K202568 · Medacta International S.A. · LPH · Dec 20, 2021 · Orthopedic
Device Facts
Record ID
K202568
Device Name
Mpact® 3D Metal Implants - DMLS Technology
Applicant
Medacta International S.A.
Product Code
LPH · Orthopedic
Decision Date
Dec 20, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Mpact® 3D Metal implants are designed for cementless use in total hip arthroplasty in primary or revision surgery. The patient should be skeletally mature. The patient's condition should be due to one or more of: · Severely painful and/or disabled joint: as a result of osteoarthritis, rheumatic arthritis, rheumatoid arthritis or psoriatic arthritis, congenital hip dysplasia, ankylosing spondylitis - · Avascular necrosis of the femoral head · Acute traumatic fracture of the femoral head or neck · Failure of previous hip surgery: joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present
Device Story
Mpact® 3D Metal Implants are sterile, cementless acetabular shells and compression polyaxial locking screws for total hip arthroplasty. Shells are manufactured via Direct Metal Laser Sintering (DMLS) using titanium alloy powder, forming complex geometries directly from CAD data. Screws provide fixation to cancellous bone, featuring a titanium alloy outer part and Co-Cr-Mo alloy inner part. Used in OR by orthopedic surgeons; implants replace damaged hip joint components to restore function and reduce pain. DMLS process ensures precise porous structures for osseointegration. Clinical benefit derived from mechanical stability and biocompatible material integration in primary or revision hip surgery.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by non-clinical bench testing, including stereological evaluation of porous coatings (ASTM F1854-15), static/fatigue shear testing (ASTM F1044-05, F1160-14), tension testing (ASTM F1147-05), ROM evaluation (ISO 21535), fatigue/deformation testing, and modular disassembly force testing (ASTM F1820-13). Biocompatibility was assessed per ISO 10993.
Technological Characteristics
Acetabular shells: Titanium alloy, DMLS additive manufacturing. Screws: Titanium alloy (ISO 5832-3) outer, Co-Cr-Mo alloy (ISO 5832-12) inner. Shell diameters 46-72mm; screws 20-70mm length, 6.5mm thread diameter. Sterile, single-use. DMLS process uses high-wattage laser to sinter metal powder from CAD data.
Indications for Use
Indicated for skeletally mature patients requiring cementless total hip arthroplasty (primary or revision) due to severely painful/disabled joints (osteoarthritis, rheumatoid/psoriatic/post-traumatic arthritis, congenital hip dysplasia, ankylosing spondylitis), avascular necrosis of the femoral head, acute traumatic fracture of the femoral head/neck, or failure of previous hip surgery (reconstruction, arthrodesis, hemiarthroplasty, surface replacement, or total hip replacement) with sufficient bone stock.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
Mpact® 3D Metal Acetabular Shells and 3D Metal Augments (K171966)
Mpact Acetabular System, Cancellous Bone Screws flat head (K103721)
Mpact Extension, Cancellous Bone Screws flat head (K132879)
{0}------------------------------------------------
December 20, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
Medacta International SA % Chris Lussier Director, Quality and Regulatory Medacta USA 3973 Delp Street Memphis, Tennessee 38118
Re: K202568
Trade/Device Name: Mpact® 3D Metal Implants - DMLS Technology Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH, LZO Dated: November 22, 2021 Received: November 23, 2021
Dear Chris Lussier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Limin Sun, PhD Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K202568
Device Name Mpact® 3D Metal Implants - DMLS Technology
## Indications for Use (Describe)
The Mpact® 3D Metal™ implants are designed for cementless use in total hip arthroplasty in primary or revision surgery. The patient should be skeletally mature.
The patient's condition should be due to one or more of:
· Severely painful and/or disabled joint: as a result of osteoarthritis, rheumatic arthritis, rheumatoid arthritis or psoriatic arthritis, congenital hip dysplasia, ankylosing spondylitis
- · Avascular necrosis of the femoral head
· Acute traumatic fracture of the femoral head or neck
· Failure of previous hip surgery: joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Medacta International SA
# 3.0 510(k) Summary
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Director of Quality and Regulatory, Medacta USA clussier@medacta.us.com Date Prepared: September 1, 2020 Date Revised: December 20, 2021
#### II. Device
| Device Proprietary Name: | Mpact® 3D Metal Implants - DMLS Technology |
|--------------------------|--------------------------------------------------------------|
| Common or Usual Name: | Total Hip Prosthesis |
| Classification Name: | Hip joint metal/polymer/metal semi-constrained porous-coated |
| | uncemented prosthesis |
| Primary Product Code: | LPH |
| Secondary Product Code: | LZO |
| Regulation Number: | 21 CFR 888.3358 and 21 CFR 888.3353 |
| Device Classification | II |
#### Predicate Device III.
Considering the Mpact® 3D Metal Implants manufactured through DMLS Technology, substantial equivalence is claimed to the following device:
- > Mpact® 3D Metal Acetabular Shells and 3D Metal Augments, K171966, Medacta International SA
Moreover, considering the Compression Polyaxial Locking Screws, substantial equivalence is claimed to the following devices:
- > Medacta International SA, Mpact Acetabular System, Cancellous Bone Screws flat head, K103721, Medacta International SA
- > Medacta International SA, Mpact Extension, Cancellous Bone Screws flat head, K132879, Medacta International SA
In addition, the following reference device is cited within the submission:
- > Medacta International SA, Medacta Shoulder System, Glenoid Polyaxial Locking Screws (K170452)
{4}------------------------------------------------
#### Device Description IV.
The Mpact® 3D Metal Implants - DMLS Technology are sterile implantable devices intended to be used during Total Hip Arthroplasty. The devices subject of this submission are:
- Mpact® 3D Metal Acetabular Shells Two-Hole Ø 46-66 mm manufactured through DMLS ● Technology;
- . Mpact® 3D Metal Acetabular Shells Multi-Hole - Ø 46-72 mm - manufactured through DMLS Technology:
- . Mpact® 3D Metal Acetabular Shells Multi-Hole Thin - Ø 48-60 mm - manufactured through DMLS Technology; and
- . Compression Polyaxial Locking Screws - 11 sizes depending on screw length: from 20 to 70mm in 5mm steps.
The Mpact® 3D Metal Acetabular Shells manufactured through DMLS Technology are a line extension to the Mpact® 3D Metal Acetabular Shells and 3D Metal Augments (K171966) and to the Mpact® Acetabular Systems Shells (K103721, K122641 and K132879) and are designed to be used with the Medacta Total Hip Prosthesis System.
The subject Mpact® 3D Metal Acetabular Shells are manufactured using a Direct Metal Laser Sintering (DMLS) process with titanium alloy powder. DMLS is a metal additive layer manufacturing process that uses a precise and high-wattage laser to "sinter" powdered metals and alloys to form accurate, complex and fully-functional metal parts directly from CAD data.
Compression Polyaxial Locking Screws have been designed to provide fixation of implants to the cancellous bone and they can be considered a line extension to the predicate devices, Cancellous Bone Screws flat head, previously cleared within K103721 and K132879. The subject screws have a thread diameter of 6.5 mm and they can be coupled with implants that have dedicated fixation holes with a spherical seat shape of Ø8mm.
The outer screw of the Compression Polyaxial Locking Screws is made of titanium alloy (Ti6-Al4-V) according to ISO 5832-3:2016, Implants for Surgery - Metallic materials - Part 3: Wrought titanium 6aluminum 4-vanadium alloy, while, the inner part, not in contact with bone, is made of Co-Cr-Mo alloy according to ISO 5832-12:2019, Implants for Surgery - Metallic materials - Part 3: Wrought cobaltchromium-molybdenum alloy.
#### V. Indications for Use
The Mpact® 3D Metal implants are designed for cementless use in total hip arthroplasty in primary or revision surgery. The patient should be skeletally mature.
The patient's condition should be due to one or more of:
- Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis, . rheumatoid arthritis or psoriatic arthritis, congenital hip dysplasia, ankylosing spondylitis
- Avascular necrosis of the femoral head .
- Acute traumatic fracture of the femoral head or neck ●
- Failure of previous hip surgery: ioint reconstruction, internal fixation, arthrodesis, . hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present
{5}------------------------------------------------
Medacta International SA
#### VI. Comparison of Technological Characteristics
#### Mpact® 3D Metal Acetabular Shells - DMLS Technology .
The Mpact® 3D Metal Acetabular Shells manufactured through DMLS Technology and the predicate devices (K171966) share the following characteristics:
- indication for use; ●
- shape and design;
- diameters:
- materials;
- biocompatibility;
- device usage;
- sterility;
- shelf life; and
- packaging.
The only difference between the Mpact® 3D Metal Acetabular Shells manufactured through DMLS Technology and the predicate devices (K171966) is the manufacturing process.
#### . Compression Polyaxial Locking Screws
The subject Compression Polyaxial Locking Screws and the predicate cancellous bone screws flat head (K103721 and K132879) share the following characteristics:
- sizes; ●
- thread outer diameter ●
- head shape ●
- outer material:
- biocompatibility;
- device usage;
- sterility;
- shelf life; and
- packaging.
The subject Compression Polyaxial Locking Screws differ from the predicate cancellous bone screws flat head (K103721 and K132879) with respect to:
- . neck diameter; and
- head height.
- thread core diameter;
- head diameter; ●
- recess; and ●
- . inner material.
{6}------------------------------------------------
#### Performance Data VII.
Based on the risk analysis, design validation and performance testing were conducted to written protocols. The following validation and tests are being provided in support of the substantial equivalence determination:
# Non-Clinical Studies
- DESIGN VALIDATION ●
- Mpact 3D Metal Two-Hole, Validation Workshop O
- Mpact 3D Metal Multi-Hole + Augment, Cadaver workshop and Evaluation forms o
- o Validation rational Mpact 3D Metal Multi-Hole Thin
- o Compression Polyaxial Locking Screw Design Validation Report (Cadaver Workshop), according to M07.85.003 rev.3. Test Report A1.
- PERFORMANCE TESTING ●
- 3D Metal DMLS Stereological Evaluation according to ASTM F1854-15 Standard Test o Method For Stereological Evaluation Of Porous Coatings On Medical Implants and EndoLab Test Report No. 167.180116.10.2760 Rev. 1
- 3D Metal DMLS Static and Fatigue Shear Test according to ASTM F1044-05 Standard Test o Method for Shear Testing of Calcium Phosphate Coatings and Metallic Coatings, ASTM F1160-14 Standard Test Method For Shear And Bending Fatigue Testing Of Calcium Phosphate And Metallic Medical And Composite Calcium Phosphate/Metallic Coatings, Medacta Test Protocol IL 07.09.476 Rev.0 and EndoLab Test Reports 167.180404.70.1049 Rev. 0 and 167.180406.70.1050 Rev. 0
- 3D Metal DMLS Tension Test according to ASTM F1147-05 Standard Test Method For o Tension Testing Of Calcium Phosphate And Metal Coatings, Medacta Test Protocol IL 07.09.477 Rev. 0 and EndoLab Test Report No. 167.180404.70.1048 Rev. 0
- Mpact 3D Metal Two-Hole, Multi-Hole and Multi-Hole Thin Evaluation of ROM according о to EN ISO 21535 Non-active Surgical implants - Joint replacement implants - Specific requirements for hip-joint replacement implants
- Mpact 3D Metal Fatigue and Deformation Test according to Medacta FEM Fatigue and о deformation test worst case Rev. 2, Test Protocols IL 07.09.516 Rev. 01 and IL 07.09.342 Rev. 02. Element Test Report No. 00812-008387-1A and Medacta Test Reports
- Mpact 3D Metal Shell with Liner Stability test according to ASTM F1820-13 Standard Test o Method For Determining the Forces for Disassembly of Modular Acetabular Devices, Medacta Test Protocol IL 07.09.596 Rev. 00 and EndoLab Test Reports 167.300310.10.3195 Rev. 0, 167.300310.10.3196-part 1 and part 2 Rev. 0
- Compression Polyaxial Locking Screw Mechanical Test according to Medacta Protocol IL o 07.09.450 rev.0 and Endolab Test report No. 167.171121.120.714 - part 1, 2 and 3. Test report A2
- PYROGENICITY
- o Bacterial endotoxin test (LAL test) according to European Pharmacopoeia §2.6.14 (which is equivalent to USP chapter <85>)
{7}------------------------------------------------
- o Pyrogen test according to USP chapter <151> for pyrogenicity determination
- o The subject devices are not labeled as non-pyrogenic or pyrogen free.
- BIOCOMPATIBILITY ●
- o Biocompatibility assessment as per ISO 10993 series and FDA Biocompatibility Guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"
# Clinical Studies:
- No clinical studies were conducted.
#### VIII. Conclusion
Based on the above information, the Mpact® 3D Metal Implants - DMLS Technology and the Compression Polyaxial Locking Screws are substantially equivalent to the identified predicate devices. Substantial equivalence has been demonstrated through a comparison of intended use, design and technological characteristics, as well as performance evaluations. The Mpact® 3D Metal Implants -DMLS Technology and the Compression Polyaxial Locking Screws are as safe and effective as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.