M.U.S.T. Pedicle Screw System

K193365 · Medacta International S.A. · NKB · Jan 27, 2020 · Orthopedic

Device Facts

Record IDK193365
Device NameM.U.S.T. Pedicle Screw System
ApplicantMedacta International S.A.
Product CodeNKB · Orthopedic
Decision DateJan 27, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The M.U.S.T. Pedicle Screw System is intended for posterior non-cervical pedicle fixation (T1-S2/ilum) and anterolateral fixation (T8-L5). These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis and failed previous fusion in skeletally mature patients.

Device Story

M.U.S.T. Pedicle Screw System is a spinal fixation construct comprising rods, polyaxial/mono-axial pedicle screws, hooks, and connectors; this submission adds M.U.S.T. MC Cross Connectors. Used by surgeons in clinical settings for stabilization and fusion of non-cervical spine. System components are assembled intraoperatively to create rigid construct to stabilize spinal segments. Output is mechanical stabilization of spine to facilitate fusion; benefits include correction of spinal deformities and stabilization of injured or degenerated segments.

Clinical Evidence

No clinical studies were conducted. Evidence consists of non-clinical bench testing, including static and dynamic compression/bending per ASTM F1717 and cadaver studies.

Technological Characteristics

System includes rods, polyaxial/mono-axial screws, hooks, and cross connectors. Rod compatibility is 5.5mm. Mechanical testing performed per ASTM F1717. Pyrogenicity testing per USP <85> and <151>. No software or energy source.

Indications for Use

Indicated for skeletally mature patients requiring posterior non-cervical pedicle fixation (T1-S2/ilium) or anterolateral fixation (T8-L5) as an adjunct to fusion for degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, spinal curvatures (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ January 27, 2020 Medacta International SA Mr. Stefano Baj Regulatory and Compliance Director Strada Regina Castel San Pietro, Switzerland CH-6874 Re: K193365 Trade/Device Name: M.U.S.T. Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB, KWP, KWQ Dated: December 3, 2019 Received: December 4, 2019 Dear Mr. Baj: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see {1}------------------------------------------------ https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for (Ronald Jean) Vacant Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K193365 Device Name M.U.S.T. Pedicle Screw System ### Indications for Use (Describe) The M.U.S.T. Pedicle Screw System is intended for posterior non-cervical pedicle fixation (T1-S2/ilum) and anterolateral fixation (T8-L5). These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis and failed previous fusion in skeletally mature patients. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------| | <span></span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 2.0 510(k) Summary #### I. Submitter Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66 Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Date Prepared: December 3, 2019 Date Revised: January 24, 2020 #### II. Device | Device Proprietary Name: | M.U.S.T. Pedicle Screw System | |--------------------------|-----------------------------------------| | Common or Usual Name: | Pedicle Screw Spinal System | | Classification Name: | Thoracolumbosacral Pedicle Screw System | | Primary Product Code: | NKB | | Secondary Product Code: | KWP, KWQ | | Regulation Number: | 21 CFR 888.3070 | | Device Classification | II | #### III. Predicate Device Substantial equivalence is claimed to the following devices: Primary predicate device: - M.U.S.T. Extension, K132878, Medacta International ● Additional predicate devices: - M.U.S.T. Extension Straight Connector and Additional Screws, K171170, Medacta International SA - M.U.S.T. Pedicle Screw System, K121115, Medacta International SA - M.U.S.T. Pedicle Screw System, K153664, Medacta International SA - M.U.S.T. Pedicle Screw System, K141988, Medacta International SA #### IV. Device Description The purpose of this submission is to add the M.U.S.T. MC Cross Connectors to the M.U.S.T. Pedicle Screw System (K132878, K171170, K121115, K153664, K141988) for the stabilization and fusion of {4}------------------------------------------------ the non-cervical spine. The M.U.S.T. Pedicle System includes a wide range of implants: Rods, Polyaxial and Mono-axial Pedicle Screws, Hooks, Connectors of various design and size, that are used in combination to compose a spinal construct. #### V. Indications for Use The M.U.S.T. Pedicle Screw System is intended for posterior non-cervical pedicle fixation (T1-S2/ilium) and anterolateral fixation (T8-L5). These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, and/or lordosis), tumor, pseudoarthrosis and failed previous fusion in skeletally mature patients. #### VI. Comparison of Technological Characteristics The M.U.S.T. MC Cross Connector implants and the predicate devices share the following characteristics: - Rod compatibility (Ø5.5mm); ● - materials of construction: ● - biocompatibility: - device usage: ● - sterility: ● - . shelf life: and - packaging The M.U.S.T. MC Cross Connector implants and the predicate devices are technologically different with respect to: - connector length; - connector configuration ● # Discussion Medacta International SA has not made any changes to the intended use, materials of construction, biocompatibility, device usage, sterility, shelf life, and packaging of the subject devices. Based on the comparison of technological characteristics and performance data provided within this submission, the M.U.S.T. MC Cross Connector implants are substantially equivalent to the identified predicate devices. {5}------------------------------------------------ #### VII. Performance Data Testing was conducted according to written protocols with acceptance criteria that were based on standards. The following mechanical studies were performed on worst-case implants in support of a substantial equivalence determination: Non-Clinical Studies: - Performance Tests - Static compression / bending per ASTM F1717 o - Dynamic compression bending per ASTM F1717 o - Cadaver studies O - Pyrogenicity - o Bacterial Endotoxin Test (LAL test) was conducted according to European Pharmacopoeia §2.6.14 (which is equivalent to USP chapter <85>) and pyrogen test according to USP chapter <151> for pyrogenicity determination - o the subject devices are not labeled as non-pyrogenic or pyrogen free Clinical Studies: - . No clinical studies were conducted. ## VIII. Conclusion The information provided above supports that the M.U.S.T. MC Cross Connector implants are substantially equivalent to the identified predicate devices. Substantial equivalence has been demonstrated through a comparison of intended use, design and technological characteristics, as well as performance evaluations.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...