BRIUS Clear Aligners

K202792 · Brius Technologies, Inc. · NXC · Jan 22, 2021 · Dental

Device Facts

Record IDK202792
Device NameBRIUS Clear Aligners
ApplicantBrius Technologies, Inc.
Product CodeNXC · Dental
Decision DateJan 22, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5470
Device ClassClass 2
AttributesTherapeutic

Indications for Use

BRIUS Clear Aligners are indicated for use in the alignment of permanent dentition (i.e. all second molars) through orthodontic treatment of misalignment and malocclusion. The aligners guide teeth to their final position by way of continuous gentle forces.

Device Story

BRIUS Clear Aligners are sequential thermoformed plastic aligners used for orthodontic tooth movement. Input consists of digital dental scans provided by a dental professional. Specialized CAD/CAM software (3Shape Ortho System) processes these scans to create a sequential treatment plan of 3D dental models. Each model is used to thermoform an aligner tray. The device is used in a clinical setting; the dental professional monitors treatment, while the patient wears the aligners 20-22 hours/day, removing them for eating and cleaning. The aligners apply continuous gentle forces to guide teeth to the desired final position. This treatment aims to correct malocclusion and improve dental alignment.

Clinical Evidence

No clinical data. Bench testing only. Validation included dimensional accuracy testing using 3rd party software and digital calipers, confirming measurements within 0.3 mm of target. Biocompatibility testing performed per ISO 10993-1 and ISO 10993-5:2009.

Technological Characteristics

Sequential thermoformed plastic aligners made of Essix thermoplastic. Non-sterile. Designed via CAD/CAM software (3Shape Ortho System) from digital intraoral scans. Manufacturing involves 3D printing molds followed by thermoforming. Biocompatible per ISO 10993-1.

Indications for Use

Indicated for alignment of permanent dentition (including second molars) to treat misalignment and malocclusion.

Regulatory Classification

Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 22, 2021 Brius Technologies, Inc % Jennifer Day Regulatory Affairs Consultant Medavice, Inc. 11218 Zest Ct. NE Blaine, Minnesota 55449 Re: K202792 Trade/Device Name: BRIUS Clear Aligners Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: November 20, 2020 Received: December 15, 2020 Dear Jennifer Day: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K202792 Device Name BRIUS Clear Aligners Indications for Use (Describe) BRIUS Clear Aligners are indicated for use in the alignment teeth (i.e. all second molars) through orthodontic treatment of misalignment and malocclusion. The aligners guide teeth to their final position by way of continuous gentle forces. | Type of Use (Select one or both, as applicable) | |------------------------------------------------------------------------------------------------| | <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | | <span style="">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Section 5. 510(k) Summary # 510(k) SUMMARY A summary of 510(k) information for K202792 in accordance with the requirements of 21 CFR 807.92. | Submitter: | BRIUS Technologies Inc. | |---------------------------|---------------------------------------------| | | 2611 Westgrove Dr. | | | Carrollton TX 75006 | | Company Contact Person: | Ted Schwarz | | Phone: | 217-778-9285 | | Email: | ted@brius.com | | Submission Correspondent: | Jennifer Day, Regulatory Affairs Consultant | | Address: | 11218 Zest Ct. NE, Blaine, MN 55449 | | Phone: | 314-809-1818 | | Email: | jday@medavice.com | | Date Prepared: | September 1, 2020 | | Proprietary Name: | BRIUS Clear Aligners | | Common Name: | Orthodontic plastic bracket. | | Product Code: | NXC - Orthodontic plastic bracket. | | Device Classification: | Class II, 21 CFR 872.5470 | | Primary Predicate Device: | ClearForm Aligners (K191838) | | Reference Predicates: | Ortho System (K180941), Essix (K062828) | # Device Description: The BRIUS Clear Aligners are thermoformed plastic aligners designed to be worn in sequence to facilitate the movement to the teeth to the final desired position. The sequential aligners introduce incremental movements that move teeth by way of gentle continuous force. The aligners are to be worn 20 to 22 hours a day and are to be removed for eating and for cleaning. BRIUS Clear Aligners are designed from digital scans of a patient's dentition submitted by a dental health professional (e.g. dentist or orthodontist). Using the scan, specialized orthodontic CAD/CAM software will be used to develop the treatment plans that consist of sequential dental models wherein the teeth are gradually realigned with each step. For this 510(k), 3Shape A/S's Ortho System (K180941) will be used for this application. Ortho System is approved for use in the management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options. The specialized orthodontic treatment planning software has a 510k clearance for the intended use under FDA Classification Product Code PNN, regulation 872.5470. {4}------------------------------------------------ Once the treatment plan is reviewed and approved by a dental health professional, each 3D model from the treatment plan is manufactured. The aligner trays are then manufactured by thermoforming a dental thermoplastic sheet over each model. The aligner trays are then delivered to the prescribing dental health professional. This dental health professional then monitors treatment from the placement of the first aligner to the delivery of the final aligner and completion of treatment. # Indications for Use: The BRIUS Clear Aligners are indicated for use in the alignment of permanent dentition (i.e. all second molars) through orthodontic treatment of misalignment and malocclusion. The aligners guide teeth to their final position by way of continuous gentle forces. # Comparison to Predicate Devices: BRIUS Clear Aligners are functionally equivalent to the following predicate device: ClearForm Aligners (Motor City Lab Works, K191838 cleared March 20th, 2020). The following table demonstrates the functional specifications of BRIUS Clear Aligners are substantially equivalent to the predicate devices. | Specification | Subject Device: | Predicate Device: | Comparison | |------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | | BRIUS Clear Aligners | ClearForm Aligners (K191838) | Result | | Regulation<br>Number | 21 CFR 872.5470 | 21 CFR 872.5470 | Same | | Classification<br>Name | Orthodontic Plastic Bracket | Orthodontic Plastic Bracket | Same | | Product Code | NXC | NXC | Same | | Classification | Class II | Class II | Same | | OTC or Rx | Rx | Rx | Same | | Material | Essix Thermoplastic | Essix Thermoplastic | Same | | Material<br>Properties | Acceptable material properties<br>established for use as an aligner. | Acceptable materials properties<br>established for use as an aligner. | Same | | Biocompatible | Yes | Yes | Same | | Sterile | Non-sterile | Non-sterile | Same | | Device<br>Description | Sequential thermoformed plastic<br>aligners | Sequential thermoformed plastic<br>aligners | Same | | Patient<br>Removable? | Yes | Yes | Same | | Indication for<br>Use | BRIUS Clear Aligners are<br>indicated for use in the<br>alignment of permanent teeth<br>(i.e. all second molars) through<br>orthodontic treatment of<br>misalignment and malocclusion.<br>The aligners guide teeth to their<br>final position by way of<br>continuous gentle forces. | ClearForm Aligners are indicated<br>for use in the alignment of<br>permanent teeth (i.e. all second<br>molars) through orthodontic<br>treatment of misalignment and<br>malocclusion. The aligners guide<br>teeth to their final position by<br>way of continuous gentle forces. | Same | | Mode of<br>Action | Continuous gentle force applied<br>to teeth following the prescribed<br>and approved treatment plan to<br>achieve orthodontic movement | Continuous gentle force applied<br>to teeth following the prescribed<br>and approved treatment plan to<br>achieve orthodontic movement | Same | #### Predicate Device Comparison Table {5}------------------------------------------------ # Comparison of Indications for Use to Predicate Devices: Based on the above comparison, the indications for use of the BRIUS Clear Aligners is similar to that of the ClearForm Aligners (K191838) as they are both indicated for use in the alignment of permanent teeth (i.e. all second molars) through orthodontic treatment of misalignment and malocclusion. The aligners guide teeth to their final position by way of continuous gentle forces. Thus, the BRIUS Clear Aligners can be considered substantially equivalent to its predicate device. # Comparison of Technological Characteristics to Predicate Devices: Based on the above comparison, the design, construction, and performance characteristics of the BRIUS Clear Aligners is similar to that of the ClearForm Aligners (K191838). Thus, the BRIUS Clear Aligners can be considered substantially equivalent to its predicate device. # Non-clinical performance testing: The use of thermoplastic materials for sequential aligners intended to treat malocclusions has been well documented in scientific literature regarding incremental tooth moving forces. However, durability testing was conducted on the aligners. Real world use was simulated to ensure that the aligner material and manufacturing process produced aligners that were suitable for their prescribed period of use. An internal manufacturing validation was performed to establish the dimensional accuracy of the manufacturing process for BRIUS Clear Aligners. The submitted intraoral scans, digital dentition models from treatment planning, 3D printed molds, and the final thermoformed aligners were all assessed quantitatively or qualitatively in the validation. Thus, each critical element in the manufacturing process was evaluated. For the validation, independent 3″ party software and digital calipers were used to perform point-to-point and critical displacement measurements, and visual inspections were performed to assess the aligner qualitatively. All measurements were within 0.3 mm of the target input value, the predefined tolerance of the manufacturing process. Furthermore, throughout the qualitative assessment of the aligners no performance, cosmetic, or other detectable issues were identified. This validation has met the preestablished acceptance criteria to demonstrate that the BRIUS manufacturing process yields dimensional accurate products that meet product specifications. The Essix thermoplastic material used for BRIUS Clear Aligners has 510(k) clearance (K062828) for use as an aligner material; the 510(k) holder (of the material) conducted the physical properties testing for the material. Biocompatibility testing for the aligner material, the only patient contacting material, was conducted by the 510(k) holder in accordance with International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". Additional cytotoxicity testing according to ISO 10993-5:2009 was performed on final manufactured BRIUS Clear Aligners. # Clinical performance testing: Clinical performance testing was not conducted. # Conclusion: Based on similarities in indications for use, technological characteristics, non-clinical performance testing, we believe that BRIUS Clear Aligners are substantially equivalent to the ClearForm Aligners.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...