NextDent Denture 3D+ is a light-cured resin indicated for the fabrication of denture bases fabricated in dental laboratories, including full and partial removable dentures. The material is an alternative to traditional heat cured and auto polymerization resins. NextDent Denture 3D+ is intended exclusively for professional dental work. Fabrication of denture bases with NextDent Denture 3D+ requires a computer-aided and manufacturing (CAD/CAM) system that includes the following scanner, design software, additive printer and post-cure unit:
Device Story
NextDent Denture 3D+ is a light-cured, methacrylate-based resin used for 3D printing denture bases. The system requires a CAD/CAM workflow: 3Shape scanner/design software, NextDent 5100 Figure 4 3D printer, and NextDent LC-3DPrint Box for post-curing. The resin is polymerized via photo-initiators using light exposure. Used in dental laboratories by technicians; output is a custom-fabricated denture base. The device benefits patients by providing an alternative to traditional heat-cured or auto-polymerizing resins for removable prosthetics. Proper mixing and post-curing are critical for material quality and biocompatibility.
Clinical Evidence
Bench testing only. Compliance with ISO 20795-1:2013 (denture base polymers) and ISO 10993-1:2009/C1:2010 (biocompatibility) demonstrated. Testing included flexural strength, flexural modulus, water sorption, water solubility, residual monomer, and comprehensive biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity).
Technological Characteristics
Methacrylate-based resin with photo-initiators, fillers, and pigments. Polymerization via visible light in 3D printer and post-cure light chamber. Compliant with ISO 20795-1 (Type 4 materials) and ISO 10993. Non-sterile; 2-year shelf life. Requires specific CAD/CAM ecosystem (3Shape scanner/software, NextDent 5100 printer, LC-3DPrint Box).
Indications for Use
Indicated for fabrication of full and partial removable denture bases in dental laboratories by professional dental personnel.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
November 1, 2019
Vertex-Dental BV % Patsy Trisler Consultant Qserve Group US Inc. 5600 Wisconsin Avenue Chevy Chase, Maryland 20815
Re: K191497
Trade/Device Name: NextDent Denture 3D+ Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI Dated: September 3, 2019 Received: October 2, 2019
Dear Patsy Trisler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191497
Device Name NextDent Denture 3D+
#### Indications for Use (Describe)
NextDent Denture 3D+ is a light-cured resin indicated for the fabrication of denture bases fabricated in dental laboratories, including full and partial removable dentures. The material is an alternative to traditional heat cured and auto polymerization resins. NextDent Denture 3D+ is intended exclusively for professional dental work. Fabrication of denture bases with NextDent Denture 3D+ requires a computer-aided and manufacturing (CAD/CAM) system that includes the following scanner, design software, additive printer and post-cure unit:
| Design: | |
|-----------------|----------------------------|
| Scanner | 3Shape |
| Type | D900 |
| Software | |
| Design software | 3Shape |
| Type | Dental-System 2016 Premium |
| Printing: | |
| Printer | 3D Systems |
| Type | NextDent 5100 Figure4® |
| Software | 3D Sprint |
| Post-curing: | |
| Post-cure unit | NextDent |
| Type | LC-3DPrint Box |
| Software | n.a. |
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K191497
## Section 5.0
### 510(k) SUMMARY-- NextDent Denture 3D+
| I. SUBMITTER | |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Name: | Vertex-Dental B.V. |
| Submitter Address: | Centurionbaan 190<br>3769 AV Soesterberg The Netherlands |
| Contact Person:<br>Telephone #: | L. Vloet-Emonts<br>+31 88 61 60 430 |
| Date Prepared: | September 27, 2019 |
| II. DEVICE | |
| Device Trade Name: | NextDent Denture 3D+ |
| Common and<br>Classification Name(s): | Resin, Denture |
| Classification #: | 21 CFR 872.3760 |
| Product Code | EBI |
| Regulatory Class | 2 |
| III. PREDICATE<br>DEVICE(s) | |
| | NextDent Denture, K162572<br>No reference devices were used in this submission. |
| IV. DEVICE<br>DESCRIPTION | |
| Device Identification: | Light-Cure Resin, provided in a container. |
| Device Characteristics: | NextDent Denture 3D+ dimethacrylic system, polymerized<br>via photo initiators in a 3D printer setting. The color of the<br>denture is determined by the addition of pigments. |
| Environment of Use: | Healthcare facility/hospital<br>Dental (technical) laboratory. |
| Summary (Description)<br>of Device: | NextDent™ Denture 3D+ must be used in combination with<br>the NextDent™ 5100 Figure4® 3D printer.<br><br>Printer and resin must be optimized to each other in order to<br>get complete and precise printed parts. If printer and resin<br>are not optimized to each other this may have an adverse<br>effect on the accuracy and physical quality of printed parts.<br><br>The NextDent™ 5100 Figure4® 3D printer and the post-<br>curing lightbox NextDent™ LC-PrintBox make use of a light<br>source to polymerize the NextDent resin. Therefore we<br>advise to wear UV protective glasses when operating a 3D<br>printer and/or lightbox. Differences in color nuance may<br>occur due to:<br>production in batches;<br>inadequate shaking and mixing of |
| | the original packaging before use; inadequate stirring in the resin Denture 3D+ before use; insufficient post-curing 3D printer is not included with the device. |
| Materials of Use: | methacrylate-based resins with photo-initiator, filler and pigments. |
{4}------------------------------------------------
| V. INDICATIONS FOR<br>USE | NextDent Denture 3D+ is a light-cured resin indicated for<br>the fabrication of denture bases fabricated in dental<br>laboratories, including full and partial removable dentures.<br>The material is an alternative to traditional heat cured and<br>auto polymerization resins. NextDent Denture 3D+ is<br>intended exclusively for professional dental work.<br>Fabrication of denture bases with NextDent Denture 3D+<br>requires a computer-aided and manufacturing (CAD/CAM)<br>system that includes the following scanner, design software,<br>additive printer and post-cure unit: | | |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
{5}------------------------------------------------
| VI. COMPARISON OF<br>TECHNOLOGICAL<br>CHARACTERISTICS<br>WITH THE<br>PREDICATE DEVICE | Both NextDent™ Denture 3D+ and the predicate device have the following similar characteristics:<br>Pre-mixed light-cure resins with photo-initiator and pigments Polymerization by visible light Automated printing of resin in multiple layers, each light-cured before adding next layer, with post curing in light chamber Post curing by visible light-curing unit The difference is in the chemical composition of the specific resin. | | | | | | |
|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|----------------------------------------|---------------------------------------|--------------------------------------|-------------------------------|---------------------------------------|
| | Technological<br>characteristics | | Predicate device:<br>NextDent™ Denture | | New device:<br>NextDent™ Denture 3D+ | | |
| | | Brand | Type | Software | Brand | Type | Software |
| | Design:<br>Scanner | 3Shape | D900 | Dental-<br>System<br>2016-<br>Premium | 3Shape | D900 | Dental-<br>System<br>2016-<br>Premium |
| | Printing: | Envision<br>Tec | DDDP<br>4 | Perfectory | | | |
| | | Rapid<br>Shape | D30 | Netfabb | | NextDent<br>5100<br>Figure 4® | 3D Sprint |
| | Printer | Miicraft | 125Y / | MiiUtility<br>Miicontroller | 3D<br>Systems | | |
| | | 3D<br>Systems | Figure<br>4® | 3D Sprint | | | |
| | | Roland DG | DWP-<br>80S | Ver1.1 | | | |
| | Post-Curing: | | | | | | |
| | Post-cure<br>unit | NextDent™ | LC-<br>3DPrint<br>box | n.a. | NextDent™ | LC-<br>3DPrint<br>Box | n.a. |
| VII. SUMMARY OF<br>TESTING<br>[PERFORMANCE<br>DATA] | NextDent™ Denture 3D+ has been tested for mechanical properties as part of the product specification. The most applicable standard for mechanical characteristics determination of denture base polymers and copolymers is the ISO 20975-1: 2013 Dentistry - Base polymers - Part 1:<br>Denture base polymers. | | | | | | |
| Biocompatibility<br>Testing: | According to ISO 7405:2008/A1:2013 NextDent™ Denture 3D+ is considered a surface device, in contact with the mucosal membrane, for > 30 days.<br>The ISO 10993-1:2009/C1:2010 standard, including parts 3, 5, 10 and 11, was followed and the following biological safety aspects have been addressed:<br>Cytotoxicity Sensitization Irritation or intracutaneous reactivity Subacute/subchronic systemic toxicity Genotoxicity | | | | | | |
{6}------------------------------------------------
| | In addition, the following risks have been considered based<br>on a risk assessment, taking into account the specific nature<br>and duration of exposure to the device Carcinogenicity Reproductive/developmental/organ toxicity Immunotoxicity Furthermore, it is demonstrated that Re-use of print resin material (stored under the<br>correct conditions) without the introduction of<br>recycling steps can be used safely without impacting<br>the biological safety No monomers remain on the surface of the 3D<br>printed parts before post-curing due to cleaning<br>steps according to IFU |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | NextDent Denture 3D+ is compliant to the requirements<br>defined in ISO 10993-1/C1:2010, including parts 3, 5, 10<br>and 11, for permanent medical devices. |
| Bench Testing | NextDent Denture 3D+ has been tested for conformity with<br>the industry standard ISO 20795-1:2013 Dentistry – Base<br>polymers - Part 1: Denture base polymers. |
| | NextDent Denture 3D+ is compliant to the requirements<br>defined in ISO 20975-1 for Type 4 materials. |
| | The following bench tests are conducted on NextDent<br>Denture 3D+: Ultimate Flexural strength Flexural modulus Water sorption Water solubility Residual monomer Biocompatibility |
| | NextDent Denture 3D+ can be repeatedly and successfully<br>printed on all NextDent 5100 Figure4® printers and<br>complies to all requirements defined in ISO 20975-1 for<br>Type 4 materials. The medical device may be printed with<br>horizontal and vertical orientation independent of the build<br>location in the build space, with the possibility to re-use<br>resin, within 80% accuracy of 100 µm, with an worst-case<br>average product density of 6.39 g/mm, at a temperature<br>range of 18°C to 28°C. |
| Reprocessing, Sterility<br>and Shelf-Life Testing | The device is provided non-sterile.<br><br>From the Shelf life testing, NextDent Denture 3D+ has a<br>shelf life of 2 years. |
| VIII. CONCLUSIONS | NextDent Denture 3D+ and the predicate have the same<br>intended use and similar technological characteristics. |
| | |
| | The results of the performed tests show that NextDent<br>Denture 3D+ meets the requirements mentioned in the<br>applicable standards, and confirm that the device performs<br>similarly to the predicate device. |
| | It is therefore concluded that NextDent Denture 3D+ is safe<br>and performs as intended, and is substantially equivalent to<br>the predicate device. |
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.