K202238 · Ge Healthcare (Tianjin) Company Limited · LNH · Sep 4, 2020 · Radiology
Device Facts
Record ID
K202238
Device Name
SIGNA Artist
Applicant
Ge Healthcare (Tianjin) Company Limited
Product Code
LNH · Radiology
Decision Date
Sep 4, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
AI/ML
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Magnetic Resonance Image Quality
Deep learning-based image reconstruction
—
Improved SNR and image sharpness while maintaining low contrast detectability
—
—
Clinical evaluation of sample images from clinically indicated scans
>1 (radiologists)
Indications for Use
The SIGNA Artist system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the SIGNA Artist system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Device Story
SIGNA Artist is a whole-body MR scanner using superconducting magnets, time-varying magnetic field gradients, and RF transmissions to acquire spatial/molecular data. System supports up to 128 receive channels. New feature, AIR Recon DL, is a deep-learning-based reconstruction algorithm that processes raw MR data to improve signal-to-noise ratio (SNR) and image sharpness, potentially enabling shorter scan times. Operated by trained clinicians in clinical settings; output is viewed as diagnostic images/spectra to assist physician diagnosis. Benefits include enhanced image quality and efficiency.
Clinical Evidence
Clinical evaluation involved analysis of in vivo images and a reader study. Radiologists compared images with and without AIR Recon DL across various anatomies, pulse sequences, and contrast-enhanced cases. Results showed equivalent or superior image quality, maintained lesion conspicuity, and radiologist preference for AIR Recon DL images.
Technological Characteristics
Whole-body MR scanner; superconducting magnet; 128-channel data acquisition system. Software includes AIR Recon DL deep-learning reconstruction. Connectivity: networked. No changes to gradient or RF subsystems compared to predicate. Technical specs (magnet stability, gradient strength/slew rate) unchanged.
Indications for Use
Indicated for use as a whole-body diagnostic MR imaging device for patients requiring axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, or dynamic images of structures/functions including head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions. Contrast agents may be used.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
September 4, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
GE Healthcare (Tianjin) Company Limited % Mr. Glen Sabin Regulatory Affairs Director GE Healthcare (GE Medical System, LLC) 3200 N Grandview Blvd. WAUKESHA WI 53188
### Re: K202238
Trade/Device Name: SIGNA Artist Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH, LNI, MOS Dated: August 6, 2020 Received: August 7, 2020
Dear Mr. Sabin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K202238
Device Name SIGNA Artist
#### Indications for Use (Describe)
The SIGNA Artist system is a whole body magnetic resonance scanner designed to support high signalto-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal. and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used.
The images produced by the SIGNA Artist system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained that may assist in diagnosis.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the alphanumeric string "K202238" in a simple, sans-serif font. The characters are uniformly sized and spaced, with a consistent stroke width. The string appears to be a code or identifier, possibly a serial number or product code.
Image /page/3/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a circular design with the letters "GE" intertwined in the center. There are three curved lines emanating from the top, left, and right sides of the circle, giving the impression of motion or energy.
510(k) Premarket Notification Submission SIGNA Artist
### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | August 06, 2020 |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare (GE Healthcare (Tianjin) Company Limited)<br>No. 266 Jingsan Road,Tianjin Airport Economic Area<br>Tianjin, China 300308 |
| Primary Contact<br>Person: | Qiang Ding<br>Regulatory Affairs Leader<br>Phone: +86 13311385163<br>Email: Ding.Qiang@ge.com |
| Secondary Contact<br>Person: | Glen Sabin<br>Regulatory Affairs Director<br>Phone: 262-894-4968<br>Email: Glen.Sabin@ge.com |
| Device Trade Name: | SIGNA Artist |
| Common/Usual Name: | Magnetic Resonance Diagnostic Device |
| Classification Names: | Magnetic Resonance Diagnostic Device |
| Regulation Number: | 21 CFR 892.1000 |
| Primary Product Code: | LNH |
| Secondary Product Code: | LNI, MOS |
| Predicate Device: | SIGNA Artist (K163331) |
| Device Description: | The SIGNA Artist system is a whole body magnetic resonance scanner<br>designed to support high resolution, high signal-to-noise ratio, and short scan<br>times. The system features a superconducting magnet. The data acquisition<br>system accommodates up to 128 independent receive channels in various<br>increments and multiple independent coil elements per channel during a<br>single acquisition series. The system uses a combination of time varying<br>magnetic fields (gradients) and RF transmissions to obtain information<br>regarding the density and position of elements exhibiting magnetic<br>resonance. The system can image in the sagittal, coronal, axial, oblique, and<br>double oblique planes, using various pulse sequences and reconstruction<br>algorithms.<br>This 510(k) submission is for the SIGNA Artist 1.5T MR system, and has<br>been triggered by the addition of the AIR Recon DL software feature. |
| | The AIR Recon DL feature has been previously cleared for use on the<br>SIGNA Premier 3T system through K193282, which is used as a reference<br>device for this submission. |
| Indications for Use: | The Indications for Use statement for the proposed device is identical to<br>that of the predicate device: |
| | The SIGNA Artist system is a whole body magnetic resonance<br>scanner designed to support high resolution, high signal-to-noise<br>ratio, and short scan times. It is indicated for use as a diagnostic<br>imaging device to produce axial, sagittal, coronal, and oblique<br>images, spectroscopic images, parametric maps, and/or spectra,<br>dynamic images of the structures and/or functions of the entire<br>body, including, but not limited to, head, neck, TMJ, spine, breast,<br>heart, abdomen, pelvis, joints, prostate, blood vessels, and<br>musculoskeletal regions of the body. Depending on the region of<br>interest being imaged, contrast agents may be used. |
| | The images produced by the SIGNA Artist system reflect the<br>spatial distribution or molecular environment of nuclei exhibiting<br>magnetic resonance. These images and/or spectra when<br>interpreted by a trained physician yield information that may assist<br>in diagnosis. |
| | The addition of the AIR Recon DL feature does not impact the intended<br>use of the SIGNA Artist system. |
| Technology<br>Characteristics: | Many of the technological characteristics of the proposed SIGNA Artist<br>system are unchanged from the predicate device. The SIGNA Artist<br>system has been introduced an additional configuration with IPM magnet.<br>There are no changes to gradient, and RF subsystems compared to the<br>predicate K163331. Key performance specifications (such as magnet<br>stability, maximum gradient strength and slew rate, etc.) for the system<br>are also unchanged. |
| | The software used on the proposed SIGNA Artist system has been<br>modified to include the AIR Recon DL feature. The User interface<br>provides operators of the system with new options for selecting AIR<br>Recon DL and adjusting the associated level of image noise reduction.<br>The resulting images can have higher SNR and improved sharpness<br>compared to images reconstructed without AIR Recon DL. |
| | AIR Recon DL has been previously cleared for use with GE Healthcare's<br>3T SIGNA Premier system through K193282. Due to the technical<br>similarities, SIGNA Premier (K193282) is used as a reference device for<br>this submission. |
| Determination of | Summary of Non-Clinical Tests: |
| Substantial<br>Equivalence: | The AIR Recon DL feature has undergone the performance testing.<br>These tests were designed to evaluate the AIR Recon DL feature and its<br>impact on image quality, including SNR, sharpness, low contrast<br>detectability, and noise spectral content. Analysis was performed to<br>confirm that the feature does not introduce significant bias that might<br>impact quantitative measurements based on signal intensity. The<br>influence of motion during image acquisition on the performance of AIR<br>Recon DL was also evaluated.<br>The nonclinical testing demonstrated that AIR Recon DL does improve |
| | SNR and image sharpness while maintaining low contrast detectability<br>and having minimal impacts to noise spectral content, average signal<br>intensity, or the appearance of motion artifacts. AIR Recon DL was also<br>able to maintain image SNR and did not sacrifice sharpness for images<br>acquired with a reduced scan time. The nonclinical testing passed the<br>defined acceptance criteria, and did not identify any adverse impacts to<br>image quality or other concerns related to safety and performance. |
| | Summary of Clinical Tests:<br>Objective measures of in vivo images were analyzed to confirm that AIR<br>Recon DL improves SNR and image sharpness for typical clinical use |
| | cases. |
| | Additionally, sample images from clinically indicated scans were<br>evaluated both with and without the AIR Recon DL feature. These<br>samples included images using exogenous contrast and images involving<br>pathology spanning a variety of anatomies and pulse sequences.<br>Radiologists were asked to rate the images, and to comment on any<br>notable aspects related to image quality. This study showed that the AIR<br>Recon DL feature provides images with equivalent or better image<br>quality, lesion conspicuity is maintained, and that the radiologists<br>preferred the AIR Recon DL images for clinical use. |
| Conclusion Drawn from<br>Performance Testing: | The nonclinical and clinical testing demonstrated that AIR Recon DL<br>satisfies the product claims of improved SNR and image sharpness, and<br>can enable shorter scan times while maintaining SNR and image<br>sharpness. |
| | The proposed SIGNA Artist system with AIR Recon DL has been<br>developed under GE Healthcare's quality system and is at least as safe<br>and effective as the legally marketed predicate. The performance testing<br>did not identify any new hazards, adverse effects, or safety and<br>performance. |
| performance concerns that are significantly different from those. | |
| associated with MR imaging in general.<br>Therefore, GE Healthcare believes that SIGNA Artist with AIR Recon DL<br>is substantially equivalent to the predicate device, and is safe and<br>effective for its intended use. | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circle. There are three decorative flourishes around the circle, resembling stylized water droplets or leaves. The logo is black and white.
510(k) Premarket Notification Submission SIGNA Artist
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized script, enclosed within a circle. There are three decorative flourishes extending outward from the circle, positioned at the top, left, and right. The logo is black and white.
510(k) Premarket Notification Submission SIGNA Artist
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are enclosed within a circle, and there are three droplet-shaped elements evenly spaced around the outside of the circle.
510(k) Premarket Notification Submission SIGNA Artist
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.