K203105 · Inion OY · HWC · Jan 13, 2021 · Orthopedic
Device Facts
Record ID
K203105
Device Name
Inion CompressOn Screw
Applicant
Inion OY
Product Code
HWC · Orthopedic
Decision Date
Jan 13, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Inion CompressOn™ products are intended for maintenance of alignment and fixation of bone fractures, comminuted fractures, osteotomies, arthrodeses or bone grafts or allografts) in the presence of appropriate additional immobilization (e.g., rigid fixation implants, cast or brace). In addition, the Inion CompressOn™ Ø 3.5/4.0/4.5/5.0/6.0 mm products are specifically intended for use in following indications: A. General indication: maintenance of reduction of cancellous bone fractures, osteotomies or arthrodeses of the upper extremity, ankle and foot in the presence of appropriate brace and/or immobilization. B. Specific indications: fractures and osteotomies of the malleoli, and ankle fractures.
Device Story
Cannulated headless compression screws; composed of degradable L-lactic and D-lactic acid co-polymers; degrade via hydrolysis into alpha-hydroxy acids; bioresorption within 2-4 years. Used for bone fixation; inserted by surgeons using specialized instrumentation (drill bits, taps, K-wires, depth gauges); can be inserted over guide wire. Provides compression to fracture/osteotomy line; maintains alignment during healing. Not intended to replace healthy bone or withstand full load-bearing. Provided sterile; single-use.
Clinical Evidence
Bench testing only. Mechanical torsion and pull-out testing performed per ASTM F2502-11. In vitro degradation testing performed per ISO 13781:2017 and ASTM F1635-16. Functional, handling, and simulated clinical use testing verified device performance and user needs.
Technological Characteristics
Degradable co-polymers (L-lactic/D-lactic acid). Cannulated headless screw form factor; 2.7-6.0 mm diameter; 12-90 mm length. Mechanical properties verified per ASTM F2502-11. Testing materials per ASTM F1839-08. Sterile, single-use. No software or electronic components.
Indications for Use
Indicated for patients requiring maintenance of alignment and fixation of bone fractures, comminuted fractures, osteotomies, arthrodeses, or bone grafts in upper extremity, ankle, and foot. Specific use for malleoli and ankle fractures. Requires additional immobilization (cast, brace, or rigid fixation). Not for full load-bearing.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
January 13, 2021
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Inion Oy Kati Marttinen VGM Quality and Regulatory Laakarinkatu 2 Tampere. 33520 Finland
Re: K203105
Trade/Device Name: Inion CompressOn Screw Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: October 9, 2020 Received: October 15, 2020
Dear Kati Marttinen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For; Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203105
Device Name Inion CompressOn™ Screw
#### Indications for Use (Describe)
The Inion CompressOn™ products are intended for maintenance of alignment and fixation of bone fractures, comminuted fractures, osteotomies, arthrodeses or bone grafts or allografts) in the presence of appropriate additional immobilization (e.g., rigid fixation implants, cast or brace).
In addition, the Inion CompressOn™ Ø 3.5/4.0/4.5/5.0/6.0 mm products are specifically intended for use in following indications:
A. General indication: maintenance of reduction of cancellous bone fractures, osteotomies or arthrodeses of the upper extremity, ankle and foot in the presence of appropriate brace and/or immobilization.
B. Specific indications: fractures and osteotomies of the malleoli, and ankle fractures.
Type of Use (Select one or both, as applicable)
| <div> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------|
| <div> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY Inion CompressOn™ Screw
| Manufacturer and submitter: | Inion Oy,<br>Lääkärinkatu 2,<br>FIN-33520 Tampere, FINLAND |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Date: | December 11, 2020 |
| Contact person | Kati Marttinen, VGM Quality and Regulatory<br>Phone: +358 10 830 6600<br>Fax: +358 10 830 6691<br>kati.marttinen@inion.com |
| Establishment registration<br>number | 9710629 |
| Trade name of the device | Inion CompressOn™ Screw |
| Device classification and<br>product code | Class II<br>Classification Panel: Orthopaedic<br>Product Code: HWC<br>Common name: Bone fixation screw<br>Regulation number: 888.3040 |
| Predicate device | Inion FreedomScrew™ (K123672) |
| Conformance with performance<br>standards | Compliance to voluntary consensus standards is<br>listed in the application. |
## Device description and principles of operation
Inion CompressOn™ Screws are cannulated headless compression screws made of degradable co-polymers composed of L-lactic acid and D-lactic acid. These polymers have a long history of safe medical use and they degrade in vivo by hydrolysis into alpha-hydroxy acids that are metabolised by the body. The implants retain sufficient strength to fulfil their intended function during the healing period of the fracture or osteotomy, and degrade gradually thereafter. Bioresorption takes place within two to four years. The implants are provided sterile to the user and are not to be re-used or re-sterilized.
Inion CompressOn™ Screws have nominal dimensions ranging from 2.7-6.0 mm in diameter and 12-90 mm in length. They are designed to be inserted using Inion® instrumentation: drill bits, bone taps, drill taps, screwdriver shafts, drill sleeves, K-wires and depth gauges. The screws can be inserted over a guide wire.
Inion CompressOn™ Screws provide fixation and are not intended to replace healthy bone or withstand the stress of full load bearing.
{4}------------------------------------------------
# 510(k) SUMMARY Inion CompressOn™ Screw
# Indications for use
The Inion CompressOn™ products are intended for maintenance of alignment and fixation of bone fractures, comminuted fractures, osteotomies, arthrodeses or bone grafts (i.e., autografts or allografts) in the presence of appropriate additional immobilization (e.g., rigid fixation implants, cast or brace).
In addition, the Inion CompressOn™ Ø 3.5/4.0/4.5/5.0/6.0 mm products are specifically intended for use in following indications:
A. General indication: maintenance of reduction and fixation of cancellous bone fractures, osteotomies or arthrodeses of the upper extremity, ankle and foot in the presence of appropriate brace and/or immobilization.
B. Specific indications: fractures and osteotomies of the malleoli, and ankle fractures.
## Comparison with predicate device
The Inion CompressOn™ Screws have the same material composition, manufacturing method and indications for use as the predicate device. The differences compared to the predicate device are:
- Inion CompressOn™ Screws include larger diameter designs (5.0/6.0 mm) -
- -Inion CompressOn™ Screws are headless to reduce the risk of palpability/irritation
- Inion CompressOn™ Screws provide compression to the fracture/osteotomy line -
- Inion CompressOn™ Screws do not include a disposable insertion adapter -
- Inion CompressOn™ Screws are not intended for plate fixation. -
## Performance testing for substantial equivalence determination
Mechanical torsion testing was conducted to determine and verify the sufficient mechanical properties of the Inion CompressOn™ screws for insertion. A mechanical pull-out test was conducted to verify that the pull-out properties of Inion CompressOn™ Screw are substantially equivalent with the predicate device Inion FreedomScrew™. The mechanical tests were conducted according to FDA recognized consensus standard ASTM F2502-11 Standard specification and test methods for absorbable plates and screws for internal fixation implants. In vitro degradation test was conducted in accordance with FDA recognized consensus standards ISO 13781:2017 Implants for surgery - Homopolymers, copolymers and blends on poly(lactide) - In vitro degradation testing and ASTM F1635-16 Standard Test Method for in vitro Degradation Testing of Hydrolytically Degradable Polymer Resins and Fabricated Forms for Surgical Implants, to verify that the degradation rate and pull-out properties of Inion CompressOn™ Screw are substantially equivalent with the predicate device Inion FreedomScrew™M initially and over the healing period. The choice of test material was done in accordance with FDA recognized consensus standard ASTM F1839-08 (Re-approved 2016) Standard specification for rigid polyurethane foam for use as a standard material for testing orthopaedic devices and instruments.
Functional and handling test and simulated clinical use test were performed to verify that the implants, accessory instruments, packaging and instructions for use are functioning together as intended, and conform to the defined user needs and intended uses.
{5}------------------------------------------------
# 510(k) SUMMARY Inion CompressOn™ Screw
# Conclusions
The data demonstrates that the intended use, material composition and scientific technology, degradation profile and mechanical properties of Inion CompressOn™ Screws are substantially equivalent with the predicate device Inion FreedomScrew™ (K123672). The devices have passed the tests for functionality and handling in simulated clinical use settings.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.