The Azer Scientific Universal Transport Medium is intended for the collection and transport of upper respiratory clinical specimens, containing respiratory viruses from the collection site to the testing laboratory. The Azer Scientific Universal Transport Medium is a culture-based media that can be processed using standard clinical laboratory operating procedures for the isolation and detection of upper respiratory viruses including Influenza A, Respiratory Syncytial Virus (RSV) and Adenovirus.
Device Story
Device consists of plastic screw-cap tube containing 3 mL isotonic, non-toxic transport medium and three 3 mm glass beads; supplied alone or in kits with flocked or foam swabs. Medium contains Hank's Balanced Salt Solution, BSA, L-cysteine, L-glutamic acid, gelatin, HEPES buffer, sucrose, Phenol red, and antibiotics (Vancomycin, Amphotericin B, Colistin). Used by clinicians to collect upper respiratory specimens; specimens transported to laboratory for viral culture. Medium maintains viral stability and inhibits competing microbial growth. Phenol red serves as visual pH indicator. Supports viral recovery for up to 48 hours at 2-25°C.
Clinical Evidence
No clinical data. Bench testing only. Viral recovery study performed using spiked pooled human nasopharyngeal swab matrix (negative by RT-PCR). Tested Influenza A, Adenovirus, and RSV at 0, 24, and 48 hours at 2-8°C and 20-25°C. Recovery measured via TCID50/mL and percent log reduction using Reed-Muench method. All results met acceptance criteria (change within one log or ≤90% of baseline). Shelf-life stability (12 months) and sterility/contamination checks also performed.
Technological Characteristics
Isotonic, non-toxic transport medium in plastic conical-bottom tubes. Formulation: Hank's Balanced Salt Solution, BSA, L-cysteine, gelatin, sucrose, L-glutamic acid, HEPES, phenol red, Vancomycin, Amphotericin B, Colistin. pH 7.3 ± 0.2. Includes 3 mm glass beads. Swabs: Flocked nylon or reticulated polyurethane foam. Shelf life: 12 months at 2-25°C. Non-sterile medium; components sterilized via moist heat or filtration; swabs provided sterile.
Indications for Use
Indicated for the collection and transport of upper respiratory clinical specimens containing respiratory viruses (Influenza A, RSV, Adenovirus) from the collection site to the laboratory for culture-based isolation and detection.
Regulatory Classification
Identification
A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.
Predicate Devices
Copan Universal Transport Medium (utm-rt) System (K042970)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K202531
B Applicant
Azer Scientific, Inc.
C Proprietary and Established Names
Azer Scientific Universal Transport Medium
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JSM | Class I, reserved | 21 CFR 866.2390 - Transport Culture Medium | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain a substantial equivalence determination for the Azer Scientific Universal Transport Medium for the collection and transport of clinical specimens containing viruses.
B Measurand:
Not Applicable.
C Type of Test:
Non-propagating Transport Device with Culture Medium
## III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K202531 - Page 2 of 10
B Indication(s) for Use:
The Azer Scientific Universal Transport Medium is intended for the collection and transport of upper respiratory clinical specimens, containing respiratory viruses from the collection site to the testing laboratory. The Azer Scientific Universal Transport Medium is a culture-based media that can be processed using standard clinical laboratory operating procedures for the isolation and detection of upper respiratory viruses including Influenza A, Respiratory Syncytial Virus (RSV) and Adenovirus.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
Not Applicable.
IV Device/System Characteristics:
A Device Description:
The Azer Scientific Universal Transport Medium includes a plastic screw-cap tube with conical bottom containing 3 mL of transport medium and three, 3 mm size glass beads. Azer Scientific Universal Transport Medium tubes can be supplied alone or in a kit format with one of three different collection swab options in a sterile peel pouch.
Swab Options:
- PurFlock Ultra 6" Sterile Standard Flock Swab w/Polystyrene Handle, 80mm Breakpoint. Regular Flock Tipped Specimen Collection Applicator and Plastic Handle (80mm Breakpoint) (SKU#: 25-3306-U).
- PurFlock Ultra 6" Sterile Revised Ultrafine Flock Swab w/Polystyrene Handle, 80mm Breakpoint. Revised Ultrafine Flock Tipped Applicator with Plastic Handle (80mm breakpoint) (SKU#: 25-3320-U EMB 80MM)
- Azer Scientific 3" Foam Swab, individually wrapped, sterile (ES45019S)
The Azer Scientific Universal Transport Medium is offered in the following configurations:
| SKU | PRODUCT DESCRIPTION | | PACKAGING |
| --- | --- | --- | --- |
| | TUBE | SWAB | |
| PFUTM-10 | 3 mL universal transport medium in 10mL screw cap tube with internal shaped conical bottom with 3 mm size glass beads | NA | 50/Pack |
| PFUTM-15 | 3 mL universal transport medium in 15mL screw cap tube with internal shaped conical bottom with 3 mm size glass beads | NA | 50/Pack |
{2}
| ES45019S | NA | Azer Scientific Foam Swab, individually wrapped, sterile | 500/pack |
| --- | --- | --- | --- |
| NPKIT-10 | PFUTM-10 | PurFlock Ultra 6" Sterile Revised Ultrafine Flock Swab w/Polystyrene Handle, 80mm Breakpoint. Revised Ultrafine Flock Tipped Applicator with Plastic Handle (80mm breakpoint) | 50/pack |
| MCKIT-10 | PFUTM-10 | PurFlock Ultra 6" Sterile Standard Flock Swab w/Polystyrene Handle, 80mm Breakpoint. Regular Flock Tipped Specimen Collection Applicator and Plastic Handle (80mm Breakpoint) | 50/pack |
| FOAMKIT-10 | PFUTM-10 | 3" Foam Specimen Collection Swab | 50/pack |
| NPKIT-15 | PFUTM-15 | PurFlock Ultra 6" Sterile Revised Ultrafine Flock Swab w/Polystyrene Handle, 80mm Breakpoint. Revised Ultrafine Flock Tipped Applicator with Plastic Handle (80mm breakpoint) | 50/pack |
| MCKIT-15 | PFUTM-15 | PurFlock Ultra 6" Sterile Standard Flock Swab w/Polystyrene Handle, 80mm Breakpoint. Regular Flock Tipped Specimen Collection Applicator and Plastic Handle (80mm Breakpoint) | 50/pack |
| FOAMKIT - 15 | PFUTM-15 | 3" Foam Specimen collection Swab | 50/pack |
B Principle of Operation:
The Azer Scientific Universal Transport Medium System is an isotonic and non-toxic medium. The medium consists of the following: Hank's Balanced Salt Solution. Bovine Serum Albumin,
K202531 - Page 3 of 10
{3}
L-cysteine, L-glutamic acid, Vancomycin, Amphotericin B, Colistin, gelatin, HEPES Buffer, sucrose and Phenol red. The Hank's buffer creates a neutral environment to help increase the stability of the virus. Bovine Serum Albumin (BSA) acts as a protein stabilizer by forming a protective film on the protein shell of the virus, making it less likely to break down and ensuring the integrity of the virus. Vancomycin, Amphotericin B, and Colistin inhibit the proliferation of competing bacteria and yeasts. L-glutamic acid serves as an auxiliary energy source to keep cell and virus stability. HEPES buffer provides additional help to maintain a stable pH. Phenol red is a pH indicator which serves as a visual quality control mechanism. The L-cysteine, gelatin, and sucrose, helps in preserving the virus.
V Substantial Equivalence Information:
A Predicate Device Name(s): Copan Universal Transport Medium (utm-rt) System
B Predicate 510(k) Number(s): K042970
C Comparison with Predicate(s):
| Device & Predicate Device(s): | Device: K202531 | Predicate: K042970 |
| --- | --- | --- |
| Device Trade Name | Azer Scientific Universal Transport Medium | Copan Universal Transport Medium (UTM-RT) System |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The Azer Scientific Universal Transport Medium is intended for the collection and transport of upper respiratory clinical specimens, containing respiratory viruses from the collection site to the testing laboratory. The Azer Scientific Universal Transport Medium is a culture-based media that can be processed using standard clinical laboratory operating procedures for the isolation and detection of upper respiratory | Copan Universal Transport Medium (UTM-RT) System is intended for the collection and transport of clinical specimens containing viruses, Chlamydiae, Mycoplasma or Ureaplasma from the collection site to the testing laboratory. UTM-RT can be processed using standard clinical laboratory operating procedures for viral, chlamydial, mycoplasma and ureaplasma culture. |
K202531 - Page 4 of 10
{4}
K202531 - Page 5 of 10
| | viruses including Influenza A, Respiratory Syncytial Virus (RSV) and Adenovirus. | |
| --- | --- | --- |
| Single-Use Device | Yes | Same |
| Medium Formulation | Hank’s Balanced Salt Solution
Bovine Serum Albumin
L-cysteine
Gelatin
Sucrose
L-glutamic acid
HEPES buffer
Vancomycin
Amphotericin B
Colistin
Phenol red | Same |
| pH | 7.3 ± 0.2 | Same |
| Storage Temperature | 2 – 25°C | Same |
| Glass Beads | 3 x 3 mm | Same |
| Container | Plastic, conical bottom | Same |
| Product Configuration | Media in tube with cap or Media Kit; Media in tube with cap and sterile swab options | Same |
| Shelf Life | 12 months | Same |
| General Device Characteristic Differences | | |
| Fill Volume | 3 mL | 1.5 mL; 3 mL; or 10 mL |
| Swab Tip | Flocked Nylon (Specimen Collection and NP Swabs)
Reticulated Polyurethane Foam (Mid-turbinate Foam Swab) | Polyester |
| Swab Shaft | Polypropylene (Specimen Collection, | Plastic; Stainless Steel - Plastic |
{5}
| | NP, and Foam Swabs) | |
| --- | --- | --- |
| Supported Strains | Viruses:
Influenza A
Adenovirus
Respiratory Syncytial Virus (RSV) | Viruses:
Adenovirus
Cytomegalovirus (CMV)
Echovirus Type 30 (Echo 30)
Herpes Simplex Virus Type 1 (HSV1)
HSV2
Influenza A
Parainfluenza Type 3
Respiratory Syncytial Virus (RSV)
Varicella Zoster Virus (VZV)
Chlamydiae:
Chlamydia pneumoniae Strain CM-1
Chlamydia trachomatis Type 1 Strain UW-12/UR
Mycoplasma:
Mycoplasma hominis
Mycoplasma pneumoniae
Ureaplasma:
Ureaplasma urealyticum |
| Sample Stability | 48 hrs. at 2-8°C or 20-25°C | 48 hrs. at 4°C or 20-25°C |
VI Standards/Guidance Documents Referenced:
Not Applicable.
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Not Applicable.
K202531 - Page 6 of 10
{6}
2. Linearity:
Not Applicable.
3. Analytical Specificity/Interference:
Not Applicable.
4. Assay Reportable Range:
Not Applicable.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Shelf-life:
The shelf-life of the Azer Scientific Universal Transport Medium was determined to be 12 months from the date of manufacture when stored at 2-25°C. Three lots of Azer Scientific Universal Transport Medium were assessed qualitatively at each time point for functionality and physical characteristics using real time ageing studies. In the real time study, media lots were held at 2-8°C, and 20-25°C for 12 months (T=0, 3, 6, 9, and 12 months).
a) Appearance
To evaluate appearance, the different lots of the Azer Scientific Universal Transport Medium were visually examined. The appearance of the product was observed to have no significant color change (pink, transparent), precipitate, particulates, or cloudy appearance. All lots tested at each time point passed the criteria for appearance when held at 2-8°C, and 20-25°C.
b) pH Stability
The pH of the media was used as one of the indicators to support product stability. For all the tubes at each time point and each lot, the pH was within the targeted range of 7.3±0.2 when held at 2-8°C and 20-25°C only.
c) Microbial contamination check
Microbial contamination was determined by incubating an appropriate number of Azer Scientific Universal Transport Medium lots overnight at 37±2°C and examined for growth contaminants. All lots tested at each time point and each temperature passed the criteria for no contamination (no growth) when held at 2-8°C, and 20-25°C.
Sterility:
The Azer Scientific Universal Transport Medium is not claimed to be sterile nor is it intended to be sterilized by the end user. To decrease the chances of contamination, components of the Azer Scientific Universal Transport Medium are either sterilized by moist heat sterilization or filter sterilization and the transport medium is filled aseptically under controlled conditions. The swabs provided with the Azer Scientific Universal Transport Medium are individually packaged and are sold as sterile.
K202531 - Page 7 of 10
{7}
# 6. Detection Limit:
# Performance Testing - Viral Recovery:
Performance of the Azer Scientific Universal Transport Medium was evaluated by performing culture-based recovery studies. The viral recovery study was conducted by spiking virus stock Influenza A/Puerto Rico-8-9VMC2/1934 (H1N1), Human Adenovirus 10 JJ (AV), and Human Respiratory Syncytial Virus, Strain A1998/3-2 (HRSV) into pooled, human nasopharyngeal swab sample matrix tested as negative using an FDA cleared RT-PCR assay.
Performance evaluation was carried out using three (3) lots of test kits each including a PurFlock Ultra $6^{\circ}$ Sterile Revised Ultrafine Flock Swab w/Polystyrene Handle, $80~\mathrm{mm}$ Breakpoint swab provided with Azer Scientific Universal Transport Medium with serial dilutions of each virus $(10^{-1}, 10^{-2}, 10^{-3}$ and $10^{-4})$ spiked into negative clinical matrix in triplicate. Contrived samples were held at both $2 - 8^{\circ}\mathrm{C}$ and $20 - 25^{\circ}\mathrm{C}$ for 0, 24, and 48 hours in a 96-well plate.
Samples in the 96-well plate were examined under the microscope for cytopathic effect (CPE) after the appropriate incubation period for each virus time point. Viral titers of samples collected at 0, 24, and 48 hours were calculated for recovery and determined by percent log reduction of $\mathrm{TCID}_{50} / \mathrm{mL}$ using the Reed-Muench method. Host cell lines used for the different viruses are outlined in Table 1. Viral Recovery study data is outlined in Table 2. Any change of virus titer that was within one log (+/-90%) from the baseline (time point 0) was considered acceptable.
Table 1: Host Cells
| Organism | Strain ID | Host Cell/Growth Medium |
| --- | --- | --- |
| Influenza A/Puerto Rico-8-9VMC2/1934 (H1N1) | NR-29027 | MDCK cells/EMEM + 10% FBS |
| Human Adenovirus 10 JJ (AV) | NR-53907 | A549 cells/EMEM + 10% FBS |
| Human Respiratory Syncytial Virus, Strain A1998/3-2 (HRSV) | NR-28529 | HEP-2 cells/EMEM + 10% FBS |
Table 2: Viral Recovery
| Viral Strain | Duration (hrs.) | 2-8°C | | 20-25°C | |
| --- | --- | --- | --- | --- | --- |
| | | Average TCID50/mL | Average Percent Change (- is a reduction) | Average TCID50/mL | Average Percent Change (- is a reduction) |
| Influenza A/Puerto Rico-8-9VMC2/1934 (H1N1) | 0 | 4.74 x 103 | N/A | 4.74 x 103 | N/A |
| | 24 | 11.03 x 103 | 133%* | 8.76 x 103 | 85% |
| | 48 | 8.38 x 103 | 77% | 6.69 x 103 | 41% |
| Adenovirus 10 JJ (AV) | 0 | 22.43 x 103 | N/A | 22.43 x 103 | N/A |
| | 24 | 14.30 x 103 | -36% | 39.47 x 103 | 76% |
| | 48 | 9.53 x 103 | -58% | 13.94 x 103 | -38% |
| Human Respiratory | 0 | 16.48 x 103 | N/A | 16.48 x 103 | N/A |
| | 24 | 10.01 x 103 | -39% | 15.50 x 103 | -6% |
K202531 - Page 8 of 10
{8}
| Viral Strain
Syncytial Virus (HRSV) | Duration (hrs.) | 2-8°C | | 20-25°C | |
| --- | --- | --- | --- | --- | --- |
| | | Average TCID_{50}/mL | Average Percent Change (- is a reduction) | Average TCID_{50}/mL | Average Percent Change (- is a reduction) |
| | 48 | 16.47 x 10^{3} | -0% | 12.63 x 10^{3} | -23% |
*Considered acceptable because subsequent timepoint, i.e., 48 h time points showed ≤ 90% increase.
## Conclusion of the culture-based viral recovery study:
The Azer Scientific Universal Transport Medium demonstrated the recovery of H1N1 Influenza BEI NR-29027, Adenovirus 10 JJ, and Human Respiratory Syncytial Virus (HRSV) in all replicates at tested incubation times and storage conditions met the acceptance criteria. This data supports the transportation of Influenza A, Adenovirus, and Respiratory Syncytial Virus (RSV) in Azer Scientific Universal Transport Medium when stored refrigerated (2-8°C) or room temperature (20-25°C) for up to 48 hours.
7. **Assay Cut-Off:**
Not Applicable.
## B Comparison Studies:
1. **Method Comparison with Predicate Device:**
Not Applicable.
2. **Matrix Comparison:**
Not Applicable.
## C Clinical Studies:
1. **Clinical Sensitivity:**
Not Applicable.
2. **Clinical Specificity:**
Not Applicable.
3. **Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):**
Not Applicable.
## D Clinical Cut-Off:
Not Applicable.
K202531 - Page 9 of 10
{9}
E Expected Values/Reference Range:
Not Applicable.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K202531 - Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.