, 2021 · General, Plastic Surgery
Device Facts
Record ID
K203665
Device Name
STOP U Model UXV Device
Applicant
Pollogen, Ltd.
Product Code
PAY · General, Plastic Surgery
Decision Date
Feb 25, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4420
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The STOP U Model UXV device is intended for use in the non-invasive treatment of mild to moderate facial wrinkles for adult users who have Fitzpatrick Skin Types II-IV.
Device Story
STOP U Model UXV is a home-use electrosurgical device for aesthetic wrinkle reduction. It delivers bipolar RF energy (1 MHz) to dermis and subcutaneous layers to generate heat via electrical impedance. Device consists of applicator unit, power supply, and preparation gel. User operates device on facial skin; integrated temperature sensor monitors skin temperature and automatically disables or reduces RF transmission upon reaching maximum threshold to prevent overheating. Output is thermal energy; intended to improve appearance of facial wrinkles. Device is intended for lay person use in home environment.
Clinical Evidence
Bench testing only. Verification testing confirmed technical specifications for max RF power output, vibration, and buzzer functionality. Electrical safety, EMC, and software validation were performed per FDA guidance and relevant IEC/ISO standards. No clinical data provided.
Technological Characteristics
RF bipolar energy; 1 MHz frequency; 5.7W max RMS power; 4 electrodes with 6mm spherical diameter. Includes temperature sensor for thermal regulation. Biocompatible per ISO 10993-1. Complies with IEC 60601-1, 60601-1-2, 60601-1-6, 60601-1-11, 60601-2-2, and ISO 14971. Power source 100-240V.
Indications for Use
Indicated for non-invasive treatment of mild to moderate facial wrinkles in adults with Fitzpatrick Skin Types II-IV.
Regulatory Classification
Identification
An electrosurgical device for over-the-counter aesthetic use is a device using radiofrequency energy to produce localized heating within tissues for non-invasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance data must demonstrate that the device meets all design specifications and performance requirements. The following performance characteristics must be tested: Over-heating, power accuracy radiofrequency, pulse cycle, waveform, pulse duration, and device characterization parameters.
(2) Label comprehension and self-selection performance evaluation must demonstrate that the intended over-the-counter users can understand the package labeling and correctly choose the device for the indicated aesthetic use.
(3) Usability performance evaluation must demonstrate that the over-the-counter user can correctly use the device, based solely on reading the directions for use, to treat the indicated aesthetic use.
(4) Clinical performance evaluation must demonstrate that the device performs as intended under anticipated conditions of use to achieve the intended aesthetic results.
(5) The patient-contacting components of the device must be demonstrated to be biocompatible.
(6) Instructions for cleaning the device must be validated.
(7) Performance data must be provided to demonstrate the electromagnetic compatibility and electrical safety, including the mechanical integrity, of the device.
(8) Software verification, validation, and hazard analysis must be performed.
(9) Labeling must include:
(i) Warnings, precautions, and contraindications to ensure the safe use of the device for the over-the-counter users.
(ii) A statement that the safety and effectiveness of the device's use for uses other than the indicated aesthetic use are not known.
(iii) A summary of the clinical information used to establish effectiveness for each indicated aesthetic usage and observed adverse events.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 25, 2021
Pollogen Ltd. % Elissa Burg Regulatory Consultant BioVision Ltd. Had Nes 183 Had Nes, 1295000 Israel
Re: K203665
Trade/Device Name: STOP U Model UXV Device Regulation Number: 21 CFR 878.4420 Regulation Name: Electrosurgical Device for Over-The-Counter Aesthetic Use Regulatory Class: Class II Product Code: PAY Dated: January 26, 2021 Received: January 28, 2021
Dear Elissa Burg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203665
Device Name STOP U Model UXV
Indications for Use (Describe)
The STOP U Model UXV device is intended for use in the non-invasive treatment of mild to moderate facial wrinkles for adult users who have Fitzpatrick Skin Types II-IV.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### VI. 510(k) SUMMARY
#### Pollogen Ltd.'s STOP U Model UXV Device - K203665
- Pollogen Ltd. Applicant's name: 6 Kaufman St. Tel Aviv ISRAEL 6801298 Tel. (972)3-510-4110 Fax (972)3-510-4112
- Elissa Burg Contact Person: Regulatory Consultant BioVision Ltd. Had Nes 183 Israel 1295000 Tel. (972) 526633572 Fax (972) 4-6827312
Date Prepared: February 24, 2021
Name of Device: STOP U Model UXV
Common or Usual Name: Electrosurgical device for over-the-counter aesthetic use
Classification: Product Code: PAY Regulation No: 21 C.F.R. §878.4420 Class: II Classification Panel: General & Plastic Surgery
Predicate Device: Pollogen Ltd., STOP U (K182774)
### Intended Use / Indications for Use
The STOP U Model UXV device is intended for use in the non-invasive treatment of mild to moderate facial wrinkles for adult users who have Fitzpatrick Skin Types II-IV.
{4}------------------------------------------------
### Modified Device Description
The STOP U Model UXV device delivers RF current into the skin to generate heat through electrical impedance in the dermis and subcutaneous layers. The device consists of the following components and accessories: The STOP U Model UXV device (applicator unit), the STOP U Model UXV Power Supply and the STOP Preparation Gel.
#### Comparison of Technological Characteristics
The TriPollar™ STOP U Model UXV device delivers RF energy at a frequency of 1 MHz (+/-10%) and a maximum output RMS power of 5.7 watts (+/- 10%) into the skin through its electrodes. The device generates heat through electrical impedance in the dermis and subcutaneous layers. The temperature sensor located in the electrodes area constantly monitors the skin temperature and disables/reduces RF transmission once the maximal skin temperature is obtained.
The following table compares the modified STOP U Model UXV device to the predicate device with respect to intended use, technological characteristics and principles of operation, providing detailed information regarding the basis for the determination of substantial equivalence.
| | Proposed Device:<br>STOP U Model UXV<br>(K203665) | Predicate Device:<br>STOP U (K182774) |
|-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Pollogen® Ltd. | Pollogen® Ltd. |
| Device Class | Class II | Class II |
| Regulation Description | Electrosurgical device for over-<br>the-counter aesthetic use | Electrosurgical device for over-the-<br>counter aesthetic use |
| Regulation Number | 21 C.F.R. 878.4420 | 21 C.F.R. 878.4420 |
| Product Code | PAY | PAY |
| Intended Use /<br>Indications for Use | The STOP U Model UXV device<br>is intended for use in the non-<br>invasive treatment of mild to<br>moderate facial wrinkles for adult<br>users who have Fitzpatrick Skin<br>Types II-IV | The STOP U device is intended for<br>use in the non-invasive treatment of<br>mild to moderate facial wrinkles for<br>adult users who have Fitzpatrick<br>Skin Types II-IV |
| Deep tissue<br>Heating<br>Electromagnetic<br>Energy | RF | RF |
| Modes of Operation | RF Bipolar Energy | RF Bipolar Energy |
| Nominal Operating RF<br>Power (200 Ohms) | 5.7W (+/- 10%) | 5.7W (+/- 10%) |
| | Proposed Device:<br>STOP U Model UXV<br>(K203665) | Predicate Device: STOP U<br>(K182774) |
| RF Carrier Frequency | 1MHz (+/- 10%) | 1MHz (+/- 10%) |
| Waveform | Sinusoid | Sinusoid |
| Applicator Effective<br>Area | 1 cm² | 1 cm² |
| Total Power Density<br>(fluence) | 5.7 W/cm² (+/- 10%) | 5.7 W/cm² (+/- 10%) |
| Output Voltage of Power<br>Source | 8V DC | 8V DC |
| Output Current of<br>Power Source | 1.5A | 2.5A |
| Dimensions | H=134mm; L=51mm; W=32mm | H=134mm; L=51mm; W=32mm |
| Weight | 85 gr | 85 gr |
| RF Energy Emission<br>Indicator | Yes (Temp. sensor) | Yes (Temp. sensor) |
| Energy Source | 100-240V, 50-60Hz, 0.4A | 100-240V, 50-60Hz, 0.6A |
| Heating Levels | 1 | 1 |
| Electrodes | 4 | 4 |
| Diameter of Spherical<br>Portion of Electrodes | 6mm | 4mm |
| Biocompatibility | All parts that are in<br>contact with patient comply with<br>the requirements of ISO<br>10993-1 | All parts that are in<br>contact with patient comply with<br>the requirements of ISO<br>10993-1 |
| Software | Verified and validated<br>according to the FDA<br>guidance | Verified and validated<br>according to the FDA<br>guidance |
| Intended Operating<br>Environment | Home Use Device | Home Use Device |
| Intended Operator | Lay Person | Lay Person |
| Testing | Electrical safety & EMC | Electrical safety, EMC, & Usability<br>Study |
Special 510(k) Submission - STOP U Model UXV Device - Pollogen Ltd.
{5}------------------------------------------------
{6}------------------------------------------------
# Performance Data
Pollogen conducted several performance tests to demonstrate that the STOP U Model UXV device complies with performance standards and that it functions as intended.
- Verification test demonstrating that the STOP U Model UXV meets the system's technical specification for the, max RF power output, vibration parameters and buzzer functionality.
- -Electrical safety and compatibility testing was performed to validate that the STOP U Model UXV's power control and accuracy is in reference to the user's input.
- The STOP U Model UXV software was validated as required. -
In all instances, the STOP U Model UXV device functioned as intended and observations were as expected.
# Performance Standards
The STOP U Model UXV device complies with the following performance standards:
- IEC/EN 60601-1 Edition 3.1 Medical Electrical Equipment Part 1: General requirements . for safety (2005) and A1:2012.
- IEC 60601-1-6 Medical Electrical Equipment Part 1-6: General requirements for basic ● safety and essential performance - Collateral standard: Usability (2010/AMD2013).
- IEC 60601-1-11:2015 (2nd edition), Medical electrical equipment Part 1-11 General . requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- IEC/EN 60601-2-2 Medical Electrical Equipment Part 2-2: Particular requirements for . the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories (Edition 6.0 2017-03).
- IEC 62304 Medical device software - Software life cycle processes (2006, Ed. 1/AMD A1:2015).
- IEC 60601-1-2 Medical Electrical Equipment Part 1-2: General requirements for basic . safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests (2014, Ed. 4).
- ISO 15223-1:2016- Medical devices -- Symbols to be used with medical device labels, . labeling and information to be supplied - Part 1: General requirement.
- ISO 14971:2007 Medical devices Application of risk management to medical devices. ●
{7}------------------------------------------------
# Conclusion
The STOP U Model UXV device is as safe and effective as its predicate, Pollogen's STOP U device (K182774) for the requested intended use. The STOP U Model UXV device has the same intended use and indications for use and similar technological characteristics and principles of operation as its predicate. Performance data demonstrated that the minor technological differences in the STOP U Model UXV device do not raise any issues of safety or effectiveness in comparison to the predicate device. Thus, the STOP U Model UXV device is substantially equivalent to its predicate, Pollogen's STOP U device (K182774) for the requested intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.