E-Denture Pro is a light-curable resin indicated for the fabrication of denture bases fabricated in dental laboratories for full removable dentures. The material is an alternative to traditional heat-curable and auto polymerizing resins. E-Denture Pro is intended exclusively for professional dental work. Fabrication of denture bases with E-Denture Pro requires a computer-aided and manufacturing (CAD/CAM) system that includes the following components: digital denture base files based on a digital impression, a digital light processing (DLP) printer, and curing light equipment.
Device Story
E-Denture Pro is a light-curable resin for additive manufacturing of denture bases. Input: digital denture base files derived from digital impressions. Process: CAD/CAM system uses DLP printer to cure resin; post-curing required. Output: custom denture base for full removable dentures. Used in dental laboratories by dental professionals. Resin supplied in lightproof bottles with programmed 'TAG' chip for printer identification. Benefits: alternative to traditional heat-curable/auto-polymerizing resins; enables digital workflow for denture fabrication.
Clinical Evidence
No clinical data. Bench testing only. Evaluated per DIN EN ISO 20795-1:2013 (base polymers) and FDA additive manufacturing guidance. Tests included surface quality, dimensional stability, color stability, translucency, flexural strength (72.7 MPa), flexural modulus, porosity, bonding to synthetic teeth, residual monomer, sorption, and solubility. Biocompatibility confirmed via ISO 10993 (cytotoxicity, sensitization, irritation, acute systemic toxicity).
Technological Characteristics
Light-curable resin composed of acrylates, methacrylates, methacrylated oligomers/monomers, photo initiators, colorants, and absorbers. Additive manufacturing (DLP). Non-sterile. Requires programmed TAG chip for printer compatibility. Validated for EnvisionOne cDLM, Vida, P4K, and D4K series printers.
Indications for Use
Indicated for fabrication of denture bases for full removable dentures in dental laboratories. Intended for professional dental use only. Requires CAD/CAM system (digital impression files, DLP printer, curing light).
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
e-DENT Temporary Resin and Extra-Oral Curing System (K102776)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
March 30, 2021
EnvisionTEC GmbH % Patsy Trisler Regulatory Consultant Qserve Group US, Inc. 7949 Beaumont Green East Drive Indianapolis, Indiana 46250
Re: K203641
Trade/Device Name: E-Denture Pro Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, or Rebasing Resin Regulatory Class: Class II Product Code: EBI Dated: March 4, 2021 Received: March 8, 2021
Dear Patsy Trisler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K203641
Device Name
EnvisionTEC E-Denture Pro
#### Indications for Use (Describe)
E-Denture Pro is a light-curable resin indicated for the fabrication of denture bases fabricated in dental laboratories for full removable dentures. The material is an alternative to traditional heat-curable and auto polymerizing resins. E-Denture Pro is intended exclusively for professional dental work. Fabrication of denture bases with E-Denture Pro requires a computer-aided and manufacturing (CAD/CAM) system that includes the following components: digital denture base files based on a digital impression, a digital light processing (DLP) printer, and curing light equipment.
Type of Use (Select one or both, as applicable)
| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
r
# 510(k) Summary – K203641
| I. SUBMITTER | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Name: | EnvisionTEC GmbH |
| Submitter Address: | Bruesseler Straße 51<br>D-45968 Gladbeck<br>Germany |
| Contact Person: | Ruediger van Bernum, Head of Application |
| Email: | ruediger.v.bernum@envisiontec.de |
| Telephone: | +49 2043 987545 |
| Date Prepared: | March 26, 2021 |
| II. DEVICE | |
| Trade Name: | E-Denture Pro |
| Common Name | Denture Resin |
| Classification:<br>Name Number<br>Product Code<br>Device Class | Denture Relining, Repairing, or Rebasing Resin 21 CFR 872.3760<br>EBI<br>II |
| III. PREDICATE DEVICE | |
| Primary<br>Predicate Device: | K191497, NextDent Denture 3D+, Vertex-Dental BV |
| Reference Device | K102776, e-DENT Temporary Resin and Extra-Oral Curing System,<br>DeltaMed GmbH |
| IV. INDICATIONS FOR USE STATEMENT | |
| E-Denture Pro is a light-curable resin indicated for the fabrication of denture bases<br>fabricated in dental laboratories for full removable dentures. The material is an alternative<br>to traditional heat-curable and auto polymerizing resins. E-Denture Pro is intended<br>exclusively for professional dental work. Fabrication of denture bases with E-Denture Pro<br>requires a computer-aided and manufacturing (CAD/CAM) system that includes the<br>following components: digital denture base files based on a digital impression, a digital<br>light processing (DLP) printer, and curing light equipment. | |
| V. DEVICE DESCRIPTION | |
| Device<br>Identification | The E-Denture Pro system combines a scanner with design<br>software, the light-curable resin, a 3D printer and a curing unit.<br>These components are used together during the manufacture of the<br>customized denture base for the removable full denture. |
| Technological<br>Characteristics | The light-curable resin is a proprietary composition of acrylates,<br>methacrylates, methacrylated oligomers and monomers, photo<br>initiators, colorants/dyes and absorbers. It is used by dental<br>laboratories and dental practices to make the denture bases for<br>removable full dentures. |
{4}------------------------------------------------
| The resin is offered in lightproof 1 kg PE bottles along with a<br>programmed chip (referred to as TAG), which is required for use<br>with the 3D printer. The TAG contains information identifying the<br>resin material, name and amount.<br>E-Denture Pro resin is an alternative material to heat-curable and<br>auto-polymerizable resins.<br>EnvisionTECs Perfactory® 3D-Printer models designed and<br>validated for use with the E-Denture Pro light cured resin are:<br>EnvisionOne cDLM, with LED<br>Vida Series, with LED<br>P4K Series, with LED<br>D4K Series, with LED | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| VI. SUBSTANTIAL EQUIVALENCE COMPARISON TABLE | | |
| | NEW DEVICE | PRIMARY PREDICATE |
| 510(k) NUMBER; | K203641 | K191497 |
| DEVICE NAME; | E-Denture Pro | NextDent Denture 3D+ |
| MANUFACTURER | EnvisionTEC GmbH | Vertex-Dental BV |
| PRODUCT CODE | EBI | EBI |
| REGULATORY NAME | Denture Relining, Repairing, or<br>Rebasing Resin | Denture Relining, Repairing, or<br>Rebasing Resin |
| CLASSIFICATION | 21 CFR 872.3760 | 21 CFR 872.3760 |
| INDICATIONS FOR<br>USE | E-Denture Pro is a light-curable<br>resin indicated for the fabrication<br>of denture bases fabricated in<br>dental laboratories for full<br>removable dentures. The<br>material is an alternative to<br>traditional heat-curable and auto<br>polymerizing resins. E-Denture<br>Pro is intended exclusively for<br>professional dental work.<br>Fabrication of denture bases with<br>E-Denture Pro requires a<br>computer-aided and<br>manufacturing (CAD/CAM)<br>system that includes the following<br>components: digital denture base<br>files based on a digital<br>impression, a digital light<br>processing (DLP) printer, and<br>curing light equipment. | NextDent Denture 3D+ is a light-<br>cured resin indicated for the<br>fabrication of denture bases<br>fabricated in dental laboratories,<br>including full and partial removable<br>dentures. The material is an<br>alternative to traditional heat-cured<br>and auto polymerization resins.<br>NextDent Denture 3D+ is intended<br>exclusively for professional dental<br>work. Fabrication of denture bases<br>with NextDent Denture 3D+ requires<br>a computer-aided and<br>manufacturing (CAD/CAM) system<br>that includes the following scanner,<br>design software, additive printer and<br>post-cure unit:<br>Design:<br>Scanner: 3Shape D900<br>Design Software: 3Shape Dental-<br>System 2016 Premium<br>Printing:<br>Printer: 3D Systems NextDent 5100<br>Figure 4®; Software: 3D Sprint<br>Post-curing:<br>Post-cure unit: NextDent LC-3D<br>Print Box |
{5}------------------------------------------------
| INGREDIENTS | Light-curable Resin | Light-curable Resin |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| MANUFACTURING<br>TECHNOLOGY TYPE | Additive | Additive |
| PRODUCT<br>CHARACTERISTICS:<br>Sterility | Non-sterile | Non-sterile |
| Flexural Strength<br>(comparative testing) | 72.7 MPa | 81.2 MPa |
| Secant Modulus<br>(comparative testing) | 2192 MPa | 3113 MPa |
| Flexural Strain<br>(comparative testing) | >5.0 % at break | 2.8 % at break |
| Biocompatibility | Biocompatible, according to ISO<br>10993 testing | Biocompatible, according to ISO<br>10993 testing |
| VII PERFORMANCE AND SAFETY TESTING | | |
| Animal Testing: | This product category does not require animal testing. | |
| Clinical Testing: | This product category does not require human clinical testing. | |
| Laboratory<br>Testing: | Testing was conducted to evaluate the performance of a<br>manufactured denture base, according to requirements of DIN EN<br>ISO 20795-1:2013, Dentistry – Base Polymers.<br><br>Including biocompatibility requirements, the following specification<br>requirements of the 3D-printed denture base material samples<br>were tested and have been met:<br>Surface qualityDimensional stabilityColor and color stabilityTranslucencyFlexural strength and flexural modulusFreedom from porosityBonding to synthetic teethResidual monomerSorptionSolubility | |
| Shelf Life Testing: | Validated real-time shelf life of the E-Denture Pro resin at time of<br>510(k) submission is 3 months. The resin is on real-time validation<br>testing for an ultimate shelf life of 24 months, stored in the original<br>packaging at temperatures at 30° C. Properties being tested<br>include material viscosity, material photoreactivity and color<br>change. Resin also was tested for good transport stability. | |
| Biocompatibility<br>Testing: | Testing, according to ISO 10993 and Good Laboratory Practices,<br>confirms that E-Denture Pro denture base is biocompatible and<br>non-toxic and meets the requirements for a device in contact with<br>mucosal membrane for >30 days. A biocompatibility risk<br>assessment was developed and presented in the 510(k). As a<br>result the following ISO 10993 tests were conducted and results | |
| | met the requirements of each test: | |
| | • Cytotoxicity Study Using ISO Elution Method (Part 5) | |
| | • Guinea Pig Maximization Sensitization Test (Part 10) | |
| | • Tests for Irritation and Skin Sensitization – Intracutaneous<br>Injection in Rabbits (Part 10) | |
| | • Acute Systemic Toxicity in Mice (Part 11) | |
| <b>Additive<br/>Manufacturing</b> | Testing, according to FDA's guidance Technical Considerations for<br>Additive Manufactured Medical Devices, was performed and<br>results were provided in the 510(k). These tests included<br>evaluation of all relevant properties of the printed resin using the<br>permitted machines. Further, tests based on considerations of the<br>orientation during manufacturing were performed. | |
{6}------------------------------------------------
### VIII COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE AND REFERENCE DEVICES
The intended use, critical specifications, and additive method of manufacturing of E-Denture Pro are substantially equivalent to the predicate device, NextDent Denture 3D+.
While the resin of the predicate is different from E-Denture Pro, both are photo-curable resins used in additive manufacturing and are of the same material category, and the software is the same as used in 3D-printing the Reference device.
The additive manufacturing processes both use a resin, scanner, printer and curing unit.
The testing performed by EnvisionTEC, directly compared to the predicate produced results that are similar to the predicate.
The noted differences, in comparison to the predicate device, raise no new questions of safety and effectiveness.
## VIX CONCLUSION
Based on the comparisons provided and the data submitted in this 510(k), it can be concluded E-Denture Pro is substantially equivalent to the predicate device. EnvisionTEC's analysis of E-Denture Pro compared to the predicate show they have the same intended use, and technological parameters that meet the requirements of ISO 20795-1:2003.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.