K132286 · Scarlet Rf, Inc. · GEX · Jan 22, 2014 · General, Plastic Surgery
Device Facts
Record ID
K132286
Device Name
HYPERION LONG PULSED ND: YAG LASER
Applicant
Scarlet Rf, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jan 22, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Hyperion is a long pulsed Nd:YAG laser intended for treatment of wrinkles, Removal of unwanted hair for stable long term or *permanent hair reduction, treatment of pseudofolliculitis barbae (PFB), surgical incision, excision, vaporization, ablation and coagulation of soft tissue, and photocoagulation and hemostasis of pigmented and vascular lesions in dermatology. *Permanent hair reduction is defined as the long-term stable reduction in the number of hairs re-grown when measured at 6, 9, and 12 months after the completion of a treatment regimen
Device Story
Hyperion is a long-pulsed Nd:YAG laser system (1064 nm) for dermatological and surgical applications. System components include a main cabinet (power supply, cooling system, microprocessor, laser generator), optical fiber with handpieces, and a footswitch. Operated by clinicians in a medical setting, the device delivers laser energy to target tissue for ablation, coagulation, or hair reduction. The microprocessor controls laser emission parameters. Clinical benefit is achieved through precise thermal interaction with soft tissue, pigmented lesions, or vascular structures. The device is used for surgical incision, excision, and various dermatological treatments.
Clinical Evidence
Bench testing only. Non-clinical testing included visual and mechanical inspection, electrical and mechanical safety testing (IEC 60601-1 and 60601-1-2), and software testing. No clinical data provided.
Technological Characteristics
Long-pulsed Nd:YAG laser; 1064 nm wavelength. System includes main cabinet, optical fiber, handpieces, and footswitch. Electrical safety per IEC 60601-1 and IEC 60601-1-2. Microprocessor-controlled emission.
Indications for Use
Indicated for patients requiring dermatological laser treatment for wrinkles, unwanted hair removal (permanent reduction), pseudofolliculitis barbae (PFB), soft tissue surgical procedures (incision, excision, vaporization, ablation, coagulation), and photocoagulation/hemostasis of pigmented and vascular lesions.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
### Section 4 510(k) Summary Statement Hyperion Long Pulsed Nd:YAG Laser
.
______________________________________________________________________________________________________________________________________________________________________________
# Section 4 - This 510(k) Summary of safety and effectiveness is submitted in accordance with the requirements of 21 CFR §807.92
| Submitter's Information | |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Applicant: | Laseroptek Co. Ltd. |
| Address: | 204 Hyundai I Valley<br>223-12 Sangdaiwon, Jungwon<br>Sungnam-Si, Gyeonggido, 462-714<br>Rep. of Korea (South Korea) |
| Contact Person: | Hong Chu, Ph.D. (President & CEO) |
| Telephone / Fax / E-mail: | Tel) +82.31.737.9885<br>E-mail) hchu@laseroptek.com |
| Submitter: | Scarlet RF, Inc. |
| Address: | 17145 Von Karman Ave, Suite 106<br>Irvine, CA 92614, USA |
| Contact Person: | Kevin J. Choi |
| Telephone / Fax / E-Mail: | Tel) 310-634-4480<br>E-mail) kjychoi@gmail.com |
| Preparation Date: | 04/22/2013 |
| Device Trade Name: | Hyperion |
| Common Name: | Long Pulsed Nd:YAG Dermatological Laser |
| Classification Name: | Instruments, Surgical, Powered, Laser 79-GEX<br>(21 CFR §878-4810) |
| Legally Marketed Predicate Devices: | Candela, Inc. - Gentle YAG Laser System<br>(K022951)<br>Fotona d.d. - XP Nd:YAG Laser System<br>(K090126) |
- Cynosure, Inc. Apogee Elite Laser (K034030) .
{1}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| Intended Use of Hyperion: | Hyperion is a long pulsed Nd:YAG laser intended for treatment<br>of wrinkles, Removal of unwanted hair for stable long term or<br>*permanent hair reduction, treatment of pseudofolliculitis barbae<br>(PFB), surgical incision, excision, vaporization, ablation and<br>coagulation of soft tissue, and photocoagulation and hemostasis<br>of pigmented and vascular lesions in dermatology. *Permanent<br>hair reduction is defined as the long-term stable reduction in the<br>number of hairs re-grown when measured at 6, 9, and 12 months<br>after the completion of a treatment regimen |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Description: | Hyperion is a long pulse Nd:YAG laser system that produces<br>laser emission at the wavelength of 1064nm. The system consists<br>of three interconnected sections: a) the main cabinet, which<br>houses the power supply, cooling system, microprocessor, and<br>the laser generator, b) the optical fiber with hand pieces which<br>can be used selectively depending on the size of the laser beam<br>exposure, c) the footswitch. |
| Performance Data: | Non-clinical testing of Hyperion included visual and mechanical<br>inspection, electrical and mechanical safety testing, software<br>testing, etc., in bench. Electrical testing was performed in<br>accordance with IEC 60601-1 and 60601-1-2 standards. |
| Substantial Equivalence<br>(Intended Use) | Hyperion's intended use is identical to the intended use of the<br>predicate devices: Candela's Gentle YAG Laser System, Fotona's<br>XP Nd:YAG Laser System, and Cynosure's Apogee Elite Laser |
| Substantial Equivalence<br>(Technical) | Hyperion is based on the Nd:YAG (1064 nm) laser technology as<br>its predicated devices listed above. Hyperion long pulsed<br>Nd:YAG laser operates in a similar fashion as its predicate<br>devices, with the same principle of operation, the same<br>wavelength (1064 nm) and essentially the same power range and<br>the same indications for uses. |
| | Based upon analysis of the overall performance characteristic for<br>the device, Laseroptek Co. Ltd. believes that no significant<br>differences exist between Hyperion and its predicate devices.<br>Therefore, Hyperion should not raise new safety or effectiveness<br>issues. |
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird-like figure with three stylized wing or feather shapes.
Public Health Service
Food and Drug Administration 10903 New Hamnshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002
Laseroptek Company, Ltd % Mr. Kevin J. Choi Scarlet RF Incorporated 28 Stafford Drive West Windsor, New Jersey 08550
January 22, 2014
Re: K132286
Trade/Device Name: Hyperion Long Pulse Nd: YAG Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: December 24, 2013 Received: December 27, 2013
Dear Mr. Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 - Mr. Kevin J. Choi
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Joshua
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director For Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K132286
Device Name Hyperion Long Pulse Nd: YAG Laser
Indications for Use (Describe)
Treatment of wrinkles
Removal of unwanted hair for stable long term or *permanent hair reduction
Treatment of PFB (pseudofolliculitis barbae)
Surgical incision, excision vaporization, ablation and coagulation of soft tissue
Photocoagulation and hemostasis of pigmented and vascular lesions in dermatology
*Permanent hair reduction is defined as the long-term stable reduction in the number of hair re-grown when measured at 6, 9, and 12 months after the completion of a treatment regimen.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
The proportion of the FOR FDA USE ONLY CONTRACT AND A PARTY CHANGE OF SCHOOL
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/4/Picture/17 description: The image shows the name "Neil R Ogden" followed by "-S". Below that is the date "2014.01.21" and the time "13:56:37". To the right of the time is a logo that is difficult to make out. The logo appears to have the numbers "05:00" in it.
FORM FDA 3881 (1/14)
PSC Publishing Services (301) 443-6740
FI
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.