M.U.S.T. Pedicle Screw Extension and Long Tab Implants
K203482 · Medacta International S.A. · NKB · Jan 26, 2021 · Orthopedic
Device Facts
Record ID
K203482
Device Name
M.U.S.T. Pedicle Screw Extension and Long Tab Implants
Applicant
Medacta International S.A.
Product Code
NKB · Orthopedic
Decision Date
Jan 26, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The M.U.S.T. Pedicle screws system is intended for posterior non-cervical pedicle fixation (T1-S2/ilium) and non-pedicle fixation, or anterolateral fixation (T8-L5). These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.
Device Story
M.U.S.T. Pedicle Screw Extension and Long Tab implants are spinal fixation components used in thoracolumbar surgery. The system includes cannulated/non-cannulated pedicle screws, rods, lateral connectors, and set screws. The Long Tab (LT) screws feature two tabs on the screw head to facilitate rod reduction during surgery; tabs are broken off post-reduction, leaving a standard tulip profile. The system is used by surgeons in clinical settings for spinal stabilization and fusion. The device provides mechanical support to the spine, aiding in the correction of deformities or stabilization of injured segments. The output is the physical fixation of the spinal construct, which promotes bone fusion and patient recovery.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by non-clinical bench testing, including geometrical analysis, cadaveric surgical simulation, and mechanical performance testing (static/dynamic axial compression, torsion, tensile force resistance) per ASTM F1717, F543, and F2193 standards. Pyrogenicity assessments were also performed.
Technological Characteristics
Materials include Titanium and Cobalt-Chrome (CoCr). Components include cannulated/non-cannulated pedicle screws, straight/pre-contoured rods, lateral connectors, and set screws. Mechanical testing performed per ASTM F1717-18, ASTM F543-17, and ASTM F2193-18a. Devices are sterile, implantable, and intended for posterior/anterolateral spinal fixation.
Indications for Use
Indicated for skeletally mature patients requiring posterior non-cervical pedicle fixation (T1-S2/ilium) or anterolateral fixation (T8-L5) as an adjunct to fusion for degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, spinal curvatures (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date January 26, 2021. The text is in a simple, sans-serif font. The date is written out in full, with the month, day, and year clearly visible. The text is centered and takes up most of the image.
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Medacta International SA % Chris Lussier Senior Director, Quality and Regulatory Medacta USA 3973 Delp Street Memphis, Tennessee 38118
Re: K203482
Trade/Device Name: M.U.S.T. Pedicle Screw Extension and Long Tab Implants Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB, KWQ, KWP Dated: November 24, 2020 Received: November 27, 2020
Dear Chris Lussier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K203482
### Device Name
M.U.S.T. Pedicle Screw Extension and Long Tab implants
### Indications for Use (Describe)
The M.U.S.T. Pedicle screws system is intended for posterior non-cervical pedicle fixation (T1-S2/ilium) and non-pedicle fixation, or anterolateral fixation (T8-L5). These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### Submitter I.
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66 Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Senior Director of Quality and Regulatory, Medacta USA Date Prepared: November 24, 2020
#### II. Device
| Device Proprietary Name: | M.U.S.T. Pedicle Screw Extension and Long Tab Implants |
|--------------------------|--------------------------------------------------------|
| Common or Usual Name: | Pedicle screw spinal system |
| Classification Name: | Thoracolumbosacral Pedicle Screw System |
| Primary Product Code: | NKB |
| Secondary Product Code: | KWQ, KWP |
| Regulation Number: | 21 CFR 888.3070, 21 CFR 888.3060, 21 CFR 888.3050 |
| Device Classification | II |
#### III. Predicate Device
Substantial equivalence is claimed to the following devices:
Primary predicate device:
- M.U.S.T. Pedicle Screw System, K153664, Medacta International SA ●
Reference predicate devices:
- M.U.S.T. Pedicle Screw System, K121115, Medacta International SA
- M.U.S.T. Pedicle Screw System, K132878, Medacta International SA ●
- M.U.S.T. Pedicle Screw System, K141988, Medacta International SA
- M.U.S.T. Pedicle Screw System, K162061, Medacta International SA
- M.U.S.T. Pedicle Screw System, K171170, Medacta International SA
- M.U.S.T. Combined Set Screws, K171758, Medacta International SA ●
- M.U.S.T. Pedicle Screw System, K193365, Medacta International SA
#### IV. Device Description
The M.U.S.T. Extension and M.U.S.T. Long Tab is intended to be used as part of the M.U.S.T. Pedicle Screw system (K121115, K132878, K141988, K153664, K162061, K171170, K171758, K193365) for
{4}------------------------------------------------
the stabilization and the fusion of the lumbar and thoracic spine. The M.U.S.T. pedicle screw system includes cannulated or non-cannulated poly-axial pedicle screws (K121115, K132878, K153664), cannulated or non-cannulated mono-axial pedicle screws (K132878), set screws (K121115, K171758), straight and pre-bent rods (K121115, K141988, K162061), lateral connectors (K162061) and cross connectors (K132878, K193365). The M.U.S.T. pedicle screw system also includes the enhanced screws and rods designed for percutaneous surgery (K141988).
The M.U.S.T. Extension and M.U.S.T. Long Tab introduce new sizes of sterile implantable devices intended to be used for posterior non-cervical pedicle fixation (T1-S2/ilium) or anterolateral fixation (T8-L5). The devices subject of this submission are:
- New size of Straight Rods (both Titanium and CoCr version): length 600mm Hex interface at . the end
- New size of Straight Rods (Anodized Titanium version): length 600mm ●
- Pre-contoured Rods (both Titanium and CoCr version) ●
- New sizes of Lateral Connectors: length 80; 100; 125 & 150mm ●
- New sizes of Enhanced Cannulated Pedicle Screws: Ø8; Ø9; Ø10 length from 30 to 100mm ●
- New cannulated pedicle screws with thread 4 leads (15 and 25 mm) ●
- Setscrews in Titanium ●
Intended purpose and the performance specification of the devices are equivalent to the ones of the current US cleared portfolio: screws, rods and lateral connectors are already used in the MUST implant construct.
#### V. Indications for Use
The M.U.S.T. Pedicle screws system is intended for posterior non-cervical pedicle fixation (T1-S2/ilum) and non-pedicle fixation, or anterolateral fixation (T8-L5). These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.
#### VI. Comparison of Technological Characteristics
#### M.U.S.T. Extension .
The new sizes of Straight Rods (Titanium and CoCr version) included in the M.U.S.T. Extension and the predicate devices (K121115) share the following characteristics:
- indication for use; ●
- design;
- diameters;
- materials; ●
{5}------------------------------------------------
- biocompatibility; ●
- device usage;
- sterility;
- shelf life;
- packaging. ●
The only difference between the Straight Rods included in the M.U.S.T. Extension and the predicate devices (K121115) is the rods length.
#### M.U.S.T. Extension ●
The new size of Straight Rod and Enhanced Straight Rod (Anodized Titanium version) included in the M.U.S.T. Extension and the predicate devices (K162061) share the following characteristics:
- indication for use;
- geometry;
- diameters; ●
- materials; ●
- biocompatibility;
- device usage;
- sterility;
- shelf life;
- packaging.
The only difference between the Straight Rods included in the M.U.S.T. Extension and the predicate devices (K162061) is the rods length.
#### M.U.S.T. Extension ●
The subject Pre-contoured Rods and Straight Enhanced Rods (both Titanium and CoCr version) included in the M.U.S.T. Extension and the predicate devices (K141988) share the following characteristics:
- indication for use; ●
- design; ●
- diameters; ●
- materials;
- biocompatibility; ●
- device usage; ●
- sterility;
- shelf life;
- packaging.
The only difference between the Pre-contoured Rods and Straight Enhanced Rods included in the M.U.S.T. Extension and the predicate devices (K141988) is the rods length.
{6}------------------------------------------------
#### M.U.S.T. Extension ●
The new sizes of Lateral connectors included in the M.U.S.T. Extension and the predicate devices (K162061) share the following characteristics:
- indication for use; ●
- geometry;
- diameters; ●
- materials; ●
- interconnection mechanism;
- biocompatibility;
- device usage;
- sterility;
- shelf life;
- packaging. ●
The only difference between the Lateral connectors included in the M.U.S.T. Extension and the predicate devices (K162061) is the connector length.
#### M.U.S.T. Extension ●
The new sizes of Enhanced Cannulated Pedicle Screws included in the M.U.S.T. Extension and the predicate devices (K171170 and K141988) share the following characteristics:
- indication for use;
- design; ●
- materials; ●
- biocompatibility; ●
- device usage;
- sterility;
- shelf life; ●
- packaging. ●
The only difference between the Enhanced Cannulated Pedicle Screws included in the M.U.S.T. Extension and the predicate devices (K141988) is the screws diameters and lengths; the thread of the new sizes is substantially equivalent to the thread shape of the predicate device.
#### M.U.S.T. Long Tab (LT) ●
The new M.U.S.T. LT Cannulated pedicle screws included in the M.U.S.T. LT and the predicate devices (K153664 and K171170) share the following characteristics:
- indication for use;
- design; ●
- materials;
{7}------------------------------------------------
- diameters;
- screw lengths; ●
- biocompatibility;
- device usage;
- sterility; ●
- shelf life;
- packaging. ●
The only difference between the M.U.S.T. LT Cannulated pedicle screws included in the M.U.S.T. LT and the predicate devices (K153664 and K171170) is the pedicle screw head that has two long tabs, that can easily allow reduction operation during the surgery (two different lengths of reduction; 15mm and 25mm). The MUST LT tulip is equivalent to the predicate device in terms of material and geometry. In fact, the two tabs are intended to the broken, during surgery and after rod reduction. The resulting tulip has a similar profile and dimensions reduced as the predicate device.
The new M.U.S.T. LT Combined setscrew in the M.U.S.T. LT and the predicate devices (K171758) share the following characteristics:
- indication for use; ●
- diameters; ●
- biocompatibility; ●
- device usage;
- sterility;
- shelf life; ●
- packaging.
The differences between the M.U.S.T. LT Combined setscrews in the M.U.S.T. LT and the predicate devices (K171758) are material and dimensions.
#### VII. Performance Data
The addition of the M.U.S.T. Extension and M.U.S.T. LT components to M.U.S.T. - Pedicle Screw System portfolio was evaluated by risk analysis to identify any new risks associated with the change. Based on the risk analysis, design verification was conducted to written protocols with pre-defined acceptance criteria. The protocols and pre-defined acceptance criteria were based on the standards, FDA guidance, and comparison to the predicate device system.
The following validation and tests are being provided in support of the substantial equivalence determination:
## Non-Clinical Studies
- DESIGN VALIDATION ●
{8}------------------------------------------------
- Geometrical Analysis MUST LT Thread Geometry: To ensure that the implant has o sufficient fixation in the bone, a geometrical analysis was performed in comparison to approved and marketed implants, related to the most important dimensions of the threads.
- Cadaver surgery performed according to the well-known MIS percutaneous (Minimally o invasive system) technique (99.PERC46.12 rev.04) by experienced surgeons and according to the M07.138.001 prepared for the Chicago Wetlab.
- Tab welding MUST LT Flexion strength test according to IL07.09.598 to characterize the o specific design of the tulip in the dedicated breaking area
- PERFORMANCE TESTING ●
- Worst Case definition for ASTM F1717 testing: discussion of implant types and sizes, chosen o as the worst case for mechanical testing of the spinal construct in a vertebrectomy model.
- Mechanical evaluation under static and dynamic axial compression according to ASTM о F1717-18 Standard test methods for Spinal Implants Construct in a Vertebrectomy model: Endolab 970 200220 30 1413 part 1 rev.0 and 970 200220 30 1414 part 1 rev.0
- Mechanical evaluation under static torsion according to ASTM F1717-18 Standard test o methods for Spinal Implants Construct in a Vertebrectomy model: Endolab 970 200220 30 1413 part 2 rev.0 and 970 200220 30 1414 part 2 rev.0
- Mechanical evaluation of axial tensile force resistance according to ASTM F543-17 Standard o specification and test methods for metallic medical bone screws and ASTM F2193-18a Standard Specification and test methods for components used in the surgical fixation of the spinal skeletal and 970.200227.120.868-part1 rev. 0
- Mechanical evaluation of torsion force resistance according to ASTM F543-17 Standard o specification and test methods for metallic medical bone screws and ASTM F2193-18a Standard Specification and test methods for components used in the surgical fixation of the spinal skeletal system: Endolab 970.200227.120.867-part2 rev. 0 and 970.200227.120.868part2 rev. 0
- o Mechanical evaluation of torsional yield strength, maximum torque and braking angle according to ASTM F543-17 Standard specification and test methods for metallic medical bone screws - A1: Endolab 970.200227.120.867-part3 rev. 0 and 970.200227.120.868-part3 rev. O
- Characterization and evaluation of risk of breakage during the insertion in the bone of MUST o screws according to ASTM F543-17 Standard specification and test methods for metallic medical bone screw: Medacta test report ASTM F543 Summary Report
{9}------------------------------------------------
- . PYROGENICITY
- 0 Pyrogenicity Assessment, ENDOTOXINS-MEDIATED PYROGENICITY ASSESSMENT REPORT FOR: "MUST PEDICLE SCREWS IMPLANTS SURGICAL KITS" TF VI-PS-01, RAS-01.008.141 Rev. 5, Dated: 17 October 2020
- 0 Pyrogenicity Assessment, ENDOTOXINS-MEDIATED PYROGENICITY ASSESSMENT REPORT FOR: "MUST LT IMPLANTS (VI-PS-01)", RAS-01.008.277, Rev. 0, Dated: 16 September 2020
Clinical Studies:
- No clinical studies were conducted.
## VIII. Conclusion
Based on the above information, the M.U.S.T. Extension and M.U.S.T. LT components are substantially equivalent to the identified predicate devices.
Substantial equivalence has been demonstrated through a comparison of intended use, design and technological characteristics, as well as performance evaluations.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.