K202288 · SNJ Co., Ltd. · GEX · Nov 19, 2021 · General, Plastic Surgery
Device Facts
Record ID
K202288
Device Name
Finebeam
Applicant
SNJ Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 19, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Finebeam Dual is a surgical laser instrument intended to provide laser energy in two different operation modes (Qswitched (QSW) and long pulse (LPS)) with 1064nm or 532 nm wavelength per use in a variety of dermatological procedures. Each function is intended for: 1) QSW mode with 1064nm single or 1064nm photoacoustic double pulse (PDP) function - Removal dark ink (black, blue, and brown) tattoos; - Treatment melasma, treatment nevus of Ota; - Treatment common nevi; - Removal of lightening of unwanted hair; and - Skin resurfacing procedures for the treatment of acne scars and wrinkles 2) QSW mode with 532nm single function - Removal of light ink (red, tan, purple, and orange) tattoos; - Treatment of vascular lesions including, but not limited to port-wine stains, spider angioma, cherry angioma; - Treatment of pigmented lesions including, but not limited to cafe-au-Lait birthmarks, solar lentigines, Becker's nevi, freckles, common nevi, and nevus spilus; - Treatment of seborrheic keratosis: - Treatment of post-inflammatory hyperpigmentation; and - Skin resurfacing procedures for the treatment of acne scars and wrinkles in the QSW mode function; 3) QSW mode with 1064nm Non-Q function - Removal of unwanted hair, for stable long term or permanent hair reduction and treatment of Pseudofolliculitis Barbae (PFB). The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin; - Photocoagulation and hemostasis of pigmented and vascular lesions, such as but not limited to port wine stains, cherry angioma, Hemangiomas, warts, telangiectasias, rosacea, leg veins, Coagulation and hemostasis of soft tissue; and - Treatment of wrinkles 4) LPS mode - Removal of unwanted hair, for stable long term or permanent hair reduction and treatment of PFB. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin; - Photocoagulation and hemostasis of pigmented and vascular lesions, such as but not limited to port wine stains, hemangiomas, warts, telangiectasias, rosacea, leg veins, and spider veins; - Coagulation and hemostasis of soft tissue; - Treatment of wrinkles and acne vulgaris
Device Story
Finebeam Dual is a Nd:YAG laser system for dermatological surgery; utilizes 1064nm and 532nm wavelengths. Operates in Q-switched (QSW) and long pulse (LPS) modes. QSW mode includes single pulse and photoacoustic double pulse (PDP) functions for tattoo removal, pigmented/vascular lesion treatment, and skin resurfacing. LPS mode and 1064nm Non-Q function used for hair reduction, PFB treatment, and soft tissue coagulation/hemostasis. Device features articulated arm beam delivery, LCD touch screen interface, and internal closed-loop water-air heat exchanger for thermal control. Operated by physicians in clinical settings. Microprocessor-controlled system regulates parameters and ensures safe operation. Output energy and pulse duration are adjustable based on procedure. Benefits include precise tissue interaction for various dermatological conditions.
Clinical Evidence
Bench testing only. Compliance verified against IEC 60601-1, IEC 60601-1-6, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1. No clinical data provided.
Indicated for patients of all skin types (Fitzpatrick I-VI, including tanned skin) requiring dermatological laser procedures including tattoo removal (dark/light ink), treatment of pigmented/vascular lesions (e.g., melasma, nevus of Ota, port-wine stains, angiomas, lentigines), hair reduction, treatment of Pseudofolliculitis Barbae (PFB), skin resurfacing (acne scars/wrinkles), and soft tissue coagulation/hemostasis.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Fotona QX Nd:YAG/KTP Laser System Family (K083889)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
November 19, 2021
SNJ Co., Ltd. % Jongrak Kim Director Med.com 2202 Gongwon-ro, Guro-gu Seoul, Seoul Teugbyeolsi 08295 Korea, South
Re: K202288
Trade/Device Name: Finebeam Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX, Dated: October 14, 2021 Received: October 18, 2021
Dear Jongrak Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K202288
Device Name Finebeam Dual
### Indications for Use (Describe)
The Finebeam Dual is a surgical laser instrument intended to provide laser energy in two different operation modes (Qswitched (QSW) and long pulse (LPS)) with 1064nm or 532 nm wavelength per use in a variety of dermatological procedures. Each function is intended for:
- 1) QSW mode with 1064nm single or 1064nm photoacoustic double pulse (PDP) function
- Removal dark ink (black, blue, and brown) tattoos;
- Treatment melasma, treatment nevus of Ota;
- Treatment common nevi;
- Removal of lightening of unwanted hair; and
- Skin resurfacing procedures for the treatment of acne scars and wrinkles
- 2) QSW mode with 532nm single function
- Removal of light ink (red, tan, purple, and orange) tattoos;
- Treatment of vascular lesions including, but not limited to port-wine stains, spider angioma, cherry angioma;
- Treatment of pigmented lesions including, but not limited to cafe-au-Lait birthmarks, solar lentigines, Becker's nevi, freckles, common nevi, and nevus spilus;
- Treatment of seborrheic keratosis:
- Treatment of post-inflammatory hyperpigmentation; and
- Skin resurfacing procedures for the treatment of acne scars and wrinkles in the QSW mode function; 3) QSW mode with 1064nm Non-Q function
- Removal of unwanted hair, for stable long term or permanent hair reduction and treatment of Pseudofolliculitis Barbae (PFB). The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin;
- Photocoagulation and hemostasis of pigmented and vascular lesions, such as but not limited to port wine stains, cherry angioma, Hemangiomas, warts, telangiectasias, rosacea, leg veins, Coagulation and hemostasis of soft tissue; and
- Treatment of wrinkles
4) LPS mode
- Removal of unwanted hair, for stable long term or permanent hair reduction and treatment of PFB. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin;
- Photocoagulation and hemostasis of pigmented and vascular lesions, such as but not limited to port wine stains, hemangiomas, warts, telangiectasias, rosacea, leg veins, and spider veins;
- Coagulation and hemostasis of soft tissue:
- Treatment of wrinkles and acne vulgaris
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
o... Itd.
No. 604, Ace-Techno Tower 2, Digital-ro 31-gil 19, Guro-gu, Seoul, Republic of Korea
### SECTION 05. 510(k) SUMMARY
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92.
- Submitter 1.
| | - | Company Name: | SNJ Co., Ltd. |
|----|---|---------------------------|--------------------------------------------------------------------------------------------|
| | - | Name | Hyang-Kee Lee |
| | - | Address: | No. 604, Ace-Techno Tower 2,<br>Digital-ro 31-gil 19, Guro-gu,<br>Seoul, Republic of Korea |
| | - | Phone Number: | +82-2-866-9005 |
| | - | FDA Registration Number: | 3004633032 |
| 2. | | Application Correspondent | |
| | - | Company Name: | MED.COM |
| | - | Name | Jongrak Kim |
| | - | Address: | 2202 Gongwon-ro, Guro-gu<br>Seoul, Republic of Korea |
| | - | E-mail: | cefdacert@gmail.com |
| 3. | | Device Information | |
| | - | Model Name: | Finebeam Dual |
| | - | Common Name: | Nd:YAG Laser System |
| | - | Device Class: | Class II under 21 CFR CFR 878.4810 |
| | - | Classification Name: | Powered Laser Surgical Instrument |
| | - | Product Code: | GEX |
| 4. | | Primary Predicate Device | |
| | - | Device Name: | Fotona QX Nd:YAG/KTP Laser System Family |
| | - | 510(k) Number: | K083889 |
| | - | Manufacturer: | Fotona d.d |
| | | | |
- Additional Predicate Devices 5.
{5}------------------------------------------------
| - | Device Name: | Hyperion Long Plused Nd:YAG Laser |
|---|----------------|-----------------------------------------|
| - | 510(k) Number: | K132286 |
| - | Manufacturer: | Laseroptek Co. Ltd. |
| - | Device Name: | StarWalker Laser System Family |
| - | 510(k) Number: | K171227 |
| - | Manufacturer: | Fotona d.o.o. |
| - | Device Name: | Long pulse Nd:YAG Laser System |
| - | 510(k) Number: | K140249 |
| - | Manufacturer: | Advanced Technology Laser Company, Ltd |
| - | Device Name: | Pastelle Q-Switched Nd:YAG Laser System |
| - | 510(k) Number: | K123293 |
| - | Manufacturer: | WON Technology Co., Ltd. |
### 6. Device Description
The Finebeam Dual is designed to use neodymium-doped yttrium aluminum garnet (Nd:YAG) to transform and generate high voltage sources for the optical resonator, and the atoms in the source are activated to amplify and emit laser light.
- 7. Indications For Use / Intended Use
The Finebeam Dual is a surgical laser instrument intended to provide laser energy in two different operation modes (Q-switched (QSW) and long pulse (LPS)) with 1064nm or 532 nm wavelength per each function for use in a variety of dermatological procedures. Each function is intended for:
- QSW mode with 1064nm single or 1064nm photoacoustic double pulse (PDP) 1) function
- Removal dark ink (black, blue, and brown) tattoos;
- Treatment melasma, treatment nevus of Ota; ı
- Treatment common nevi; -
- Removal of lightening of unwanted hair; and -
- Skin resurfacing procedures for the treatment of acne scars and wrinkles -
- 2) QSW mode with 532nm single function
{6}------------------------------------------------
### Co.. Ltd.
No. 604, Ace-Techno Tower 2, Digital-ro 31-gil 19, Guro-gu, Seoul, Republic of Korea
- Removal of light ink (red, tan, purple, and orange) tattoos;
- Treatment of vascular lesions including, but not limited to port-wine stains, telangiectasias, spider angioma, cherry angioma;
- i Treatment of pigmented lesions including, but not limited to cafe-au-Lait birthmarks, solar lentigines, senile lentigines, Becker's nevi, freckles, common nevi, and nevus spilus;
- Treatment of seborrheic keratosis; -
- Treatment of post-inflammatory hyperpigmentation; and i
- Skin resurfacing procedures for the treatment of acne scars and wrinkles in the -QSW mode with 532nm single function;
- 3) QSW mode with 1064nm Non-Q function
- Removal of unwanted hair, for stable long term or permanent hair reduction and treatment of Pseudofolliculitis Barbae (PFB). The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin;
- Photocoagulation and hemostasis of pigmented and vascular lesions, such as but not limited to port wine stains, cherry angioma, Hemangiomas, warts, telangiectasias, rosacea, leg veins, and spider veins, Coagulation and hemostasis of soft tissue; and
- Treatment of wrinkles -
- LPS mode 4)
- Removal of unwanted hair, for stable long term or permanent hair reduction and i treatment of PFB. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin;
- Photocoagulation and hemostasis of pigmented and vascular lesions, such as but ı not limited to port wine stains, hemangiomas, warts, telangiectasias, rosacea, leg veins, and spider veins;
- Coagulation and hemostasis of soft tissue; i
- 8. Summary of Technological Characteristics
The Finebeam Dual has the same technological and design characteristics (design, chemical composition, energy source; wavelength, active medium, cooling system, power supply, beam delivery, controls, housing) as the previously cleared devices. The output characteristics are for the intended use the same as those of the predicate devices. All
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for SNJ Co., Ltd. The logo consists of the letters SNJ in a stylized, sans-serif font, with the letters in a light blue color. To the right of the letters is the company name, "SNJ Co., Ltd.", in a bold, black font. Below the company name is the address, "No. 604, Ace-Techno Tower,", also in black.
systems are based on variable pulse duration power supply technology. All lasers utilize class I aiming beams which pose no hazard to the user. All systems are microprocessor controlled devices. The microprocessor control regulates normal operation, permits parameter selection, and avoids hazard incidence. All systems utilize an internal closed loop water-air heat exchanger circuit for optimal thermal control of the laser cavity. The risk and benefits for the Finebeam Dual are identical to the predicate devices when used for similar clinical applications.
A comparison of the technical specifications for the intended use of the Finebeam Dual with the previously cleared devices is provided the table in section 10.
### Non-Clinical Test Conclusion
Nonclinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- IEC 60601-1:2005/A1:2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- · IEC 60601-1-6:2010/A1:2013 Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-2-22:2014 Medical electrical equipment Part 2: Particular requirements for the safety of diagnostic and therapeutic laser equipment
- IEC 60825-1:2014 Safety of laser products Part 1: Equipment classification and requirements
{8}------------------------------------------------
## SNJ Co., Ltd.
No. 604, Ace-Techno Tower 2, Digital-ro 31-gil 19, Guro-gu, Seoul, Republic of Korea
- Comparison of technological characteristics with the predicate device 9.
- 9.1. Intended use comparison of 532nm single and 1064nm single function in the Q-switched mode
| Device Name | Fotona QX Nd:YAG/KTP Laser System Family<br>(Predicate) | Finebeam (Proposed) |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | GEX | Same |
| Regulation Number | 21 CFR 878.4810 | Same |
| K number | K083889 | K202288 |
| Applicant | Fotona d.o.o. | SNJ Co., Ltd. |
| Wavelength | 1064nm and 532nm | Same |
| Intended use | 1064 nm wavelength in Q-switched mode:<br>- Removal of dark ink (black, blue and brown)<br>tattoos<br>- Treatment melasma<br>- Treatment of nevus of Ota<br>- Treatment of common nevi<br>- Removal or lightening of unwanted hair<br>- Skin resurfacing procedures for the treatment<br>of acne scars and wrinkles<br><br>532 nm wavelength in Q-switched mode (nominal<br>delivered energy of 585 nm and 650 nm with the<br>optional 585 nm and 650 nm dye converter<br>handpieces):<br>- Removal of light ink (red, sky blue, green, tan,<br>purple, and orange) tattoos | With 1064nm single or 1064nm PDP function<br>- Removal dark ink (black, blue and brown)<br>tattoos<br>- Treatment melasma<br>- Treatment nevus of Ota<br>- Treatment common nevi<br>- Removal of lightening of unwanted hair<br>- Skin resurfacing procedures for the treatment<br>of acne scars and wrinkles<br><br>With 532nm single function<br>- Removal of light ink (red, tan, purple, and<br>orange) tattoos<br>- Treatment of vascular lesions including, but not<br>limited to port-wine stains, telangiectasias,<br>spider angioma, cherry angioma |
| Finebeam Dual – 510(k)<br>Summary | - SNJ Co., Ltd. - | Page 05-5 of 15 |
| Device Name | Fotona QX Nd:YAG/KTP Laser System Family<br>(Predicate) | Finebeam (Proposed) |
| | - Treatment of vascular lesions including, but not<br>limited to:<br>- port wine birthmarks<br>- telangiectasias<br>- spider angioma<br>- Cherry angioma<br>- Spider nevi<br>- Treatment of pigmented lesions including, but<br>not limited to:<br>- cafe-au-lait birthmarks<br>- solar lentigines<br>- senile lentigines<br>- Becker's nevi<br>- freckles<br>- common nevi<br>- nevus spilus<br>- Treatment of seborrheic keratosis<br>- Treatment of post inflammatory<br>hyperpigmentation<br>- Skin resurfacing procedures for the treatment<br>of acne scars and wrinkles. | - Treatment of pigmented lesions including, but<br>not limited to cafe-au-Lait birthmarks, solar<br>lentigines, senile lentigines, Becker's nevi,<br>freckles, common nevi, and nevus spilus<br>- Treatment of seborrheic keratosis<br>- Treatment of post-inflammatory<br>hyperpigmentation<br>- Skin resurfacing procedures for the treatment<br>of acne scars and wrinkles |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the SNJ logo. The logo consists of the letters SNJ in a stylized font. The letters are in blue. To the right of the logo is the text "SNJ" with a line underneath it, followed by the text "No. 604,"
SNJ Co., Ltd.
No. 604, Ace-Techno Tower 2, Digital-ro 31-gil 19, Guro-gu, Seoul, Republic of Korea
### 9.2. Intended use comparison of Non-Q-Switched function in the Q-Switched mode
| Device Name | Fotona QX Nd:YAG/KTP Laser System Family<br>(Predicate) | Finebeam (Proposed) |
|--------------|---------------------------------------------------------|---------------------|
| Product Code | GEX | Same |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for SNJ Co., Ltd. The logo consists of the letters "SNJ" in a stylized font, with the words "SNJ Co., Ltd." written in a more traditional font to the right of the letters. Below the company name is the text "No. 604, Ace-Techno To". The letters "SNJ" are in blue, while the company name and address are in black.
| Device Name | Fotona QX Nd:YAG/KTP Laser System Family<br>(Predicate) | Finebeam (Proposed) |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number | 21 CFR 878.4810 | Same |
| K number | K083889 | K202288 |
| Applicant | Fotona d.o.o. | SNJ Co., Ltd. |
| Wavelength | 1064nm | Same |
| Intended use | 1064 nm wavelength in non-Q-switched mode:<br>- Removal of unwanted hair, for stable long term<br>or permanent hair reduction and for treatment<br>of PFB. The laser is indicated for all skin types,<br>Fitzpatrick I-VI, including tanned skin<br>- Photocoagulation and hemostasis of pigmented<br>and vascular lesions, such as, but not limited<br>to,port wine stains, hemaongiomae, warts,<br>telangiectasiae, rosacea, venus lake, leg veins<br>and spider veins<br>- Coagulation and hemostasis of soft tissue<br>- Treatment of wrinkles<br>- Treatment of mild to moderate inflammatory<br>acne vulgaris | With 1064nm Non-QSW function<br>- Removal of unwanted hair, for stable long term<br>or permanent hair reduction and treatment of<br>PFB (Pseudofolliculitis Barbae). The laser is<br>indicated for all skin types, Fitzpatrick I-VI,<br>including tanned skin<br>- Photocoagulation and hemostasis of pigmented<br>and vascular lesions, such as but not limited to<br>port wine stains, cherry angioma,<br>Hemangiomas, warts, telangiectasias, rosacea,<br>leg veins, and spider veins<br>- Coagulation and hemostasis of soft tissue<br>- Treatment of wrinkles |
#### Intended use comparison of PDP function in the QSW mode 9.3.
| Device Name | Pastelle<br>Q-Switched<br>(Predicate) | Nd:YAG Laser System | Finebeam (Proposed) |
|-------------------|---------------------------------------|---------------------|---------------------|
| Product Code | GEX | | Same |
| Regulation Number | 21 CFR 878.4810 | | Same |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for SNJ Co., Ltd. The logo consists of the letters "SNJ" in a stylized font on the left side of the image. To the right of the letters is the company name, "SNJ Co., Ltd." The text is in a bold, sans-serif font.
| Device Name | Pastelle Q-Switched Nd:YAG Laser System<br>(Predicate) | Finebeam (Proposed) |
|--------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K number | K123293 | K202288 |
| Applicant | WON Technology Co., Ltd. | SNJ Co., Ltd. |
| Wavelength | 1064nm and 532nm | Same |
| Intended use | 1064 nm wavelength in PTP mode:<br>- Removal of dark ink (black, blue and brown)<br>tattoos<br>- Treatment of melasma<br>- Treatment of nevus of Ota<br>- Treatment of common nevi<br>- Removal or lightening of unwanted hair<br>- Skin resurfacing procedures for the treatment<br>of acne scars and wrinkles | With 1064nm PDP function<br>- Removal of dark ink (black, blue and brown)<br>tattoos<br>- Treatment of melasma<br>- Treatment of nevus of Ota<br>- Treatment of common nevi<br>- Removal or lightening of unwanted hair<br>- Skin resurfacing procedures for the treatment<br>of acne scars and wrinkles |
9.4. Intended use comparison of LPS mode
| Device Name | Fotona XP System Family | Nd:YAG Laser | Family of Lasers | CoolGlide Aesthetic | Finebeam (Proposed) |
|-------------------|----------------------------------------------------------------------------------|--------------|----------------------------------------|---------------------|----------------------------------------|
| Product Code | GEX | | Same | | Same |
| Regulation Number | 21 CFR 878.4810 | | Same | | Same |
| K number | K090126 | | K132185 | | K202288 |
| Applicant | Fotona d.o.o. | | Cutera, Inc. | | SNJ Co., Ltd. |
| Wavelength | 1064nm | | Same | | Same |
| Intended use | - Removal of unwanted hair, for stable long term or permanent hair reduction and | | 1064 nm wavelength in long pulse mode: | | 1064 nm wavelength in long pulse mode: |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for SNJ Co., Ltd. The logo consists of the letters "SNJ" in a stylized font on the left, followed by the text "SNJ Co., Ltd." on the right. Below the company name is the text "No. 604, Ace-Techno Tower".
| Device Name | Fotona XP Nd:YAG Laser System Family | Family of CoolGlide Aesthetic Lasers | Finebeam (Proposed) | |
|-------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| | for treatment of PFB. The<br>lasers are indicated on all<br>skin types Fitzpatrick I-VI<br>including tanned skin.<br>- Photocoagulation and<br>hemostasis of pigmented and<br>vascular lesions, such as but<br>not limited to port wine stains,<br>hemangiomas, warts,<br>telangiectasias, rosacea, leg<br>veins and spider veins.<br>- Coagulation and hemostasis<br>of soft tissue.<br>- Incision/excision of soft body<br>tissue in dermatology.<br>- Soft tissue general surgery<br>applications-skin incision;<br>tissue dissection; cornpelte or<br>partial resection of internal<br>organs, tumors, lesions;<br>tissue ablation; vessel<br>coagulation.<br>- Benign pigmented lesions<br>such as, but not limited to:<br>lentigos (age spots), solar<br>lentigos (sun spots), cafe au<br>lait macules, seborrheic | - Coagulation and hemostasis<br>of benign vascular lesions<br>such as, but not limited to,<br>port wine stains,<br>hemangiomas, warts,<br>telangiectasias, rosacea,<br>venus lake, leg veins, spider<br>veins and poikilodlerma of<br>civatte; and treatment of<br>benign cutaneous lesions<br>such as, but not limited to,<br>warts, scars, striae and<br>psoriasis. The lasers are also<br>intended for the treatment of<br>benign pigmented lesions<br>such as, but not limited to,<br>lentigos (age spots), solar<br>lentigos (sun spots), cafe and au<br>lait macules, seborrheic<br>keratoses, nevi, chloasma,<br>verrucae, skin tags,<br>keratoses, tattoos (significant<br>reduction in the intensity of<br>black and/or blue/black<br>tattoos) and plaques.<br>- Pigmented lesions to reduce<br>lesion size, for patients with | - Removal of unwanted hair,<br>for stable long term or<br>permanent hair reduction and<br>treatment of PFB. The lasers<br>are indicated on all skin types<br>Fitzpatrick I-VI including<br>tanned skin.<br>- Photocoagulation and<br>hemostasis of pigmented and<br>vascular lesions, such as but<br>not limited to port wine stains<br>hemangiomas, warts,<br>telangiectasias, rosacea, leg<br>veins, and spider veins.<br>- Coagulation and hemostasis<br>of soft tissue.<br>- Incision/excision of soft body<br>tissue.<br>- Treatment of wrinkles and<br>acne vulgaris | |
| Device Name | Fotona XP Nd:YAG Laser System Family | Family of CoolGlide Aesthetic Lasers | Finebeam (Proposed) | |
| | keratosis, nevi, chloasma (melisma), verruca, skin tags, keratosis, tattoo (significant reduction in the intensity of black and/or blue-black tattoo) and plaques.<br>- Treatment of wrinkles and acne vulgaris. | lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.<br>- Treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles.<br>- Temporary and permanent hair reduction. Permanent hair reduction is defined as long-term, stable reduction in hair counts observed at 6, 9, and 12 months after the end of a treatment regime.<br>- Also indicated for the treatment for pseudofolliculitis barbae (PFB)<br>- Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.<br>- All skin types (Fitzpatrick I-VI), including tanned skin. | | |
| Device Name | Fotona XP Nd:YAG Laser<br>System Family | | Family of CoolGlide Aesthetic<br>Lasers | Finebeam (Proposed) |
| | | | - Incision/excision and cutting,<br>ablation,<br>coagulation/hemostasis of<br>soft tissue in the performance<br>of surgical applications in<br>endoscopy/laparoscopy,<br>gastroenterology, general<br>surgery, head and<br>necklotorhinolaryngology<br>(ENT), neurosurgery,<br>oculoplastics, orthopedics,<br>plastic surgery,<br>pulmonary/thoracic surgery,<br>gynecology (e.g.<br>menorrhagia) and urology. | |
{13}------------------------------------------------
### SIZ SNJ Co., Ltd.
No. 604, Ace-Techno Tower 2, Digital-ro 31-gil 19, Guro-gu, Seoul, Republic of Korea
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for SNJ Co., Ltd. The logo consists of the letters "SNJ" in a stylized font on the left, followed by the words "SNJ Co., Ltd." on the right. Below the company name is the text "No. 604, Ace-Techno Tower". The letters "SNJ" on the left are in a light blue color, while the rest of the text is in black.
9.5. Technical characteristics comparison
- Q-Siwtch mode י
| Predicate | Fotona QX Nd:YAGIKTP Laser<br>System Family | Pastelle Q-Switched<br>Laser System | Nd:YAG<br>Finebeam |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | GEX | Same | Same |
| Regulation Number | 21CFR 878.4810 | Same | Same |
| K number | K083889 | K123293 | K202288 |
| Applicant | Fotona d.o.o. | WON Technology Co., Ltd. | SNJ Co., Ltd. |
| Predicate | Fotona QX Nd:YAGIKTP Laser<br>System Family | Pastelle Q-Switched<br>Laser System | Nd:YAG<br>Finebeam |
| Laser type | Nd:YAG Laser | Same | Same |
| Wavelength | 532nm and 1064nm | Same | Same |
| Pulse frequency | up to 10 Hz | 1-10 Hz | 1-10 Hz |
| Pulse duration | 1064nm Q-switched mode: 5-<br>20ns<br>532nm Q-switched mode: 5-<br>20ns<br>non-Q switched mode: 0.25ms | Q-switched Mode: 4 to 48 ns<br>Genesis Mode (Non-Q): 0.08 -<br>0.48ms | 1064 single (Q-switched mode):<br>< 25ns<br>532 single (Q-switched mode): <<br>25ns<br>PDP mode: < 25ns<br>non-Q switched mode: 0.3ms |
| Energy (max) | 1600mJ at 1064nm<br>600mJ at 532nm<br>5000mJ at Non-Q mode | 1300mJ at 1064nm<br>500mJ at 532nm<br>1600mJ at PTP<br>3500mJ at Non-Q mode | 1600mJ at 1064nm<br>500mJ at 532nm<br>2200mJ at PTP<br>3500mJ at Non-Q |
| Spot size | - R28 Handpiece<br>: 1064nm = 2 ~ 8 mm<br>: 532nm = 2 ~ 8 mm<br>- R585, 650 Handpiece<br>: 585nm = 2 ~ 4<br>: 650nm = 2 ~ 4 | - Zoom Handpiece<br>: 1064 nm, 532 nm = 2~10 mm<br>- Collimator Handpiece<br>: 1064 nm = 8 mm<br>- Fractional Handpiece (Option)<br>: 1064 nm = unknown the size per<br>spot<br>Area: 5x5 mm² (DOE,<br>Diffractive optical element),<br>Spacing between micro spot:<br>unknown | - Zoom Handpiece<br>: 1064 nm = 2~10 mm<br>: 532 nm = 1~7 mm<br>- Collimator Handpiece<br>: 1064 nm = 8 mm |
| Aiming laser | 650nm diode laser (<5mW),<br>Class III | 650nm diode laser (<5mW), Class<br>III | 655nm diode laser (<5mW),<br>Class III |
| Beam Delivery | Articulated arm | same | same |
| Predicate | Fotona QX Nd:YAGIKTP Laser System Family | Pastelle Q-Switched Laser System Nd:YAG | Finebeam |
| User interface | Push button control | LCD Touch screen | LCD Touch screen |
| - Long Pulse mode | | | |
| Predicate | Fotona XP Nd:YAG Laser System Family | Family of CoolGlide Aesthetic Lasers | Finebeam |
| Product Code | GEX | Same | Same |
| Regulation Number | 21CFR 878.4810 | Same | Same |
| K number | K090126 | K132185 | K202288 |
| Applicant | Fotona d.o.o. | Cutera, Inc. | SNJ Co., Ltd. |
| Laser type | Nd:YAG Laser | Same | Same |
| Wavelength | 1064nm | 1064 nm | 532nm and 1064nm |
| Pulse frequency | up to 75 Hz | 1-10 Hz and single shot | 1-10 Hz |
| Pulse duration | 0.1-300ms | 0.1-300ms | Long Pulse mode: 0.5-300ms |
| Energy (max) | Up to 120 J | up to 300J/cm2 | 60J at LPS (Long pulse) |
| Spot size | R33 Handpiece<br>: 2 - 10 mm<br>R34 Handpiece<br>: 15 - 20 mm | Integrated 3, 5, 8, 10 mm | - Zoom Handpiece<br>: 1064 nm = 2~10 mm<br>: 532 nm = 1~7 mm<br>- Collimator Handpiece<br>: 1064 nm = 8 mm |
| Aiming laser | 650nm diode laser (<5mW),<br>Class III | 630-680 nm | 655nm diode laser (<5mW),<br>Class III |
| Predicate | Fotona XP Nd:YAG Laser System Family | Family of CoolGlide Aesthetic Lasers | Finebeam…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.