STOP U (Packed Black USA), STOP U (Packed White USA)

K182774 · Pollogen, Ltd. · PAY · Jun 19, 2019 · General, Plastic Surgery

Device Facts

Record IDK182774
Device NameSTOP U (Packed Black USA), STOP U (Packed White USA)
ApplicantPollogen, Ltd.
Product CodePAY · General, Plastic Surgery
Decision DateJun 19, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4420
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The STOP U device is intended for use in the non-invasive treatment of mild to moderate facial wrinkles for adult users who have Fitzpatrick Skin Types II-IV.

Device Story

Device delivers RF current to skin; generates heat via electrical impedance in dermis and subcutaneous layers. Components: applicator unit, power supply, preparation gel. Used in home environment by lay person. Temperature sensor between electrodes monitors skin temperature; disables RF transmission upon reaching target temperature. Output: thermal energy for aesthetic wrinkle reduction. Benefits: non-invasive wrinkle improvement. Clinical decision-making: user self-selects based on packaging/instructions; device provides automated thermal regulation to ensure safety during treatment.

Clinical Evidence

Clinical trial (n=40) for prescription version (K140255) supported OTC clearance. Subjects followed 3 months post-treatment. Primary endpoint: improvement in facial wrinkles via blinded evaluation using Fitzpatrick Wrinkle and Elastosis scale. >80% of subjects showed >=1 grade improvement. No adverse events; mild erythema/edema resolved spontaneously. Usability study (n=61) demonstrated 100% success rate in human factors validation and effective self-selection.

Technological Characteristics

RF energy at 1 MHz; max RMS power 5.7W. Bipolar RF mode. 4 electrodes. Temperature sensor for thermal monitoring/cutoff. Dimensions: 134mm x 51mm x 32mm; weight 85g. Biocompatibility per ISO 10993-1. Standards: IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-1-6, ISO 14971.

Indications for Use

Indicated for non-invasive treatment of mild to moderate facial wrinkles in adults with Fitzpatrick Skin Types II-IV.

Regulatory Classification

Identification

An electrosurgical device for over-the-counter aesthetic use is a device using radiofrequency energy to produce localized heating within tissues for non-invasive aesthetic use.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance data must demonstrate that the device meets all design specifications and performance requirements. The following performance characteristics must be tested: Over-heating, power accuracy radiofrequency, pulse cycle, waveform, pulse duration, and device characterization parameters. (2) Label comprehension and self-selection performance evaluation must demonstrate that the intended over-the-counter users can understand the package labeling and correctly choose the device for the indicated aesthetic use. (3) Usability performance evaluation must demonstrate that the over-the-counter user can correctly use the device, based solely on reading the directions for use, to treat the indicated aesthetic use. (4) Clinical performance evaluation must demonstrate that the device performs as intended under anticipated conditions of use to achieve the intended aesthetic results. (5) The patient-contacting components of the device must be demonstrated to be biocompatible. (6) Instructions for cleaning the device must be validated. (7) Performance data must be provided to demonstrate the electromagnetic compatibility and electrical safety, including the mechanical integrity, of the device. (8) Software verification, validation, and hazard analysis must be performed. (9) Labeling must include: (i) Warnings, precautions, and contraindications to ensure the safe use of the device for the over-the-counter users. (ii) A statement that the safety and effectiveness of the device's use for uses other than the indicated aesthetic use are not known. (iii) A summary of the clinical information used to establish effectiveness for each indicated aesthetic usage and observed adverse events.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 19, 2019 Pollogen Ltd. % Elissa Burg Regulatory Consultant BioVision Ltd Had Nes 183 Had Nes, Israel 1295000 #### Re: K182774 Trade/Device Name: STOP U (Packed Black USA), STOP U (Packed White USA) Regulation Number: 21 CFR 878.4420 Regulation Name: Electrosurgical Device for Over-The-Counter Aesthetic Use Regulatory Class: Class II Product Code: PAY Dated: May 8, 2019 Received: May 9, 2019 Dear Elissa Burg: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### V. INDICATIONS FOR USE STATEMENT | DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. | |-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------| |-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------| **Indications for Use** | 510(k) Number ( <i>if known</i> ) | K182774 | |-----------------------------------|---------| | Device Name | STOP U | Indications for Use (Describe) The STOP U device is intended for use in the non-invasive treatment of mild to moderate facial wrinkles for adult users who have Fitzpatrick Skin Types II-IV. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (7/17) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF Traditional 510(k) Submission - STOP U - Pollogen Ltd. {3}------------------------------------------------ #### VI. #### 510(k) SUMMARY #### Pollogen Ltd.'s STOP U Device | Applicant's name: | Pollogen Ltd.<br>6 Kaufman St.<br>Gibor House, P.O.B. 50320<br>Tel Aviv<br>ISRAEL 6801298<br>Tel. (972)3-510-4110<br>Fax (972)3-510-4112 | |-------------------|------------------------------------------------------------------------------------------------------------------------------------------| |-------------------|------------------------------------------------------------------------------------------------------------------------------------------| - Contact Person: Elissa Burg Regulatory Consultant BioVision Ltd. Had Nes 183 Israel 1295000 Tel. (972) 526633572 Fax (972) 4-6827312 - Date Prepared: September 25, 2018 - Name of Device: STOP U Common or Usual Name: Electrosurgical device for over-the-counter aesthetic use - Product Code: PAY Classification: Regulation No: 21 C.F.R. §878.4420 Class: II Classification Panel: General & Plastic Surgery #### Predicate Devices - Pollogen Ltd., STOP U (K140255) । - -Endymed Ltd., NEWA (DEN150005) {4}------------------------------------------------ # Intended Use / Indications for Use The STOP U device is intended for use in the non-invasive treatment of mild to moderate facial wrinkles for adult users who have Fitzpatrick Skin Types II-IV. # Device Description The STOP U device delivers RF current into the skin to generate heat through electrical impedance in the dermis and subcutaneous layers. The device consists of the following components and accessories: The STOP U device (applicator unit), the STOP U Power Supply and the STOP Preparation Gel. # Technological Characteristics The STOP U device delivers RF energy at a frequency of 1 MHz and a maximum output RMS power of 5.7 watts into the skin through its electrodes. The device generates heat through electrical impedance in the dermis and subcutaneous layers. The temperature sensor, located between the electrodes constantly monitors the skin temperature and disables RF transmission once the desired skin temperature is obtained. ## Performance Data Pollogen conducted several performance tests to demonstrate that the STOP U device complies with performance standards and that it functions as intended. - STOP U Electrical safety and compatibility testing was performed to validate the STOP U power control and accuracy in reference to the user's input. - STOP U over-heating safety testing was performed to validate the conformity with the STOPU's design requirements and specifications for its temperature sensor and profile in reference to the user's input. - -The STOP U software was validated as required. In all instances, the STOP U device functioned as intended and observations were as expected. ## Performance Standards The STOP U device complies with the following performance standards: - IEC/EN 60601-1 Edition 3.1 Medical Electrical Equipment Part 1: General . requirements for safety (2005) and A1:2012. {5}------------------------------------------------ - IEC 62304 Medical device software Software life cycle processes (2006, ● Ed. 1/AMD A1:2015) - IEC/EN 60601-2-2 Medical Electrical Equipment Part 2-2: Particular requirements . for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories (2009 Ed.5) sections 202.6.1 (Emission) & 202.6.2 (Immunity). - IEC 60601-1-2 Medical Electrical Equipment Part 1-2: General requirements for ● basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests (2014, Ed. 4). - IEC 60601-1-11:2015 (2nd edition), Medical electrical equipment Part 1-11 -General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment - IEC 60601-1-6 Medical Electrical Equipment Part 1-6: General requirements for ● basic safety and essential performance – Collateral standard: Usability (2010/AMD2013) - ISO 10993-1:2009 Biological evaluation of medical devices – Part 1: Evaluation and testing - ISO 15223-1:2016- Medical devices -- Symbols to be used with medical device ● labels, labeling and information to be supplied -- Part 1: General requirement - ISO 14971:2012 Medical devices Application of risk management to medical ● devices ## Clinical performance data #### Safety & Effectiveness The effect of treatment of mild to moderate facial wrinkles using the STOP U was tested in a clinical trial that was conducted to support the clearance of the prescription version of the STOP U device (K140255). Since the over-the-counter (OTC) version of the STOP U device is identical to the prescription version, this data is also applicable to support the safety and effectiveness of the OTC device. Altogether, 40 subjects (37 female and 3 male) were enrolled in the study. Subjects were treated for improvement of facial wrinkles appearance and were followed for 3 months post last {6}------------------------------------------------ # K182774 treatment. In order to assess safety, adverse events occurrence was monitored before and after each treatment and at follow up visits. In order to evaluate treatment efficacy, pre and post treatment photos were introduced to three uninvolved physicians for blinded evaluation based on Fitzpatrick Wrinkle and Elastosis scale. Over 80% of the subjects showed at least one grade improvement in Fitzpatrick wrinkle score at three months follow-up post treatment based on objective evaluations of the baseline and three months follow-up photographs. There were no incidences of adverse effects or complications. As expected, mild to moderate erythema and mild edema were detected at the site of treatment immediately after treatment. All cases resolved without treatment within a few hours. Treatment was well tolerated with minimal to no pain in the majority of study subjects. The data reported in this study clearly indicates that the Stop U provides a safe and effective treatment for facial wrinkles. ## Usability, Self-selection & Labeling Comprehension Furthermore, Pollogen conducted a usability study for the STOP U device which was divided into two stages: - Stage one Self-Selection Study of the STOP U Device; 1. - 2. Stage two - Human Factors Validation of the Pollogen STOP U Device. The Self -Selection study using the final STOP U packaging design produced a correct self-selection rate that met Pollogen's goal. The final packaging design promotes correct self-selection and adequately explains user eligibility to potential users in the real world. 61 subjects (39 female and 22 male) which had successfully identified themselves as potential device users participated in the Human Factors validation with 100% success rate. These results indicated that the design of the STOP U and its associated instructional materials facilitated safe use of the device. {7}------------------------------------------------ # K182774 # Substantial Equivalence The following table compares the STOP U device to the predicate devices with respect to intended use, technological characteristics and principles of operation, providing detailed information regarding the basis for the determination of substantial equivalence. | | STOP U<br>Proposed Device | STOP U<br>(K140255) | Newa<br>(DEN150005) | |-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | Pollogen® Ltd. | Pollogen® Ltd. | Endymed™ | | Device Class | Class II | Class II | Class II | | Regulation Description | Electrosurgical device for<br>over-the-counter aesthetic<br>use | Electrosurgical cutting<br>and coagulation device<br>and accessories | Electrosurgical device<br>for over-the-counter<br>aesthetic use | | Regulation Number | 21 C.F.R. 878.4420 | 21 C.F.R. §878.4400 | 21 C.F.R. 878.4420 | | Product Code | PAY | GEI | PAY | | Intended<br>Use/Indications for Use | The STOP U device is<br>intended for use in the<br>non-invasive treatment of<br>mild to moderate facial<br>wrinkles for adult users<br>who have Fitzpatrick Skin<br>Types II-IV. | The STOP U device is<br>intended for use in the<br>non-invasive treatment<br>of mild to moderate<br>facial wrinkles and<br>rhytides. | The EndyMed Newa™<br>is an over-the-counter<br>home use device<br>intended for non-<br>invasive treatment of<br>mild to moderate facial<br>wrinkles for adult<br>women with Fitzpatrick<br>Skin Types I-IV | | Deep tissue<br>Heating<br>Electromagnetic<br>Energy | RF | RF | RF | | Modes of Operation | RF Bipolar Energy | RF Bipolar Energy | RF Bipolar Energy | | Nominal Operating<br>RF Power (200 Ohms) | 5.7W | 5.7W | 10W | | RF Carrier<br>Frequency | 1 MHz | 1 MHz | 675 KHz | | | STOP U<br>Proposed Device | STOP U<br>(K140255) | Newa<br>(DEN150005) | | | | | | | Waveform | Sinusoid | Sinusoid | Square | | Applicator Effective<br>Area | 1 cm² | 1 cm² | $0.9 X 1.5 = 1.35 cm²$ | | Total Power Density<br>(fluence) | 5.7 W/cm² | 5.7 W/cm² | 10 W/cm² | | Output Voltage | 8V DC | 8V DC | 9V DC | | Dimensions | H=134mm; L=51mm;<br>W=32mm | H=134mm;<br>L=51mm; W=32mm | 120mm X 73mm X<br>37mm | | Weight | 85 gr | 85 gr | 70 gr | | RF Energy Emission<br>Indicator | Yes (Temp. sensor) | Yes (Temp. sensor) | Yes (Temp. sensor &<br>motion sensor) | | Energy Source | 100-240V, 50-60Hz,<br>600mA | 100-240V, 50-60Hz,<br>600mA | 100-240V, 50-60Hz,<br>800mA | | Heating Levels | 1 | 1 | 2 | | Electrodes | 4 | 4 | 6 | | Biocompatibility | All parts that are in<br>contact with patient<br>comply with the<br>requirements of ISO<br>10993-1 | All parts that are in<br>contact with patient<br>comply with the<br>requirements of ISO<br>10993-1 | All parts that are in<br>contact with patient<br>comply with the<br>requirements of ISO<br>10993-1 | | Software | Verified and validated<br>according to the FDA<br>guidance | Verified and validated<br>according to the FDA<br>guidance | Verified and validated<br>according to the FDA<br>guidance | | Intended Operating<br>Environment | Home Use Device | Prescription Use<br>Device | Home Use Device | | Intended Operator | Lay Person | Healthcare<br>professionals | Lay Person | | Testing | Electrical safety, EMC,<br>& Usability Study | Electrical safety, EMC<br>& Clinical Study | Electrical safety,<br>EMC, Clinical &<br>Usability Study | {8}------------------------------------------------ {9}------------------------------------------------ ## K182774 The subject STOP U device is as safe and effective as Pollogen's STOP U device (K140225) and Endymed's NEWA (DEN150005). The Pollogen STOP U device has the same intended use and indications for use and similar technological characteristics and principles of operation as its predicate devices. Performance data demonstrate that the STOP U device is as safe and effective as its predicate devices. Thus, the STOP U device is substantially equivalent to its predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...