COPAN UNIVERSAL TRANSPORT MEDIUM (UTM-RT) SYSTEM

K042970 · Copan Diagnostics, Inc. · JSM · Dec 7, 2004 · Microbiology

Device Facts

Record IDK042970
Device NameCOPAN UNIVERSAL TRANSPORT MEDIUM (UTM-RT) SYSTEM
ApplicantCopan Diagnostics, Inc.
Product CodeJSM · Microbiology
Decision DateDec 7, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.2390
Device ClassClass 1

Indications for Use

Copan Universal Transport Medium (UTM-RT) System is intended for the collection and transport of clinical specimens containing viruses, chlamydiae, mycoplasma or ureaplasma from the collection site to the testing laboratory. UTM-RT can be processed using standard clinical laboratory operating procedures for viral, chlamydial, mycoplasma and ureaplasma culture.

Device Story

System consists of room-temperature stable transport medium in screw-cap tubes, optionally packaged with sterile specimen collection swabs. Medium contains protein for stabilization, antibiotics to inhibit bacterial/fungal growth, and buffer for neutral pH. Used by clinicians to collect patient specimens; transported to laboratory for standard culture processing. Maintains viability and infectivity of viruses, chlamydiae, mycoplasma, and ureaplasma during transit. Facilitates diagnostic testing by preserving sample integrity.

Clinical Evidence

Bench testing only. Recovery studies compared Copan UTM-RT to a comparative product to assess viability maintenance of viruses, chlamydiae, mycoplasma, and ureaplasma during storage. Stability testing confirmed a 12-month expiration date.

Technological Characteristics

Single-use transport medium system. Components: screw-cap tube with liquid medium (protein, antibiotics, buffer) and optional sterile swabs. Room-temperature stable. No electronic or software components.

Indications for Use

Indicated for the collection and transport of clinical specimens containing viruses, chlamydiae, mycoplasma, or ureaplasma from the collection site to the laboratory for culture.

Regulatory Classification

Identification

A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K042970 DEC - 7 2004 ## 510(k) Summary ### for ## Copan Universal Transport Medium (UTM-RT) System #### 1. SPONSOR Copan Diagnostics Inc. 2175 Sampson Avenue, Suite 124 Corona, CA 92879 Contact Person: Norman Sharples Telephone: 800-216-4016 Date Prepared: October 27, 2004 #### 2. DEVICE NAME Proprietary Name: Copan Universal Transport Medium (UTM-RT) System Common/Usual Name: Transport Culture Medium Devices Classification Name: Transport Culture Medium Devices #### 3. PREDICATE DEVICES - . Multi-Microbe Collection & Transport System, M4 Medium MicroTest Inc. K910526 - MicroTest™ Multi-Microbe Collection & Transport System, M4 Medium . REMEL - Apogent [Covered by K910526 due to acquisition of MicroTest Inc. by REMEL-Apogent] #### 4. DEVICE DESCRIPTION The Copan UTM-RT System includes a universal transport medium that is room temperature stable and that can sustain viability (and infectivity) of a plurality of organisms that include clinically important viruses, chlamydiae, mycoplasma and ureaplasma during transit to the testing laboratory. The formulation of UTM-RT medium includes protein for stabilization, antibiotics to minimize bacterial and fungal contamination, and a buffer to maintain a neutral pH. {1}------------------------------------------------ Copan UTM-RT System medium is provided in screw-cap tubes designed for transport of the clinical sample. Copan UTM-RT System is also supplied as a sample collection kit that comprises a package which contains one screw-cap tube of UTM-RT medium and a peel pouch incorporating one or two sterile specimen collection swabs. A range of UTM-RT sample collection kits are available which incorporate different types of shaft swabs which facilitate the collection of specimens from different sites of the patient. #### ડ. INTENDED USE Copan Universal Transport Medium (UTM-RT) System is intended for the collection and transport of clinical specimens containing viruses, chlamydiae, mycoplasma or ureaplasma from the collection site to the testing laboratory. UTM-RT can be processed using standard clinical laboratory operating procedures for viral, chlamydial, mycoplasma and ureaplasma culture. #### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE 6. The Copan Universal Transport Medium (UTM-RT) System products are substantially equivalent to the predicate transport medium devices. The Copan UTM-RT System product and the predicate devices are similar in design, intended use, and overall function. The Copan and predicate devices are single-use products intended for the collection and transport of clinical specimens containing viruses, chlamydiae, mycoplasma or ureaplasma. Both the Copan and predicate devices are offered in product configurations with medium supplied alone or in kit formats with medium and specimen collection swab options. #### 7. PERFORMANCE TESTING Studies were conducted to evaluate the performance characteristics of the Copan Universal Transport Medium (UTM-RT) System. Recovery studies were performed using the Copan UTM-RT System and a comparative product to determine the ability of the products to maintain viability of various strains of viruses, chlamydiae, mycoplasma and ureaplasma during storage and use. Stability testing was performed on aged Copan UTM-RT products to support the 12-month expiration date. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 DEC - 7 2004 Copan Diagnostics, Inc. c/o Cynthia A. Sinclair, RAC Senior Staff Consultant Medical Device Consultants, Inc. 49 Plain Street North Attleboro, MA 02760 Re: k042970 Trade/Device Name: Copan Universal Transport Medium (UTM-RT) System Regulation Number: 21 CFR 866.2390 Regulation Name: Transport Culture Medium Regulatory Class: Class I Product Code: JSM Dated: October 27, 2004 Received: October 28, 2004 Dear Ms. Sinclair: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ ### Page 2 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Salazar Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Indications for Use ### K042970 510(k) Number (if known): Device Name: Copan Universal Transport Medium (UTM-RT) System Copan Universal Transport Medium (UTM-RT) System is intended for the collection and Copan Onliversal Transport Neemaning viruses, chlamydiae, myoplasma or transport of cinnear specificals venus the testing laboratory. UTM-RT can be processed ureaplasma from the Il laboratory operating procedures for viral, chlamydial, mycoplasma and ureaplasma culture. Prescription Use __ X (Part 21 CFR 801 Subpart D) AND/OR Over-the-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Las wth Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety k042970 510(k) -
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...