K202414 · Hyperfine Research, Inc. · LLZ · Jan 7, 2021 · Radiology
Device Facts
Record ID
K202414
Device Name
BrainInsight
Applicant
Hyperfine Research, Inc.
Product Code
LLZ · Radiology
Decision Date
Jan 7, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Brain structure segmentation
Machine learning tools
—
—
—
—
—
—
Midline shift measurement
Machine learning tools
—
—
—
—
—
—
Indications for Use
BrainInsight is intended for automatic labeling, spatial measurement, and volumetric quantification of brain structures from a set of low-field MR images and returns annotated and segmented images, color overlays, and reports.
Device Story
BrainInsight is automated MR imaging post-processing software for low-field (64mT) structural MRI. It accepts 3D T1 and T2-weighted DICOM images as input. The device uses a proprietary machine learning-based pipeline to perform whole brain segmentation, ventricle segmentation, and midline shift measurements. It includes automated quality control functions, such as tissue contrast checks and scan protocol verification. The system operates in a serverless cloud environment and returns annotated images, color overlays, and morphometric reports in DICOM format. These outputs are displayed on third-party workstations or PACS for review by a trained physician. The device supports clinicians in the assessment of structural MRIs by providing quantitative data and visualizations, potentially aiding in clinical decision-making for adult patients.
Clinical Evidence
No clinical data was required to demonstrate substantial equivalence. Evidence consists of bench testing and software evaluations confirming cybersecurity, PHI protection, midline shift measurement accuracy, 3D coordinate alignment, and segmentation performance.
Technological Characteristics
Software-based image processing system for 64mT MRI. Operates in a serverless cloud environment. Inputs/outputs via DICOM. Features automated segmentation, volume calculation, and distance measurement using machine learning. Includes automated quality control (tissue contrast, scan protocol, atlas alignment).
Indications for Use
Indicated for automatic labeling, spatial measurement, and volumetric quantification of brain structures from low-field (64mT) MR images in patients aged 18 or older.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
January 7, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Hyperfine Research, Inc. % Robert W. Fasciano, Ph.D. Head of Quality Assurance & Regulatory Affairs 530 Old Whitfield Street GUILFORD CT 06437
Re: K202414
Trade/Device Name: BrainInsight Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: December 4, 2020 Received: December 7, 2020
Dear Dr. Fasciano:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202414
Device Name BrainInsight
Indications for Use (Describe)
BrainInsight is intended for automatic labeling, spatial measurement, and volumetric quantification of brain structures from a set of low-field MR images and returns annotated images, color overlays, and reports.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Page 1 of 4 Section 5 – 510(k) Summary
K202414
| 510(k) Summary<br>(As required by 21 CFR 807.92) | | |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| Date Summary Prepared: | August 21, 2020 | |
| Company Name:<br>As required by 807.92(a)(1) | Hyperfine Research, Inc. | |
| Robert W. Fasciano, PhD | | |
| Head of Quality Assurance & Regulatory Affairs | | |
| | 530 Old Whitfield St. | |
| | Guilford, CT 06437 | |
| | (617) 435-9098 | |
| | rfasciano@hyperfine-research.com | |
| Device Name:<br>As required by 807.92(a)(2) | Device/Trade Name: | BrainInsight |
| | Device Common Name: | System, Image Processing,<br>Radiological |
| | Regulation Number: | 21 CFR 892.2050 |
| | Regulation Name: | System, Image Processing,<br>Radiological |
| | Regulation Description: | Picture archive and<br>communications system |
| | Class: | II |
| | Product Code: | LLZ |
| Predicate Device(s):<br>As required by 807.92(a)(3) | NeuroQuant, K170981 | |
| Device Description:<br>As required by 807.92(a)(4) | BrainInsight is a fully automated MR imaging post-<br>processing medical software that image alignment, whole<br>brain segmentation, ventricle segmentation, and midline<br>shift measurements of brain structures from a set of MR<br>images from patients aged 18 or older. The output annotated<br>and segmented images are provided in a standard image<br>format using segmented color overlays and reports that can<br>be displayed on third-party workstations and FDA cleared<br>Picture Archive and Communications Systems (PACS). The<br>high throughput capability makes the software suitable for<br>use in routine patient care as a support tool for clinicians ir<br>assessment of low-field (64mT) structural MRIs.<br>BrainInsight provides overlays and reports based on 64mT<br>3D MRI series of a T1 and T2-weighted sequence. The<br>outputs of the software are DICOM images which include | |
{4}------------------------------------------------
volumes that have been annotated with color overlays, with each color representing a particular segmented region, spatial measurement of anatomical structures, and information reports computed from the image data, segmentations, and measurements. The BrainInsight processing architecture includes a proprietary automated internal pipeline that performs whole brain segmentation, ventricle segmentation, and midline shift measurements based on machine learning tools. Additionally, the system's automated safety measures include automated quality control functions, such as tissue contrast check and scan protocol verification. The system is installed on a standard computing platform, e.g. server that may be in the cloud, and is designed to support file transfer for input and output of results.
Statement of Intended Use: BrainInsight is intended for automatic labeling, spatial As required by 807.92(a)(5) measurement, and volumetric quantification of brain structures from a set of low-field MR images and returns annotated and segmented images, color overlays, and reports.
| Device | Proposed Device | Predicate Device |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | BrainInsight | NeuroQuant, K170981 |
| Classification | Class II, LLZ, 21 CFR 892.2050 | Class II, LLZ, 21 CFR 892.2050 |
| Intended Use | Automatic labeling, spatial<br>measurement, and volumetric<br>quantification of brain structures<br>from a set of low-field MR<br>images and returns annotated<br>and segmented images, color<br>overlays, and reports. | Automatic labeling, visualization<br>and volumetric quantification of<br>segmentable brain structures and<br>lesions from a set of MR images.<br>Volumetric data may be<br>compared to reference percentile<br>data |
| Target Anatomical Sites | Brain | Brain |
| Technology | • Automated measurement of<br>brain tissue volumes and<br>structures<br>• Automatic segmentation and<br>quantification of brain structures<br>using machine learning | • Automated measurement of<br>brain tissue volumes and<br>structures and lesions<br>• Automatic segmentation and<br>quantification of brain structures<br>using a dynamic probabilistic<br>neuroanatomical atlas, with age<br>and gender specificity, based on<br>the MR image intensity |
| Method of Use | MR images are automatically<br>sent to BrainInsight and<br>processed images are<br>automatically returned in<br>approximately 7 minutes. | User manually sends MR images<br>to NeuroQuant and processed<br>images are automatically<br>returned in approximately 7<br>minutes. |
| Device | Proposed Device<br>BrainInsight | Predicate Device<br>NeuroQuant, K170981 |
| User Interface / Physical<br>Characteristics | • No software required<br>• Operates in a serverless cloud environment<br>• User interface through PACS<br>(multiple vendors) | • Software package installed on<br>User hardware<br>• Operates on off-the-shelf<br>hardware (multiple vendors)<br>• User interface through the<br>software package |
| Operating System | Supports Linux | Supports Linux, Mac OS X and<br>Windows |
| Processing Architecture | Automated internal pipeline that<br>performs:<br>- segmentation<br>- volume calculation<br>- distance measurement<br>- numerical information display | Automated internal pipeline that<br>performs:<br>- artifact correction<br>- segmentation<br>- lesion quantification<br>- volume calculation<br>- report generation |
| Data Source | • MRI scanner: Hyperfine FSE<br>MRI scans acquired with<br>specified protocols<br>• Supports DICOM format as<br>input | • MRI scanner: 3D T1 MRI<br>scans acquired with specified<br>protocols<br>• NeuroQuant Supports DICOM<br>format as input |
| Output | Provides volumetric<br>measurements of brain structures<br>• Includes segmented color<br>overlays and morphometric<br>reports<br>• Supports DICOM format as<br>output of results that can be<br>displayed on DICOM<br>workstations and Picture<br>Archive and Communications<br>Systems | Provides volumetric<br>measurements of brain structures<br>and lesions<br>• Includes segmented color<br>overlays and morphometric<br>reports<br>• Automatically compares results<br>to reference percentile data and<br>to prior scans when available<br>• Supports DICOM format as<br>output of results that can be<br>displayed on DICOM<br>workstations and Picture<br>Archive and Communications<br>Systems |
| Safety | Automated quality control<br>functions<br>• Tissue contrast check<br>• Scan protocol verification<br>• Atlas alignment check<br>• Results must be reviewed by a<br>trained physician | Automated quality control<br>functions<br>• Tissue contrast check<br>• Scan protocol verification<br>• Atlas alignment check<br>• Results must be reviewed by a<br>trained physician |
## Comparison of Technological Characteristics with Predicate Devices: As required by 807.92(a)(6)
{5}------------------------------------------------
# Comparison of Technological Characteristics with Predicate Devices: As required by 807.92(a)(6)
Performance data was limited to software evaluations to Non-clinical Performance Data: As required by 807.92(b)(1) confirm:
- Cybersecurity and PHI protection •
- . Midline shift
- 3D Coordinates and alignment •
- . Segmentation
{6}------------------------------------------------
| Data Quality Control Audit trail User Manual information Software control Ventricle segmentation Midline shift measurement Skull stripping |
|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Assessment of Clinical Data: No clinical data was required to demonstrate substantial<br>As required by 807.92(b)(2) equivalence. |
| Overall Conclusions: Based on the indications for use, technologic<br>As required by 807.92(b)(3) characteristics, and comparison to predicate device |
ical ice, BrainInsight has been shown to be substantially equivalent to the predicate and is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.