The Inion FreedomScrew™ products are intended for maintenance of alignment and fixation of bone fractures, comminuted fractures, osteotomies, arthrodeses or bone grafts (i.e., autografts or allografts) in the presence of appropriate additional immobilization (e.g., rigid fixation implants, cast or brace). In addition, the Inion FreedomScrew™ Ø3.5/4.0/4.5 mm products are specifically intended for use in following indications: A. General indications: maintenance of reduction of cancellous bone fractures, osteotomies or arthrodeses of the upper extremity, ankle and foot in the presence of appropriate brace and/or immobilization. B. Specific indications: fractures and osteotomies of the malleoli, and ankle fractures.
Device Story
Inion FreedomScrew™ is a degradable internal bone fixation device; composed of 100% poly(L-lactide-co-D,L-lactide). Device provides fixation for bone fractures, osteotomies, and arthrodeses; requires additional immobilization (cast/brace). Available in solid or cannulated designs (2.0-4.5mm diameter; 10-90mm length). Used by orthopedic surgeons in clinical settings; often with Inion OTPS™ plating systems or customized instrumentation (drill bits, taps, holding sleeves). Implants retain strength for 12 weeks; bioresorb within 2-4 years. Not intended to replace healthy bone or withstand full load-bearing stress. Provided sterile for single use.
Clinical Evidence
Bench testing only. Mechanical testing (shear, torsion, pull-out) verified strength properties. In vitro degradation testing confirmed material stability over the healing period and bioresorption profile. Functional, handling, and simulated clinical use testing confirmed device performance and compatibility with instrumentation compared to predicate devices.
Technological Characteristics
Material: 100% poly(L-lactide-co-D,L-lactide) degradable co-polymer. Form factor: Solid or cannulated screws (2.0-4.5mm diameter, 10-90mm length). Principle: Internal bone fixation via mechanical grip; bioresorbable. Compatibility: ISO 5835/ISO 9714 compatible instrumentation. Sterilization: Provided sterile.
Indications for Use
Indicated for maintenance of alignment and fixation of bone fractures, comminuted fractures, osteotomies, arthrodeses, or bone grafts (autograft/allograft) with additional immobilization. Specific 3.5/4.0/4.5mm sizes indicated for cancellous bone fractures, osteotomies, or arthrodeses of upper extremity, ankle, and foot; specifically malleoli fractures and ankle fractures.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
K123672 (pg 1/2
#### 510(k) SUMMARY Inion FreedomScrewTM
# FEB 2 7 2013
Image /page/0/Picture/3 description: The image shows the word "INION" in bold, white letters against a black background. The letters are large and fill most of the frame. The background has a grainy texture, which contrasts with the smooth, solid letters. The overall impression is simple and graphic.
| Manufacturer and submitter: | Inion Oy,<br>Lääkärinkatu 2,<br>FIN-33520 Tampere, FINLAND |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date of this summary:<br>Contact person | November 23, 2012<br>Kati Marttinen, Quality and Regulatory Director<br>Phone: +358 10 830 6600<br>Fax: +358 10 830 6691<br>kati.marttinen@inion.com |
| Establishment registration<br>number | 9710629 |
| Trade name of the device | Inion FreedomScrewTM |
| Device classification and<br>product code | Class II<br>Classification Panel: Orthopaedic<br>Product Code: HWC<br>Common name: Bone screw<br>Regulation number: 888.3040 |
| Predicate device | K031961<br>K043142<br>K030900 |
| Conformance with performance<br>standards | No applicable mandatory performance standards exist for<br>this device. Compliance to voluntary consensus standards is<br>listed in the application. |
#### Device description and principles of operation
Inion FreedomScrew™ implants are internal fracture fixation devices made of degradable co-polymer composed of 100 % poly(L-lactide-co-D,L-lactide). Based on in vitro testing, the implants retain their initial strength up to 12 weeks after implantation, and gradually lose their strength thereafter. Bioresorption takes place within two to four years.
Inion FreedomScrews™ are in the range of diameter: 2.0 - 4.5 mm and length: 10 - 90 mm. The designs include fully and partially threaded screws, as well as solid and cannulated screws. Inion FreedomScrew™ implants are used together with the previously cleared Inion OTPS™ plating systems, as well as alone according to their indications for use. The Inion FreedomScrew™ products provide fixation and are not intended to replace healthy bone or withstand the stress of full load bearing.
Inion FreedomScrews™ are provided sterile to the user and are non-collagenous. The shelf life of the device is 3 years and they are intended for single use only.
The Inion FreedomScrews™ are primarily designed to be used with customized Inion instrumentation, such as drill bits, bone taps, holding sleeves, screwdriver shafts, countersinks, depth gauges, K-wires and cutting instruments, as well as with ISO 5835/ISO 9714 compatible instrumentation.
{1}------------------------------------------------
## 510(k) SUMMARY Inion FreedomScrew™
Image /page/1/Picture/2 description: The image shows the word "INION" in white letters against a black background. The letters are bold and sans-serif. The background appears to have a grainy or textured quality, possibly due to noise or artifacts in the image.
#### Indications for use
The Inion FreedomScrew™M products are intended for maintenance of alignment and fixation of bone fractures, comminuted fractures, osteotomies, arthrodeses or bone grafts (i.e., autografts or allografts) in the presence of appropriate additional immobilization (e.g., rigid fixation implants, cast or brace).
In addition, the Inion FreedomScrew™ Ø3.5/4.0/4.5 mm products are specifically intended for use in following indications:
A. General indications: maintenance of reduction of cancellous bone fractures, osteotomies or arthrodeses of the upper extremity, ankle and foot in the presence of appropriate brace and/or immobilization.
B. Specific indications: fractures and osteotomies of the malleoli, and ankle fractures.
## Performance testing for substantial equivalence determination
Mechanical shear, torsion and pull-out tests were performed to verify the strength properties of Inion FreedomScrew™ implants and to compare them to the predicate devices. Mechanical fixation test was performed to verify the properties of Inion FreedomScrew™ with the Inion FreedomPlate™ and to compare them to the fixation properties of the predicate devices. Fixation tests with Inion FreedomPlate™ included the technique of removing the screw heads after insertion by cutting forceps or heat loop device, where plate-screw interface creates appropriate grip and secures fixation.
In vitro degradation testing was carried out to determine the degradation profile (i.e., change in material and mechanical properties) and verify the sufficiency of the mechanical stability over healing period as the polymer degrades during in vitro degradation and to ensure the degradation of the Inion FreedomScrews™.
- Functional and handling test and simulated clinical use test were performed to verify that the implants, accessory instruments, packaging and instructions for use are functioning together as intended, and conform to the defined user needs and intended uses. The tests were done also in comparison to the predicate devices.
The data demonstrates that the intended use, material composition and scientific technology, degradation profile and mechanical properties of Inion FreedomScrew" are substantially equivalent with the predicate devices Inion OTPS_"" Mesh System (K031961), Inion OTPS™ Mini Screws (K043142) and Inion OTPS™ Fixation System (K030900). The devices have passed the tests for functionality and handling in simulated clinical use settings and compare favourably to the predicate devices.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 27, 2013
Inion Oy % Ms. Kati Marttinen Quality and Regulatory Director Lääkärinkatu 2 335200 Tampere Finland
Re: K123672
Trade/Device Name: Inion FreedomScrewTM Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC Dated: November 23, 2012 Received: November 29, 2012
Dear Ms. Marttinen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
## Page 2 - Ms. Kati Marttinen
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Erin I.D.Keith
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Statement of Indications for Use
510(k) Number:
K123672 (pg 1/1)
Device Name:
Inion FreedomScrew™M
The Inion FreedomScrew™ products are intended for maintenance of alignment and fixation of bone fractures, comminuted fractures, osteotomies, arthrodeses or bone grafts (i.e., autografts or allografts) in the presence of appropriate additional immobilization (e.g., rigid fixation implants, cast or brace).
In addition, the Inion FreedomScrew™ Ø3.5/4.0/4.5 mm products are specifically intended for use in following indications:
A. General indications: maintenance of reduction and fixation of cancellous bone fractures, osteotomies or arthrodeses of the upper extremity, ankle and foot in the presence of appropriate brace and/or immobilization.
B. Specific indications: fractures and osteotomies of the malleoli, and ankle fractures.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
の日本の出版が出来ないというとして、そして、そして、日本の間に出来ている。 1981年には、 1981年には、 1992年には、 1972年(昭和47年)、 1972年(昭和47年) 1972年) 1972年(昭和47年) 1982年(昭和47年) 1982年(昭和47年) 1982年(昭和47年) 1982年(昭和47年)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ElizabethLMFrank-S
Divison of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.