K153364 · Kerecis Limited · OXE · Aug 19, 2016 · General, Plastic Surgery
Device Facts
Record ID
K153364
Device Name
Kerecis SecureMesh
Applicant
Kerecis Limited
Product Code
OXE · General, Plastic Surgery
Decision Date
Aug 19, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Kerecis SecureMesh is intended for use as a prosthesis when staple line reinforcement is needed in surgical repair of soft tissue deficiencies using surgical staplers. Kerecis SecureMesh can be used for reinforcement of staple lines during: - Bariatric surgical procedures - Colorectal and Colon surgical procedures - Gastric, small bowels and mesentery procedures - Lung and bronchus resections
Device Story
Kerecis SecureMesh is an acellular, lyophilized fish skin (Atlantic Cod) sheet used as a staple line buttress. Supplied in sterile, double-pouch packaging with a delivery system; healthcare professionals load the buttress onto the jaws of a surgical stapler in the operating theatre. During stapling, the mesh reinforces the staple line to prevent leaks and tissue damage. The device is non-cross-linked and absorbable. It functions as a mechanical reinforcement for soft tissue deficiencies. Benefits include improved staple line integrity and reduced leakage risk. Used by surgeons in clinical settings.
Clinical Evidence
No clinical testing was included in this submission. Evidence consists of bench testing (biocompatibility, tensile strength, usability, leak pressure, seal burst) and an acute porcine ex-vivo study comparing the subject device to Peri-Strips Dry Staple Line Reinforcement (K040415).
Technological Characteristics
Material: Decellularized, lyophilized Atlantic Cod fish skin. Form factor: Sheet-based buttress with delivery system. Energy source: None (mechanical). Connectivity: None. Sterilization: Ethylene Oxide (EtO). Software: None.
Indications for Use
Indicated for patients undergoing bariatric, colorectal, colon, gastric, small bowel, mesentery, lung, or bronchus surgical procedures requiring staple line reinforcement during soft tissue repair.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Peri-Strips Dry Staple Line Reinforcement (K040415)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name encircling a stylized image of three human profiles facing right. The profiles are stacked on top of each other, creating a sense of depth and unity. The overall design is simple and professional, reflecting the department's mission of promoting health and well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 19, 2016
Kerecis Limited Gudmundur Sigurjonsson CEO Eyrargata 2 Isafjordur, 400 Iceland
Re: K153364
Trade/Device Name: Kerecis SecureMesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: OXE Dated: July 15, 2016 Received: July 21, 2016
Dear Gudmundur Sigurjonsson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K153364
Device Name
Kerecis SecureMesh
Indications for Use (Describe)
Kerecis SecureMesh is intended for use as a prosthesis when staple line reinforcement is needed in surgical repair of soft tissue deficiencies using surgical staplers.
Kerecis SecureMesh can be used for reinforcement of staple lines during:
- Bariatric surgical procedures
- Colorectal and Colon surgical procedures
- Gastric, small bowels and mesentery procedures
- Lung and bronchus resections
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510k Summary
#### SUBMITTER/510K HOLDER 1
| Address: | Kerecis Limited<br>Eyrargata 2<br>400 Isafjordur<br>Iceland |
|-----------------|-------------------------------------------------------------|
| Contact Person: | Gudmundur Fertram Sigurjonsson<br>Executive Chairman |
| Telephone: | 011 354 562 2601 |
| Date Prepared: | November 9th, 2015 |
#### 2 DEVICE NAME
| Proprietary Name: | Kerecis SecureMes |
|------------------------------|-------------------|
| Common/Usual Name: | Surgical Mesh |
| Classification Name: | Mesh, Surgical |
| Classification Product Code: | OXE |
| Regulatory Number: | 878.3300 |
#### 3 PREDICATE DEVICES
- Veritas Dry Collagen Matrix (K041669)- PrimaryPredicate
- . MariGen Wound (K132343) - Supporting Predicate
#### DEVICE DESCRIPTION বা
Subject Device is formed of intact acellular fish skin and is intended to be used as a staple line buttress with surgical staplers. The buttress material is supplied in a double pouch sterile packaging (EtO), where a pair of buttresses is provided on a pair of carriers to be used by a healthcare professional, to transfer the buttresses to the opposed jaws of a surgical stapler in the operating theatre prior to a stapling procedure. One pair of carrier and a buttress is used for each stapler firing.
{4}------------------------------------------------
#### INTENDED USE ഗ
Subject Device is intended for use as a prosthesis when staple line reinforcement is needed in surgical repair of soft tissue deficiencies using surgical staplers.
Subject Device can be used for reinforcement of staple lines during:
- . Bariatric surgical procedures
- . Colo-rectal and colon surgical procedures
- . Gastric, small bowels and mesentery procedures
- Lung, and bronchus resections
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE 6 PREDICATE DEVICES
Subject Device is substantially equivalent to the Primary Predicate, where the intended use and indication list of the Subject Device falls within the intended use and indication list of the Primary Predicate.
The Subject Device, and the Supporting Predicate, both are fish skin sheets that have been decellularized and lyophilized. Both devices consist of a terminally sterilized sheet of the provided in prescribed size configuration and thicknesses. Both are packaged in a pouch configuration.
| 510(k) | Subject Device K153364 | Primary Predicate<br>(K041669) |
|---------------|--------------------------------------------------|--------------------------------------------------|
| Device name | Kerecis SecureMesh | Veritas Dry Collagen Matrix |
| Product Codes | OXE | FTM |
| Intended use | Intended to be used as a staple<br>line buttress | Intended to be used as a staple<br>line buttress |
| Indications | Reinforcement of staple lines<br>during | Reinforcement of staple lines<br>during |
| 1 | Bariatric surgical procedures | Bariatric surgical procedures |
{5}------------------------------------------------
| 2 | Gastric, small bowels and<br>mesentery procedures | Gastric, small bowel, mesentery |
|---------------|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| 3 | Colorectal and colon procedures | Colon and colorectal procedures |
| 4 | Lung, and bronchus resections | Lung and bronchus resections |
| 5 | Not claimed | Occlusion of the left atrial<br>appendage during open chest<br>procedures during cardiac |
| Animal Origin | Atlantic Cod Fish | Bovine |
| Tissue type | Skin | Pericardium |
| Presentation | A delivery system with an<br>absorbable buttress<br>(dehydrated, non-cross- linked | A delivery system with a<br>buttress (comprised of<br>dehydrated, non-cross- linked |
Table 1 Comparing the predicate device and subject device Intended Use and Indication for Use. The the underlying functional design and principles of operation
#### PERFORMANCE 7
## 7.1 BENCH TESTING
The Supporting Predicate, MariGen Wound (K132343) has undergone appropriate biocompatibility testing and viral-inactivation evaluation. The information provided demonstrate that the device is biocompatible and that viral risks are not present. The Subject Device, (K153364) and the Supporting Predicate, MariGen Wound (K132343) utilize the same raw materials (fish skin), processing components/solutions and manufacturing processes, and are identical. Therefore, no further biocompatibility testing or anti-viral considerations were made.
The Subject Device is substantially equivalent to the Primary Predicate in terms of physical characteristics and tests results. The following tests have been performed to substantiate the equivalence.
| 510(k) | Subject Device K153364 |
|------------------|------------------------------------------------------------------------------------------------------------|
| Device name | Kerecis SecureMesh |
| Tensile strength | A Uniaxial strength test was performed to compare the tensile strength of the<br>product and the predicate |
{6}------------------------------------------------
| Usability | Loading time, Transfer of buttress to stapler, Ease of Use, Force to fire and<br>Slippage |
|-------------|-------------------------------------------------------------------------------------------|
| Performance | Leak at Staple line, Leak Pressure and Seal burst |
Table 2 - Subject Device Bench testing
### 7.2 ANIMAL STUDIES
An acute porcine ex-vivo study was performed on the Subject Device and Peri-Strips Dry Staple Line Reinforcement (K040415) to support substantial equivalence. Peri-Strips Dry Staple Line Reinforcement (K040415) is the Predicate Device for the Veritas Dry Collagen Matrix (K041669) and the two materials are identical, from the material source, with the same treatment indications as well as mechanical and clinical properties. Comparison with un-buttressed stapling was also undertaken (negative control).
There was no significant difference between the Subject Device and the Primary Predicate. The Subject Device technological and physical characteristics are equivalent to the Predicate Device.
## 7.3 CLINICAL STUDIES
No clinical testing was included in this submission
#### CONCLUSION 8
The subject device, Kereis SecureMesh, meets the requirements to perform its intended use as a staple line sealer and is substantially equivalent to the legally marketed predicate device, Veritas Dry Collagen Matrix (K041669).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.